|
|
||||||||
Articles |
a Address correspondence to this author at: Andrology Unit, Department of Medicine (D02), University of Sydney, NSW 2006, Australia. Fax +61 2 9351 4560; e-mail djh{at}med.su.oz.au
| Abstract |
|---|
|
|
|---|
Key Words: indexing terms: fingerprick blood vs plasma sample androgen testosterone nandrolone radioimmunoassay sample handling
| Introduction |
|---|
|
|
|---|
Rapid, early sequence sampling also requires an indwelling venous cannula and detaining the subject for a period of time. Therefore a trained staff for venipuncture as well as sample handling, storage, and transport facilities is needed. These invasive procedures also artificially confine subjects' activities, with unpredictable effects on steroid absorption and metabolism, which limits the generalization of such findings to real-world experience. Furthermore, the problems of sample collection, storage, and transport restrict most studies to specialized centers. Studies of prototype androgen-based hormonal male contraceptives have raised questions of ethnopharmacological differences in response of different populations to androgens (1)(2)(3).
Recent advances in population pharmacokinetic estimation techniques have the potential to liberalize the stringency of sampling frequency. However, a simpler and less invasive mechanism to obtain blood samples would be highly desirable for such field studies. The major noninvasive alternative to repeated blood sampling has been the use of saliva. Salivary immunoassays have been described for steroids including testosterone (4), 17-hydroxyprogesterone (5), other androgens (6), and cortisol (7). However, repeated, frequent collection of saliva may be difficult and suboptimal for remote sampling because salivary samples remain perishable and require handling of bulky tubes. In addition, sonication or centrifugation of samples may increase the apparent concentration of testosterone and other steroids in saliva (8) and, critically, results are invalidated by admixture of even tiny amounts of blood arising from oral abrasions caused by toothbrushing, eating, or dental disease (9)(10). Schramm et al. (10) and Kathol et al. (11) describe oral diffusion sink (ODS) devices for collection of testosterone and cortisol, respectively.1 The study by Schramm et al. (10) in particular clearly shows that their ODS is designed to circumvent contamination with binding proteins. Dabbs et al. (12) has demonstrated problems of reproducibility between centers in salivary testosterone measurement. Finally, the lack of circulating binding proteins in saliva means that salivary steroid concentrations are very low and reflect only free but not total testosterone concentrations. This makes studies of androgen pharmacology, where suppression of endogenous testosterone may be expected, difficult. Thus, saliva is not ideal for remote sampling studies.
An alternative sample matrix is the filter paper spot. Filter spots have been widely used for large centralized population screening programs for neonatal detection of rare but treatable genetic disorders such as phenylketonuria, hypothyroidism, and congenital adrenal hyperplasia (CAH) (5)(13)(14). They have not, however, been evaluated for use in steroid pharmacological studies, although a filter spot testosterone assay with limited validation has been reported (15). We proposed that testosterone and nandrolone concentrations from pharmacological studies could be measured with filter paper blood spots because concentrations would be comparable with those of 17-hydroxyprogesterone concentrations in the screening and therapeutic monitoring studies of CAH (13). Thus we developed and evaluated a system to collect, store, transport, extract, and measure androgens from 15-µL spots of blood dried onto filter paper. In particular, we evaluated the effects on steroid concentrations under adverse environmental conditions simulated in the laboratory as well as in the field.
| Materials and Methods |
|---|
|
|
|---|
Venous blood was collected by venipuncture and plasma stored frozen at -20 °C. In some studies, fresh, unhemolyzed venous blood was spotted directly onto filter paper, allowed to air-dry, and stored with desiccant in sealed containers at -20 °C. Capillary blood spots were collected by fingerprick (Autolet II Lancet; Owen Mumford, Oxon, UK), and the blood spotted directly onto filter paper.
extraction of steroids from plasma and blood spots
Microfuged plasma was extracted by a solid-phase method
(16). If tracer recoveries were required, tritiated
steroid in 10 µL of ethanol was added to the plasma and allowed to
equilibrate for 1 h at 37 °C before extraction. An aliquot of
50 µL (testosterone) or 250 µL (nandrolone) was applied to a glass
Pasteur pipette three-quarter filled with dry kieselguhr
(Extrelut®; E. Merck, Kilsyth, Victoria, Australia) and
the outlet constricted with a 3-mm glass bead. After distribution for
20 min, steroids were eluted with four 750-µL aliquots of
hexane:ethyl acetate (3:2), applied at 5-min intervals. The pooled
eluates were air-dried in borosilicate glass tubes, with gentle warming
(~18 h, 30 °C) in a fume hood.
Steroids were extracted from dried blood spots by punching out one or more spots of 7.9 mm diameter and soaking for 4 h in 3 mL of methanol in a 25 x 100 mm round-bottom glass tube at 38 °C with occasional mixing. The spot was then discarded and the extract air-dried overnight at 30 °C in a fume hood. The extract was redissolved in 1 mL of warm assay buffer and then reextracted on a kieselguhr column before air-drying overnight in a fume hood. The extract was again dissolved in 500 µL of assay buffer with duplicate 200-µL aliquots for assay or for liquid scintillation counting for tracer recovery.
testosterone and nandrolone rias
We used an assay buffer of 100 mmol/L PBS, pH 7.2, containing 1.0
g/L swine gelatin (Sigma, St. Louis, MO) and 1.0 g/L sodium azide.
Dextran-coated charcoal suspension contained 2.5 g/L activated charcoal
(WHO Matched Reagent Program) and 25 mg/L dextran T70 (WHO Matched
Reagent Program) in PBS. Steroid calibrators were obtained from
Steraloids (Wilton, NH) and made up into stock solutions at 1 g/L in
ethanol (Spectrosol grade, Ajax Chemicals, Auburn, NSW, Australia)
stored at 4 °C in a tightly stoppered glass container.
[1,2,6,7,16,17-(N)3H]-Testosterone, 9.25 MBq,
specific activity 5.006.66 TBq/mmol, was purchased from Dupont
Australia, North Ryde, NSW, Australia. [19-3H]Nandrolone,
9.25 MBq, specific activity 1.37 TBq/mmol, was purchased from Amersham
Australia, North Ryde, NSW, Australia. Polyclonal rabbit
antitestosterone-3-(O-carboxymethyl)oxime:bovine serum
albumin (BSA) (SGT-1) from B. Caldwell, Yale University School of
Medicine, New Haven, CT (17), used at a final dilution of
1:14 300, had cross-reactivities [defined as relative molar potency
at ~50% effective dose (ED50)] with nandrolone
(20.5%), 5
-dihydrotestosterone (26.7%), androstenedione (2.5%),
estradiol (0.15%), nandrolone phenylpropionate (0.05%), and
nandrolone decanoate (<0.001%). A rabbit polyclonal antibody to
19-nortestosterone 17-hemisuccinate (18)(19)(20) used at a
final dilution of 1:39 000 had cross-reactivities with testosterone
(0.04%), 5
-dihydrotestosterone (0.01%), estradiol (0.01%),
androstenedione (0.06%), nandrolone phenylpropionate (11.8%), and
nandrolone decanoate (0.3%). The very low cross-reactivity of the
nandrolone antibody with testosterone allowed elimination of
chromatography before immunoassay and was based on the calculation that
the maximum endogenous testosterone concentration likely to be
encountered in this study (40 nmol/L) would contribute only a
negligible apparent nandrolone concentration (0.016 nmol/L) through
cross-reaction compared with the expected nandrolone concentrations of
115 nmol/L (21).
Dried extracts were reconstituted in 0.5 mL of assay buffer with intermittent vortex-mixing at 37 °C for 30 min. Duplicate aliquots (0.1 mL) were transferred into 12 x 75 mm glass assay tubes with additional aliquots taken for internal recovery estimation. Steroid stock calibrator (10 µg/L in ethanol, stored at -20 °C) was diluted serially in assay buffer to make a calibration curve (0.5 to 1000 pg of testosterone/tube; 4 to 1000 pg of nandrolone/tube). Antibodies diluted in assay buffer and tracer [10 000 dpm/tube; ~8 pg (30 fmol) of testosterone or ~30 pg (120 fmol) of nandrolone] were each added in 0.1 mL of buffer. The assay tubes in a total volume of 350 µL were mixed and incubated at 4 °C overnight (16 h) before separation by the addition of 500 µL of ice-cold, constantly stirred dextran-coated charcoal for exactly 15 min at 4 °C. The tubes were then centrifuged (2000g, 30 min, 4 °C) and the supernatant decanted into plastic 7-mL scintillation vials containing 4 mL of scintillation fluid (Ultima Gold XR; Packard Instruments, Meriden, CT). The capped, shaken vials were counted by an automated liquid scintillation counter (RackBeta; LKB, Bromma, Sweden) and RIA data were processed with RIACALC software (Wallac, Turku, Finland).
Assay precision, defined as the within-assay and between-assay CVs, were monitored by the inclusion of control samples containing low, medium, and high concentrations of the analyte. The nandrolone assay had a detection limit of 1.8 pg/tube (equivalent to 0.9 nmol/L) and an ED50 of 64 ± 4 pg/tube (n = 8; equivalent to 28 nmol/L) with tracer recovery of 64.8% ± 8.6% (n = 8). The CVs at low (6.5 pg/tube), mid-range (37 pg/tube), and high (121 pg/tube) concentrations (n = 5 assays) were 66%, 5.2%, and 3.3% (within-assay) and 126%,14.2%, and 3.9% (between-assay). The testosterone assay had a detection limit of 0.9 pg/tube (equivalent to 0.4 nmol/L) and an ED50 of 22 ± 3 pg/tube (n = 5, equivalent to 5.5 nmol/L) with a tracer recovery of 63.0% ± 1.9% (n = 9). The CVs at low (16 pg/tube), mid-range (65 pg/tube), and high (140 pg/tube) concentrations (n = 4 assays) were 86%, 21.3%, and 16.4% (within-assay) and 30.1%,13.1%, and 11.6% (between-assay).
data analysis
An unpaired t-test was used to compare the two paper
types. Unprocessed, filtered, and centrifuged extracts from each paper
type were compared by one-way ANOVA with Dunnett's multiple
comparisons test.
Results of assays in laboratory and field studies were examined with unpaired t-tests or one-way ANOVA with Tukey's multiple comparisons test.
Data from matched series of plasma and venous and capillary blood spot samples obtained during pharmacokinetic studies were examined by either repeated-measures ANOVA (nandrolone; plasma and venous and capillary blood spots) or a paired t-test (testosterone; plasma and venous blood spots). Additionally, the results obtained for the plasma samples were compared with the corresponding blood spot samples with BlandAltman plots (22). Deviation in the BlandAltman plot was calculated as the difference between the values for the blood spot sample and the simultaneous plasma sample, divided by the mean of the two values. Evidence of bias in the BlandAltman plots was sought with McNemar's test (23).
Statistical analyses were conducted with NCSS (Kaysville, UT) statistical software or Instat (GraphPad, San Diego, CA).
| Results |
|---|
|
|
|---|
|
To calibrate the volume of blood impregnated in a filter spot, known volumes of venous blood (30 or 50 µL) were applied to filter paper and air-dried. The area of the blood spot was calculated by cutting out an enlarged copy of the image and weight-comparing with a standard known area (400 mm2). The measured diameter of the punch was 7.9 mm, giving an expected punch area of 49.5 mm2, and the measured area was 52.4 ± 0.4 (n = 5; 106% ± 1% of expected). On the basis of measuring the volume of blood contained in punched spots in 15 separate assays (n = 515 replicates per assay, 9 different blood samples), the volume of blood in a punched spot was 14.8 ± 1.2 µL with a CV of 3.9% (median 4%, range 1.47.5%).
stability of androgens in dried blood spots under laboratory
conditions
The stability of androgens in dried blood spots during transport
was evaluated in the laboratory over 1 or 2 weeks at high temperatures
(35 °C and 50 °C) compared with control samples stored at
-20 °C over silica gel desiccant. Samples for testing at 35 °C
and 50 °C were maintained in closed glass 20-mL vials and stored at
the appropriate temperature in drying ovens. The 1- and 2-week
experiments were staggered to allow extraction and assay of all the
samples together. The uncorrected nandrolone and testosterone values
(pg recovered) did not change significantly during exposure to high
temperature (Table 2
; 35 °C and 50 °C) for up to 2 weeks compared with samples
stored at -20 °C over desiccant gel. In contrast, recovery of
tritiated steroid tracers from blood spots declined progressively
during storage, especially at the higher temperatures, compared with
the control spots stored at -20 °C. These effects were more evident
for nandrolone than for testosterone. As a result of differential
tracer recovery, correcting the androgen concentrations for tracer
recovery led to apparent increases in blood spot nandrolone
concentration within 1 week of storage at 50 °C but not at 35 °C.
The corrected testosterone concentration did not increase at 35 °C
and only increased significantly after 2 weeks at 50 °C.
|
stability of androgens in dried blood spots under field
transport conditions
Stability during transport was assessed by sending samples through
the regular postal system to and from destinations in the north of
Australia. Blood spots sealed individually in plastic packets were sent
by regular post to the destinations and, after several days of storage
at ambient room temperature, were posted back from the destination
through an open-air post box. The parcels were sent in midsummer season
(February) to Townsville (TV), Queensland (146°49'E, 19°16'S) and
to Alice Springs (AS), Northern Territory (133°52'E, 23°42'S).
Weather conditions at the destinations in the monsoon season were, in
TV, hot (daytime >30 °C) and humid; AS was very hot (daytime
>40 °C) and dry. A third set of samples was also sent in midwinter
(June) to AS when the weather conditions were then warm (daytime
~20 °C, <0 °C nighttime) and dry. Conditions in Sydney during
each experiment were similar to AS, although temperatures never fell
below 0 °C. After return between 7 and 20 days, the packets were
stored unopened at -20 °C until steroid assay, together with
controls kept at -20 °C in a dry atmosphere in the laboratory.
In the summer experiment, nandrolone content remained stable after
return from TV but apparently increased after return from AS.
Nandrolone tracer recovery was significantly lower after travel to TV
but not AS. Consequently, after correction for blank and external
tracer recovery, corrected nandrolone concentrations were significantly
higher than controls in spots sent to both destinations. Neither the
raw nor the corrected testosterone values obtained after transportation
differed from controls, although testosterone tracer recovery appeared
to increase after travel to both destinations. The winter experiment,
with milder climatic conditions, demonstrated no significant changes in
cold, tracer, or corrected values for either androgen (Table 3
).
|
correlation of blood spot with plasma androgen concentrations in
a pharmacokinetic study
Venous and capillary blood samples were obtained serially from a
single subject participating in a pharmacokinetic study involving the
intramuscular injection of 100 mg of nandrolone decanoate in 1 mL of
arachis oil (Decadurabolin®; Organon, Sydney, Australia).
At each sampling time point, heparinized venous blood was pipetted onto
filter paper and capillary blood was collected by fingerprick.
Nandrolone (Fig. 1
) and testosterone (Fig. 2
) results were corrected for blank and external tracer recovery.
Blood spot results were corrected for venous hematocrit, to assist in
comparing blood spot and plasma results. Testosterone values were
corrected for the concurrent nandrolone concentration because of the
20.5% cross-reactivity of nandrolone in the testosterone assay. Both
venous and capillary blood spots gave significantly higher nandrolone
readings than plasma samples (P <0.001, repeated-measures
ANOVA), whereas blood spots from venous or capillary blood gave similar
results (P >0.05). BlandAltman plots showed significant
bias only in venous blood spots compared with plasma nandrolone
concentrations (P = 0.0005, McNemar's test).
Testosterone concentrations obtained from plasma and venous blood spots
showed no significant difference (P = 0.11, paired
t-test) nor any systematic bias identified by a
BlandAltman plot.
|
|
| Discussion |
|---|
|
|
|---|
-hydroxyprogesterone immunoassays with filter paper spots, widely
used in screening programs for CAH due to 21-hydroxylase deficiency
(13)(24)(25)(26)(27)(28). This technology, however, has
not yet been evaluated for application to field pharmacological
studies. We anticipated that the expected androgen concentrations in
pharmacological studies were lower than 17
-hydroxyprogesterone
concentrations encountered during diagnostic screening or therapeutic
monitoring for CAH, both of which have been reportedly achieved with
filter spot assays. Our findings with one commercial filter paper, which demonstrated intractable assay blanking problems, indicate that some filter papers may be unsuitable. Extracts from this paper were refractory to filtration or microfugation. The problem probably resulted from liberation of loose cellulose fibers that nonspecifically adsorbed steroids. With suitable filter paper and an organic solvent extraction, impregnation of blood to cover two 7.9-mm diameter circles containing ~30 µL provided sufficient sample to adequately characterize plasma nandrolone or testosterone concentrations. The nandrolone results were less satisfactory, presumably because of the lower sensitivity of that assay; such limitations might be resolved by more sensitive immunoassay techniques such as nonisotopic time-resolved fluorescence labeling (29).
Beyond the validity of the filter spot androgen assays, the stability of the samples to storage and transport under adverse environmental conditions is an important issue regarding the feasibility of applying such a technology to field studies. The stability of androgen concentrations recovered from filter spots was analyzed after short-term laboratory storage at high temperatures as well as under actual field conditions of postage to and from destinations in tropical Northern regions of Australia in the summer monsoon season. In the laboratory, under relatively mild, constant conditions of high and dry heat, the nandrolone and testosterone concentrations extracted from blood spots were stable, although tracer nandrolone recovery declined slightly in spots stored at 35 °C and more so at the higher temperature (50 °C). We have been unable to find an explanation for the lower recovery and presumably higher susceptibility of tritiated steroid tracers to deterioration, but it most probably represents a differential thermal stability of tritiated compared with unlabeled steroids. This differential tracer stability indicates that nandrolone or testosterone concentrations should not be corrected for apparent tracer recovery, and for other analogous systems, the tracer stability should be evaluated critically.
The stability of the androgens in blood spots were also studied under
actual field transport conditions within the regular postal system.
Again we found that unlabeled nandrolone and testosterone were
recovered in relatively constant amounts, but recovery of tritiated
androgens was reduced, especially at the more extreme adverse
environmental conditions. As with the laboratory experiments,
correction for apparent tracer recovery led to artifactual inflation of
apparent steroid concentrations and should be avoided. In contrast,
field transport under milder environmental conditions did not lead to
any deterioration in tracer or sample estimates. It is clear that
androgen concentrations are stable under normal conditions of storage
and transport, but that exposure to extreme conditions over a week or
more have detrimental effects on the stability of the steroids, or on
the sample matrix, leading to measurement inconsistencies. In general,
it appears that if samples can be recovered within a week or maintained
at lower temperatures, stability of the androgens can be expected.
These findings are consistent with the observation that
17
-hydroxyprogesterone concentrations in dried blood spots are
stable in storage at room temperature for 3 weeks (30),
although neither longer duration nor more extreme environmental
conditions were studied. The most extensive studies with dried blood
spots for steroid estimations (13)(14)
involved transport of samples from Alaska to New York, but exposure to
high temperature seems unlikely in that setting. Progesterone in dried
blood spots was found to be stable for up to 15 weeks under dry
conditions at up to 25 °C, and up to 9 weeks at 37 °C, although
experimentally produced humidity was very deleterious
(31). Hofman et al. (32) found
17
-hydroxyprogesterone in dried blood spots to be stable for up to 9
months at similarly moderate temperatures. In the present study,
humidity was excluded as the samples were sealed in air-tight plastic
bags, and sealed containers containing desiccant were used for routine
long-term storage. However, because in the field humidity can influence
dried blood spots in neonatal screening programs for phenylketonuria
(33) and hypothyroidism (34), deleterious
effects on steroid stability cannot be excluded.
These filter spot techniques were then applied in a pilot study of a pharmacokinetic study of androgens. Concurrent samples of plasma, venous blood spots, and capillary blood spots from a volunteer showed a good correspondence between nandrolone concentrations in the venous and capillary spot blood samples. Capillary blood spot results, however, were not significantly different from the plasma concentrations. The testosterone profile obtained from the plasma and the venous spots showed even stronger agreement. The superiority of the testosterone over the nandrolone assay was presumably a reflection of its greater sensitivity; comparable performance could be expected from the nandrolone assay if its sensitivity could be improved while still retaining its high specificity.
We conclude that, under such field conditions as pharmacological or anthropological studies, impregnation of a cellulose-based filter paper with blood is a potentially useful technique for sample collection, storage, and transport of blood to a centralized laboratory for subsequent androgen immunoassay.
| Footnotes |
|---|
1 Nonstandard abbreviations: ODS, oral diffusion sink; CAH, congenital adrenal hyperplasia; and ED50, 50% effective dose. ![]()
| References |
|---|
|
|
|---|
-hydroxyprogesterone assay. J Endocrinol Invest 1982;5:87-90.
[ISI][Medline]
[Order article via Infotrieve]
-hydroxyprogesterone in dried blood samples on filter paper and its application to mass screening for congenital adrenal hyperplasia. Chem Pharm Bull 1983;31:2724-2731.
-hydroxyprogesterone, immunoreactive trypsin, and creatinine kinase isoenzyme in dried blood spots. Clin Chem 1992;38:2038-2043.
[Abstract]
-hydroxyprogesterone radioimmunoassay: its application for rapid screening for congenital adrenal hyperplasia. J Clin Endocrinol Metabol 1977;45:1003-1008.
[ISI][Medline]
[Order article via Infotrieve]
-hydroxyprogesterone in whole-blood samples from newborns. Clin Chem 1985;31:1127-1130.
The following articles in journals at HighWire Press have cited this article:
![]() |
D. Torok, A. Muhl, F. Votava, G. Heinze, J. Solyom, J. Crone, S. Stockler-Ipsiroglu, and F. Waldhauser Stability of 17{alpha}-Hydroxyprogesterone in Dried Blood Spots after Autoclaving and Prolonged Storage Clin. Chem., February 1, 2002; 48(2): 370 - 372. [Full Text] [PDF] |
||||
![]() |
L. Ramakrishnan, K. S. Reddy, and B. L. Jailkhani Measurement of Cholesterol and Triglycerides in Dried Serum and the Effect of Storage Clin. Chem., June 1, 2001; 47(6): 1113 - 1115. [Full Text] [PDF] |
||||
![]() |
S.-H. Peng, J. Segura, M. Farre, and X. de la Torre Oral Testosterone Administration Detected by Testosterone Glucuronidation Measured in Blood Spots Dried on Filter Paper Clin. Chem., April 1, 2000; 46(4): 515 - 522. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |