Background: The International Sensitivity Index (ISI) calibration of whole-blood prothrombin time (PT) monitors for point-of-care testing (POCT) described by Tripodi et al. (Thromb Haemost 1993;70:9214) has been shown to be dependable but is too complex and demanding. The use of plasma would simplify calibration of whole-blood POCT PT monitors, but important differences may exist between the ISI for whole blood and plasma calibrations.
Methods: In a 10-center calibration study of two POCT whole-blood monitoring systems (CoaguChek Mini and TAS PT-NC), we characterized the relationship between the log PT for whole blood and fresh plasma with use of single lots of test strips/cards. This relationship (linear) was used to correct the difference between the whole-blood and plasma ISI. The reliability of the correction with different lots of test strips/cards was assessed at three centers. The linear relationship was used to correct the difference in the whole-blood and plasma ISI with four other lots of TAS PT-NC cards and with two additional lots of CoaguChek Mini test strips.
Results: The correction decreased the ISI difference from 13.3% to 0.9% for the TAS PT-NC and from 5.7% to 0.6% for the CoaguChek Mini. In assessments at three centers, which included different lots of test strips/cards, the mean ISI difference was markedly decreased with the TAS PT-NC but not with the CoaguChek Mini, for which the mean ISI difference increased slightly.
Conclusions: The proposed correction resolves the discrepancy between whole-blood and fresh plasma ISI calibrations with TAS PT-NC test cards. The CoaguChek Mini systems could be calibrated without this correction.
The following articles in journals at HighWire Press have cited this article:
![]() |
L. Poller, J. Jespersen, M. Keown, S. Ibrahim, and A. Tripodi The authors of the article cited above respond: Clin. Chem., August 1, 2007; 53(8): 1556 - 1557. [Full Text] [PDF] |
||||
![]() |
L. Poller, M. Keown, N. Chauhan, A. M.H.P. van den Besselaar, A. Tripodi, C. Shiach, and J. Jespersen European Concerted Action on Anticoagulation. Quality Assessment of the CoaguChek Mini and TAS PT-NC Point-of-Care Whole-Blood Prothrombin Time Monitors Clin. Chem., March 1, 2004; 50(3): 537 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Poller, M. Keown, N. Chauhan, A. M. van den Besselaar, A. Tripodi, C. Shiach, and J. Jespersen Reliability of international normalised ratios from two point of care test systems: comparison with conventional methods BMJ, July 3, 2003; 327(7405): 30. [Abstract] [Full Text] [PDF] |
||||
![]() |
L Poller, M Keown, N Chauhan, A M H P van den Besselaar, A Tripodi, C Shiach, and J Jespersen European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors J. Clin. Pathol., February 1, 2003; 56(2): 114 - 119. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |