|
|
||||||||
Articles |
Departments of
1
Endocrinology and
2
Clinical Chemistry, University of Lund, Malmö University Hospital, S-205 05 Malmö, Sweden.
a Address for correspondence: Wallenberg Laboratory, Entrance 46, 2nd Fl., Malmö University Hospital, S-205 05 Malmö, Sweden. Fax +46 40 337041; e-mail Henrik.Borg{at}medforsk.mas.lu.se
| Abstract |
|---|
|
|
|---|
Key Words: indexing terms: diabetes autoimmunity islet cell antibodies method comparison ROC curve analysis
| Introduction |
|---|
|
|
|---|
In IDDM, markers of the autoimmune process may appear in blood. Islet cell antibodies (ICA; antibodies reacting against the islets of Langerhans), are the standard among such markers. Although not all ICA antigens have been defined, the ICA assay is at present the most reliable test for identifying autoimmune diabetes. ICA can be demonstrated in ~80% of the children at the time of diagnosis with diabetes (3). ICA may also be detected in ~10% of patients diagnosed as NIDDM and in such cases indicate an on-going progressive ß cell destruction leading to insulin dependency within 3 years; i.e., these patients must be regarded as having a slow-onset form of IDDM (4). Accordingly, analysis for autoantibodies is a valuable tool in the classification of diabetes.
Although regarded as the standard for detecting the autoimmune process of IDDM, the ICA test, which utilizes indirect immunofluorescence, is laborious and difficult to standardize. A need for simpler alternatives is apparent. Glutamic acid decarboxylase antibodies (GADA) are another autoimmune marker. These antibodies are associated with ß cell failure and IDDM in patients with adult-onset diabetes (5)(6)(7) but are also often detected at the time of diagnosis of diabetes in children (8). After the 65-kDa glutamic acid decarboxylase antigen (GAD 65) was characterized (9), radioligand assays were developed to measure antibodies to this antigen. 35S-labeled GAD 65, synthesized by in vitro translation, has been established as the preferred reagent for GADA assays, showing high sensitivity (10)(11). These assays, however, also have drawbacks, being expensive and technically demanding. For routine analysis, therefore, an assay using 125I-labeled GAD 65 would be more convenient.
To evaluate two new commercial assays that use 125I-labeled GAD 65 as antigen, we tested the assays on samples from 100 children with newly diagnosed diabetes and on samples from 100 control children. Although the primary aim of this study was to evaluate the two 125I assays of GADA in relation to the 35S assay, the nature of the test samples also gave us the opportunity to examine the association between GADA and ICA among children with recently diagnosed diabetes mellitus.
| Materials and Methods |
|---|
|
|
|---|
reagents
The GAD 65 cDNA used in the comparison assay was obtained from the
University of Washington, courtesy of Thomas Dyrberg, Catherine E.
Grubin, Allan E. Karlsen, Åke Lernmark, and Karen L. Deyerle. It was
provided as a plasmid (pEx9) in DH5 224 Escherichia coli
cells. pEx9 is a recombinant of the pcDNAII vector (from Invitrogen,
Carlsbad, CA), with the GAD 65 cDNA inserted between the
BamHI and XbaI restriction sites
(11). The construction of the cDNA has been described
before (13). Minipreps of the plasmid were done with
accepted methods (14), and the plasmid DNA obtained was
checked by cutting with BamHI and XbaI, which
gave a fragment of the expected size (1807 bp), and by sequencing the
insert from both ends with vector-specific primers (Universal and M13
reverse primer; Pharmacia, Uppsala, Sweden). The sequence was in
accordance with the published cDNA sequence for human GAD 65
(15) except for a modification at the 5'-end; this did not
change the amino acid sequence but made the DNA more suitable for in
vitro expression (11). Because the cDNA is positioned
downstream of the Sp6 promoter, the mRNA formed should thus give a
complete protein, including the signal peptide.
TNT Coupled Reticulocyte Lysate System was obtained from Promega (Madison, WI), RNAsin from Appligen (Illkirch, France), L-[35S]methionine from Amersham Ltd. (Bucks, UK), Protein ASepharose CL-4B from Pharmacia, and a Multiscreen 96-well filtration system from Millipore (Bedford, MA). Buffers used in the comparison assay were "plain" buffer (NaCl 150 mmol/L, Tris 20 mmol/L, pH 7.4, and NaN3, 2 g/L), coating buffer (plain buffer plus bovine serum albumin, 10 g/L), and washing buffer (plain buffer plus bovine serum albumin, 1 g/L, and Tween 20, 1.5 mL/L).
assay methods
35S assay.
The comparison assay was a slight
modification of the assay described by Petersen et al.
(11). The 35S-labeled human recombinant GAD 65
was synthesized with circular pEx9 in the TNT Coupled Reticulocyte
Lysate System. The yield of the translation product, determined by
precipitating the product with trichloroacetic acid and measuring the
radioactivity of the precipitate, was 2030% of the total
[35S]methionine added. We did not separate the
translation product from the remaining free labeled-methionine.
Overnight incubations at 4 °C with [35S]GAD 65 were
made in duplicate for each sample. Two aliquots from each incubation
were then further incubated with Protein ASepharose on a 96-well
filtration plate to collect the immunocomplexes formed. After the
filtration and washing, the bottom of each well was punched into a
scintillation bottle, and the radioactivity was counted with a Wallac
1410 (Pharmacia) liquid scintillation counter.
Sera pooled from three blood donors served as a negative control, whereas plasma from a patient with high concentrations of GADA (diluted in negative control serum to give a more suitable concentration) served as a positive control. The controls were stored as single-use aliquots at -70 °C. The results obtained with the [35S]GADA assay are presented as a GADA index: 100 x (u - n)/(p - n), where u = counts per minute (cpm) of the unknown sample, n = cpm of the negative control, and p = cpm of the positive control. (For cpm values, we used the mean activity of all four measurements for a sample.)
At the time of the study, the assay was evaluated in the International Diabetes Workshop GADA65 Proficiency Program ((2)) and showed a sensitivity and a specificity of 100% (24 samples evaluated).
125I assay 1.
The first new assay tested was
the GAD 65-antibody assay from RSR (Cardiff, UK), which included
125I-labeled human recombinant GAD 65. The GAD 65 had been
produced by expression in yeast (Saccharomyces cerevisiae)
and was purified by ion-exchange chromatography (on DEAE) and affinity
chromatography with a monoclonal antibody to GAD 65. The assay kit also
included solid-phase Protein A; GADA assay buffer; assay calibrators
with GADA concentrations of 0, 1, 3, 10, 30, and 300 arbitrary units
(A1-units) per liter; and two assay control sera.
The assay was carried out according to the instructions of the manufacturer. Briefly: Sample and 125I-labeled GAD 65 were incubated for 120 min at room temperature to form immune complexes, which were adsorbed onto solid-phase Protein A. After addition of assay buffer and centrifugation, the radioactivity of the sediment was counted in a gamma-counter. Each sample was assayed in duplicate. Results were read from a calibration curve constructed in the same run with the calibrators (also in duplicate), and expressed in A1-units. The total time for the assay was 6 h.
125I assay 2.
Assay 2, the GAD II antibody
IRMA from Elias (Freiburg, Germany), included 125I-labeled
human recombinant GAD 65. The GAD 65 had been obtained by expression of
human full-length cDNA in a baculovirus/Sf9 insect cell system and had
6 histidyl residues at the carboxy-terminus (to facilitate purification
of the antigen on Ni2+-resin). The assay kit also included
anti-human IgG; assay buffer; assay calibrators with GADA
concentrations of 25, 100, 250, 500, 1000, and 5000 arbitrary units
(A2-units) per liter; positive and negative controls; and dilution
sera.
The assay was carried out according to the instructions of the manufacturer. Briefly: Sample and 125I-labeled GAD 65 were incubated for 120 min at room temperature to form immune complexes, which were then precipitated by anti-human IgG. After addition of assay buffer and centrifugation, the radioactivity of the sediment was counted in a gamma-counter. Each sample was assayed in duplicate. Results were read from a calibration curve constructed in the same run with the calibrators (also in duplicate), and expressed in A2-units. Total time for this assay was also 6 h.
Assay performance
. The precision of the
assays was monitored by including in each run two control samples in
duplicates, one with a low concentration of GADA and one with a high
concentration. The intra- and interassay variation was estimated by
two-way analysis of variance of the pooled values for the controls.
statistics
The nonparametric MannWhitney test was used to analyze
differences in antibody concentrations between patient and control
groups, the nonparametric Spearman test to analyze the degree of
correlation between the different GADA assays, and the Fisher test to
analyze differences in frequency. ROC (receiver operating
characteristic) curves were constructed with SAS (Cary, NC) 6.10/OS/2
Warp Connect software. The significance of differences of areas under
various ROC curves was calculated as described elsewhere
(16).
| Results |
|---|
|
|
|---|
|
Figure 2
(top) shows the ROC curves obtained for the three methods when
the sensitivities were determined from the results for the diabetic
children (n = 100) and specificities were based on the control
children's results (n = 100). To compare the three assays, we
took the area under the curve as a measure of assay efficiency, i.e.,
0.83 for the 35S assay, 0.90 for assay 1, and 0.66 for
assay 2. By this measure, the comparison assay and assay 1 were clearly
more efficient than assay 2 (P <0.001). However, the
difference between the comparison assay and assay 1 was also
statistically significant (P = 0.01). Including ICA
results to define patients and controls (i.e., excluding ICA-negative
patients and ICA-positive controls) increased the efficiency (area
under the curve) of all three assays (Fig. 2
, bottom) but did not
change the relations between the assays.
|
Because ROC curves compare methods over the whole range of
specificities, most of which are not relevant in clinical work, we also
calculated the sensitivities of the three methods at a few selected
clinically relevant specificities (Table 1
). At specificities of 97.5% and 99%, respectively, the
sensitivity of the comparison assay appeared to be higher than those of
assay 1 and assay 2; however, the difference was significant only with
regard to assay 2 (P <0.01). For assay 1, the manufacturer
suggested a cutoff limit of 1.0 A1-units. In the current subject
material, this cutoff value corresponded to a specificity of 97% and a
sensitivity of 59%. For assay 2, the manufacturer's suggested cutoff
was 70 A2-units, corresponding to a specificity of 97.7% and a
sensitivity of 40% in the current study material.
|
The ROC curve for assay 2 deviates from the usual shape, showing a
constant upward convexity (Fig. 2
, both panels). The sections with
upward concave shapes (one major and one minor) were attributable to a
subgroup of the patientsmainly those who were ICA-negative, who gave
very low GADA results, even in comparison with the majority of the
control subjects (Fig. 3
).
|
In any case, the correlations between the different GADA assays were high: comparison assay vs assay 1, rS = 0.93; comparison assay vs assay 2, rS = 0.85; assay 1 vs assay 2, rS = 0.88.
The frequency of ICA and GADA positivity at 97.5% specificity for GADA
are shown in Table 2
. All but four GADA-positive patients were also ICA-positive.
Therefore, at this specificity (97.5%) for GADA, ICA was more
sensitive than GADA: 81 of 100 ICA-positive vs 63 of 100 GADA-positive
by the comparison assay (P <0.01) and 53 of 100
GADA-positive by assay 1(P <0.001) (Table 3
).
|
|
The reproducibility of the assays was determined from the results for two control samples run in duplicate in each run of each assay. In the comparison assay, the intraassay CV was 17% at the low concentration and 20% at the high concentration. The interassay CV for the comparison assay at both values seemed small, because the estimate by analysis of variance gave negative values for the variances. In assay 1, the intraassay CV was 13% at both the low and the high concentrations. This assay also led to a negative value for the variance when estimating the interassay CV at the low concentration; at the high concentration, the intraassay CV was 4%. In assay 2, the intraassay CV was 12% at the low concentration and 8% at the high concentration. The estimated interassay CV was 15% at the low concentration and 4% at the high concentration.
| Discussion |
|---|
|
|
|---|
Assay 2 was less efficient than both the comparison assay and assay 1, but the reason for the discrepancy is unclear. Perhaps the technique for the collection of the immune complexes has an effect. Anti-human IgG might be less efficient than Protein A in capturing all of the different types of autoantibodies that could appear and that might differ between patients. Also, assay 2 used a smaller sample volume than the other two assays, which might influence the limit of detection.
Assay 2 gave very low GADA values for those patients who were ICA-negative, values that were quite low even in comparison with the majority of the values for the control subjects. We had similar findings in preliminary experiments using the same GAD 65 antigen as in assay 2 in a radioligand assay developed in our laboratory. We found (Fernlund, Borg, Sundkvist, unpublished) a subgroup of IDDM patients with lower GADA values than the majority of the control subjects. We were unable to explain these findings but believe this indicates that the antigen used in assay 2 is in some respect more similar to the antigens displayed in the ICA assay than are the antigens in the two other GADA assays.
In studies comparing the clinical performance of various assays, one must select the subjects carefully, both the patients for the sensitivity determination and the controls for the specificity determination. Ideally, the test subjects should not differ from those with whom the assay is intended to be used. It is also important to establish well-defined criteria for how the between-assays comparison should be done. One general approach is to construct ROC curves for each assay and let the areas under the curves show which assay is the best, as we have done here. This means, however, that the comparison is integrated over all possible specificities, many of which are of minor interest in clinical situations. Therefore, we have also given the sensitivities of the assays when the cutoff values correspond to some of the conventional specificities, i.e., 95%, 97.5%, and 99%. What cutoff specificity to use depends on the purpose of the investigation in which the assay is to be applied. At the specificities given (95%, 97.5%, and 99%), the comparison assay had the highest sensitivity (although this was not significantly different from that of assay 1), whereas assay 2 seemed to be the least efficient.
Almost all GADA-positive patients were ICA-positive as well. On the other hand, one-third of the ICA-positive patients were GADA-negative. Thus, adding GADA measurements to ICA measurements does not seem to increase the frequency of detection of autoimmunity markers in young diabetic patients. The observation that GADA-positive patients almost always also are ICA-positive is in accordance with the concept that GAD 65 is just one of the antigens ICA react with. Further studies of patients who were ICA-positive but GADA-negative might reveal new ICA antigens, one putative candidate being IA2 (17).
Our results in this study are in accordance with previous studies of
GADA and ICA in large populations of Caucasians with recent-onset IDDM.
The studies have shown a frequency of 6080% for GADA, whereas the
ICA frequency has generally been higher, 8090% (Table 4
). In the past, children with diabetes have almost always been
regarded as having IDDM, but the increasing incidence of NIDDM in
adolescents (18) is making the differential diagnosis
between IDDM and NIDDM more uncertain. In fact, ICA-negative children
with diabetes have features (e.g., no high-risk HLA haplotypes) that
are inconsistent with IDDM (19). ICA and GADA may be
important also for the differential diagnosis of diabetes in children,
not just in adults (4)(8).
|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
V. Granberg, N. Ejskjaer, M. Peakman, and G. Sundkvist Autoantibodies to Autonomic Nerves Associated With Cardiac and Peripheral Autonomic Neuropathy Diabetes Care, August 1, 2005; 28(8): 1959 - 1964. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Alvarsson, G. Sundkvist, I. Lager, M. Henricsson, K. Berntorp, E. Fernqvist-Forbes, L. Steen, G. Westermark, P. Westermark, T. Orn, et al. Beneficial Effects of Insulin Versus Sulphonylurea on Insulin Secretion and Metabolic Control in Recently Diagnosed Type 2 Diabetic Patients Diabetes Care, August 1, 2003; 26(8): 2231 - 2237. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ankelo, A. Westerlund-Karlsson, J. Ilonen, M. Knip, K. Savola, P. Kankaanpaa, L. Merio, H. Siitari, and A. Hinkkanen Time-resolved Fluorometric Assay for Detection of Autoantibodies to Glutamic Acid Decarboxylase (GAD65) Clin. Chem., June 1, 2003; 49(6): 908 - 915. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Henricsson, L. Nystrom, G. Blohme, J. Ostman, C. Kullberg, M. Svensson, A. Scholin, H. J. Arnqvist, E. Bjork, J. Bolinder, et al. The Incidence of Retinopathy 10 Years After Diagnosis in Young Adult People With Diabetes: Results from the nationwide population-based Diabetes Incidence Study in Sweden (DISS) Diabetes Care, February 1, 2003; 26(2): 349 - 354. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Stenstrom, B. Berger, H. Borg, P. Fernlund, J. S. Dorman, and G. Sundkvist HLA-DQ Genotypes in Classic Type 1 Diabetes and in Latent Autoimmune Diabetes of the Adult Am. J. Epidemiol., November 1, 2002; 156(9): 787 - 796. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Borg, A. Gottsater, P. Fernlund, and G. Sundkvist A 12-Year Prospective Study of the Relationship Between Islet Antibodies and {beta}-Cell Function At and After the Diagnosis in Patients With Adult-Onset Diabetes Diabetes, June 1, 2002; 51(6): 1754 - 1762. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Borg, A. Gottsater, M. Landin-Olsson, P. Fernlund, and G. Sundkvist High Levels of Antigen-Specific Islet Antibodies Predict Future {beta}-Cell Failure in Patients with Onset of Diabetes in Adult Age J. Clin. Endocrinol. Metab., July 1, 2001; 86(7): 3032 - 3038. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Borg, P. Fernlund, and G. Sundkvist Protein tyrosine phosphatase-like protein IA2-antibodies plus glutamic acid decarboxylase 65 antibodies (GADA) indicates autoimmunity as frequently as islet cell antibodies assay in children with recently diagnosed diabetes mellitus Clin. Chem., December 1, 1997; 43(12): 2358 - 2363. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Miedema Electrospray Mass Spectrometry for Measurement of Glycohemoglobin Clin. Chem., May 1, 1997; 43(5): 705 - 707. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |