Background: Sialic acid storage diseases (SSDs) are severe autosomal recessive neurodegenerative disorders caused by a transport defect across the lysosomal membrane, which leads to accumulation of sialic acid in tissues, fibroblasts, and urine. Defective free sialic acid transport can be established by quantification of free sialic acid in urine.
Methods: Urine sample size was adjusted to the equivalent of 100 nmol of creatinine. After addition of 2-keto-3-deoxy-D-glycero-D-galactonononic acid as internal standard, samples were diluted with water to an end volume of 250 µL. We used 10 µL for HPLCtandem mass spectrometric analysis in the negative electrospray ionization mode, monitoring transitions m/z 308.3
m/z 86.9 (sialic acid) and m/z 267.2
m/z 86.9 (internal standard). The overall method was validated and studied for ion suppression, interfering compounds, and pH effects. Samples from controls (n = 72) and SSD patients (n = 3) were analyzed.
Results: The limit of detection was 3 µmol/L. Intraassay imprecision (CV; n = 10) was 6%, 3%, and 2% at 30, 130, and 1000 mmol/mol creatinine, respectively; corresponding interassay CV (n = 10) were 5%, 5%, and 2%. Recovery was 109% (1001000 mmol/mol creatinine). The mean (SD) [range] excretion rates (mmol/mol creatinine) were 31.3 (16.6) [0.756.9] at 01 year (n = 20), 21.2 (9.8) [6.338.3] at 13 years (n = 15), 14.4 (8.2) [1.732.9] at 310 years (n = 25), and 4.6 (2.6) [09.8] above age 10 years (n = 12). SSD patients 1.2, 3.9, and 12 years of age had concentrations of 111.5, 54.2, and 36.1 mmol/mol creatinine, respectively.
Conclusions: The HPLC-tandem MS method for free sialic acid in urine is more rapid, accurate, sensitive, selective, and robust than earlier methods and may serve as a candidate reference method for free sialic acid in diagnosis of SSD.
The following articles in journals at HighWire Press have cited this article:
![]() |
E. Roura, C. Andres-Lacueva, R. Estruch, and R. M. Lamuela-Raventos Total Polyphenol Intake Estimated by a Modified Folin-Ciocalteu Assay of Urine Clin. Chem., April 1, 2006; 52(4): 749 - 752. [Abstract] [Full Text] [PDF] |
||||
![]() |
R Froissart, D Cheillan, R Bouvier, S Tourret, V Bonnet, M Piraud, and I Maire Clinical, morphological, and molecular aspects of sialic acid storage disease manifesting in utero J. Med. Genet., November 1, 2005; 42(11): 829 - 836. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Ma, F. M. Vaz, Z. Gu, R. J. A. Wanders, and M. L. Greenberg The Human TAZ Gene Complements Mitochondrial Dysfunction in the Yeast taz1{Delta} Mutant: IMPLICATIONS FOR BARTH SYNDROME J. Biol. Chem., October 22, 2004; 279(43): 44394 - 44399. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |