Laboratoire Claude Bernard Hôpital de Pontoise, F-95300 Pontoise, France.
2 Laboratoire de Toxicologie (EA207), Faculté
de Pharmacie 4 av. de L'Observatoire, F-75006 Paris, France.
a Author for correspondence. Fax Int + 331 30 75 53 69.
| Abstract |
|---|
|
|
|---|
Key Words: indexing terms: drug screening chromatography, thin- layer fluorescence polarization immunoassay urine gastric lavage
| Introduction |
|---|
|
|
|---|
Some authors have developed methods for toxicological screening (3)(4) based on thin-layer chromatography (TLC) separation and ultraviolet (UV) spectral detection, respectively.1 Toxi-Lab (Irvine, CA) has developed a drug-screening procedure based on liquidliquid extraction, TLC separation, and reagent-spray detection (sulfuric acid, Draggendorf reagent).
Recently, Bio-Rad (Hercules, CA) Labs. developed Remedi, an automated liquid-chromatographic analyzer for detecting drugs in urine, serum, and gastric lavage fluid by on-line sample cleanup and isocratic multicolumn separation with full-scan UV detection. Coupled with a computer memory, Remedi allows identification of 450 drugs in one analysis. Two reports have been published on the analytical evaluation of Remedi (5)(6) on urines and stomach contents.
In our study, we screened urine, serum, and gastric contents for drugs. The aim of our work was to assess the clinical utility of Remedi after 1 year of toxicological screening at the hospital of Pontoise. Although Remedi allows the determination of >450 drugs, a few drugs could not be detected, e.g., carbamates and acidic drugs (aspirin, acetaminophen, chloroquine), so two complementary methods were systematically used in conjunction with Remedi: fluorescence polarization immunoassay (FPIA) and TLC.
| Materials and Methods |
|---|
|
|
|---|
assays
FPIA
Tricyclic antidepressants, benzodiazepines, and
barbiturates were analyzed in urine and serum by FPIA with a TDx
analyzer (Abbott, Abbott Park, IL), with scrupulous application of the
manufacturer's instructions.
Remedi.
The Remedi broad-spectrum drug identification
system used a multicolumn approach to extract, purify, and analyze
drugs in patients' samples by full-scan UV detection. For sample
processing, the samples are diluted with an internal standard mixture
and centrifuged. Because amphoteric or weakly basic drugs are eluted
from the separation cartridges early, whereas basic drugs are eluted
later, two internal standards are used to monitor the chromatographic
behavior of the cartridges: N-ethyldiazepam and
chlorpheniramine. The prepared sample is then combined with a buffer
and passed through four cartridges. The purification cartridge extracts
and concentrates the drugs while allowing proteins and salt to pass
through. Once the drugs are eluted from this cartridge, a mobile phase
is introduced that sends the drugs through the extraction cartridge.
Here, endogenous organic acids are retained while weakly acidic,
neutral, and basic drugs pass through. The third cartridge, Separation
I cartridge, is a reversed-phase cartridge that separates weakly basic
compounds. The fourth cartridge, Separation II, differentiates
basic compounds by cation exchange. All separations are isocratic. A
conditioning cartridge is used to saturate the mobile phase with
silica, thereby protecting the two separation cartridges from any
dissolution.
Drug identification is performed by a full-scan UV detector coupled with a sophisticated computer algorithm. As each drug enters the detector from the last cartridge, a UV scan from 200 nm to 300 nm is made. Sample spectra are then automatically compared with the library of known drug spectra stored in memory. This, combined with the chromatographic data, allows identification of the drug. The procedure is completed in ~20 min. A very thorough analytical study of this technique has been published by Binder et al. (8).
All reagents used with the Remedi were supplied by Bio-Rad; however, the composition of the mobile phase and other reagents and the exact characteristics of the stationary phase of the cartridges were not disclosed. Sample volume was 1 mL for serum, urine, and gastric content.
Other techniques.
Alcohol (ethanol) was assayed by an
enzyme method (Sigma), acetaminophen (paracetamol) by TDx, and
salicylic acid by the Trinder reaction.
subjects
Patients (n = 354) with symptoms of poisoning
(retrospective study) were admitted to the medical emergency unit
(Pontoise Hospital) during 1994. The specimens analyzed were 184 paired
urine and blood samples, 35 urine only, and 135 blood only; we also
analyzed 117 gastric contents (not paired).
statistical method for clinical diagnosis
The statistical study was based on screening tests described by
Laplanche et al. (9) for determining the diagnostic power
of medical apparatus.
| Results |
|---|
|
|
|---|
]2 = 10.26, P <0.01).
|
To summarize the clinical assessment of the Remedi data, we used a retrospective statistical study in which the status of patients included in this study had been well established after various investigations (biology, radiology, examination, history). Of 55 cases of poisoning, Remedi identified 36 true positives (TP), 3 false positives, 4 false negatives (FN), and 12 true negatives (TN). The sensitivity or probability that a poisoned patient gave a positive result by Remedi, determined by the formula 100x TP/(TP + FN), was 90%. The specificity, or probability that a nonpoisoned patient gave a negative result by Remedi, determined by the formula 100x TN/(TN + FP), was 80%. The positive predictive value, or probability that a patient with a positive result is poisoned, determined by the formula 100x TP/(TP + FP), was 92%. The negative predictive value, or probability that a patient with a negative result is not poisoned, determined by the formula 100x TN/(TN + FN), was 75%.
Table 2
summarizes the drugs detected in specimens from all 354
patients admitted with symptoms of poisoning. The screening was done by
TLC, FPIA, Remedi, and other techniques (e.g., Trinder reaction for
salicylates). As Table 2
shows, a great number of therapeutic drugs
were involved in poisoning. The main drugs involved in suicide attempts
(benzodiazepines, antidepressant agents, neuroleptic agents) accounted
for nearly 59.6% (211 cases) of all drugs detected, and the proportion
of drug-positive results in suicide attempts was ~ 93%. In
our study, digoxin and theophylline overdoses were found only in
therapeutic drug monitoring and not in suicide attempts.
|
Benzodiazepines were often associated with antidepressants or other
psychotropic drugs (phenothiazines, meprobamate, hypnotics) in di- or
polypoisoning (Table 3
).
|
| Discussion |
|---|
|
|
|---|
2 = 65.6 for serum, P
<0.01;
2 = 140.1 for urine, P <0.01).
Benzodiazepines with low therapeutic and nontoxic concentrations (such
as clonazepam and alprazolam) were not identified by Remedi. The FPIA
technique must be preferred to Remedi in benzodiazepine screening.
Benzodiazepine detection limits were 0.02 mg/L for FPIA and 0.2 mg/L
for Remedi. Because the upper limit of the therapeutic range for some
benzodiazepines is ~0.1 mg/L, Remedi is not useful in the
identification of benzodiazepines with low toxic concentrations, e.g.,
loprazolam, flunitrazepam, clonazepam, alprazolam, triazolam,
estazolam, and nitrazepam. In urine, moreover, benzodiazepines excreted
in conjugated form were not often recognized by Remedi.
Barbiturates.
Of five cases of phenobarbital
poisoning, three were identified by Remedi, whereas TLC identified all
five. Remedi recognition of phenobarbital was incomplete because the UV
spectrum of this drug is not characteristic. The Remedi also had
problems identifying other drugs that lack a specific spectrum, e.g.,
phenytoin, phenobarbital, and primidone. We therefore use our TLC
procedure to identify some barbiturates as well as meprobamate.
Tricyclic antidepressants.
Some antidepressants
such as maprotiline are identified by FPIA only at relatively high
concentrations (TDx system operation manual, Abbott). Although the FPIA
(because of its good analytical detection limit) is more sensitive than
Remedi for some antidepressants such as amitryptiline and
nortryptiline, the FPIA antidepressant kit showed interference from
phenothiazines. Recent antidepressants such as viloxazine and
fluoxetine were not identified by FPIA. The TLC technique we used did
not identify the metabolites, whereas Remedi did. For these reasons,
Remedi is preferred to FPIA and TLC for screening for antidepressants
(see Table 1
). Remedi identified more products in urine than in serum
(59 in urines vs 51 in serum), again making urine the sample of choice.
Phenothiazines.
Phenothiazines are widely
metabolized and their identification in urine is not always easy by
Remedi. Indeed, their identification by FPIA is not possible. Thus
Remedi, as well as TLC, would be suitable for phenothiazine screening
(Table 1
).
Carbamates.
We screened for meprobamate, which
in France is still a cause of drug poisonings, by TLC only. No
immunological method for meprobamate screening is commercially
available. Moreover, meprobamate does not absorb UV energy and thus is
not identified by Remedi.
Ethanol.
Ethanol-containing drugs were prescribed by
physicians to 28% of the patients screened, and all samples from these
patients were found to be positive.
Others products.
Some cardiac, psychotropic,
antihistaminic, and other drugs were identified only by Remedi (see
Table 2
).
urine, serum, or gastric wash
As noted above, we found more substances in urine than in serum.
However, identification of drugs is easier in serum than in urine by
Remedi, because the unchanged drug is generally excreted in small
amounts in urine. The metabolites generally have the same UV spectra as
the parent drug and are eluted earlier.
Serum samples allowed semiquantification of the drugs, but assay of urine samples allowed us to find some drugs not identified in the serum. In one case of poisoning by fenspiride, the drug was found in serum and not in urine. Perhaps the fenspiride was not quantitatively eliminated in urine by the time of the sampling. For a good toxicological report, analysis from both urine and blood is needed.
For 169 gastric washes received, 33.7% were positive for drugs. The positivity of gastric content analysis allows physicians (a) to determine whether the gastric washing was "productive," and (b) to evaluate the duration of poisoning.
clinical assessment of remedi
As a clinical tool, Remedi seemed sensitive and specific, and its
positive predictive value (92%) and negative predictive value (75%)
were suitable. The high detection limit for some poisons by Remedi (in
some cases, as great as 0.5 mg/L) explains the low negative predictive
value.
The success of a general unknown analysis, i.e., the detection and unequivocal identification of an unknown toxic substance, depends on the quality of the identification system used for the analysis. According to Maier and Bogusz (10), such a system should meet several requirements: It should cover the broadest possible range of relevant substances detectable by means of a given technique; it should be standardized to enable the establishment of an applicable, common, interlaboratory database of identification markers; and the identification markers used should be independent and uncorrelated. Identification systems have found broad application in systematic toxicological screening (11). Bailey (12) studied the use of limited toxicological screening vs comprehensive toxicological screening. He found only 71% agreement between the two approaches. Remedi allows comprehensive toxicological screening with rapid turnaround time and can be used 24 h per day, the same as in limited toxicological screening.
incidence and products involved in poisoning
Nonsteroidal antiinflammatory drugs are not detected by the
methods (FPIA, Remedi, TLC) we used. Moreover, we did not use mass
spectrometry to detect these drugs and to verify their identification.
Although we regularly performed drugs-of-abuse screening with Remedi
within the framework of the methadone substitution program, no drug of
abuse was observed during the study. Thus, these substances are not
listed in Table 2
.
In limited toxicological screening, four groups of drugs were usually screened: antidepressants, benzodiazepines, ethanol, and barbiturates. However, as the number of drug overdoses in individuals with suicidal intent has increased, the number of drugs involved in poisoning has also increased, to now include ß-adrenergic blocking agents (propranolol and acebutolol), new hypnotic drugs (zolpidem and zopiclone), and new antidepressants (amoxapine, mianserin, and viloxazine).
Overdoses from barbiturates were more frequent before 1980 (13). Phenobarbital was the only barbiturate we found in this study.
For certain compounds such as propoxyphene and its metabolites, diphenhydramine, promethazine, and lidocaine and its major metabolite monoethylglycinexylidide, Remedi scored remarkably high. Other compounds, such as histamine H2-receptor antagonists cimetidine and ranitidine, and ß-adrenergic blocking agents acebutolol and propranolol, were detected exclusively by Remedi. This is in agreement with the data reported by Harris et al. (14) and Demedts et al. (5).
In the UK, antidepressants (dothiepin, amitriptyline, and imipramine) accounted for ~15% of all drug overdoses, and older tricyclic antidepressants are reported to have a fatal toxicity index 5 to 8 times higher than that of new antidepressants such as mianserin (15). The following widely prescribed drugs were commonly identified by Remedi: amoxapine, viloxazine, maprotiline, zopiclone, zolpidem, fluoxetine, and cyamemazine.
Lidocaine, an anesthetic agent, is commonly used in overdose patients who require intubation and was often found by Remedi, as was caffeine.
Among drugs involved in suicidal and poisoning cases, analgesic drugs
were the first cause of drug poisoning reported in poison control
centers (16); they were involved in few poisoning cases in
our study. Ethanol is commonly associated with other drugs in suicide
attempts (17). Poisoning situations were sometimes caused
by associations of different drugs, which aggravated the clinical
status of patients. Table 3
shows that main combinations were of
ethanol and psychotropic drugs. The 1994 Annual Report of the American
Association of Poison Control Centers (16) showed that a
single substance was implicated in 93.2% of reports, whereas 1.6% of
patients were exposed to more than two possibly poisonous drugs or
products.
Guitton et al. (6) evaluated Remedi and assessed its ability to detect in urine and gastric lavage fluids the main psychotropic drugs ingested with suicidal intent. Our findings agree with theirs. Antidepressants and most phenothiazines were well identified; benzodiazepine identification was less certain. Because of their low sensitivity in Remedi, some benzodiazepines must be screened by an immunoassay such as FPIA or an immunoenzymatic assay (Emit; Behring, Palo Alto, CA). Our results concerning benzodiazepines are in agreement with those of Demedts et al. (5). Compared with Emit, Remedi was less sensitive but allowed precise identification of the offending poison, quantified the amount, and allowed broad toxicological screening of pharmaceutical classes inaccessible to Emit.
Remedi offered another advantage: semiquantification of the main metabolites of the drug involved in poisoning and evaluation of some toxicokinetic characteristics of this drug. The forensic application of Remedi, including analysis of whole blood and tissue and comparisons with traditional laboratory methods, has been recently reported (18).
We have already applied the Remedi system to analyses in different toxicological cases. Even after the injection of >5000 samples (from whole blood, stomach contents, and urine), we observed no drift in retention behavior and no loss of efficiency, demonstrating the robustness of the system. The ability to identify the metabolites of some drugs (generally metabolites show the same spectrum and are eluted earlier than parent drugs) and to analyze urine, serum, and gastric content allowed us to determine the toxicokinetics of some drugs.
In conclusion, by using in combination several analytical techniques (e.g., FPIA for benzodiazepines and narcotics, TLC for carbamate derivatives and chloroquine, colorimetric reactions such as Trinder reaction for salicylates, and Remedi), we obtained positive results for 93% of the drug-containing samplescompared with 65% obtained by Bailey (19), for exampleand confirmed some results found by different techniques. We use these techniques routinely and suggest that a combination of different analytical techniques, including a broad-spectrum screening method such as Remedi, is the best strategy for toxicology screening in a hospital laboratory.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A Fabbri, S Ruggeri, G Marchesini, and A Vandelli A combined HPLC-immunoenzymatic comprehensive screening for suspected drug poisoning in the emergency department Emerg. Med. J., May 1, 2004; 21(3): 317 - 322. [Abstract] [Full Text] [PDF] |
||||
![]() |
A Fabbri, G Marchesini, A M Morselli-Labate, S Ruggeri, M Fallani, R Melandri, V Bua, A Pasquale, and A Vandelli Comprehensive drug screening in decision making of patients attending the emergency department for suspected drug overdose Emerg. Med. J., January 1, 2003; 20(1): 25 - 28. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Fitzgerald, J. D. Rivera, and D. A. Herold Broad Spectrum Drug Identification Directly from Urine, Using Liquid Chromatography-Tandem Mass Spectrometry Clin. Chem., August 1, 1999; 45(8): 1224 - 1234. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||