Background: Inappropriate quality management of reverse transcription-PCR (RT-PCR) assays for the detection of blood-borne prostate cancer (PCa) cells hampers clinical conclusions. Improvement of the RT-PCR methodology for prostate-specific antigen (PSA) mRNA should focus on an appropriate numeric definition of the performance of the assay and correction for PSA mRNA that is not associated with PCa cells.
Methods and Results: Repeated (RT-)PCR tests for PSA mRNA in single blood specimens from PCa patients and PCa-free controls, performed by four international institutions, showed a large percentage (
50%) of divergent test results. The best estimates of the mean,
(SD), of the expected Poisson frequency distributions of the number of positive tests among five replicate assays of samples from PCa-free individuals were 1.0 (0.2) for 2 x 35 PCR cycles and 0.2 (0.1) for 2 x 25 PCR cycles. Assessment of the numeric value of the mean can be considered as a new indicator of the performance of a RT-PCR assay for PSA mRNA under clinical conditions. Moreover, it determines the required number of positive test repetitions to differentiate between true and false positives for circulating prostate cells. At a predefined diagnostic specificity of
98%, repeated PCRs with
of either 1.0 or 0.2 require, respectively, more than three or more than one positive tests to support the conclusion that PSA mRNA-containing cells are present.
Conclusions: Repeated nested PCR tests for PSA and appropriate handling of the data allow numeric quantification of the performance of the assay and differentiation between analytical false and true positives at a predefined accuracy. This new approach may contribute to introduction of PSA RT-PCR assays in clinical practice.
The following articles in journals at HighWire Press have cited this article:
![]() |
G. Giribaldi, S. Procida, D. Ulliers, F. Mannu, R. Volpatto, G. Mandili, L. Fanchini, O. Bertetto, G. Fronda, L. Simula, et al. Specific Detection of Cytokeratin 20-Positive Cells in Blood of Colorectal and Breast Cancer Patients by a High Sensitivity Real-Time Reverse Transcriptase-Polymerase Chain Reaction Method J. Mol. Diagn., February 1, 2006; 8(1): 105 - 112. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. Ringel Molecular Detection of Thyroid Cancer: Differentiating "Signal" and "Noise" in Clinical Assays J. Clin. Endocrinol. Metab., January 1, 2004; 89(1): 29 - 32. [Full Text] [PDF] |
||||
![]() |
I. H.N. Wong The Sporadic Nature of Shedding Cells in Blood: Multiple RNA Diagnostic Testing and Prognostication of Cancer Progression Clin. Chem., September 1, 2003; 49(9): 1429 - 1431. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |