Technical Briefs |
1
Lab. Biochim. Clin. Ematol., Ospedale Niguarda Ca'Granda, I20162 Milano, Italy;
2
Electronics Design Center, Case Western Reserve Univ., Cleveland, OH 44106-7200;
a author for correspondence: fax 39-2-64442901, e-mail marta.melotti@galactica.it
In blood, free ("ionized") magnesium (Mg2) is in equilibrium with complexed (protein-bound, and organic or inorganic complexed) species. This equilibrium is influenced by pH and by protein and ligand concentration, both in vivo and in vitro. The phenomenon is well known for calcium; pH influences the serum concentration of free calcium ion (S-cCa2) through the competition between hydrogen ions and calcium ions toward binding sites of protein. Fogh-Andersen (1) proposed the relationship dpcCa2/dpH, correcting the measured S-cCa2 to the standard pH of 7.4, that, although controversial, is widely used in commercial analyzers (2)(3).
Recently, synthetic neutral carriers for the determination of free magnesium ion concentration (cMg2) have been developed (4)(5)(6)(7), the selectivity of which toward calcium allows the chemometric correction of the free magnesium measurement in the presence of pathophysiological calcium concentrations (8)(9). Magnesium bound to protein represents 3035% of the total magnesium ion concentration in serum (6)(7)(8)(10), and, as with calcium, an empirical correction of measured serum free magnesium ion concentration (S-cMg2) to a standard pH can be proposed. The slope of the regression log S-cMg2 vs pH is ~-0.1 (6)(11), half that for S-cCa2. In the S-cMg2 range of 0.310.76 mmol/L, we found an average value of -0.117 (8), which was reproducible during the life-span of the tested electrodes.
By contrast, Ising et al. (12), evaluating the performance
of the Kone Microlyte Magnesium assay (magnesium ionophore ETH 5220),
found a dependence of the dpcMg2/dpH
relation on the electrode life-time. According to these authors, the
built-in value of -0.07 was incorrect, and
Acknowledgments
Footnotes
References
The following articles in journals at HighWire Press have cited this article:
![]() |
K. Dewitte, D. Stockl, and L. M. Thienpont Reliability of Measurement of Ionized Magnesium in Ultrafiltrate Clin. Chem., April 1, 1999; 45(4): 588 - 589. [Full Text] [PDF] |
||||
![]() |
K. Dewitte, D. Stockl, and L. M. Thienpont Reliability of Measurement of Ionized Magnesium in Ultrafiltrate Clin. Chem., January 1, 1999; 45(1): 157 - 158. [Full Text] [PDF] |
||||