The University of Virginia Health Sciences Center, Medical Automation Research Center, Box 168, Charlottesville, VA 22908.
a Author for correspondence. Fax 804-924-5718; e-mail cae4q{at}virginia.edu
| Abstract |
|---|
|
|
|---|
Key Words: indexing terms: automation centrifugation hemolysis turnaround time axial separation
| Introduction |
|---|
|
|
|---|
Our previous evaluation (2) had three significant findings: (a) axial separation of heparin-containing whole blood gave equivalent chemistry values when compared with conventional specimen separation; (b) axial separation yielded a 69% time savings at the centrifugation bench (time on the centrifugation bench is a small portion of turnaround time); (c) the ASM, because of the increased speed and efficiency of axial separation, was calculated to allow a labor savings of up to one-half a full-time laboratory employee. The ASM separates one tube of whole blood in ~1 min with minimum operator effort.
In the previous study we did not measure test result turnaround time (previously, only time spent on the centrifuge bench top was considered). We studied turnaround time at UVA HSC main clinical laboratory for specimens separated at point of care. We chose to study specimens originating from a typical outpatient center because we anticipate that the UVA HSC will serve up to 18 remote facilities in the next year. The ease of use and compact design of the ASM was well-suited for use at our North Ridge outpatient clinic (located 5 miles away from the main hospital laboratory).
Conventionally, specimens arrive at the main laboratory accessioning area and are subjected to preanalytical delays caused by bottlenecks at the centrifugation station. We hypothesized that moving blood separation to point of care would reduce turnaround time because separated specimens could bypass the centrifugation step of the preanalytical process in the main laboratory.
| Materials and Methods |
|---|
|
|
|---|
Blood drawn into an APC was separated in the ASM at the North Ridge Clinic. Blood drawn into a Vacutainer Tube was separated in a conventional centrifuge (AccuspinTM; Beckman Instruments, Palo Alto, CA) at the main laboratory. Therefore, blood in an APC arrived to the main laboratory separated, whereas a Vacutainer Tube arrived containing whole blood. Turnaround time was calculated for the "chem 17" test from files stored in our laboratory information system (LIS) (Sunquest, Tucson, AZ).
Specimen transportation to the main laboratory from the North Ridge
Clinic was performed by human messengers. Average turnaround time was
calculated for both the conventional system and the ASM/APC at the
North Ridge Clinic. The time period of interest (turnaround time, Fig. 1
) for this study started with specimen arrival at the
accessioning area of the main laboratory and ended with result
availability to the LIS.
|
A time comparison between blood specimens serially separated at point of care vs conventional batch centrifugation at the main laboratory was performed. We adopted sequential time periods of 3 weeks. The ASM was placed in the laboratory 1 week before the test period to allow technologists to become accustomed to the new technology. All data points were collected from the 3-week periods to obviate any bias from data selection. The two systems could not be operated in parallel because sorting out ASM tubes from conventional tubes after transportation would have biased the timing studies.
The turnaround times were compared by the Student's t-test, and P <0.01 was taken to be significant.
| Results |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
The technologists indicated that processing specimens serially in the ASM did not require appreciable effort or delay their daily routine, unlike conventional centrifugation. In fact, in the past the North Ridge Clinic centrifuged their own specimens (conventionally), but that responsibility was eliminated when it was found to be too laborious. Previously, the technologists often batched specimens when separating by conventional centrifugation. Those batches that were in the process of centrifugation when the messenger arrived had to wait until the next delivery time (at least 1 h later).
We observed that our present laboratory protocol (in the main laboratory, after delivery) requires medical technologists to transport routine specimens from the accessioning/centrifugation area to the analytical area approximately every 30 min. APCs and Vacutainer Tubes arrive at the accessioning/centrifugation area in the same manner. Although blood in an APC was already separated, it remained in the accessioning area queue until delivery to the analyzer. So, at our main laboratory, turnaround time could be further reduced up to 30 min if the APCs could be streamlined directly to the analyzers after accessioning is complete. The process then is analyzer time dependent and nonanalytical time is minimized.
Hemolysis causes analytical problems (3) and is a common problem with patient specimens originating from another remote facility at the UVA HSC, our Cancer Treatment Center. Hemolysis requires a redraw of a patient specimen. With conventional centrifugation, hemolysis is not discovered until preanalytical specimen processing is complete at the main laboratory. A lengthy time delay results before a patient can receive medication. Thus, the ability to immediately detect hemolysis (and possibly other problems such as lipemia and icterus) at the time of phlebotomy is desirable. This can be accomplished with point-of-care centrifugation.
The ASM processes blood specimens serially. Its design is well suited for point of care. Conventional centrifuges are much more labor intensive (2). The APC technology allows for quick visual inspection of the plasma at point of care within minutes of phlebotomy completion. The need for a redraw is determined immediately. This can potentially save a patient hours of time. For example, a Cancer Treatment Center patient cannot receive chemotherapy until the analytical results are obtained. A new study being conducted at the UVA HSC Cancer Center clinic will seek to quantify the effect of point-of-care blood separation on patient care.
Point-of-care blood separation has another potential benefit: Specimens originating from remote locations are subject to analytical problems associated with delays of specimen separation (3)(4). Therefore, blood specimens should be centrifuged as soon as possible. Electrolyte concentrations decrease over time in whole blood because of the concentration gradients that exist between cells and serum (or plasma) (4). Glucose is metabolized by leukocytes and erythrocytes in whole blood (3). The use of antiglycolytic agents is commonplace for glucose measurements. However, immediate blood separation may help eliminate analytical error due to whole-blood separation delays and may obviate the need for metabolic inhibitors. Furthermore, plasma is permanently separated from cells by a plastic separator in the APCs. We believe this to be superior to traditional gel tubes, especially during transportation in hot weather.
In conclusion, prior studies demonstrated that the serial nature of the ASM system more efficiently separates whole-blood specimens on the bench top. Our clinical evidence now demonstrates that serial point-of-care blood separation reduces specimen turnaround time at the main laboratory. The ASM/APC was found to be better suited for point-of-care blood separation than a conventional centrifuge. Furthermore, we speculate that this technology has the potential to help improve the quality of analytical results by facilitating centrifugation immediately after phlebotomy at point of care. A study to assess the effect point-of-care centrifugation has on patient care is ongoing.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
J. W. Holman, T. E. Mifflin, R. A. Felder, and L. M. Demers Evaluation of an Automated Preanalytical Robotic Workstation at Two Academic Health Centers Clin. Chem., March 1, 2002; 48(3): 540 - 548. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Graves, B. Holman, and R. A. Felder Modular Robotic Workcell for Coagulation Analysis Clin. Chem., May 1, 2000; 46(5): 772 - 777. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |