|
|
||||||||
Articles |
a Author for correspondence. Fax 81-6-879-6635.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Among the several methods for measurement of serum iron, the direct colorimetric methods, which do not involve sample deproteinization, are widely used and have been modified as a fully automated assay for routine analysis (2)(3). On the other hand, the measurement of TIBC consists of three steps: saturation of transferrin by addition of an excess amount of iron, removal of unbound iron by absorption with magnesium carbonate (4)(5)(6) or ion-exchange resin (7), and finally determination of iron that is dissociated from transferrin at acidic pH.
However, the step for removal of unbound iron, requiring centrifugation, is an impediment to adaptation of the TIBC assay to a fully automated analysis. Moreover, it is difficult to generate a calibration factor necessary for the assay. When iron solution without binding protein is used as a calibrator, the unbound iron is removed in the elimination step, and therefore the reaction to obtain a calibration factor cannot be monitored in the next step. The ideal calibrator needs to bind a known amount of added iron at alkaline pH and dissociate completely at acidic pH, as serum transferrin does. To overcome this problem, we used as a calibrator a chelating agent that has a large stability constant with iron at an alkaline condition and succeeded in establishing a fully automated method for measurement of TIBC.
| Materials and Methods |
|---|
|
|
|---|
preparation of reagents
The first reagent (R1) was made up with 300 mmol/L Tris, 150
mmol/L sodium hydrogen carbonate, and 4.2 g/L Triton X-100 (pH 8.4).
The second reagent (R2) was made up with iron calibrator solution (180
µmol/L Fe3+ in 10 mmol/L HCl), which was diluted
10-fold with R1. The third reagent (R3) was made up with 10 mmol/L
ferrozine and 40 mmol/L L-ascorbic acid in Tris buffer (50
mmol/L, pH 5.0). The fourth reagent (R4) was made up with 600 mmol/L
citric acid and 25.6 mmol/L thiourea, adjusted to pH 2.0 with 300
mmol/L sodium dihydrogen phosphate. Physiological saline (NaCl 150
mmol/L) and 20 mmol/L HEDTA in 0.03 mmol/L NaOH were used as zero and
high calibrators, respectively.
apparatus
The data obtained by the fully automated TIBC assay were compared
with those obtained with conventional methods by use of the Model 7070
automated analyzer from Hitachi (also known as the Hitachi 911).
assay principle
Our fully automated assay consisted of three steps of reactions.
Table 1
shows the scheme of each step in our method. We then used HEDTA
as a calibrator to generate a calibration factor. This chelating agent
has a large stability constant with Fe3+ metal ligand
(logKML = 19.8) at pH 712 (8).
|
In the first step, serum transferrin was saturated by the iron
calibrator solution (180 µmol/L). Then the unbound iron was reduced
to Fe2+ by ascorbic acid and eliminated by formation
of a complex with ferrozine (9) used as a chromogenic
reagent (second step).
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() |
assay procedure
The calibrator solution or a serum sample (50 µL) was mixed with
200 µL of R1, then incubated with 50 µL of R2, and kept at 37 °C
for 3.3 min. At 4.9 min after addition of 50 µL of R3, 100 µL of R4
was added. The absorbance change was recorded for 5.9 min at 570 nm
with reference at 660 nm.
method for comparison
We used TIBC CUPTM (ESA) for saturation of serum
transferrin (7), and the transferrin-bound iron was
determined by UnimateTM (Hoffmann-La Roche)
(9), a direct colorimetric method that has ferrozine as
the color reagent. TIBC CUP contains an ion-exchange resin that
includes bound iron. When serum is added to the TIBC CUP, transferrin
in the serum is saturated with the resin-bound iron, and all excess
iron remains bound to the resin beads. We also calculated the TIBC
value from the serum iron concentration and the unsaturated
iron-binding capacity (UIBC) value obtained by Unimate and Unimate UIBC
(Hoffmann-La Roche).
| Results |
|---|
|
|
|---|
We also examined the effect of the HEDTA concentration used as a calibrator, which behaved in the same manner as transferrin. When the HEDTA concentration was <5 mmol/L, a slight increase in absorbance was observed after addition of R3 (elimination step). When the HEDTA concentration was 20 mmol/L, there was no increase of absorbance, indicating that dissociated iron was not present. We thus used this concentration as the calibrator.
The pH of the mixture of the sample and calibrator after addition of R3
was 8.39 ± 0.03, and the pH of the mixture was 3.90 ± 0.05
at the final step. The typical time course of the calibrator and human
serum is shown in Fig. 1
. The reaction between unbound iron and ferrozine ended within 1
min, and dissociation of iron from transferrin and formation of the
ferrozineiron complex ended within 2 min after addition of R4.
|
To examine dissociation of iron from transferrin after addition of R4, we determined the serum iron concentration by our method but without saturation of transferrin and compared the results with those from the Unimate method. The correlation between values obtained with our method (y) and the Unimate method (x) was: y = 0.965x - 0.99 µmol/L (r = 0.995, Sy|x = 0.93, n = 40).
assay evaluation
To study the linearity of the calibration curve, we assayed the
calibrator in duplicate with six concentrations of R2 (0144
µmol/L). The calibration curve was straight for iron concentrations
up to 180 µmol/L. We also assayed several human sera that were
diluted with physiological saline. The dilution curve was linear for
TIBC values up to 90 µmol/L (Fig. 2
).
|
We also examined the detection limit of our method by assaying the zero calibrator (physiological saline) 10 times; the result (mean ± SD) was 0.63 ± 0.40 µmol/L. The detection limit, defined as the mean TIBC value of the zero calibrator + 3 SD, was 1.83 µmol/L.
To examine the precision of our method, we assayed pooled human sera.
As shown in Table 2
, the within-assay CVs ranged from 0.66% to 2.43% and the
day-to-day CVs were from 1.06% to 1.57%. Our method was compared with
the TIBC CUP method as well as the Unimate method.
|
As shown in Fig. 3
, the correlation between values obtained with our method
(y) and the TIBC CUP method (x) was:
y = 0.963x + 0.29 µmol/L
(r = 0.973, Sy|x = 2.83, n =
59) and with the Unimate method (x) was: y =
1.01x - 1.06 µmol/L (r = 0.994,
Sy|x = 1.66, n = 51).
|
We found no interference with TIBC ranging from 27.2 to 58.7 µmol/L
by addition of 150 mg/L bilirubin, 100 µmol/L Mn2+,
100 µmol/L Cu2+, and 2460 formazine turbidity units of
the lipemic material to pooled human sera. Table 3
shows the negative effects of hemolysis on this method. We then
tested the absorption spectrum of the overall reaction in a hemolytic
serum sample (100 mg/L hemoglobin). The reaction with unbound iron and
ferrozine showed two peaks at 542 and 576 nm. After addition of R4, the
increase of absorbance occurred in the range of 600 and 700 nm with a
peak at 630 nm. The cause of this peak is unknown. When the absorbance
at 660 nm (reference) was subtracted from the absorbance at 570 nm with
the hemolytic sample, the absorbance change was about -25%, which was
a negative error as compared with the nonhemolytic sample. This unknown
peak was not observed when physiological saline was used as the sample
and hemolysate was added.
|
We also examined the interference by ferritin iron. Serum samples containing ferritin concentrations of 200012 000 µg/L were measured by our method, and the results were compared with those obtained by the TIBC CUP method and the Unimate method. The values obtained by our method showed no positive error as compared with two other methods.
| Discussion |
|---|
|
|
|---|
A calibration factor could be generated by the use of HEDTA as a
calibrator. The final absorbance of the calibrator after addition of R4
was the same as that of the zero calibrator (Fig. 1
). This indicated
that iron formed a complex with ferrozine completely in the presence of
HEDTA.
The values obtained by our method without saturation of serum transferrin indicated that the serum iron concentrations correlated well with the values obtained with the Unimate method. There also was a good correlation between our method and the TIBC CUP method or other methods. These data suggested that the binding sites of serum transferrin were well saturated with added iron and that iron was dissociated from transferrin after addition of R4.
Recently, we reported in patients with hyperferritinemia that the serum iron concentration measured by the method proposed by the International Committee for Standardization in Haematology (12)(13) and a constant-potential coulometric method had positive errors caused by liberation of iron from circulating ferritin by deproteinization or the acidic solvent (14)(15). However, no interference from ferritin was observed with the present method. This result means that transferrin-bound iron was released specifically from transferrin in the acidic condition, even at high concentrations of ferritin.
In summary, our fully automated method does not require any absorbents for removal of unbound iron or centrifugation. Moreover, the total amount of iron bound to serum transferrin is measured as TIBC without interference from other cations, bilirubin, lipemic material, or ferritin. We thus believe that our fully automated method for serum TIBC may be suitable for routine clinical use in the laboratory.
| Footnotes |
|---|
1 Nonstandard abbreviations: TIBC, total iron-binding capacity; UIBC, unsaturated iron-binding capacity; HEDTA, N-(2-hydroxyethyl)ethylenediamine-N,N',N'-triacetic acid. ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
T. Shan, Y. Wang, Y. Wang, J. Liu, and Z. Xu Effect of dietary lactoferrin on the immune functions and serum iron level of weanling piglets J Anim Sci, September 1, 2007; 85(9): 2140 - 2146. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Yamanishi, S. Iyama, Y. Yamaguchi, Y. Kanakura, and Y. Iwatani Total Iron-binding Capacity Calculated from Serum Transferrin Concentration or Serum Iron Concentration and Unsaturated Iron-binding Capacity Clin. Chem., January 1, 2003; 49(1): 175 - 178. [Full Text] [PDF] |
||||
![]() |
H. Yamanishi, S. Iyama, Y. Yamaguchi, Y. Kanakura, and Y. Iwatani Modification of Fully Automated Total Iron-binding Capacity (TIBC) Assay in Serum and Comparison with Dimension TIBC Method Clin. Chem., September 1, 2002; 48(9): 1565 - 1570. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Siek, J. Lawlor, D. Pelczar, M. Sane, and J. Musto Direct Serum Total Iron-binding Capacity Assay Suitable for Automated Analyzers Clin. Chem., January 1, 2002; 48(1): 161 - 166. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |