| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Point-of-Care Testing |
1 Department of Clinical Chemistry, Isala klinieken, Zwolle, The Netherlands; 2 European Reference Laboratory for Glycohemoglobin, Zwolle, The Netherlands.
aAddress correspondence to this author at: Isala klinieken, Department of Clinical Chemistry, Groot Wezenland 20, 8011 JW Zwolle, The Netherlands. Fax +31-38-4242-676; e-mail w.b.lenters{at}isala.nl.
Background: Hemoglobin A1c (Hb A1c) point-of-care (POC) instruments are widely used to provide rapid-turnaround results in diabetic care centers. We investigated the conformance of various Hb A1c POC instruments (In2it from Bio-Rad, DCA Vantage from Siemens, Afinion and Nycocard from Axis-Shield, Clover from Infopia, InnovaStar from DiaSys, A1CNow from Bayer, and Quo-Test from Quotient Diagnostics) with generally accepted performance criteria for Hb A1c.
Methods: The CLSI protocols EP-10, EP-5, and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using 3 certified secondary reference measurement procedures and the mean of the 3 reference methods. Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria, as calculated from analyses with 2 different reagent lot numbers for each Hb A1c method, was also evaluated.
Results: Because of disappointing EP-10 results, 2 of the 8 manufacturers decided not to continue the evaluation. The total CVs from EP-5 evaluations for the different instruments with a low and high Hb A1c value were: In2it 4.9% and 3.3%, DCA Vantage 1.8% and 3.7%, Clover 4.0% and 3.5%, InnovaStar 3.2% and 3.9%, Nycocard 4.8% and 5.2%, and Afinion 2.4% and 1.8%. Only the Afinion and the DCA Vantage passed the NGSP criteria with 2 different reagent lot numbers.
Conclusions: Only the Afinion and the DCA Vantage met the acceptance criteria of having a total CV <3% in the clinically relevant range. The EP-9 results and the calculations of the NGSP certification showed significant differences in analytical performance between different reagent lot numbers for all Hb A1c POC instruments.
The following articles in journals at HighWire Press have cited this article:
![]() |
B. Irvin, J. Knaebel, and D. Simmons Assessing the Performance of Point-of-Care Hemoglobin A1c Systems Clin. Chem., August 1, 2010; 56(8): 1359 - 1360. [Full Text] [PDF] |
||||
![]() |
E. Lenters-Westra and R. J. Slingerland In Reply Clin. Chem., August 1, 2010; 56(8): 1360 - 1361. [Full Text] [PDF] |
||||
![]() |
E. Lenters-Westra and R. J. Slingerland Evaluation of the Quo-Test Hemoglobin A1c Point-of-Care Instrument: Second Chance Clin. Chem., July 1, 2010; 56(7): 1191 - 1193. [Full Text] [PDF] |
||||
![]() |
D. E. Bruns and J. C. Boyd Few Point-of-Care Hemoglobin A1c Assay Methods Meet Clinical Needs Clin. Chem., January 1, 2010; 56(1): 4 - 6. [Full Text] [PDF] |
||||