Clinical Chemistry 43: 2164-2168, 1997;
(Clinical Chemistry. 1997;43:2164-2168.)
© 1997 American Association for Clinical Chemistry, Inc.
Comparison of NCEP performance specifications for triglycerides, HDL-, and LDL-cholesterol with operating specifications based on NCEP clinical and analytical goals
Patricia C. Fallest-Strobl1,
Elin Olafsdottir2,
Donald A. Wiebe1 and
James O. Westgard1,a
1
Department of Pathology and Laboratory Medicine, University of Wisconsin Medical School, Madison, WI 53792.
2
Department of Clinical Biochemistry, University
Hospital, Reykjavik, Iceland.
3
4
Nonstandard abbreviations: NCEP, National
Cholesterol Education Program; OPSpecs, operating specifications for
allowable imprecision (CV), allowable inaccuracy (bias), and necessary
QC; TEa, total allowable analytical error;
Dint, clinical decision interval; swsub,
within-subject biological variation; TRIG, triglycerides; LDL-C,
low-density lipoprotein cholesterol; HDL-C, high-density lipoprotein
cholesterol
a Author for correspondence. Fax (608)263-1568; e-mail jo.westgard{at}hosp.wisc.edu
 |
Abstract
|
|---|
The National Cholesterol Education Program (NCEP) performance
specifications for methods that measure triglycerides, HDL-cholesterol,
and LDL-cholesterol have been evaluated by deriving operating
specifications from the NCEP analytical total error requirements and
the clinical requirements for interpretation of the tests. We
determined the maximum imprecision and inaccuracy that would be
allowable to control routine methods with commonly used single and
multirule quality-control procedures having 2 and 4 control
measurements per run, and then compared these estimates with the NCEP
guidelines. The NCEP imprecision specifications meet the operating
imprecision necessary to assure meeting the NCEP clinical quality
requirements for triglycerides and HDL-cholesterol but not for
LDL-cholesterol. More importantly, the NCEP imprecision specifications
are not adequate to assure meeting the NCEP analytical total error
requirements for any of these three tests. Our findings indicate that
the NCEP recommendations fail to adequately consider the
quality-control requirements necessary to detect medically important
systematic errors.
 |
Introduction
|
|---|
Three different types of performance criteria have been
established for lipid tests performed in healthcare laboratories. For
example, cholesterol tests are regulated by the Clinical Laboratory
Improvement Act (CLIA-88), which requires laboratories to provide test
values that are correct within 10% of target values for proficiency
testing materials (a total error criterion). The US National
Cholesterol Education Program
(NCEP) provides physicians with explicit guidelines to interpret a
patient's cholesterol values (a clinical criterion) and also provides
laboratories with specifications for the imprecision and inaccuracy of
their routine cholesterol methods (analytical criteria). Guidelines
published by NCEP recommended that by 1992 the performance of
cholesterol tests should have a 3.0% or less CV (or imprecision) and
3.0% or less bias (inaccuracy) (1). We evaluated these
recommendations in earlier studies (2)(3)(4) and expressed
concern that QC issues had not been adequately considered, in that a CV
of 2.4% or less was actually required to assure that routine methods
would achieve the performance required for the clinical use of the
test.
Because cholesterol is a key risk factor for cardiovascular disease,
the importance for establishing both clinical and analytical goals
cannot be argued. However, these goals are interdependent (i.e., a
method's analytical performance will often determine the clinical
usefulness of a given test and, likewise, clinical demands often
dictate the stringency required of a method's analytical performance).
The two groups that established these guidelines, the Adult Treatment
Panel and the Laboratory Standardization Panel, failed to interconnect
the different performance criteria and address them together.
Nonetheless, the primary goals of NCEP have been overwhelmingly
achieved. Family physicians appropriately treat hypercholesterolemic
patients with aggressive therapy to lower cholesterol to lessen their
cardiovascular risk. Laboratories have improved their performance and
are providing more-accurate cholesterol test values. The laboratory
success was driven in part by manufacturers that provided improved
instruments and techniques designed to meet the NCEP guidelines.
Cholesterol, however, is only one of several lipid components of
clinical importance for which appropriate guidelines are needed. In
1995, the NCEP Laboratory Standardization Panel published
recommendations of performance specifications for triglycerides (TRIG),
HDL-cholesterol (HDL-C), and LDL-cholesterol (LDL-C)
(5)(6)(7). These recommendations define specifications for
the imprecision and inaccuracy of the respective methods, the total
analytical error that is allowable, and the medical decision values.
However, lack of guidance for the routine operation and QC of these
methods prompted us to evaluate the operating specifications necessary
to assure that the NCEP clinical and analytical goals could be achieved
in routine lipid testing, taking into account the QC performance that
would be necessary to detect analytical runs in which method
performance is unstable. Detailed specifications for QC are essential
for these methods, which are prone to many analytical variables that
could affect routine laboratory performance and result in inappropriate
patient data transmitted to the patient.
 |
Materials and Methods
|
|---|
OPSpecsTM charts (8) were prepared by
using the QC ValidatorTM computer program (9)
(Westgard Quality Corp., Ogunquit, ME). The principles of the OPSpecs
methodology have been described in detail elsewhere (10).
OPSpecs charts plot the allowable inaccuracy on the y-axis
vs the allowable imprecision on the x-axis for a defined
quality requirement and a specified requirement of error detection, or
analytical quality assurance. The observed imprecision and inaccuracy
of an individual method can be plotted to locate the operating point of
the method. Appropriate QC procedures are provided by the control rules
and number of control measurements per run (N), the operating limits of
which are above the method's observed operating point
(11). Estimates of maximum allowable imprecision can be
obtained from the x-intercepts of the operating limits
(12).
In this study, OPSpecs charts were prepared to show the imprecision and
inaccuracy that are allowable for common QC procedures with N = 2
to 4, 90% analytical quality assurance, and the stated NCEP analytical
total error requirements and the clinical decision interval
(Dint) requirements implied by the guidelines for test
interpretation (5)(6)(7). We calculated Dint for
each of the tests according to the medical decision values for
treatment of patients as recommended by the Adult Treatment Panel. For
TRIG, concentrations for healthy subjects are <2000 mg/L and high
concentrations are
4000 mg/L; therefore, the Dint is
100%: (2000/2000) x 100%. For HDL-C, desirable concentrations are
>600 mg/L and major risk concentrations are <350 mg/L; therefore,
Dint is 42%: (250/600) x 100%. For LDL-C, desirable
values are <1300 mg/L and high-risk concentrations are
1600 mg/L;
therefore, Dint = 23.1%: (300/1300) x 100%. Estimates of
the within-subject biological variation were taken as 23.7% for TRIG
(6), 7.5% for HDL-C (7), and 8.2% for LDL-C
(5), as recommended in the NCEP documents. The range of
allowable imprecision (operating imprecision in Table 1
) was determined at the bias stated for each test.
View this table:
[in this window]
[in a new window]
|
Table 1. Comparison of NCEP clinical and analytical requirements
with operating imprecision derived from OPSpecs charts.
|
|
 |
Results
|
|---|
Clinical quality requirements.
The OPSpecs charts for
the NCEP clinical quality requirements for TRIG, HDL-C, and LDL-C are
shown in Fig. 1
. For TRIG, the operating point that represents the
NCEP-recommended specifications (inaccuracy of 5%, imprecision of 5%)
is well below the allowable range of imprecision (14.022.5%) at the
stated value of 5% for inaccuracy, as denoted by the double-headed
arrow in Fig. 1A
. Therefore, the TRIG specifications for imprecision
and inaccuracy will assure the quality defined by the NCEP clinical
requirement (Dint = 100%; swsub = 23.7%). The
operating point for HDL-C falls within the allowable range of
imprecision (5.07.8%) for the control rules with N = 24 (Fig. 1B
); thus, the HDL-C specifications for accuracy (10.0%) and precision
(6.0%) will also assure the quality defined by the NCEP clinical
requirement (Dint = 42.0%, swsub = 7.5%).
However, the LDL-C operating point of 4.0% imprecision and 4.0%
inaccuracy exceeds the allowable range of imprecision (1.52.6%), as
shown in Fig. 1C
. Therefore, the NCEP method performance specifications
for LDL-C will not assure the quality defined by the NCEP clinical
requirement (Dint = 23.1%, swsub = 8.2%).

View larger version (22K):
[in this window]
[in a new window]
|
Figure 1. OPSpecs chart for clinical quality requirement for
(A) triglycerides (Dint = 100%,
swsub = 23.7%), (B) HDL-cholesterol
(Dint = 42.0%, swsub = 7.5%), and
(C) LDL-cholesterol (Dint = 23.1%,
swsub = 8.2%).
Curves (top to bottom in each panel) represent
control procedures for the following rules: 12 s with
N = 4,
1 3 s/22 s/R4 s/41 s
with N = 4, 12.5 s with N = 4, 12 s
with N = 2, 13 s with N = 4, 12.5 s
with N = 2, 13 s/22 s/R4 s
with N = 2, and 13 s with N = 2). , range of
imprecision at the bias allowed by NCEP.
|
|
Analytical quality requirements.
The operating point for
TRIG (inaccuracy 5%, imprecision 5%) exceeds the allowable range of
imprecision (2.23.2%) for analytical quality requirements
(TEa = 15.0%), as depicted by the OPSpecs chart in Fig. 2
A. Therefore, common QC procedures with N = 24 will not
be able to detect unstable performance. Similarly, the OPSpecs charts
for HDL-C and LDL-C (Fig. 2B
and C) show operating points that also
exceed the allowable range of imprecision for detection of unstable
performance. Therefore, the analytical quality requirements for
precision of test methods for TRIG, HDL-C, and LDL-C will not assure
the analytical quality defined by NCEP.

View larger version (23K):
[in this window]
[in a new window]
|
Figure 2. OPSpecs chart for analytical quality
requirements for (A) triglycerides (TEa =
15.0%), (B) HDL-cholesterol (TEa = 22.0%), and
(C) LDL-cholesterol (TEa = 12.0%).
See Fig. 1
for QC rules.
|
|
NCEP's recommendations for HDL-C (7) also included
performance specifications (TEa = 13%, CV = 4.0%,
bias = 5.0%) to be achieved by 1998. Although these
specifications are more stringent than the current criteria for HDL-C,
analytical quality requirements are still not satisfied. OPSpecs charts
prepared for use of common QC procedures (not shown) indicate that the
allowable imprecision for HDL-C methods still needs to be <4.0%
(within the range 1.52.3%) at a bias of 5.0% and TEa =
13.0%.
Table 1
summarizes the ranges of allowable imprecision derived from
OPSpecs charts, based on the NCEP clinical and analytical quality
requirements for all of these tests. Also shown are the allowable
ranges of imprecision for each test if bias were eliminated as a
consideration (bias = 0). In this scenario, the allowable ranges
of imprecision are increased for all of the tests, but NCEP analytical
quality requirements are still not assured for TRIG or LDL-C (Table 1
).
 |
Discussion
|
|---|
Methods for determination of TRIG, HDL-C, and LDL-C are complex
and can be problematic, which means it is important to select optimal
QC procedures capable of detecting unstable performance. TRIG
measurements are often automated, whereas HDL-C and LDL-C methods
usually include manual steps in test protocols. LDL-C test results may
be the most problematic because common practice is to report estimated
LDL-C by calculation with the Friedewald equation [estimated LDL
= total cholesterol - HDL - (TRIG/5)]. Therefore, each LDL-C
result relies on accurate and precise measurements of three other
lipids: cholesterol, HDL-C, and TRIG. The ability to detect unstable
performance in routine laboratory testing for lipids through use of
statistical QC procedures is crucial to avoid further propagation of
error in generating LDL-C test results.
Each laboratory should pay special attention to the operating
specifications needed to assure that the desired analytical or clinical
quality is achieved in routine service. OPSpecs charts provide a
graphical display of the imprecision and inaccuracy that are allowable
when different QC procedures are used. In this approach, statistical QC
is factored in as a component of the error budget for a test, thus
providing a more realistic assessment of the imprecision and inaccuracy
limits that are necessary if the QC procedure is to detect medically
important errors. OPSpecs charts, therefore, are a very useful tool for
defining the specifications needed for lipid tests in a routine service
laboratory.
The NCEP analytical recommendations for imprecision and inaccuracy seem
to be based on the method acceptance criterion of TEa
= bias + 2 s. For methods that just meet this criterion, commonly
used control procedures are not expected to provide reliable detection
of medically important errors (13). Even if bias could be
eliminated from lipid test methods, the derived data for operating
precision (when bias = 0) show that NCEP-recommended precision
goals for TRIG and LDL-C are not satisfactory to meet the stated
analytical total error requirements. Laboratories should select methods
for these tests for which operating imprecision is within the ranges
indicated in Table 1
to assure adequate detection of unstable
performance. In general, further reducing by a factor of 2 the limits
set by NCEP for allowable imprecision would be desirable to provide
more-controllable testing processes.
In conclusion, improvements in lipid test performance and
interpretation have continued over the years because of efforts by
NCEP, but our findings indicate that additional guidelines are needed
for statistical QC to assure detection of unstable performance.
Although NCEP's 1998 goals for HDL-C are a step in a right direction,
they will still fail to provide a high degree of analytical quality
assurance. In the future, development of performance specifications for
analytical methods should take into account the performance
characteristics of the statistical QC procedures that are expected to
be used in monitoring routine laboratory performance. The NCEP Adult
Treatment Panel and Laboratory Standardization Panel independently
generated clinical and analytical goals that have conflicting or
unobtainable targets. Clearly, laboratorians will need guidance to
establish acceptable performance standards for these routine, yet
demanding lipid tests. However, if performance goals result in
unrealistic objectives, the final outcome could be misuse or
inappropriate handling of proficiency testing material to ensure that
the analytical goals are met, while patients' needs are ignored. These
activities must be coordinated with clearer QC operations for
laboratories to properly manage these tests.
 |
Acknowledgments
|
|---|
The Department of Pathology and Laboratory Medicine, University of
Wisconsin Medical School, provided partial support for E.O. during a
sabbatical period and full support for P.F.-S. during a Clinical
Chemistry Fellowship.
 |
References
|
|---|
-
. National Cholesterol Education Program Laboratory Standardization Panel. Current status of blood cholesterol measurement in clinical laboratories in the United States. Clin Chem 1988;34:193-201.
[Free Full Text]
-
Westgard JO, Hyltoft Petersen P, Wiebe DA. Laboratory process specifications for assuring quality in the US National Cholesterol Education Program. Clin Chem 1991;37:656-661.
[Abstract/Free Full Text]
-
Westgard JO, Wiebe DA. Cholesterol operational process specifications for assuring quality required by CLIA proficiency testing. Clin Chem 1991;37:1938-1944.
[Abstract/Free Full Text]
-
Wiebe DA, Westgard JO. Cholesterola model system to relate medical needs with analytical performance. Clin Chem 1993;39:1504-1513.
[Abstract]
-
. for the NCEP Working Group on
Lipoprotein MeasurementBachorik PS, Ross JW. National Cholesterol Education Program recommendations for measurement of low-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1414-1420.
[Free Full Text]
-
. for the NCEP Working Group on
Lipoprotein MeasurementStein EA, Myers GL. National Cholesterol Education Program recommendations for triglyceride measurement: executive summary. Clin Chem 1995;41:1421-1426.
[Free Full Text]
-
. for the NCEP Working Group on
Lipoprotein MeasurementWarnick GR, Wood PD. National Cholesterol Education Program recommendations for measurement of high-density lipoprotein cholesterol: executive summary. Clin Chem 1995;41:1427-1433.
[Free Full Text]
-
Westgard JO. Charts of operational process specifications ("OPSpecs charts") for assessing the precision, accuracy, and quality control needed to satisfy proficiency testing criteria. Clin Chem 1992;38:1226-1233.
[Abstract/Free Full Text]
-
Westgard JO, Stein B. Automated selection of statistical quality-control procedures to assure meeting clinical or analytical quality requirements [Tech Brief]. Clin Chem 1997;43:400-403.
[Free Full Text]
-
Westgard JO. Error budgets for quality management: practical tools for planning and assuring the analytical quality of laboratory testing processes. Clin Lab Manage Rev 1996;10:377-403.
[Medline]
[Order article via Infotrieve]
-
Mugan K, Carlson IH, Westgard JO. Planning QC procedures for immunoassays. J Clin Immunoassay 1994;17:216-222.
-
Westgard JO, Seehafer JJ, Barry PL. European specifications for imprecision and inaccuracy compared with operating specifications that assure the quality required by US CLIA proficiency-testing criteria. Clin Chem 1994;40:1228-1232.
[Abstract/Free Full Text]
-
Westgard JO, Burnett RW. Precision requirements for cost-effective operation of analytical processes. Clin Chem 1990;35:1629-1632.
The following articles in journals at HighWire Press have cited this article:

|
 |

|
 |
 
E. J Parks
A targeted goal for energy-restricted diets in the management of coronary risk?
Am. J. Clinical Nutrition,
February 1, 2001;
73(2):
147 - 148.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. P. Caudill, G. R. Cooper, S. J. Smith, and G. L. Myers
Assessment of current National Cholesterol Education Program guidelines for total cholesterol, triglyceride, HDL-cholesterol, and LDL-cholesterol measurements
Clin. Chem.,
August 1, 1998;
44(8):
1650 - 1658.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. P. Caudill, S. J. Smith, G. R. Cooper, G. L. Myers, J. O. Westgard, and D. A. Wiebe
Adequacy of NCEP Recommendations for Total Cholesterol, Triglycerides, HDLC, and LDLC Measurements Two of the authors of the above- mentioned article reply:
Clin. Chem.,
May 1, 1998;
44(5):
1063 - 1066.
[Full Text]
[PDF]
|
 |
|