Brief Communications |
1 Division of Molecular Medicine, Wadsworth Center for Laboratories and Research, New York State Department of Health, Albany, NY; 2 Department of Biomedical Sciences, School of Public Health, University at Albany, Albany, NY;
aaddress correspondence to this author at: Division of Molecular Medicine, Wadsworth Center for Laboratories and Research, New York State Department of Health, Albany, NY 12201-0509. Fax 518-474-9185; e-mail tim{at}wadsworth.org.
Abstract
Background: Human chorionic gonadotropin (hCG) is a heterodimeric glycoprotein hormone that exists in multiple forms. Immunoassays commonly used in clinical laboratories measure intact hCG, total β hCG (intact hCG + hCG free β-subunit), and/or hCG free β-subunit. Measurement of serum concentrations of hCG is useful for confirmation and monitoring of pregnancy, diagnosis of trophoblastic diseases and monitoring of the efficacy of treatment, and prenatal screening. Correctly reporting results for the various forms of hCG is clinically important.
Method: We prepared samples by addition of intact hCG and hCG free β-subunit to an essentially hCG-free human serum matrix. The samples were analyzed by participant laboratories using various immunoassay methods.
Results: We identified errors in participant reporting of intact hCG results as total β hCG (9.3%; 22 of 235 laboratories) and total β hCG as intact hCG (13.1%; 8 of 61 laboratories).
Conclusions: Many factors contribute to the erroneous reporting of hCG results, including (a) the complexity of hCG molecule and confusion of nomenclature on the various forms of hCG; (b) laboratory personnels lack of awareness of the distinctions of the forms of hCG and failure to recognize the specificity of assays for their measurement; (c) lack of clarity and uniformity in manufacturers reagent labeling; and (d) most product inserts lack of information on the specificity of each method to the various forms of hCG.
The following articles in journals at HighWire Press have cited this article:
![]() |
C. M. Sturgeon, P. Berger, J.-M. Bidart, S. Birken, C. Burns, R. J. Norman, U.-H. Stenman, and on behalf of the IFCC Working Group on hCG Differences in Recognition of the 1st WHO International Reference Reagents for hCG-Related Isoforms by Diagnostic Immunoassays for Human Chorionic Gonadotropin Clin. Chem., August 1, 2009; 55(8): 1484 - 1491. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gronowski and D. G. Grenache Characterization of the hCG Variants Recognized by Different hCG Immunoassays: An Important Step Toward Standardization of hCG Measurements Clin. Chem., August 1, 2009; 55(8): 1447 - 1449. [Full Text] [PDF] |
||||