Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 47: 1279-1286, 2001;
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (11)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Eddy, C. V.
Right arrow Articles by Arnold, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Eddy, C. V.
Right arrow Articles by Arnold, M. A.
Related Collections
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 2001;47:1279-1286.)
© 2001 American Association for Clinical Chemistry, Inc.


Articles

Near-Infrared Spectroscopy for Measuring Urea in Hemodialysis Fluids

Christopher V. Eddy1 and Mark A. Arnold1a

1 Department of Chemistry and Optical Science and Technology Center, University of Iowa, Iowa City, IA 52242.


aAuthor for correspondence. Fax 319-335-1115; e-mail mark-arnold{at}uiowa.edu.

Background: Near-infrared spectroscopy is proposed as a method for providing real-time urea concentrations during hemodialysis treatments. The feasibility of such noninvasive urea measurements is evaluated in undiluted dialysate fluid.

Methods: Near-infrared spectra were collected from calibration solutions of urea prepared in dialysate fluid. Spectra were collected over three distinct spectral regions, and partial least-squares calibration models were optimized and compared for each. Selectivity for urea was demonstrated with two-component samples composed of urea and glucose in the dialysate matrix. The clinical significance of this approach was assessed by measuring urea in real hemodialysate samples.

Results: Urea absorptions within the combination and short-wavelength, near-infrared spectral regions provided sufficient spectral information for sound calibration models in the dialysate matrix. The combination spectral region had SEs of calibration (SEC) and prediction (SEP) of 0.38 mmol/L and 0.26 mmol/L, respectively, over the 4720–4600 cm-1 spectral range with 5 partial least-square factors. A second calibration model was established over the combination region from a series of solutions prepared with independently variable concentrations of urea and glucose. The best calibration model for urea in the presence of variable glucose concentrations had a SEC of 0.6 mmol/L and a SEP of 0.4 mmol/L for a 5-factor model over the 4600–4350 cm-1 spectral range. There was no significant decrease in SEP when the 4720–4600 cm-1 calibration model was used to measure urea in real samples collected during actual hemodialysis.

Conclusions: Urea can be determined with sufficient sensitivity and selectivity for clinical measurements within the matrix of the hemodialysis fluid.




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
Q.-B. Li, X.-J. Sun, Y.-Z. Xu, L.-M. Yang, Y.-F. Zhang, S.-F. Weng, J.-S. Shi, and J.-G. Wu
Diagnosis of Gastric Inflammation and Malignancy in Endoscopic Biopsies Based on Fourier Transform Infrared Spectroscopy
Clin. Chem., February 1, 2005; 51(2): 346 - 350.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
J. T. Olesberg, M. A. Arnold, and M. J. Flanigan
Online Measurement of Urea Concentration in Spent Dialysate during Hemodialysis
Clin. Chem., January 1, 2004; 50(1): 175 - 181.
[Abstract] [Full Text] [PDF]