Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 54: 1076-1079, 2008; 10.1373/clinchem.2007.101568
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Supplemental Data
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shih, J.
Right arrow Articles by Möckel, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shih, J.
Right arrow Articles by Möckel, M.
(Clinical Chemistry. 2008;54:1076-1079.)
© 2008 American Association for Clinical Chemistry, Inc.


Brief Communications

Effect of Collection Tube Type and Preanalytical Handling on Myeloperoxidase Concentrations

Jessie Shih1,a, Saul A. Datwyler1, Stephen C. Hsu1, Matthew S. Matias1, David P. Pacenti1, Christian Lueders2, Christian Mueller3, Oliver Danne4 and Martin Möckel2

1 Abbott Diagnostics, Abbott Park, IL; Departments of 2 Cardiology, 3 Clinical Chemistry, and 4 Nephrology, Charité-Universitätsmedizin, Campus Virchow-Clinikum, Berlin, Germany

aaddress correspondence to this author at: Abbott Laboratories, D-9L5 AP6C, 100 Abbott Park Road, Abbott Park, IL, 60064. Fax 847-937-5350; e-mail Jessie.Shih{at}abbott.com


Abstract

Background: Myeloperoxidase (MPO) has shown potential as a marker for cardiovascular disease. Limited studies have been published with a variety of sample types, resulting in a wide range of MPO values. Little is known or understood about the impact of collection tube type and preanalytical handling of specimens for MPO determination.

Method: MPO concentration was determined by use of the ARCHITECT® MPO research use assay, which is currently under development. Samples were collected into multiple anticoagulant collection tubes from donors and patients presenting to the emergency department with symptoms of acute coronary syndromes. Whole blood was stored on ice or at room temperature for predetermined time periods. We also evaluated serum and plasma after centrifugation followed by storage at room temperature, 2–8 °C, and below –10 °C.

Results: Baseline sample concentrations were dependent on collection tube type as well as handling conditions. MPO concentrations were consistently higher in samples collected in serum and heparin plasma tubes than in samples in EDTA or citrate tubes. Spike recovery was acceptable in all sera and plasma tested, indicating that the increased MPO concentrations were not due directly to an anticoagulant interference.

Conclusions: The collection tube type and preanalytical handling are critical for accurate and consistent MPO measurement. The preferred anticoagulant and tubes are the EDTA or EDTA plasma preparation tube. MPO concentrations in samples collected in these tubes are stable before centrifugation as whole blood as well as plasma after processing.




The following articles in journals at HighWire Press have cited this article:


Home page
J Am Coll Cardiol ImgHome page
W.H. W. Tang
Enhancing the Prognostic Value of Cardiac Imaging With Multimodal Risk Assessment
J. Am. Coll. Cardiol. Img., September 1, 2009; 2(9): 1100 - 1102.
[Full Text] [PDF]


Home page
Clin. Chem.Home page
R. K. Schindhelm, L. P. van der Zwan, T. Teerlink, and P. G. Scheffer
Myeloperoxidase: A Useful Biomarker for Cardiovascular Disease Risk Stratification?
Clin. Chem., August 1, 2009; 55(8): 1462 - 1470.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
A. H.B. Wu
Novel Biomarkers of Cardiovascular Disease: Myeloperoxidase for Acute and/or Chronic Heart Failure?
Clin. Chem., January 1, 2009; 55(1): 12 - 14.
[Full Text] [PDF]