Clinical Chemistry, Vol 6, 345-351, Copyright © 1960 by the American Association for Clinical Chemistry
1 Medical College of Virginia, Richmond, Va.
A method is described for arriving at a calibration equation expressing concentration in terms of absorbancy. This equation is based on the assumption of a constant stray radiant energy or a constant dark current compensation error. The deviation from linearity was found to correspond to 0.1 per cent stray radiant energy at 273 mµ in a Beckman spectrophotometer.