| Abstract |
|---|
|
|
|---|
6.4% and for 5.54103.95 nmol/L
progesterone were
7.7%, thus meeting published analytical goals. The
detection limits (2 SDs from mean of zero calibrator) were 27.1 pmol/L
for estradiol (n = 72 over 20 days) and 0.51 nmol/L for
progesterone (n = 47 over 20 days). The assays were linear to 9766
pmol/L and 113.0 nmol/L, respectively. Estradiol results agreed well
with the Diagnostic Products Corporation (DPC) assays, except for serum
samples from patients receiving estrogen replacement therapy; results
for these samples agreed closely with the DPC estradiol-6 assay. The
progesterone assay agreed closely with the DPC assay, except for
samples from uremic patients. Reference values were estimated by the
study of 29 women throughout the menstrual cycle with 15 samples per
subject. We concluded that both assays demonstrate suitable precision,
linearity, and intermethod agreement to allow their use in the clinical
laboratory. | Introduction |
|---|
|
|
|---|
Serum estradiol assays are used for the evaluation of gynecomastia or feminization states caused by estrogen-secreting tumors that occur rarely in both males and females of all ages (6)(7). Measurement of this analyte is also useful for the evaluation of precocious puberty and sexual maturation in females (8). Monitoring of estradiol concentration for hormonal replacement therapy is sometimes practiced but has limited use, as other estrogen metabolites are predominant (9)(10).
Assessment of progesterone is important for monitoring ovulation and luteal function and has been proposed as a marker for pregnancy outcome (11). The diagnostic value of progesterone is in the detection of ovulation and the functional evaluation of the corpus luteum. It is also useful for monitoring patient ovulation during induction therapy (11)(12). Progesterone assays can also be used for the surveillance of gestational trophoblastic neoplasia (13) and the prediction of success of in vitro fertilization (14)(15).
Considering the clinical applications of estradiol and progesterone measurements, the assays measuring these analytes need a high sensitivity, an excellent precision, and preferably a wide analytical range.
The Bayer Immuno 1TM system is an automated random access immunoassay analyzer. The estradiol and progesterone assays performed on this analyzer are based on the magnetic separationsequential competitive immunoassay principle. These assays involve a conjugate of analyte and alkaline phosphatase (ALP) that competes with the sample analyte for the binding sites on an anti-analyte antibody conjugated to fluorescein isothiocyanate (FITC).1 Anti-FITC antibodies coupled to magnetic particles allow separation of the immune complexes.
In this study, we evaluated the precision, linearity, and minimum detectable concentration of the Immuno 1 estradiol and progesterone assays. These methods were compared with the Diagnostic Products Corp. (DPC) RIAs. The method comparison study included various categories of serum samples that could interfere with these assays. We also determined the reference ranges for the Immuno 1 estradiol, progesterone, follitropin (FSH), and lutropin (LH) assays throughout the menstrual cycle.
| Materials and Methods |
|---|
|
|
|---|
immuno 1 assays
The Immuno 1 estradiol and progesterone assays are magnetic
separationsequential competitive immunoassays that involve an
anti-analyte antibodyFITC conjugate and the analyte labeled with ALP.
The estradiol assay involves a rabbit polyclonal antibody and the
progesterone assay involves a monoclonal antibody. The Immuno 1 assays
for FSH and LH are magnetic separation sandwich immunoassays that
involve an anti-analyte antibodyFITC conjugate and a second
anti-analyte antibody conjugated to ALP. The Immuno 1 FSH and LH assays
were previously evaluated (16).
The reagent boats (Bayer) for these analytes contain reagents for 100 tests. All the assay steps are automated and performed at 37 °C. For the estradiol and progesterone assays, the samples are incubated with the anti-analyte antibodyFITC conjugate. After 15 min, the analyteALP conjugate is added and incubated for an additional 5 min. For the FSH and LH assays, the samples are incubated simultaneously for 20 min with an anti-analyte antibodyFITC conjugate and an anti-analyte antibodyALP conjugate.
In all assays, the immune complexes formed are incubated for 8 min with magnetic particles coated with anti-FITC antibody. The immune complexes bound to the particles are then washed and incubated with p-nitrophenyl phosphate substrate (pNPP). The absorbance is measured at 405 nm every 30 s for 5 min. The pNPP hydrolysis produces a color inversely proportional to the analyte concentration for the competitive assays and directly proportional for the sandwich assays. The time required to obtain the first sample result is 38 min and 30 additional seconds for each subsequent result.
Six concentrations of calibrators are used for the assay calibrations. The Bayer SETpoint calibrators are supplied in solution for estradiol and must be reconstituted before use for progesterone, FSH, and LH. Common calibrators are supplied for FSH and LH. Each analyte is measured in duplicate for the calibrations. The SETpoint calibrators are stable for at least 30 days at 4 °C when opened. The ranges for the calibration curves are: 0127 nmol/L for progesterone, 011 030 pmol/L for estradiol, 0150 IU/L for FSH, and 0200 IU/L for LH.
dpc estradiol, estradiol-6, and progesterone assays
The following DPC Coat-a-Count assays were used for the method
comparison studies: progesterone, new progesterone, estradiol, and
estradiol-6 assays. DPC changed the antibody used in their progesterone
assay during the course of this study. We compared the Immuno 1
progesterone assay with both kits, which we call "new" and
"old" DPC progesterone assays.
The procedure is the same for all methods. Seven calibrators are tested with each assay. The calibrators, three Lyphochek immunoassay controls (lot 93 000; Bio-Rad Labs.), and serum samples are incubated in anti-analyte antibody-coated tubes in the presence of 125I-labeled progesterone or 125I-labeled estradiol according to the assay. The tubes are incubated at room temperature for 3 h, the supernatants decanted, and the remaining radioactivity in each tube counted for 1 min.
precision studies
The precision of the Immuno 1 estradiol and progesterone assays
was evaluated according to the NCCLS protocol EP5-T2 (17).
Three Bayer TESTpoint ligand controls, three human serum pools at
clinically important estradiol or progesterone concentrations (Bayer),
and the six respective SETpoint calibrators were analyzed in duplicate
over 20 days.
minimum detectable concentration
The minimum detectable concentrations of the Immuno 1 estradiol
and progesterone assays were evaluated with the respective Bayer
SETpoint calibrator 1 (0 nmol/L progesterone or 0 pmol/L estradiol)
during the precision study. These values were obtained by calculating,
from the calibration curve equations, the concentrations corresponding
to the means of the reaction rates minus two total SDs. According to
recent guidelines, these concentrations should correspond to the means
of the reaction rates minus three total SDs; these values were also
calculated.
dilution linearity studies
Six serum samples with high concentrations of analyte were diluted
with the respective SETpoint calibrator 1 at 0%, 20%, 25%, 33%,
50%, and 100% of serum. Each dilution was tested in duplicate. The
statistical evaluations were made by linear regression analysis.
method comparison studies
We used serum samples with increased bilirubin (up to 304
µmol/L), creatinine (up to 1544 µmol/L), rheumatoid factor (>320
kIU/L), immunoglobulins (IgG up to 34.5 g/L or IgA up to 58.5 g/L),
triglycerides (up to 19.8 mmol/L), and hemolysis (up to 12 g/L
hemoglobin), and serum samples from postmenopausal women,
postmenopausal women taking estrogens with or without progestins,
healthy men, diabetics, hemodialysis patients, pregnant women, and
women receiving ovarian stimulation treatment for infertility. The
remaining serum samples tested were chosen by their progesterone or
estradiol concentration to cover the assay analytical range.
The Immuno 1 estradiol assay was compared with the DPC estradiol assay by using 176 samples. The samples that showed the greatest differences between the two assays were also analyzed by the DPC estradiol-6 assay (n = 61). The Immuno 1 progesterone assay has been compared with the old and new DPC progesterone assays by using 185 and 48 samples, respectively. The serum samples were tested by the Immuno 1 assays and by the comparative methods (in single analyses) on the same day. The statistical evaluations were made by linear regression and bias plot analyses.
reference values
Serum from 29 menstrual-cycling women were used to determine the
reference values for estradiol, progesterone, FSH, and LH. Fifteen
blood samples were obtained during a 35-day period at a frequency of
three collections per week. Each volunteer signed a consent form
conforming to the hospital ethics committee criteria and completed a
detailed health questionnaire. The women were not taking anovulant
medication and their ages ranged from 20 to 45 years. All cycles showed
a normal profile ranging from 21 to 31 days.
The longitudinal cycle samples were tested in duplicate by the Immuno 1 estradiol, progesterone, FSH, and LH assays. The follicular, midcycle, and luteal phases were established by designating the highest LH value as day zero. The midcycle phase was represented by the samples containing the highest LH values. The follicular phase included the samples after the onset of menses until the LH surge; the luteal phase followed the LH surge until the first day of the menses. All cycle samples from each woman were used for the statistical determination, except samples that overlap the next cycle. Statistical analyses were carried out by the parametric method according to the NCCLS protocol C28-P (18). The data were previously transformed to a natural logarithm or square root to obtain the coefficient of skewness and kurtosis nearest from zero, determining the best gaussian distribution.
| Results |
|---|
|
|
|---|
|
The within-run CVs ranged from 1.5% to 10.1% and the between-day CVs from 1.7% to 10.0% for progesterone concentrations from 2.66 to 103.95 nmol/L. At the low progesterone concentration of 1.15 nmol/L, within-run and between-day CVs of 23.5% and 22.3% were obtained, and at 1.28 nmol/L, CVs of 15.2% and 15.6% were observed. This concentration is close to the first calibrator containing progesterone (calibrator 2 with 1.15 nmol/L). The observed CVs are consistent with the manufacturer's claims.
minimum detectable concentration
The minimum detectable concentration of the Immuno 1 estradiol
assay was estimated with 72 reaction rates of the SETpoint estradiol
calibrator 1 at 27.1 pmol/L (mean - 2 SDs) and at 35.6 pmol/L
(mean - 3 SDs). For the Immuno 1 progesterone assay, the minimum
detectable concentration was estimated with 47 reaction rates of the
SETpoint progesterone calibrator 1 at 0.51 nmol/L (mean - 2 SDs)
and at 0.75 nmol/L (mean - 3 SDs). As noted above, these
estimates are based on total imprecision (including both within- and
between-run components).
linearity
The results of the dilution linearity studies are shown in Table 2
. The tests to verify the linearity quality were made according
to the NCCLS protocol EP6-P (19). Both assays met
acceptance criteria proposed by this NCCLS protocol.
|
method comparison studies
The relation between the Immuno 1 estradiol and DPC estradiol
assays is shown in Fig. 1
. The regression analysis (Fig. 1A
) shows a slope of 1.20, an
intercept of -6.4 pmol/L, a correlation coefficient of 0.957, and a
Sy|x of 147.5 pmol/L. The 95%
confidence interval shows that the slope is statistically different
from 1.0 (Table 3
). Serum samples from patients receiving estrogen replacement
therapy gave DPC results higher than those measured by the Immuno 1
assay and are excluded from the linear regression analysis. The
cross-reactivity of the antibody used in the DPC estradiol assay with
estrone is well-known (20)(21). Patients
receiving estrogen replacement therapy have already been reported to
have higher estradiol results by the DPC estradiol assay as compared
with other methods (20)(21). The bias plot
representing the relation between the difference and the mean of
measurements obtained with both methods is shown in Fig. 1B
.
|
|
The serum samples that showed a higher residual between Immuno 1 and
DPC estradiol results have been tested with the DPC estradiol-6 assay,
which offers a lower cross-reactivity with estrone than the DPC
estradiol method. The correlation done with 61 samples, including 21
samples from patients receiving estrogen replacement therapy, is
presented in Fig. 2
. The Immuno 1 results obtained for these samples correlate
better with the DPC estradiol-6 results than with DPC estradiol assay.
The regression analysis gives a slope of 0.97, an intercept of 110.6
pmol/L, a correlation coefficient of 0.956, and a
Sy|x of 161.0 pmol/L. The 95%
confidence interval of the intercept is statistically different from
zero (Table 3
).
|
The correlation between the Immuno 1 progesterone and DPC progesterone
assays is shown in Fig. 3
. An excellent correlation is observed after excluding 14 serum
samples from uremic patients that give discrepant results between the
methods. These results are excluded from the linear regression
analysis. The Immuno 1 assay gave slightly higher results than those
measured by DPC assay, with a regression slope of 1.23 and an intercept
of 1.00 nmol/L (Fig. 3A
). A correlation coefficient of 0.982 and a
Sy|x of 6.74 nmol/L were observed.
|
The Immuno 1 progesterone assay was also compared with the new DPC
progesterone method (Fig. 4
). The regression line obtained shows a slope of 1.28 and an
intercept of 1.21 nmol/L (Fig. 4A
). The results from uremic patients
were also excluded from the regression analysis. The correlation
coefficient (r = 0.985) and the
Sy|x (4.77 nmol/L) were similar to
those observed with the old DPC progesterone assay. Both correlation
studies gave slopes that are different from 1.0 according to the 95%
confidence intervals (Table 3
).
|
reference values
The individual hormone profiles from the 29 healthy women showed a
typical menstrual cycle pattern. Two representative normal menstrual
cycle hormone profiles are presented in Figs. 5
and
6. The reference values estimated by the parametric method are
shown in Table 4
.
|
|
| Discussion |
|---|
|
|
|---|
Among the important criteria that influence the desirable analytical performance is the biological variation of a given analyte (32)(33)(34). The recommendation is that an analytical goal for imprecision should be a CV equal to or less than half the within-subject variation for individual single and multipoint measurements. It is interesting to examine the reported studies of biological variation for estradiol and progesterone, even though such studies do not include the possible biological variation for each different clinical circumstance in which the assay might be used. The intraindividual CV for estradiol has been shown to vary from 21.7% to 61.4%, suggesting an analytical goal for CV of 10.9% (34)(35). All estradiol concentrations evaluated except the lower concentration (54.5 pmol/L) gave CVs within this value. The analytical requirements proposed at the Second Estradiol International Workshop are that the CV be <25% for a working range 1501000 pmol/L and <10% for a range of 100010 000 pmol/L (36). The intraindividual CV for progesterone has been reported as 13.09% (37), suggesting an analytical goal of 6.5%. The within-run and between-day CVs are lower than this value for progesterone concentrations >5.5 nmol/L.
The minimum detectable concentration for estradiol is equivalent to (24)(26)(38) or better (25)(26)(38) than other commercial automated estradiol assays. For the progesterone assay, the minimum detectable concentration is also equivalent to other commercial automated progesterone assays (27)(28)(30)(31).
The dilution studies performed for the Immuno 1 estradiol or progesterone assays with calibrator 1 are compatible with acceptance criteria proposed by the NCCLS protocol EP6-P (19). The percentages of recovery and the linear regression analysis demonstrate the excellent linearity of the assays throughout the entire analytical range. No matrix effect was apparent with the manufacturer's recommended diluent.
The comparison between the Immuno 1 and DPC estradiol assays shows a good correlation between the two methods, although the results obtained from the Immuno 1 assay are slightly higher than those measured by the DPC assay (regression analysis slope statistically different from 1.0). Samples from patients taking estrogens gave DPC results higher than those obtained from the Immuno 1 assay. Estrone has been shown to interfere with the DPC method (20)(21)(39). Because estradiol undergoes first-pass hepatic metabolism to estrone and its conjugated metabolites in patients receiving replacement therapy, the increased DPC values represent cross-reactivity with nonphysiological concentrations of estrone and these metabolites. The results obtained from the Immuno 1 assay correlate better with the DPC estradiol-6 results. This assay is commercialized as having a low cross-reactivity with estrone (40). These results confirm that the Immuno 1 assay has a low cross-reactivity with estrone.
The results obtained from the Immuno 1 progesterone assay correlate well with those measured by both old and new DPC assays, although the results obtained from the Immuno 1 assay are higher than those obtained by the DPC methods (regression analysis slopes statistically different from 1.0). The serum samples from uremic patients give higher values with the Immuno 1 progesterone assay. Interference by these serum samples with the Immuno 1 assay is suggested by finding progesterone values in males more than three times the upper limit of the reference interval.
Many other categories of serum samples were tested in these correlation studies. For both analytes, no significant difference between the Immuno 1 assays and comparative assays was observed in patient serum samples with increased bilirubin, rheumatoid factor, immunoglobulins or triglycerides, presence of hemolysis, serum samples from postmenopausal women without hormone replacement therapy, healthy men, diabetic patients, pregnant women, or patients receiving stimulation treatment for infertility. In addition, no significant difference was observed with serum samples from uremic patients between Immuno 1 and DPC estradiol methods, and with serum samples from postmenopausal women with hormone replacement therapy between Immuno 1 and DPC progesterone methods.
The regression analysis of correlation studies gave slopes significantly different from 1.0, except for the correlation performed with the DPC estradiol-6 assay, suggesting differences in assay calibration. Indeed, large differences between commercial methods used in routine laboratories for the measurements of these analytes have been reported (38)(41)(42). Consequently, great variability currently exists among laboratories (43)(44). There is a need for a better calibration of these hormone assays between different commercial methods for the same analyte. In addition, the establishment of reference values specific for each method is important for an optimal result interpretation, especially in view of the differences that can exist between commercially available assays. The reference ranges determined for the Immuno 1 assays by the statistical parametric analysis are consistent with literature values (39)(41)(45).
The fully automated Immuno 1 assays for progesterone and estradiol in serum have the technical requirements for their utilization in the routine laboratory. These assays are rapid, with the first result obtained in 38 min, and 120 additional results able to be processed in the following hour. In addition to allowing the use of nonradioactive reagents, the calibration curve stability of the Immuno 1 assay is at least 2 months, in contrast to the RIA, which requires a new calibration curve in each protocol. Both Immuno 1 assays offer an extended range, allowing minimal number of sample dilutions.
In conclusion, this evaluation of the Immuno 1 estradiol and progesterone assays demonstrates several performance characteristics that contribute to increasing the reliability of these assays. Excellent precision and reproducibility, very good linearity, intermethod correlation, and minimal interference are important criteria for their use in the clinical laboratory.
|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
P. M. Lam, C. Briton-Jones, C. K. Cheung, L. S. Po, L. P. Cheung, and C. Haines Increased mRNA expression of vascular endothelial growth factor and its receptor (flt-1) in the hydrosalpinx Hum. Reprod., November 1, 2003; 18(11): 2264 - 2269. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Lam, C. Briton-Jones, C. K. Cheung, I. H. Lok, P. M. Yuen, L. P. Cheung, and C. Haines Vascular Endothelial Growth Factor in the Human Oviduct: Localization and Regulation of Messenger RNA Expression In Vivo Biol Reprod, May 1, 2003; 68(5): 1870 - 1876. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Rinaldi, H. Dechaud, C. Biessy, V. Morin-Raverot, P. Toniolo, A. Zeleniuch-Jacquotte, A. Akhmedkhanov, R. E. Shore, G. Secreto, A. Ciampi, et al. Reliability and Validity of Commercially Available, Direct Radioimmunoassays for Measurement of Blood Androgens and Estrogens in Postmenopausal Women Cancer Epidemiol. Biomarkers Prev., July 1, 2001; 10(7): 757 - 765. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |