|
|
||||||||
Articles |
1
Institute of Analytical Chemistry, Chemo- and Biosensors, University of Regensburg, 93040 Regensburg, Germany.
2
Faculty of Medicine, Departments of Clinical Chemistry
and Laboratory Medicine, BL I, Karl-Franzens University, 8010 Graz,
Austria.
a Author for correspondence. Fax 49-941-943-4064; e-mail Otto.Wolfbeis{at}chemie.uni-regensburg.de
| Abstract |
|---|
|
|
|---|
Key Words: indexing terms: urine diabetes renal function hypertension atherosclerosis nephropathy cardiovascular disease
| Introduction |
|---|
|
|
|---|
MAU is an early sign of incipient renal disease and a marker of its progression. Furthermore, MAU is a key indicator of the need for intensified treatment in diabetes mellitus and hypertension (3) and is one of the most powerful predictors of cardiovascular disease in nondiabetic individuals (1)(4)(5).
Patients with insulin-dependent diabetes mellitus are susceptible to developing nephropathy, the peak incidence occurring as late as 1520 years after the onset of diabetes (6). As is now well established, even a slightly increased excretion of albumin (1530 mg/24 h) is both diagnostic for incipient renal disease and prognostic for the development of end-stage renal disease and mortality (7)(8)(9). Recent studies show that also in patients with non-insulin-dependent diabetes mellitus, persistent MAU is clearly associated with increased mortality (10)(11).
Apart from its use in monitoring renal function loss and renoprotection in diabetes mellitus, MAU plays a key role as a potential cardiovascular and atherosclerotic risk factor (12)(13). In association with endothelial dysfunction, it is considered a generalized indicator of cardiovascular vulnerability. Hence, the detection of MAU is an early alert to undertake efforts to reduce other cardiovascular risk factors (14).
Broad screening for MAU is justified by a favorable cost-benefit ratio, because it can help identify patients who would benefit from renoprotective therapy and antihypertensive treatment (15). Moreover, evidence indicates that, if detected at an early stage, MAU can be stabilized or even reduced with antihypertensive intervention (16). The American Diabetes Association (17) and the American Kidney Foundation (18) are in consensus that all persons with diabetes should have their urine tested annually for MAU.
In addition to the qualitative detection of overt MAU by dipstick methods, precise quantitative determination of albumin is essential for diagnosis of incipient MAU, for assessing the renal state, for optimizing diabetes care, and for monitoring the success of therapy. Therefore, precise assays for urinary albumin are now becoming inevitable in laboratory medicine (19).
For determination of MAU, the urinary albumin concentration range of interest is 2200 mg/L, particularly the 1540 mg/L region that covers the usual cutoff limits between normal and increased albumin excretion (20). Several formats of immunoassays are available that are sufficiently responsive and reliable for albumin determinations at these concentrations, and some are fully automated to analyze large numbers of samples. However, using these methods to assay only a few samples is impractical and expensive, which limits their application in local laboratories. In recent years, handy immunoassay-based test kits have emerged but are designed for screening rather than precise quantification of MAU (21); routine quantifying methods are not yet common.
Recently, the Albumin Blue (AB) series of fluorescent albumin dyes has been described (22). These dyes are capable of targeting albumin highly specifically by forming a strongly fluorescent complex. Unlike the case with other common albumin probes, the fluorescence of the AB complex is enhanced by more than an order of magnitude upon binding to albumin and is not subject to interference by other urinary proteins. The properties of derivatives AB 633 and AB 670 have been exploited to assay urinary albumin at trace concentrations (23)(24). Although that method is comparable with immunonephelometry in terms of detection limits and performance, the rather short shelf lives of AB 633 and AB 670 stock solutions obviate their routine application.
To overcome poor reagent stability, we developed a new derivative (AB 580) with that has better stability without compromising assay performance. The derivative has already been used to determine serum albumin (25). In this work we have implemented the AB 580 assay on the Cobas FaraTM II (Hoffmann-La Roche, Basel, Switzerland) centrifugal analyzer for urine analysis and report our evaluation of this combination as a fast and simple low-cost assay especially designed for precise quantification of MAU.
| Materials and Methods |
|---|
|
|
|---|
reagents
We obtained AB 580,
([2-chloro-3-(2,2-dicyanoethenyl)-2-cyclohexen-1-ylidene]methylpropanedinitrile
potassium salt (Mr = 306.8), from Molecular
Probes (Eugene, OR).
The dye stock (AB 580 71 µmol/L in isopropanol) was prepared as follows: Dissolve solid AB 580 in isopropanol (cat. no. 405-7; Sigma Chemical Co., St. Louis, MO) to give a concentration of ~30 mg/L in a dark blue solution (crude stock). Check the concentration of AB 580 by diluting an aliquot 10-fold with isopropanol and measuring the absorbance at 580 nm. If A580 = 1.0 ± 0.05 (1-cm cuvette, blanked against isopropanol), use the crude stock as the dye stock. Otherwise, dilute the crude stock with isopropanol or add more solid AB 580 and recheck the concentration. The dye stock remains stable for at least 2 years when stored at 4 °C in the dark.
To prepare the working dye solution, dilute 5.0 mL of dye stock to a final volume of 100 mL with an equivolume solution of isopropanol/water. The A580 of this solution is 0.5 ± 0.05, and the reagent is stable for at least 6 months when stored at -20 °C in the dark.
The working buffer was prepared by mixing 3.0 g of 3-N-morpholinopropanesulfonic acid (MOPS free acid; Sigma M-1254), 9.0 g of MOPS sodium salt (Sigma M-9381), 12.0 g of sodium chloride, and 1.0 g of EDTA disodium salt (Sigma E-4884) with 900 mL of distilled water, and 100 mL of isopropanol. The resulting pH is 7.4 ± 0.2.
calibrators and controls
The calibrator diluent was prepared by mixing 2.7 g of
KH2PO4, 0.9 g of
K2HPO4, 4.5 g of sodium chloride, 0.5
g of EDTA disodium salt, and 50 mg of human IgG (Sigma G-4386), in 500
mL of distilled water. The resulting pH is 6.0 ± 0.5. IgG is a
stabilizer, preventing nonspecific adsorption of albumin to the wall of
the container. The albumin stock calibrator (2000 mg/L) was prepared by
dissolving 10 mg of human serum albumin (essentially fatty acid- and
globulin-free, prepared from Cohn Fraction V; Sigma 3782) in 5 mL of
calibrator diluent. The concentration was finely adjusted until the
A280 was 1.080 ± 0.010 (26)
blanked against calibrator diluent. Calibrators were prepared by
diluting calibrator stock with calibrator diluent to final
concentrations of 2.0, 10, 30, 100, and 200 mg/L (working range 2200
mg/L) and 1.0, 10, 30, and 50 mg/L (working range 150 mg/L),
respectively.
N-Protein Standard (SY/OWXI 12/13, lot no. 083604 D; Behring) was used as a calibrator in the nephelometric assay. N/T-Protein Control Serum (SY/M OWXK 12/13, lot no. 083703; Behring) was diluted 1:1000 with calibrator diluent to contain 42.0 mg/L; this solution was used as control material for all assays.
urine samples
We selected centrifuged urine samples from a bulk of samples
submitted to routine analysis. After initial measurement by the Cobas
Fara, the samples were divided into five categories according to
albumin concentrations: A: 030 mg/L; B: 3060 mg/L; C: 60100 mg/L;
D: 100200 mg/L; and E: >200 mg/L. For method comparison, we drew 30
samples each of categories A and B and 20 samples each of categories C
and D by random selection and analyzed with the BNA.
assay procedure
The AB 580 working dye solution and the working buffer were placed
in the start reagent cup and the main reagent cup of the Cobas Fara
reagent rack, respectively. Before the first assay run, we adjusted the
lowest detection limit of the photomultiplier according to the
instructions, with the 100 mg/L calibrator. The assay program is
summarized in the
Appendix.
| Results |
|---|
|
|
|---|
|
Compared with derivatives AB 633 and AB 670, dye AB 580 has much better stability as indicated by the times to decomposition of one-half of the amount of dye in pH 7.4 aqueous buffer: 3 days for AB 633, 3 h for AB 670, and 50 days for AB 580 at 20 °C. Moreover, the degradation product of AB 580 does not interfere in the assay. Nonetheless, the improved stability of aqueous stock solutions is still insufficient for routine application. Fortunately, isopropanol solutions are much more stable. Thus, an isopropanol stock of AB 580 did not show any notable degradation after 2 years of storage at 4 °C.
assay performance
Working range and calibration curve.
Two Cobas Fara programs
were developed with working ranges of (a) 2200 mg/L
(injected sample volume 15 µL) for a general test covering
normoalbuminuria and the full concentration range of MAU, and
(b) 150 mg/L (injected sample volume 60 µL), for a test
with an enhanced detection limit and greater precision at low
concentrations, covering normoalbuminuria and albumin concentrations in
early MAU (Fig. 2
). The greater precision is advantageous for
determining increased albumin excretion.
|
Precision.
For the working range 2200 mg/L,
within-run CVs (n = 10) were 8.9% (7.2 mg/L), 3.0% (26 mg/L),
3.1% (52.2 mg/L), 2.5% (98.9 mg/L), and 1.7% (189 mg/L). Between-run
CVs (n = 10; calibration with each run) were 6.4% (15 mg/L),
2.6% (39 mg/L), 3.2% (62 mg/L), and 3.1% (104 mg/L). The detection
limit for albumin (3 SD of the blank signal, n = 10, within-run)
was 1.4 mg/L in the sample, equivalent to 0.07 mg/L in the cuvette
solution.
For the working range 150 mg/L, within-run CVs (n = 10) were 5.7% (2.3 mg/L), 2.5% (6.5 mg/L), 3.2% (24.9 mg/L), and 2.9% (43.3 mg/L). Between-run CVs (n = 10; calibration with each run) were 10.1% (2.3 mg/L), 7.1% (6.5 mg/L), 3.8% (24.9 mg/L), and 3.7% (43.3 mg/L). The detection limit for albumin, defined as above, was 0.4 mg/L in the sample.
Recovery.
Values obtained for controls were within the
specified reference range, with recoveries between 95% and 100% of
the reference value. For analysis of supplemented samples, we added
various volumes of calibrator stock to aliquots of urine pools.
Analytical recovery of albumin is summarized in Table 1
.\
|
Method comparison.
Comparison of the AB 580 fluorescence assay
(y) with the BNA immunonephelometric assay (x)
gave y = 0.922(± 0.010) x + 4.16 (± 0.78)
(n = 100, r = 0.989) with mean y =
63.8 mg/L and mean x = 63.0 mg/L (Fig. 3
).
|
Interference.
We tested assay interference by supplementing
aliquots of urine pools (15 mg/L albumin) with common drugs, vitamins,
proteins, and urine metabolites (Table 2
). The
concentrations added were well above the maximum concentrations
encountered in clinical practice. The overall mean albumin
concentrations of 71 samples tested was 14.95 mg/L (CV 7.9%). Because
the albumin concentrations measured were all within 2 SD of the
reference value, i.e., 15 ± 1.9 mg/L, we considered that none of
the substances interfered significantly. Metformin gave the greatest
interference of all substances tested, the albumin concentration in its
presence being 13.1 mg/L.
|
Human proteins globulin, transferrin, Bence Jones protein, plasmin, plasminogen, antitrypsin, fibrinogen, trypsin, and chymotrypsin, as well as soybean inhibitor, bovine pancreatic trypsin inhibitor, and poly(D-lysine) did not interact substantially with AB 580. The presence of hemoglobin at concentrations >100 mg/L (between first visible and overt hematuria) lowered the measured albumin values by as much as 20%. Similarities in amino acid sequence and tertiary structure mean that human and bovine serum albumin as well as other albumin species (except ovalbumin) responded almost equally. Also, micellar solutions of detergents strongly enhanced the fluorescence of AB 580 and therefore strongly interfered in the assay.
Fatty acids displace AB 580 (and vice versa) by competitive binding. Nevertheless, because of the low solubility of free fatty acids in aqueous solution, free AB 580 is always in excess, and the equilibrium is shifted towards bound AB 580. This is indicated by only minor signal changes upon transition from essentially fatty acid free to fatty acid saturated albumin.
| Discussion |
|---|
|
|
|---|
We specify the concentration of the dye stock by A580 rather than weight per volume. The former is more accurate and practical and is independent of the purity of solid AB 580. The concentration of stock solution and the protocol presented here may be changed to improve the detection limit of the assay or to get a wider working range. The assay is performed at pH 7.4, which is within the pH region of optimum performance. Although the assay is not extremely pH-sensitive, the strong buffer used prevents pH shifts from samples with extreme pH values.
Calibration and assay procedure.
The binding of AB 580 to
albumin obeys the Law of Mass Action. Therefore, calibration curves are
nonlinear and approach a point of saturation at high albumin
concentrations. When performed in the upper flat region of the
calibration curve, the assay gave unreliable results with unacceptable
high CVs. To prevent this situation, we use excess AB 580, which makes
the calibration curve steeper and provides an extended reliable working
range up to 200 mg/L. Moreover, excess AB 580 suppresses the
competitive binding of potential interferents at the albumin binding
sites. Apart from the logit-log4 function used by the Cobas Fara, the
calibration curve may be approximated by the function y
= [Ax/(1 + Bx)] + C, in which x
= concentration of albumin in the sample; y = relative
fluorescence intensity; and A, B, and C are parameters obtained from a
curve fit.
Because the assay was found to depend strongly on temperature, good temperature control is essential. This is achieved by the thermostating feature of the Cobas Fara. Errors arising from minor temperature deviations are compensated for by running calibrators with each run.
Method comparison.
Given the different principles of the
methods compared, the fluorescence assay agrees well with the
independent BNA immunonephelometric method. In addition, the
correlation coefficient (0.989) is comparable with that obtained from
comparing two immunological methods (27). Regression
analysis gives a y-intercept of 4.2 ± 0.78, which is
significantly different from zero, and a slope (0.922 ± 0.010)
significantly different from 1.0. Both deviations may be attributed to
the different calibrators used in the two methods.
In some cases, low values obtained with the BNA were determined to be falsely negative after dilution and reexamination. These samples were easily identified by the Cobas Fara from their high fluorescence reading and the out-of-range indicator (>200 mg/L).
Features and advantages of the method.
Given a detection limit
and a selectivity about equal to those of immunological methods, the AB
580 assay performed with the Cobas Fara has several advantages.
(a) It is fast and easy to perform without incubation or
pretreatment of urine samples. (b) The extended working
range of 2200 mg/L covers the entire scale of
MAU without dilution
of samples. (c) With no need for expensive antibodies, reagent costs
are low, and no laboratory animals are needed for antibody production.
(d) Unlike immunoassays, the AB 580 method does not report
false negatives if albumin concentrations highly exceed the upper limit
of the working range. (e) There is no lot-to-lot variation
of the reagent, and the quality of the reagent can be easily checked by
measuring its absorption spectrum.
In conclusion, the method presented here is ideally suited for assaying urinary albumin in determining MAU. Furthermore, this method also appears potentially useful for determining albumin at low concentrations in other biological fluids. It is easy to perform, fully automated, and comparable with existing immunoassays in terms of detection limits and performance. Besides the Cobas Fara, the assay may also be performed with other fluorescence instruments, including microplate readers, an approach we are currently investigating.
|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
G. C. Shearer, J. W. Newman, B. D. Hammock, and G. A. Kaysen Graded Effects of Proteinuria on HDL Structure in Nephrotic Rats J. Am. Soc. Nephrol., May 1, 2005; 16(5): 1309 - 1319. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Choi, E. Y. Choi, H. S. Kim, and S. W. Oh On-Site Quantification of Human Urinary Albumin by a Fluorescence Immunoassay Clin. Chem., June 1, 2004; 50(6): 1052 - 1055. [Full Text] [PDF] |
||||
![]() |
D. P. Basile, D. Donohoe, K. Roethe, and J. L. Osborn Renal ischemic injury results in permanent damage to peritubular capillaries and influences long-term function Am J Physiol Renal Physiol, November 1, 2001; 281(5): F887 - F899. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |