Clinical Chemistry 43: 1601-1609, 1997;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (12)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lévesque, A.
Right arrow Articles by Grant, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lévesque, A.
Right arrow Articles by Grant, A.
Related Collections
Right arrow Laboratory Management
Right arrow Endocrinology and Metabolism
Right arrow Automation and Analytical Techniques
(Clinical Chemistry. 1997;43:1601-1609.)
© 1997 American Association for Clinical Chemistry, Inc.


Articles

Analytical performance of Bayer Immuno 1TM estradiol and progesterone assays

Ann Lévesque1,a, Marc Letellier2, Paul W. Dillon3 and Andrew Grant1

1 Centre for Research and Evaluation in Immunodiagnostics, Department of Clinical Biochemistry, and
2 Department of Clinical Biochemistry, Centre universitaire de santé de l'Estrie, Site Fleurimont, 3001, 12e Ave. Nord, Sherbrooke, QC, Canada J1H 5N4.

3 Bayer Corporation, Business Group Diagnostics, 511 Benedict Ave., Tarrytown, NY 10591-5097.
a Address correspondence to this author at: Immunova Ltd., 2750 Einstein, Suite 110, Sainte-Foy, QC, Canada G1P 4R1. Fax 418-656-0219.


   Abstract
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
We evaluated the analytical performance of new estradiol and progesterone assays performed on the Bayer Immuno 1TMsystem. Within-run and between-day CVs for estradiol at concentrations of 116.8–6645.8 pmol/L were <=6.4% and for 5.54–103.95 nmol/L progesterone were <=7.7%, thus meeting published analytical goals. The detection limits (2 SDs from mean of zero calibrator) were 27.1 pmol/L for estradiol (n = 72 over 20 days) and 0.51 nmol/L for progesterone (n = 47 over 20 days). The assays were linear to 9766 pmol/L and 113.0 nmol/L, respectively. Estradiol results agreed well with the Diagnostic Products Corporation (DPC) assays, except for serum samples from patients receiving estrogen replacement therapy; results for these samples agreed closely with the DPC estradiol-6 assay. The progesterone assay agreed closely with the DPC assay, except for samples from uremic patients. Reference values were estimated by the study of 29 women throughout the menstrual cycle with 15 samples per subject. We concluded that both assays demonstrate suitable precision, linearity, and intermethod agreement to allow their use in the clinical laboratory.


   Introduction
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
The determination of estradiol and progesterone concentrations in serum is useful for the evaluation of fertility and menstrual problems in women. Serial measurements of estradiol allow the monitoring of follicular growth and predict the day of ovulation during a regular menstrual cycle (1) and ovarian stimulation therapy (2). Serum estradiol has been shown to correlate with the diameter of the leading follicle (1)(3). A relation between serum estradiol and the number of preovulatory follicles, oocytes, and embryos has also been found (4). The measurement of serum estradiol is used to prevent ovarian overstimulation syndrome (5).

Serum estradiol assays are used for the evaluation of gynecomastia or feminization states caused by estrogen-secreting tumors that occur rarely in both males and females of all ages (6)(7). Measurement of this analyte is also useful for the evaluation of precocious puberty and sexual maturation in females (8). Monitoring of estradiol concentration for hormonal replacement therapy is sometimes practiced but has limited use, as other estrogen metabolites are predominant (9)(10).

Assessment of progesterone is important for monitoring ovulation and luteal function and has been proposed as a marker for pregnancy outcome (11). The diagnostic value of progesterone is in the detection of ovulation and the functional evaluation of the corpus luteum. It is also useful for monitoring patient ovulation during induction therapy (11)(12). Progesterone assays can also be used for the surveillance of gestational trophoblastic neoplasia (13) and the prediction of success of in vitro fertilization (14)(15).

Considering the clinical applications of estradiol and progesterone measurements, the assays measuring these analytes need a high sensitivity, an excellent precision, and preferably a wide analytical range.

The Bayer Immuno 1TM system is an automated random access immunoassay analyzer. The estradiol and progesterone assays performed on this analyzer are based on the magnetic separation–sequential competitive immunoassay principle. These assays involve a conjugate of analyte and alkaline phosphatase (ALP) that competes with the sample analyte for the binding sites on an anti-analyte antibody conjugated to fluorescein isothiocyanate (FITC).1 Anti-FITC antibodies coupled to magnetic particles allow separation of the immune complexes.

In this study, we evaluated the precision, linearity, and minimum detectable concentration of the Immuno 1 estradiol and progesterone assays. These methods were compared with the Diagnostic Products Corp. (DPC) RIAs. The method comparison study included various categories of serum samples that could interfere with these assays. We also determined the reference ranges for the Immuno 1 estradiol, progesterone, follitropin (FSH), and lutropin (LH) assays throughout the menstrual cycle.


   Materials and Methods
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
immuno 1 system
The Bayer (Business Group Diagnostics) Immuno 1 system is a fully automated immunoassay analyzer that can analyze in batch, random, or stat modes. This analyzer is composed of a reagent, a substrate, a sample, a buffer and immunomagnetic particle, and a reaction compartment. The Immuno 1 system also includes a workstation for system programming.

immuno 1 assays
The Immuno 1 estradiol and progesterone assays are magnetic separation–sequential competitive immunoassays that involve an anti-analyte antibody–FITC conjugate and the analyte labeled with ALP. The estradiol assay involves a rabbit polyclonal antibody and the progesterone assay involves a monoclonal antibody. The Immuno 1 assays for FSH and LH are magnetic separation sandwich immunoassays that involve an anti-analyte antibody–FITC conjugate and a second anti-analyte antibody conjugated to ALP. The Immuno 1 FSH and LH assays were previously evaluated (16).

The reagent boats (Bayer) for these analytes contain reagents for 100 tests. All the assay steps are automated and performed at 37 °C. For the estradiol and progesterone assays, the samples are incubated with the anti-analyte antibody–FITC conjugate. After 15 min, the analyte–ALP conjugate is added and incubated for an additional 5 min. For the FSH and LH assays, the samples are incubated simultaneously for 20 min with an anti-analyte antibody–FITC conjugate and an anti-analyte antibody–ALP conjugate.

In all assays, the immune complexes formed are incubated for 8 min with magnetic particles coated with anti-FITC antibody. The immune complexes bound to the particles are then washed and incubated with p-nitrophenyl phosphate substrate (pNPP). The absorbance is measured at 405 nm every 30 s for 5 min. The pNPP hydrolysis produces a color inversely proportional to the analyte concentration for the competitive assays and directly proportional for the sandwich assays. The time required to obtain the first sample result is 38 min and 30 additional seconds for each subsequent result.

Six concentrations of calibrators are used for the assay calibrations. The Bayer SETpoint calibrators are supplied in solution for estradiol and must be reconstituted before use for progesterone, FSH, and LH. Common calibrators are supplied for FSH and LH. Each analyte is measured in duplicate for the calibrations. The SETpoint calibrators are stable for at least 30 days at 4 °C when opened. The ranges for the calibration curves are: 0–127 nmol/L for progesterone, 0–11 030 pmol/L for estradiol, 0–150 IU/L for FSH, and 0–200 IU/L for LH.

dpc estradiol, estradiol-6, and progesterone assays
The following DPC Coat-a-Count assays were used for the method comparison studies: progesterone, new progesterone, estradiol, and estradiol-6 assays. DPC changed the antibody used in their progesterone assay during the course of this study. We compared the Immuno 1 progesterone assay with both kits, which we call "new" and "old" DPC progesterone assays.

The procedure is the same for all methods. Seven calibrators are tested with each assay. The calibrators, three Lyphochek immunoassay controls (lot 93 000; Bio-Rad Labs.), and serum samples are incubated in anti-analyte antibody-coated tubes in the presence of 125I-labeled progesterone or 125I-labeled estradiol according to the assay. The tubes are incubated at room temperature for 3 h, the supernatants decanted, and the remaining radioactivity in each tube counted for 1 min.

precision studies
The precision of the Immuno 1 estradiol and progesterone assays was evaluated according to the NCCLS protocol EP5-T2 (17). Three Bayer TESTpoint ligand controls, three human serum pools at clinically important estradiol or progesterone concentrations (Bayer), and the six respective SETpoint calibrators were analyzed in duplicate over 20 days.

minimum detectable concentration
The minimum detectable concentrations of the Immuno 1 estradiol and progesterone assays were evaluated with the respective Bayer SETpoint calibrator 1 (0 nmol/L progesterone or 0 pmol/L estradiol) during the precision study. These values were obtained by calculating, from the calibration curve equations, the concentrations corresponding to the means of the reaction rates minus two total SDs. According to recent guidelines, these concentrations should correspond to the means of the reaction rates minus three total SDs; these values were also calculated.

dilution linearity studies
Six serum samples with high concentrations of analyte were diluted with the respective SETpoint calibrator 1 at 0%, 20%, 25%, 33%, 50%, and 100% of serum. Each dilution was tested in duplicate. The statistical evaluations were made by linear regression analysis.

method comparison studies
We used serum samples with increased bilirubin (up to 304 µmol/L), creatinine (up to 1544 µmol/L), rheumatoid factor (>320 kIU/L), immunoglobulins (IgG up to 34.5 g/L or IgA up to 58.5 g/L), triglycerides (up to 19.8 mmol/L), and hemolysis (up to 12 g/L hemoglobin), and serum samples from postmenopausal women, postmenopausal women taking estrogens with or without progestins, healthy men, diabetics, hemodialysis patients, pregnant women, and women receiving ovarian stimulation treatment for infertility. The remaining serum samples tested were chosen by their progesterone or estradiol concentration to cover the assay analytical range.

The Immuno 1 estradiol assay was compared with the DPC estradiol assay by using 176 samples. The samples that showed the greatest differences between the two assays were also analyzed by the DPC estradiol-6 assay (n = 61). The Immuno 1 progesterone assay has been compared with the old and new DPC progesterone assays by using 185 and 48 samples, respectively. The serum samples were tested by the Immuno 1 assays and by the comparative methods (in single analyses) on the same day. The statistical evaluations were made by linear regression and bias plot analyses.

reference values
Serum from 29 menstrual-cycling women were used to determine the reference values for estradiol, progesterone, FSH, and LH. Fifteen blood samples were obtained during a 35-day period at a frequency of three collections per week. Each volunteer signed a consent form conforming to the hospital ethics committee criteria and completed a detailed health questionnaire. The women were not taking anovulant medication and their ages ranged from 20 to 45 years. All cycles showed a normal profile ranging from 21 to 31 days.

The longitudinal cycle samples were tested in duplicate by the Immuno 1 estradiol, progesterone, FSH, and LH assays. The follicular, midcycle, and luteal phases were established by designating the highest LH value as day zero. The midcycle phase was represented by the samples containing the highest LH values. The follicular phase included the samples after the onset of menses until the LH surge; the luteal phase followed the LH surge until the first day of the menses. All cycle samples from each woman were used for the statistical determination, except samples that overlap the next cycle. Statistical analyses were carried out by the parametric method according to the NCCLS protocol C28-P (18). The data were previously transformed to a natural logarithm or square root to obtain the coefficient of skewness and kurtosis nearest from zero, determining the best gaussian distribution.


   Results
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
precision studies
The statistical analyses of the precision studies are shown in Table 1 . The within-run CVs ranged from 1.9% to 6.4% and the between-day CVs from 1.7% to 5.0% for estradiol concentrations from 116.8 to 6645.8 pmol/L. At the lowest evaluated concentration of 54.5 pmol/L, within-run and between-day CVs of 17.9% and 13.8% were observed. This concentration is below the first calibrator containing estradiol (calibrator 2 with 116.8 pmol/L), and thus was not used further in evaluation.


View this table:
[in this window]
[in a new window]
 
Table 1. Within-run and between-day precision of the Immuno 1 estradiol and progesterone assays.

The within-run CVs ranged from 1.5% to 10.1% and the between-day CVs from 1.7% to 10.0% for progesterone concentrations from 2.66 to 103.95 nmol/L. At the low progesterone concentration of 1.15 nmol/L, within-run and between-day CVs of 23.5% and 22.3% were obtained, and at 1.28 nmol/L, CVs of 15.2% and 15.6% were observed. This concentration is close to the first calibrator containing progesterone (calibrator 2 with 1.15 nmol/L). The observed CVs are consistent with the manufacturer's claims.

minimum detectable concentration
The minimum detectable concentration of the Immuno 1 estradiol assay was estimated with 72 reaction rates of the SETpoint estradiol calibrator 1 at 27.1 pmol/L (mean - 2 SDs) and at 35.6 pmol/L (mean - 3 SDs). For the Immuno 1 progesterone assay, the minimum detectable concentration was estimated with 47 reaction rates of the SETpoint progesterone calibrator 1 at 0.51 nmol/L (mean - 2 SDs) and at 0.75 nmol/L (mean - 3 SDs). As noted above, these estimates are based on total imprecision (including both within- and between-run components).

linearity
The results of the dilution linearity studies are shown in Table 2 . The tests to verify the linearity quality were made according to the NCCLS protocol EP6-P (19). Both assays met acceptance criteria proposed by this NCCLS protocol.


View this table:
[in this window]
[in a new window]
 
Table 2. Linearity of the Immuno 1 estradiol and progesterone assays.

method comparison studies
The relation between the Immuno 1 estradiol and DPC estradiol assays is shown in Fig. 1 . The regression analysis (Fig. 1A ) shows a slope of 1.20, an intercept of -6.4 pmol/L, a correlation coefficient of 0.957, and a Sy|x of 147.5 pmol/L. The 95% confidence interval shows that the slope is statistically different from 1.0 (Table 3 ). Serum samples from patients receiving estrogen replacement therapy gave DPC results higher than those measured by the Immuno 1 assay and are excluded from the linear regression analysis. The cross-reactivity of the antibody used in the DPC estradiol assay with estrone is well-known (20)(21). Patients receiving estrogen replacement therapy have already been reported to have higher estradiol results by the DPC estradiol assay as compared with other methods (20)(21). The bias plot representing the relation between the difference and the mean of measurements obtained with both methods is shown in Fig. 1B .



View larger version (23K):
[in this window]
[in a new window]
 
Figure 1. Comparison between Immuno 1 estradiol and DPC estradiol assays.

{circ}, serum samples from patients taking estrogen therapy (n = 26). These results are excluded from the analyses: (A) linear regression and (B) bias plot.


View this table:
[in this window]
[in a new window]
 
Table 3. Regression analyses of relations between the Immuno 1 and comparative estradiol and progesterone assays.

The serum samples that showed a higher residual between Immuno 1 and DPC estradiol results have been tested with the DPC estradiol-6 assay, which offers a lower cross-reactivity with estrone than the DPC estradiol method. The correlation done with 61 samples, including 21 samples from patients receiving estrogen replacement therapy, is presented in Fig. 2 . The Immuno 1 results obtained for these samples correlate better with the DPC estradiol-6 results than with DPC estradiol assay. The regression analysis gives a slope of 0.97, an intercept of 110.6 pmol/L, a correlation coefficient of 0.956, and a Sy|x of 161.0 pmol/L. The 95% confidence interval of the intercept is statistically different from zero (Table 3Up ).



View larger version (21K):
[in this window]
[in a new window]
 
Figure 2. Comparison between Immuno 1 estradiol and DPC estradiol-6 assays.

{circ}, serum samples from patients taking estrogen therapy (n = 21). These results are included in the analyses: (A) linear regression and (B) bias plot.

The correlation between the Immuno 1 progesterone and DPC progesterone assays is shown in Fig. 3 . An excellent correlation is observed after excluding 14 serum samples from uremic patients that give discrepant results between the methods. These results are excluded from the linear regression analysis. The Immuno 1 assay gave slightly higher results than those measured by DPC assay, with a regression slope of 1.23 and an intercept of 1.00 nmol/L (Fig. 3A ). A correlation coefficient of 0.982 and a Sy|x of 6.74 nmol/L were observed.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 3. Comparison between Immuno 1 progesterone and old DPC progesterone assays.

Inset shows the correlation at low progesterone concentrations. {circ}, serum samples from uremic patients (n = 14). These results are excluded from the analyses: (A) linear regression and (B) bias plot.

The Immuno 1 progesterone assay was also compared with the new DPC progesterone method (Fig. 4 ). The regression line obtained shows a slope of 1.28 and an intercept of 1.21 nmol/L (Fig. 4A ). The results from uremic patients were also excluded from the regression analysis. The correlation coefficient (r = 0.985) and the Sy|x (4.77 nmol/L) were similar to those observed with the old DPC progesterone assay. Both correlation studies gave slopes that are different from 1.0 according to the 95% confidence intervals (Table 3Up ).



View larger version (19K):
[in this window]
[in a new window]
 
Figure 4. Comparison between Immuno 1 progesterone and new DPC progesterone assays.

{circ}, serum samples from uremic patients (n = 3). These results are excluded from the analyses: (A) linear regression and (B) bias plot.

reference values
The individual hormone profiles from the 29 healthy women showed a typical menstrual cycle pattern. Two representative normal menstrual cycle hormone profiles are presented in Figs. 5 and 6. The reference values estimated by the parametric method are shown in Table 4 .



View larger version (18K):
[in this window]
[in a new window]
 
Figure 5. Changes in serum concentrations of FSH (•, IU/L), LH ({circ}, IU/L), estradiol ({blacksquare}, pmol/L), and progesterone ({square}, nmol/L) throughout the regular menstrual cycle of patient 1.


View this table:
[in this window]
[in a new window]
 
Table 4. Reference values for Immuno 1 FSH, LH, estradiol, and progesterone assays.


   Discussion
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 
The within-run and between-day CVs were excellent for both estradiol and progesterone assays, except for concentrations <2.66 nmol/L for progesterone and <54.5 pmol/L for estradiol, which gave CVs >10%. The precision performance of the Immuno 1 estradiol assay is superior to the semi- or fully automated assays reported (22)(23)(24)(25)(26). The performance of the Immuno 1 progesterone assay is also better than other reported automated progesterone assays (27)(28)(29)(30)(31).

Among the important criteria that influence the desirable analytical performance is the biological variation of a given analyte (32)(33)(34). The recommendation is that an analytical goal for imprecision should be a CV equal to or less than half the within-subject variation for individual single and multipoint measurements. It is interesting to examine the reported studies of biological variation for estradiol and progesterone, even though such studies do not include the possible biological variation for each different clinical circumstance in which the assay might be used. The intraindividual CV for estradiol has been shown to vary from 21.7% to 61.4%, suggesting an analytical goal for CV of 10.9% (34)(35). All estradiol concentrations evaluated except the lower concentration (54.5 pmol/L) gave CVs within this value. The analytical requirements proposed at the Second Estradiol International Workshop are that the CV be <25% for a working range 150–1000 pmol/L and <10% for a range of 1000–10 000 pmol/L (36). The intraindividual CV for progesterone has been reported as 13.09% (37), suggesting an analytical goal of 6.5%. The within-run and between-day CVs are lower than this value for progesterone concentrations >5.5 nmol/L.

The minimum detectable concentration for estradiol is equivalent to (24)(26)(38) or better (25)(26)(38) than other commercial automated estradiol assays. For the progesterone assay, the minimum detectable concentration is also equivalent to other commercial automated progesterone assays (27)(28)(30)(31).

The dilution studies performed for the Immuno 1 estradiol or progesterone assays with calibrator 1 are compatible with acceptance criteria proposed by the NCCLS protocol EP6-P (19). The percentages of recovery and the linear regression analysis demonstrate the excellent linearity of the assays throughout the entire analytical range. No matrix effect was apparent with the manufacturer's recommended diluent.

The comparison between the Immuno 1 and DPC estradiol assays shows a good correlation between the two methods, although the results obtained from the Immuno 1 assay are slightly higher than those measured by the DPC assay (regression analysis slope statistically different from 1.0). Samples from patients taking estrogens gave DPC results higher than those obtained from the Immuno 1 assay. Estrone has been shown to interfere with the DPC method (20)(21)(39). Because estradiol undergoes first-pass hepatic metabolism to estrone and its conjugated metabolites in patients receiving replacement therapy, the increased DPC values represent cross-reactivity with nonphysiological concentrations of estrone and these metabolites. The results obtained from the Immuno 1 assay correlate better with the DPC estradiol-6 results. This assay is commercialized as having a low cross-reactivity with estrone (40). These results confirm that the Immuno 1 assay has a low cross-reactivity with estrone.

The results obtained from the Immuno 1 progesterone assay correlate well with those measured by both old and new DPC assays, although the results obtained from the Immuno 1 assay are higher than those obtained by the DPC methods (regression analysis slopes statistically different from 1.0). The serum samples from uremic patients give higher values with the Immuno 1 progesterone assay. Interference by these serum samples with the Immuno 1 assay is suggested by finding progesterone values in males more than three times the upper limit of the reference interval.

Many other categories of serum samples were tested in these correlation studies. For both analytes, no significant difference between the Immuno 1 assays and comparative assays was observed in patient serum samples with increased bilirubin, rheumatoid factor, immunoglobulins or triglycerides, presence of hemolysis, serum samples from postmenopausal women without hormone replacement therapy, healthy men, diabetic patients, pregnant women, or patients receiving stimulation treatment for infertility. In addition, no significant difference was observed with serum samples from uremic patients between Immuno 1 and DPC estradiol methods, and with serum samples from postmenopausal women with hormone replacement therapy between Immuno 1 and DPC progesterone methods.

The regression analysis of correlation studies gave slopes significantly different from 1.0, except for the correlation performed with the DPC estradiol-6 assay, suggesting differences in assay calibration. Indeed, large differences between commercial methods used in routine laboratories for the measurements of these analytes have been reported (38)(41)(42). Consequently, great variability currently exists among laboratories (43)(44). There is a need for a better calibration of these hormone assays between different commercial methods for the same analyte. In addition, the establishment of reference values specific for each method is important for an optimal result interpretation, especially in view of the differences that can exist between commercially available assays. The reference ranges determined for the Immuno 1 assays by the statistical parametric analysis are consistent with literature values (39)(41)(45).

The fully automated Immuno 1 assays for progesterone and estradiol in serum have the technical requirements for their utilization in the routine laboratory. These assays are rapid, with the first result obtained in 38 min, and 120 additional results able to be processed in the following hour. In addition to allowing the use of nonradioactive reagents, the calibration curve stability of the Immuno 1 assay is at least 2 months, in contrast to the RIA, which requires a new calibration curve in each protocol. Both Immuno 1 assays offer an extended range, allowing minimal number of sample dilutions.

In conclusion, this evaluation of the Immuno 1 estradiol and progesterone assays demonstrates several performance characteristics that contribute to increasing the reliability of these assays. Excellent precision and reproducibility, very good linearity, intermethod correlation, and minimal interference are important criteria for their use in the clinical laboratory.



View larger version (18K):
[in this window]
[in a new window]
 
Figure 6. Changes in serum concentrations of FSH (•, IU/L), LH ({circ}, IU/L), estradiol ({blacksquare}, pmol/L), and progesterone ({square}, nmol/L) throughout the regular menstrual cycle of patient 2.


   Acknowledgments
 
We thank Bayer Corporation, Business Group Diagnostics, for providing the reagents and the financial support for this work, through the industry partnership program of the Fonds de la Recherche en Santé du Québec.


   Footnotes
 
1 Nonstandard abbreviations: ALP, alkaline phosphatase; FITC, fluorescein isothiocyanate; FSH, follitropin; LH, lutropin; and pNPP, p-nitrophenyl phosphate.


   References
Top
Abstract
Introduction
Materials and Methods
Results
Discussion
References
 

  1. Balos O, Lundkvist O, Wide L, Bergh T. Ultrasonographical and hormonal description of the normal ovulatory menstrual cycle. Acta Obstet Gynecol Scand 1994;73:790-796. [ISI][Medline] [Order article via Infotrieve]
  2. Muse MD, Wilson EA. Monitoring ovulation: use of biochemical and biophysical parameters. Semin Reprod Endocrinol 1986;4:301-309.
  3. Fossum Gt, Vermesh M, Kletzky OA. Biochemical and biophysical indices of follicular development in spontaneous and stimulated ovulatory cycles. Obstet Gynecol 1990;75:407-411. [Abstract/Free Full Text]
  4. Sushanek E, Huderer K, Dobec D, Hlavati V, Simunic V, Grizelj V. Number of follicles, oocytes and embryos in human in vitro fertilization is relative to serum estradiol and progesterone patterns during different types of ovarian hyperstimulation. Eur J Obstet Gynecol Reprod Biol 1994;56:121-127. [ISI][Medline] [Order article via Infotrieve]
  5. Navot D, Bergh PA, Laufer N. Ovarian hyperstimulation syndrome in novel reproductive technologies: prevention and treatment. Fertil Steril 1992;58:249-261. [ISI][Medline] [Order article via Infotrieve]
  6. Kuhn JM, Mahoudeau JA, Billaud L, Joly J, Rieu M, Gancel A, et al. Evaluation of diagnostic criteria for Leydig cell tumours in adult men revealed by gynaecomastia. Clin Endocrinol (Oxf) 1987;26:407-416. [Medline] [Order article via Infotrieve]
  7. Valensi P, Coussieu C, Pauwels A, Attali JR, Kemeny JL, Amouroux J, Sebaoun J. Feminizing Leydig cell tumor: endocrine and incubation studies. J Endocrinol Invest 1987;10:187-193. [ISI][Medline] [Order article via Infotrieve]
  8. Garibaldi LR, Aceto T, Weber C. The relationship between luteinizing hormone and estradiol secretion in female precocious puberty: evaluation by sensitive gonadotropin assays and the leuprolide stimulation test. J Clin Endocrinol Metab 1993;76:851-856. [Abstract]
  9. Jamin C. Surveillance biologique du traitement hormonal substitutif: se fier aux dosages plasmatiques d'estradiol ou aux produits utilisés?. J Gynecol Obstet Biol Reprod 1995;24:467-474. [Medline] [Order article via Infotrieve]
  10. Cook NJ, Read GF. Oestradiol measurement in women on oral hormone replacement therapy: the validity of commercial test kits. Br J Biomed Sci 1995;52:97-101. [ISI][Medline] [Order article via Infotrieve]
  11. Cowan BD, Vandermolen DT, Long CA, Whitworth NS. Receiver-operator characteristic, efficiency analysis, and predictive value of serum progesterone concentration as a test for abnormal gestations. Am J Obstet Gynecol 1992;166:1729-1734. [ISI][Medline] [Order article via Infotrieve]
  12. Long CA, Lincoln SR, Whitworth NS, Cowan BD. Serum progesterone predicts abnormal gestations in chlomiphene citrate conception cycles as well as in spontaneous conception cycles. Fertil Steril 1994;61:838-842. [ISI][Medline] [Order article via Infotrieve]
  13. Vasilev SA, Schlaerth JB, Ahn C, Sauer MV. Serum progesterone monitoring in post-molar surveillance. Gynecol Oncol 1994;54:288-291. [ISI][Medline] [Order article via Infotrieve]
  14. Silverberg KM, Burns WN, Olive DL, Riehl RM, Schenken RS. Serum progesterone levels predict success of in vitro fertilization/embryo transfer in patients stimulated with leuprolide acetate and human menopausal gonadotropins. J Clin Endocrinol Metab 1991;73:797-803. [Abstract]
  15. Prien SD, Canez MS, Messer RH. Increases in progesterone after human chorionic gonadotropin administration may predict cycle outcome in patients undergoing in vitro fertilization–embryo transfer. Fertil Steril 1994;62:1066-1068. [ISI][Medline] [Order article via Infotrieve]
  16. Letellier M, Lévesque A, Daigle F, Grant A. Performance evaluation of automated immunoassays on the Technicon Immuno 1 system. Clin Chem 1996;42:1695-1701. [Abstract/Free Full Text]
  17. NCCLS document EP5–T2. Evaluation of precision performance of clinical chemistry devices. Villanova, PA: National Committee for Clinical Laboratory Standards, 1992..
  18. NCCLS Document C28-P. How to define, determine and utilize reference intervals in the clinical laboratory. Villanova, PA: National Committee for Clinical Laboratory Standards, 1992..
  19. NCCLS proposed guideline EP6-P. Evaluation of the linearity of quantitative analytical methods. Villanova, PA: National Committee for Clinical Laboratory Standards, 1992..
  20. Swanson KH, Arnold LL, Haven MC. Evaluation of estradiol by the IMxTM microparticle enzyme immunoassay [Abstract]. Clin Chem 1993;39:1164.
  21. Pierard D, Van Waesberghe L, Smitz J, Van Steirteghem AC. Evaluation of five commercially available direct radio-immunoassays for the determination of 17ß-estradiol in serum for the hormonal monitoring of ovarian stimulation. Ann Biol Clin 1987;45:444-448.
  22. Noel SA. Analytical characteristics and clinical applications for a chemiluminescent immunoassay (Amerlite) for serum estradiol [Abstract]. Clin Chem 1992;38:947.
  23. Ooi DS, Laxdal V, Pronovost C. Two commercial automated immunoenzymometric assays for serum estradiol evaluated [Tech Brief]. Clin Chem 1994;40:147-148. [Free Full Text]
  24. Rodriguez LC, Card S, Nassar R, Wongsam S, Knez R, Cheng P, Bauer R. Performance of a direct estradiol assay on the Stratus® II analyzer [Abstract]. Clin Chem 1993;39:1204.
  25. Jiang V, Hoke C, Ettinger B. Comparison of estradiol levels in serum by RIA and ELISA methods [Abstract]. Clin Chem 1995;41:S48.
  26. Uettwiller-Geiger DL, Grand S, Gray J, O'Mara D, Malchiodi L. An evaluation of two automated nonisotopic estradiols as compared to the Pantex radio-immunoassay in an IVF setting [Abstract]. Clin Chem 1993;39:1265.
  27. Mills RA, Cheever E, Stoiloff S, Castro D, Johnson R, DeMarco C, Houle P. Development of a rapid, automated procedure for the determination of progesterone on the Stratus® II analyzer [Abstract]. Clin Chem 1993;39:1161.
  28. Pesce MA, Bodourian SH, Sheehan M, Henkel CF. An automated biotin–streptavidin procedure for progesterone evaluation. Clin Biochem 1992;25:451-455. [ISI][Medline] [Order article via Infotrieve]
  29. Bymes S, Branson J, Palosaari N, Eiler M, Schmidt J, Kapsner K. Development of an automated progesterone assay [Abstract]. Clin Chem 1995;41:S85.
  30. Wilson H, El Shami AS, Lei J-D. An automated chemiluminescent immunoassay for progesterone on the Immulite system [Abstract]. Clin Chem 1994;40:1035.
  31. Ford K, Caplan DE, Tobias DD, Turn JE, Osikowicz E. Automated direct enzyme immunoassay of progesterone on the Abbott IMx® analyzer [Abstract]. Clin Chem 1990;36:1099.
  32. Ross JW, Lawson NS. Analytic goals, concentration relationships, and the state of the art for clinical laboratory precision. Arch Pathol Lab Med 1995;119:495-513. [ISI][Medline] [Order article via Infotrieve]
  33. Fraser CG, Petersen PH. Desirable performance standards for imprecision and bias in alternate sites. Arch Pathol Lab Med 1995;119:909-913. [ISI][Medline] [Order article via Infotrieve]
  34. Ricos C, Arbos MA. Quality goals for hormone testing. Ann Clin Biochem 1990;27:353-358.
  35. Ooi LSM, Panesar NS, Masarei JRL. Within- and between-subject variation in, and associations between, serum concentrations and urinary excretion of testosterone and estradiol in Chinese men. Clin Chim Acta 1995;236:87-92. [ISI][Medline] [Order article via Infotrieve]
  36. Thienpont L. Meeting report: First and Second Estradiol International Workshops. Clin Chem 1996;42:1122-1124. [Free Full Text]
  37. Nakajima ST, McAuliffe T, Gibson M. The 24-hour pattern of the levels of serum progesterone and immunoreactive human chorionic gonadotropin in normal early pregnancy. J Clin Endocrinol Metab 1990;71:345-353. [Abstract]
  38. Mikkelsen AL, Borggaard B, Lebech PE. Results of serial measurement of estradiol in serum with six different methods during ovarian stimulation. Gynec Obstet Invest 1996;41:35-40.
  39. Estradiol. Coat-A-Count kit insert, Diagnostic Products Corporation, Los Angeles, CA, 1994..
  40. Estradiol-6. Coat-A-Count kit insert, Diagnostic Products Corporation, Los Angeles, CA, 1995..
  41. Anttila L, Koskinen P, Irjala K, Kaihola H-L. Reference intervals for serum sex steroids and gonadotropins in regularly menstruating women. Acta Obstet Gynecol Scand 1991;70:475-481. [Medline] [Order article via Infotrieve]
  42. Lee CS, Smith NM, Kohn SN. Analytic variability and clinical significance of different assays for serum estradiol. J Reprod Med 1991;36:156-160. [ISI][Medline] [Order article via Infotrieve]
  43. Patricot MC, Badonnel Y, Bugugnani M-J, Collignon I, Delvigne L, Lacroix I, Mathian B. Validité des dosages immunochimiques de l'estradiol sanguin—étude réalisée en 1994. Ann Biol Clin 1995;53:399-406.
  44. Potischman N, Falk RT, Laiming VA, Siiteri PK, Hoover RN. Reproducibility of laboratory assays for steroid hormones and sex hormone-binding globulin. Cancer Res 1994;54:5363-5367. [Abstract/Free Full Text]
  45. Progesterone. Coat-A-Count kit insert, Diagnostic Products Corporation, Los Angeles, CA, 1993..



The following articles in journals at HighWire Press have cited this article:


Home page
Hum ReprodHome page
P. M. Lam, C. Briton-Jones, C. K. Cheung, L. S. Po, L. P. Cheung, and C. Haines
Increased mRNA expression of vascular endothelial growth factor and its receptor (flt-1) in the hydrosalpinx
Hum. Reprod., November 1, 2003; 18(11): 2264 - 2269.
[Abstract] [Full Text] [PDF]


Home page
Biol. Reprod.Home page
P. M. Lam, C. Briton-Jones, C. K. Cheung, I. H. Lok, P. M. Yuen, L. P. Cheung, and C. Haines
Vascular Endothelial Growth Factor in the Human Oviduct: Localization and Regulation of Messenger RNA Expression In Vivo
Biol Reprod, May 1, 2003; 68(5): 1870 - 1876.
[Abstract] [Full Text] [PDF]


Home page
Cancer Epidemiol. Biomarkers Prev.Home page
S. Rinaldi, H. Dechaud, C. Biessy, V. Morin-Raverot, P. Toniolo, A. Zeleniuch-Jacquotte, A. Akhmedkhanov, R. E. Shore, G. Secreto, A. Ciampi, et al.
Reliability and Validity of Commercially Available, Direct Radioimmunoassays for Measurement of Blood Androgens and Estrogens in Postmenopausal Women
Cancer Epidemiol. Biomarkers Prev., July 1, 2001; 10(7): 757 - 765.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (12)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lévesque, A.
Right arrow Articles by Grant, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lévesque, A.
Right arrow Articles by Grant, A.
Related Collections
Right arrow Laboratory Management
Right arrow Endocrinology and Metabolism
Right arrow Automation and Analytical Techniques


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS