|
|
||||||||
Articles |
Division of Clinical Chemistry, Parkland Memorial Hospital, and Department of Pathology, The University of Texas Southwestern Medical Center at Dallas, 5323 Harry Hines Blvd., Dallas, TX 75235.
a Author for correspondence. Fax 214-648-2037; email jialal.i{at}pathology.swmed.edu
| Abstract |
|---|
|
|
|---|
4 g/L, compared with 6.2% and 12.5%, respectively,
for D-LDL. L-LDL correctly classified only 68% of patients with LDL-C
<1.30 g/L and 57% of patients with LDL-C between 1.301.59 g/L as
compared with 98% and 93%, repectively, for D-LDL (P
<0.001). In patients with type III hyperlipidemia, L-LDL had a 130%
positive bias with BQ-LDL as compared with a 14% negative bias for
D-LDL. With all three methods there were no significant differences
between samples from fasting and nonfasting individuals. On the basis
of these findings, the D-LDL assay appears to be superior to the L-LDL
assay.
Key Words: indexing terms: coronary artery disease beta quantification hypertriglyceridemia immunoseparation
| Introduction |
|---|
|
|
|---|
Given the importance of LDL-C in the diagnosis, classification, and subsequent management protocols for hyperlipidemias, there exists a need for reliable methodologies for determining concentrations of LDL-C in serum for routine use in the clinical laboratory. However, as yet, there is no consensus-approved and consensus-validated "user friendly" method for measuring LDL-C as exists for total cholesterol. Most laboratories use indirect methods for the measurement of LDL-C. The currently accepted "gold standard," beta quantification (BQ-LDL), involves ultracentrifugation of the serum sample for 18 h (7). This is a cumbersome procedure, and is labor intensive and technique dependent. It requires expensive instrumentation that preempts its extension to the routine laboratory. In spite of being the reference method, it still involves an indirect measurement of serum LDL-C.
The Friedewald formula is still used by most laboratories to estimate
LDL-C concentrations (8). The equation has been clearly
shown to be invalid in hypertriglyceridemic patients [triglycerides
(Tg)
4 g/L] and in type III hyperlipidemias (9)(10)(11)(12).
Also, a fasting sample is required to avoid a Tg bias. In light of
this, Genzyme Diagnostics developed an assay for the direct measurement
of LDL-C (available from Sigma Diagnostics), the Direct
LDLTM (D-LDL). It is an immunoseparation method that
includes antibodies against apolipoprotein (apo) AI and apo E to remove
HDL and VLDL fractions, respectively, allowing LDL-C to be directly
measured in the filtrate. Studies have demonstrated that this method
has a good correlation with BQ-LDL for measurement of LDL-C in normo-
and hypertriglyceridemic patients as well as in type III hyperlipidemic
patients (10)(13)(14). The method
is rapid, cost effective, and suitable for routine testing. Recently,
Polymedco introduced another method for the measurement of LDL-C called
LipiDirect (L-LDL). This method involves a buffered heparin reagent to
precipitate LDL, leaving HDL and VLDL in the supernatant. LDL-C is then
obtained indirectly by subtracting the cholesterol concentration of the
supernatant from the total cholesterol. This method claims to have the
same advantages of being rapid, cost effective, and suited for the
routine laboratory.
Since there appears to be no published reports comparing the new L-LDL assay with BQ-LDL or the D-LDL immunoseparation assay, the present study was undertaken to evaluate the validity of L-LDL against BQ-LDL (reference method) as well as compare it with the D-LDL immunoseparation assay in patients with normo- and hypertriglyceridemia and in type III hyperlipidemia.
| Materials and Methods |
|---|
|
|
|---|
patients
Serum and potassium EDTA plasma samples obtained from 156 fasting
(12-h) patients at the Lipid Clinic at Parkland Memorial Hospital were
tested for LDL-C concentrations by three different methodsL-LDL,
D-LDL, and BQ-LDL. Each sample was analyzed by all three assays within
the same week. Since the Friedewald equation is not valid for samples
with Tg concentrations
4 g/L, this concentration was defined as
"hypertriglyceridemia" for the purpose of this study. Of the 156
patients, 106 had Tg <4 g/L (0.613.96 g/L) and the remaining 50
patients had Tg concentrations
4 g/L (4.009.95 g/L). Patients with
fasting Tg concentrations >10 g/L were excluded from the study because
the L-LDL package insert did not recommend the assay in patients with
fasting Tg above this concentration. In addition, seven patients with
type III hyperlipidemia were also evaluated by the three methods (type
III hyperlipidemia was diagnosed both clinically and if there was a
combined hyperlipidemia and VLDL-C/total Tg ratio was
0.30)
(15). To determine the effect of the postprandial state on
the L-LDL assay, paired fasting and nonfasting samples from 25 subjects
were evaluated for LDL-C by all three methods.
procedures
Quantification of lipids.
Total Tg and cholesterol were
measured enzymatically on the Paramax Rx (unless otherwise indicated)
with the appropriate reagents from Dade International as described
previously (10). Any sample not analyzed the same day was
refrigerated at 4 °C. This laboratory is accredited by the College
of American Pathologists and participates in the ALERT proficiency
program (Pacific Biometrics) for lipoprotein analysis.
L-LDL assay.
The assay was performed as per the
manufacturer's protocol. Serum samples and controls measuring 50 µL
were added to microcentrifuge tubes containing 400 µL of LDL
precipitating reagent. After immediate vortex-mixing, the specimens
were allowed to incubate for 20 min at room temperature. The tubes were
then centrifuged for 10 min at 2000g and the clear
supernatant was transferred to a separate tube. The cholesterol content
in the supernatant was determined on the Cobas Mira with the
cholesterol reagent and calibrator from Sigma. LDL-C was determined by
subtracting the supernatant cholesterol from the total serum
cholesterol.
D-LDL immunoseparation assay.
The assay was performed as
previously described (10). Briefly, LDL-C reagent (200
µL) was pipetted into separation tubes provided in the kit. Controls
and serum samples (30 µL) were added to the tubes containing latex
beads. After immediate vortex-mixing, the tubes were incubated for 10
min and centrifuged at 6000g for 5 min. The cholesterol in
the filtrate was measured on the Cobas Mira with the cholesterol
reagent and calibrator from Sigma. Cholesterol calibration was closely
monitored with the ALERT standardization program for LDL-C. An average
bias of 3.6% was attained with the ALERT proficiency program for
D-LDL-C.
BQ-LDL by ultracentrifugation.
Two milliliters of plasma
at a density of 1006 g/L was centrifuged at 4 °C in a fixed-angle
rotor for 18 h at 109 000g (10). The top
and bottom fractions were collected into separate tubes and the bottom
fraction was brought up to 2 mL with saline. Cholesterol was determined
in the plasma and the two fractions. Percent recoveries, determined
from the sum of the cholesterol in the two fractions and compared with
the total plasma cholesterol, were within 100.2% ± 2.32%. LDL-C from
BQ-LDL was obtained by subtracting HDL-C (obtained by precipitation)
from the bottom fraction cholesterol. An average bias of 3.5% was
attained with the ALERT proficiency program for LDL-C by BQ-LDL.
Precision studies.
For the L-LDL intraassay precision
tests, three serum samples with LDL-C concentrations of 0.84, 1.61, and
2.26 g/L were each assayed 20 times with a single reagent lot. For the
interassay precision tests, three samples with 0.77, 1.52, and 2.10 g/L
LDL-C were assayed in duplicate for 8 days over a 10-day period. The
precision studies for the D-LDL immunoseparation assay and BQ-LDL have
been reported previously (10).
Fasting and postprandial samples.
To determine the
effect of the postprandial state on LDL-C measurement by L-LDL, 25
healthy subjects were asked to fast overnight for at least 12 h.
Their blood was drawn in the morning by venipuncture into EDTA and
serum separator tubes. The subjects were requested to consume a
standard high-fat meal (available from an in-house fast-food outlet and
containing 37 g of fat and 0.26 g of cholesterol); a second
blood sample was drawn 3.5 h later. Fasting and postprandial
samples were then assayed for LDL-C by the three methods.
Statistical analysis.
Linear regression analysis was
used to compare LDL-C values obtained by the L-LDL assay and D-LDL
method against the reference method (BQ-LDL) in normo- and
hypertriglyceridemic (Tg
4 g/L) samples. The Wilcoxon test was used
to compare data that were not normally distributed, whereas the paired
t-test was used for data that was normally distributed. The
mean absolute bias [
(|xi -
|)/n] was calculated for L-LDL and D-LDL methods
compared with BQ-LDL. The intraclass correlation coefficient (ICC) was
used in evaluating the performance of L-LDL against D-LDL in type III
hyperlipidemic patients. The McNemar test was used to compare
appropriate classification of LDL-C at the NCEP cutoff values of 1.30
and 1.60 g/L for L-LDL and D-LDL.
| Results |
|---|
|
|
|---|
|
comparison of ldl-c values obtained from l-ldl, d-ldl, and bq-ldl
Linear regression analysis was performed for L-LDL and D-LDL vs
BQ-LDL for normo- and hypertriglyceridemic subgroups. As seen in Fig. 1
a and 1b, both L-LDL and D-LDL show good correlation with BQ-LDL
for Tg concentrations <4 g/L (r = 0.95 and 0.97,
respectively). Fig. 2
a and 2b shows the good correlation of L-LDL and D-LDL with
BQ-LDL for Tg concentrations
4 g/L (r = 0.91 and
0.94, respectively). For the comparison studies, LDL-C concentrations
were measured in 156 samples (Tg range 0.619.95 g/L) by the three
assays. Table 2
shows the mean ± SD LDL-C obtained by each method for the
overall Tg range as well as in normo- (<4 g/L) and
hypertriglyceridemic (
4 g/L) samples. The mean LDL-C value of L-LDL
was significantly higher when compared with D-LDL as well as BQ-LDL for
the overall Tg range and in both normo- and hypertriglyceridemic
subgroups (P <0.001). The mean LDL-C value of D-LDL did not
show any statistical difference with BQ-LDL for the overall Tg range
and in samples with Tg <4 g/L but was significantly higher than BQ-LDL
in samples with Tg
4 g/L (P <0.001).
|
|
|
Bias plots of L-LDL and D-LDL against BQ-LDL concentrations and
increasing Tg concentrations are shown in Figs. 3
and 4, respectively. As seen in Fig. 3
, there is a mean absolute bias
of 0.22 g/L for L-LDL, whereas the mean absolute bias for D-LDL is only
0.10 g/L. A larger positive bias is seen with L-LDL than with D-LDL in
Fig. 4
, which becomes more pronounced in hypertriglyceridemic samples
(including some biases >0.50 g/L for L-LDL). The mean absolute
percentage bias of L-LDL vs BQ-LDL was 18.5% for the entire range of
Tg values, 12.7% for Tg <4 g/L, and 30.6% for Tg
4 g/L, whereas
the mean absolute bias for D-LDL vs BQ-LDL was 8.3% for the entire
range of Tg values, 6.2% for Tg <4 g/L, and 12.5% for Tg
4 g/L.
|
|
appropriate classification of patients by l-ldl and d-ldl assays
with the ncep ldl-c cutoff points
The NCEP has laid down LDL-C values of <1.30, 1.301.59, and
160 g/L as cutoff points for classifying patients without CAD with
normal, borderline, and high LDL-C. Since these cutoffs are crucial in
therapeutic decision making, we studied the appropriate classification
by L-LDL and D-LDL assays into the three categories with LDL-C values
coinciding within ± 10% of BQ-LDL (reference cholesterol value).
As seen in Table 3
, with BQ-LDL as the point of reference, L-LDL correctly
classified only 68% of patients with LDL-C <1.30 g/L, whereas D-LDL
values coincided within ± 10% of BQ-LDL in 98% of these
patients. Similarly for concentrations 1.301.59 g/L, L-LDL and D-LDL
correctly classified 57% and 93% of the patients, respectively. The
McNemar test was used to compare the appropriate classification of the
two methods for patients with LDL-C <1.30 g/L (P =
0.0001) and with LDL-C 1.301.59 g/L (P = 0.0006). For
LDL-C values >1.60 g/L, the appropriate classification by both assays
was 100%.
|
comparison of mean ldl-c determined by l-ldl, d-ldl, and bq-ldl
assays in type iii hyperlipidemics
Because the Friedewald equation is invalid in type III
hyperlipidemia, having an assay that can reliably determine LDL-C
concentrations in these patients is essential. To test the validity of
the L-LDL test in type III hyperlipidemia, LDL-C was determined in
seven patients by all three methods. As seen in Table 4
, the median LDL-C determined by L-LDL was significantly higher
than BQ-LDL (2.72 vs 1.03 g/L), with a mean positive bias of 130%. In
contrast, D-LDL showed a mean negative bias of 14% from the BQ-LDL. We
also compared the mean LDL-C values for L-LDL and D-LDL with BQ-LDL by
using an ICC to express and compare the reliability index of the test
result between the two assays. The ICC for L-LDL was 0.82 in type III
hyperlipidemics as compared with 0.99 for D-LDL, clearly demonstrating
the superiority of the latter assay.
|
validity of l-ldl and d-ldl assays in the postprandial state
Table 5
shows the effect of postprandial state on the concentrations of
total cholesterol, Tg, and LDL-C (as measured by the three methods) in
25 subjects. Total cholesterol showed no statistical difference between
fasting and nonfasting samples, whereas the Tg was significantly higher
in postprandial samples. The results for L-LDL, D-LDL, and BQ-LDL were
not significantly different in the fasting and nonfasting states.
|
| Discussion |
|---|
|
|
|---|
Currently, most routine laboratories continue to use the Friedewald equation to calculate the concentrations of serum LDL-C (8). The traditional approach has been to use the equation only in fasting patients with Tg <4 g/L because the results are unreliable under the following circumstances: (a) when chylomicrons are present; (b) when plasma Tg exceeds 4 g/L; and (c) in patients with type III hyperlipidemia (9)(10)(11)(12)(16)(17). This occurs because the equation is based on the assumption that the majority of circulating Tg resides in the VLDL fraction and that the relation between Tg and cholesterol in this fraction is constant. Particles found in hypertriglyceridemic samples that are generally called "VLDL" are often, in fact, a heterogenous mix of VLDL and chylomicron remnants. It would be more precise to call such particles Tg-rich lipoproteins (TRL) (13). The percentage of cholesterol is very different across this range of particles and as the Tg concentration rises, fewer of them may contain the usual 20% cholesterol. Errors in calculating the TRL cholesterol will ultimately reflect as an error in LDL-C. Type III hyperlipidemics have increased concentrations of IDL and VLDL remnants in their serum. This fraction (beta VLDL on electrophoresis) contains proportionately more cholesterol than normal VLDL and use of the Friedewald equation in these patients clearly leads to inaccurate LDL-C values (15). The NCEP recommendation states that in the above circumstances, LDL-C should be measured with the combined ultracentrifugationpolyanion precipitation method (BQ-LDL). Unfortunately, this method is both time consuming and labor intensive, and requires ultracentrifugation that is not available for routine use in most laboratories. In spite of being a reference method, it still calculates LDL-C indirectly. Also, what we commonly refer to as "LDL-C" by BQ-LDL actually represents cholesterol contained in LDL plus IDL plus lipoprotein(a) and the measurement might better be considered to represent the cholesterol contained in several potentially atherogenic particles (7)(16).
In the search for an assay for the direct quantification of LDL-C,
Genzyme Diagnostics (Cambridge, MA) developed an immunoseparation
method (available through Sigma Diagnostics)the D-LDL assayin 1994.
Jialal et al. (10), Pisani et al. (13), and
McNamara et al. (14) have all compared this
immunoseparation method for direct LDL-C measurement with BQ-LDL and
found good correlation between the two methods for all Tg
concentrations. No significant differences were found in LDL-C
concentrations between the D-LDL assay and BQ-LDL in patients with type
III hyperlipidemia in the study by Jialal et al. (10). The
L-LDL assay marketed by Polymedco claims to be another simple, rapid,
and cost-effective test for measuring LDL-C. The NCEP has laid down
clear analytical goals for the acceptibility of any new assay measuring
LDL-C (17). Previous studies have clearly shown D-LDL to
have precision values that fall within these guidelines
(10)(13)(14). For L-LDL, the
intraassay precision was excellent (<3%), but the interassay
precision, especially at lower concentrations of LDL-C, was less than
desirable (CV <8%). When we compared LDL-C values determined by L-LDL
against the reference method, we found significant differences across
all Tg ranges (both normo- and hypertriglyceridemic, P
<0.001). With L-LDL, 103 of the total 156 patients (including 47 of 50
with Tg
4 g/L) had LDL-C values that showed a >10% bias as compared
with BQ-LDL. In contrast, D-LDL values did not have any significant
difference with the BQ-LDL values across the overall Tg range and in
normotriglyceridemic samples. Although the D-LDL assay showed
significantly increased values in hypertriglyceridemic sera, this was
not as great as with the L-LDL assay. Moreover, other authors have also
shown a definite positive bias for D-LDL estimation compared with
BQ-LDL in hypertriglyceridemia (13)(14), but
in all cases the bias was less than the NCEP total error goal of
<12%. In this study we found the mean absolute bias for D-LDL in
samples with Tg
4 g/L to be 12.5%. For the same patients, L-LDL gave
a mean absolute bias of 30.6% (with two of the values actually having
>100% bias compared with BQ-LDL). For samples with Tg <4 g/L, D-LDL
had a mean absolute bias of 6.2% as compared with 12.7% for L-LDL.
However, because L-LDL and D-LDL were performed on serum samples and
BQ-LDL was performed on EDTA plasma samples, the actual biases for the
two methods are 3% lower than that reported. Thus, our comparison
studies clearly demonstrate that L-LDL gives less precise and less
accurate results in normotriglyceridemic patients as compared with
D-LDL, and this difference becomes accentuated in cases with
hypertriglyceridemia.
The NCEP LDL-C cutoff concentrations of 1.30 and 1.60 g/L are very
important laboratory parameters in therapeutic decision making. By
reporting a patient's LDL-C above or below the conventional cutoffs,
the laboratorian is conveying a definite message to the clinician about
risk stratification. Hence, having an assay that can reliably and
consistently classify patients in the NCEP categories of LDL-C
concentrations <1.30, 1.301.59, and
1.60 g/L is absolutely
mandatory. Comparison studies for the two methods show D-LDL to be
obviously superior to L-LDL as it appropriately classified >90% of
patients with LDL-C concentrations <1.30 and 1.301.59 g/L. L-LDL
performed rather poorly in these two categories. The only occasion when
L-LDL did compare with D-LDL was in patients with LDL-C
1.60 g/L
(100% appropriate classification for both), which we believe is
because of the tendency of L-LDL to overestimate LDL-C in most cases.
If these percentages were translated into numbers it would mean that 48
of the total 156 patients were misclassified by L-LDL (with two
patients actually skipping categories) as compared with 7 of 156
patients being misclassified by D-LDL (with no patient skipping
categories). It is obvious from the above figures that any laboratory
using L-LDL for risk assessment as well as follow-up would have to
contend with these major discrepencies.
Having established the superiority of D-LDL over L-LDL as an assay with regard to analytical precison, accuracy, and appropriate classification, we examined their usefulness in the subset of patients with type III hyperlipidemia. Our results show that L-LDL is an extremely inaccurate test in these patients, with the median LDL-C being 2.74 ± 1.62 as compared with 1.23 ± 0.88 g/L for BQ-LDL. One of the possible reasons to explain this discrepancy could be that the LDL precipitating reagent used in L-LDL is also precipitating other cholesterol-rich fractions into the filtrate. In contrast, D-LDL, being an immunoseparation assay, performed better as reported previously (10).
Calculation of LDL-C with the Friedewald equation requires patients to fast for 1012 h (12). Any assay that could dispense with fasting samples would not only be convenient for the patient at large but would also prove beneficial in subsets such as diabetics. The L-LDL and D-LDL both showed good correlation between fasting and postprandial samples. Other authors who have worked with D-LDL have shown a similar correlation and recommended its use irrespective of the patient's fasting status (10)(13)(14). We felt it was paradoxical that L-LDL, which showed poor results at higher Tg concentrations, should show such good correlation in the fast-feed study. A plausible reason for this observation is that only one of the 25 subjects attained a Tg value >4 g/L after the test meal.
One criterion by which some laboratories could adopt the L-LDL in the calculation of LDL-C is that it claims to be a cheaper assay. The immunoprecipitation method used by D-LDL costs $750 for a 100-test kit. In contrast, the L-LDL assay has been priced at $620 for a 125-sample kit. However, the L-LDL controls, which have to be bought separately, cost $75 to the laboratory and are stable for only 10 days. Obviously, a more detailed scrutiny of the "fiscal benefits" in using the L-LDL assay needs to be done.
In conclusion, the purpose of this study was to compare two methods
for measuring LDL-C in the serum against the reference method and
establish the incremental benefit (if at all) of one over the other. We
feel that D-LDL has definitely emerged as a superior assay over L-LDL
in terms of precision, accuracy, and correct classification of
patients. However, we would caution at this point that any laboratory
that adopts a new cholesterol assay must participate in a proficiency
program such as ALERT to keep a constant vigilance on the quality of
the results being reported. We also feel the traditional cutoff of Tg
<4 g/L for using the Friedewald equation may need to be revisited. In
this context, McNamara et al. (12) have clearly shown that
even in samples with Tg
2 g/L the Friedewald equation results in
significant inaccuracies (>10% bias in 23% and 41% of samples with
Tg 2.013 g/L and 3.014 g/L, respectively). In addition, a recent
study from this center has demonstrated the greater accuracy of D-LDL
over Friedewald equation in the diabetic population with Tg
concentrations <4 g/L (18). Normotriglyceridemic
nonfasting patients are the only subset of patients where either the
D-LDL or L-LDL assay can be used. However, on the basis of the findings
of this study, the authors would favor the D-LDL assay over L-LDL. In
hypertriglyceridemic patients (fasting or nonfasting), the D-LDL assay
could be used in most clinical laboratories. The L-LDL is too
inaccurate to be a reliable assay in these patients. We did not
evaluate patients with Tg >10 g/L in this study for two reasons.
First, the L-LDL manufacturer does not recommend the assay in samples
with Tg >10 g/L. Second, and more importantly, we strongly believe
that in patients with such high Tg concentrations, instead of spending
time and effort in determining LDL-C, the prudent clinician would be
more concerned in lowering the Tg value and with it the attendant
immediate risk of pancreatitis before assessing cardiovascular risk.
Thus, on the basis of our data, we believe that the D-LDL
immunoseparation assay is the preferred method for measurement of LDL-C
over the L-LDL method.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
E. T. Bairaktari, K. I. Seferiadis, and M. S. Elisaf Evaluation of Methods for the Measurement of Low-Density Lipoprotein Cholesterol Journal of Cardiovascular Pharmacology and Therapeutics, January 1, 2005; 10(1): 45 - 54. [Abstract] [PDF] |
||||
![]() |
M. Nauck, G. R. Warnick, and N. Rifai Methods for Measurement of LDL-Cholesterol: A Critical Assessment of Direct Measurement by Homogeneous Assays versus Calculation Clin. Chem., February 1, 2002; 48(2): 236 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Jialal, D. Stein, D. Balis, S. M. Grundy, B. Adams-Huet, and S. Devaraj Effect of Hydroxymethyl Glutaryl Coenzyme A Reductase Inhibitor Therapy on High Sensitive C-Reactive Protein Levels Circulation, April 17, 2001; 103(15): 1933 - 1935. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Esteban-Salan, A. Guimon-Bardesi, J. M. de la Viuda-Unzueta, M. N. Azcarate-Ania, P. Pascual-Usandizaga, and E. Amoroto-Del-Rio Analytical and Clinical Evaluation of Two Homogeneous Assays for LDL-Cholesterol in Hyperlipidemic Patients Clin. Chem., August 1, 2000; 46(8): 1121 - 1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Gomez, J. Camps, J. M. Simo, N. Ferre, and J. Joven Agreement Study of Methods Based on the Elimination Principle for the Measurement of LDL- and HDL-Cholesterol Compared with Ultracentrifugation in Patients with Liver Cirrhosis Clin. Chem., August 1, 2000; 46(8): 1188 - 1191. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |