Clinical Chemistry Link to Randox Laboratories Web Site
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 48: 1011-1015, 2002;
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via ISI Web of Science (6)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Meijer, P.
Right arrow Articles by van Houwelingen, H. C.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Meijer, P.
Right arrow Articles by van Houwelingen, H. C.
Related Collections
Right arrow Laboratory Management
(Clinical Chemistry. 2002;48:1011-1015.)
© 2002 American Association for Clinical Chemistry, Inc.

Long-Term Analytical Performance of Hemostasis Field Methods as Assessed by Evaluation of the Results of an External Quality Assessment Program for Antithrombin

Piet Meijer1,2a, Moniek P.M. de Maat1,2, Cornelis Kluft1,2, Frits Haverkate1 and Hans C. van Houwelingen1,3

1 ECAT Foundation, 2301 CE Leiden, The Netherlands

2 Gaubius Laboratory, TNO Prevention and Health, 2333 CK Leiden, The Netherlands

3 Leiden University Medical Center, Department of Medical Statistics, 2300 RC Leiden, The Netherlands

aAddress correspondence to this author at: ECAT Foundation, PO Box 2215, 2301 CE Leiden, The Netherlands. Fax 31-71-5181330; e-mail P.Meijer{at}ecat.nl.

Background: It is important for a laboratory to know the stability of performance of laboratory tests over time. The aim of this study was to adapt from the field of clinical chemistry a method to assess the long-term analytical performance of hemostasis field methods.

Methods: The linear regression model was used to compare the laboratory results with the consensus mean value of a survey. This model was applied to plasma antithrombin activity using the data for 82 laboratories, collected between 1996 and 1999 in the European Concerted Action on Thrombosis (ECAT) external quality assessment program. The long-term total, random, and systematic error were calculated. The variables introduced to define the long-term performance in this model were the long-term analytical CV (LCVa) and the analytical critical difference (ACD), which indicates the minimum difference necessary between two samples measured on a long-term time-scale to consider them statistically significantly different.

Results: The systematic error (bias) ranged from 4.5 to 103 units/L. The random error ranged from 24.4 to 242 units/L. For the majority of the laboratories, random error was the main component (>75%) of the total error. The LCVa, after adjustment for the contribution of the bias, ranged from 2.8% to 48%. The ACD ranged from 78 to 1290 units/L with a median value of 190 units/L. No statistically significant differences were observed for either LCVa or ACD between the two different measurement principles for antithrombin activity based on the inhibition of either thrombin or factor Xa.

Conclusions: This linear regression model is useful for assessing the total error, random error, and bias for hemostasis field methods. The LCVa and ACD for measurement on a long-term time-scale appear to be useful for assessing the long-term analytical performance.







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS