Clinical Chemistry 43: 1618-1621, 1997;
(Clinical Chemistry. 1997;43:1618-1621.)
© 1997 American Association for Clinical Chemistry, Inc.
Quality control for qualitative assays: quantitative QC procedure designed to assure analytical quality required for an ELISA of hepatitis B surface antigen
George A. Green, IV1,a,
R. Neill Carey2,
James O. Westgard3,
Tami Carten1,
LeeAnn Shablesky1,
Daniel Achord1,
Eileen Page1 and
Anh Van Le1
1
Ortho Diagnostic Systems, 1001 US Hwy. 202, Raritan, NJ 08869.
2
Peninsula Regional Medical Center, 100 East Carroll
St., Salisbury, MD 21801.
3
Department of Pathology and Laboratory Medicine,
University of Wisconsin Medical School, Madison, WI 53792.
a Author for correspondence. Fax (908) 704-3153; e-mail ggreen{at}odsus jnj.com.
 |
Abstract
|
|---|
An assay for hepatitis B surface antigen (HBsAg) should reliably detect
0.2 µg/L, the lowest reported concentration in an asymptomatic blood
donor. The difference between this concentration and the assay cutoff
defines the analytical quality requirement in a total error format. The
design of a statistical QC procedure is critically dependent on the
precision of the assay. The precision of a developmental ELISA of HBsAg
under study ranged from 17.5% to 9.6% for controls containing 0.07 to
1.50 µg/L, respectively. Use of one positive control with the
13s QC rule provided an 85% chance of detecting a critical
loss of assay sensitivity; use of two positive controls increased the
chance of detecting critical loss of assay sensitivity to nearly 100%.
These rules are based on the precision of this developmental assay, and
must be developed individually for other assays. The development of the
proposed QC procedures illustrates how quantitative QC can be provided
for qualitative assays.
 |
Introduction
|
|---|
Conventional QC procedures for ELISAs of hepatitis B
surface antigen
(HBsAg)1
rely on testing strongly positive controls included in the
reagent kit. Often the acceptability ranges for the resulting
absorbance values are broad, to be more widely applicable. QC
procedures operating with these controls and wide ranges may not be
able to detect a clinically significant loss of assay sensitivity. The
Clinical Laboratory Improvement Act of 1988 requires that laboratories
use positive controls separate from those used to calculate the cutoff
(1), and some states require use of a positive control in
addition to the reagent kit manufacturer's test kit controls
(2). Without guidance on how to implement these new
controls, there is no guarantee that QC of the assay will be improved
by these regulations. Here, using an ELISA of HBsAg capable of
detecting <0.1 µg/L, we considered how a user laboratory could
design an improved QC procedure and what level of performance could be
expected from implementing it.
The first step in the design of a QC procedure is to define the quality
requirements of the assay (3). For HBsAg, the clinical
decision value is at the cutoff concentration; however, the lowest
reported HBsAg concentration in an asymptomatic blood donor (0.2
µg/L) is far greater than the cutoff in this assay (4).
Virtually all subjects who are infected with hepatitis B will test very
strongly positive for HBsAg soon after infection, at concentrations
tens to hundreds of times the cutoff. Thus, the difference between 0.2
µg/L and the cutoff defines the analytical quality requirement in a
total error format. Any run in which 0.2 µg/L HBsAg fails to produce
a signal above cutoff should be rejected.
The next step in the design of a QC procedure is to evaluate the
precision of the assay. We used an analysis of variance experiment to
study assay precision at concentrations from 0.07 to 1.50 µg/L.
Finally, we selected appropriate control rules and numbers of control
measurements on the basis of the information about the quality
requirement, the observed method performance characteristics, and the
performance characteristics of the candidate QC procedures, which were
determined by computer simulation studies.
 |
Materials and Methods
|
|---|
ELISA for HBsAg and positive controls.
We used the Ortho
Antibody to HBsAg ELISA Test System 3 (Ortho Diagnostic Systems). The
assay was automated with the Ortho SummitTM sample-handling
system, operated with Ortho ELISA ver. 5.2CTTM software.
Plates were incubated in Ortho Model 120 forced-air incubators, washed
with an Ortho Auto Wash 96TM plate-washer, and read with an
Ortho AutoReader IITM. We made positive controls with HBsAg
concentrations of 0.07, 0.10, 1.00, and 1.50 µg/L by diluting an
HBsAg-positive serum pool made up of three characterized positive blood
donor units. Aliquots of the pool were diluted into an HBsAg-negative
normal human serum containing preservatives.
Experimental protocol.
Ten replicates of each control
were tested on each 96-well microwell plate, together with a reagent
blank and three replicates of the negative standard (used to calculate
the cutoff absorbance value). Plate layout is shown in Table 1
. We tested 40 plates in 8 runs over 40 days, for a total of 400
replicates for each control. We performed the assay according to the
manufacturer's instructions.
Calculations, computer simulation studies, and charts of
operating specifications.
We studied control data both expressed
in terms of absorbance and as the ratio of sample absorbance to cutoff
absorbance (S/C). As with most qualitative blood-screening ELISAs, the
cutoff absorbance was derived from the mean absorbance of three
negative calibrators by adding a fixed value (0.030) that was
predetermined from performance data of the assay. We used Minitab
(State College, PA) statistical software to calculate the total
standard deviation (s) for each control, using the equations given in
NCCLS EP5 (5).
To facilitate the selection of appropriate statistical control rules
and numbers of control measurements, we determined power functions
accounting for between-run variations by computer simulation with
Minitab as previously described by Cembrowski and Carey
(6). These power functions were then entered into the QC
Validator program (WesTgard QC, Ogunquit, ME) by using the program's
utility for editing the candidate QC file. Critical-error graphs and
charts of operating specifications (OPSpecs charts) were then prepared
with the QC Validator program. For candidate QC procedures, we selected
rules that are relatively easy to implement and have reasonably low
false-rejection rates, e.g., 12.5s, 10.01,
13s, 13.5s,
X0.01/R0.01, and
X0.002/R0.002 rules. For calculation of control
limits for these rules, see Westgard et al. (7).
 |
Results
|
|---|
Mean absorbance values, within-plate CVs (CVw),
total CVs (CVt), and ratios of between-run to within-run
CVs are given in Table 2
for each control. When control data are expressed in terms of
absorbance, total CVs decrease with increasing concentration, from
17.2% at 0.07 µg/L HBsAg to 9.6% at 1.50 µg/L. Expressing data in
terms of S/C has the effect of reducing the between-run variations for
lower-concentration controls and reduces CVt (13.3% at
0.07 µg/L HBsAg). For the two moderately positive controls, however,
CVt was nearly 1% greater when results were expressed as
S/C (10.4% at 1.50 µg/L HBsAg). CVt calculated from S/C
data is lower for low-positive controls because absorbance values of
similar magnitude are being ratioed, and dividing by the cutoff
absorbance tends to compensate for between-run variations. For
moderately positive controls, however, the sample absorbance is
>10-fold the cutoff absorbance and has a lower CV, and between-run
variation is a smaller proportion of the total variation. Dividing the
relatively more precise moderately positive control absorbance by the
relatively less precise cutoff absorbance leads to the increase in CV.
Although we did not test a control exactly at the medical decision
concentration (0.2 µg/L), the CVt of the
closest-concentration control, 0.1 µg/L, is a suitable estimate of
the CVt at 0.2 µg/L for the purpose of designing a QC
procedure. The CVt at 0.2 µg/L expressed in absorbance
should be slightly less than that at 0.1 µg/L, whereas the
CVt at 0.2 µg/L expressed in S/C should be nearly equal
to that at 0.1 µg/L.
Because the response of the assay is linear, we calculated the expected
absorbance value for 0.2 µg/L HBsAg from the absorbance of the 0.1
µg/L control and obtained 0.118. The mean cutoff absorbance was
0.034. Thus, the total allowable error is a loss of assay sensitivity
from 0.2 µg/L (0.118 A) down to cutoff (0.034
A), or a 71.2% loss of sensitivity.
We first assessed the QC design for data expressed in terms of
absorbance. The critical systematic error,
SEcrit,
was calculated (8) from the allowable error and the
CVt of the 0.1 µg/L control to be a shift equivalent to
3.26 s. Power function curves for systematic error (loss of
sensitivity) are shown for several candidate control rules, with use of
1 or 2 positive controls per run, in Fig. 1
. All rules with Ped near 0.90 had
Pfr of 0.02 or greater. For a reasonable
tradeoff of QC rule sensitivity to error and low false rejections, the
best choice is probably the 13s rule with 2 positive
controls, which has Ped of 0.81 and
Pfr near 0.00.
Mean and range rules demonstrated high sensitivity to shifts; however,
they are not shown on Fig. 1
because of their high Pfr
of 0.04 and 0.02, respectively, for the
X0.01/R0.01 and
X0.002/R0.002 rules. Between-run variations
increase the Pfr for control rules that rely on the mean,
and generally reduce sensitivity to systematic error.
Expression of control data in terms of S/C reduced the
CVt, increasing the critical systematic error from
3.26 s to 4.09 s. In Fig. 1
, a critical systematic error of
4.09 is off-scale to the right; all rules have slightly higher
Ped than shown. Even the 13.5s rule
has very high sensitivity to the critical systematic error when 2
positive controls are run. Acceptable sensitivity to systematic error
is now provided by the 13s rule with only 1 positive
control per run.
 |
Discussion
|
|---|
We used four controls to evaluate the precision performance of the
assay. The 0.07 and 0.10 µg/L HBsAg controls demonstrated the
precision between the assay cutoff and the 0.20 µg/L decision
concentration. Concentrations of 1.00 and 1.50 µg/L were typical of
the moderately positive controls included in commercial HBsAg reagent
kits. To obtain optimal performance from QC procedures when
low-positive controls are used, one should express control results in
S/C. For moderately positive controls, control results can be expressed
in either S/C or simple absorbance because the between-run variations
have proportionately less impact on simple absorbance as the
concentration of the control increases. Moderately positive controls
may be preferable in practice because QC ranges calculated from
absorbance are simpler to implement than those calculated from S/C.
When results were expressed in terms of S/C, high probability to detect
loss of assay sensitivity was demonstrated by both the
13.5s rule with 2 positive controls per run and the
13s rule with 1 positive control per run. Both rules have
Pfr of <0.01 and Ped
>0.85, even when simulations take into account between-run variations.
With 2 positive controls per run, the 13s rule is virtually
certain to detect critical loss of assay sensitivity;
Pfr is <0.01. Because the 0.1 µg/L control
tested positive on all replicates on all plates, the frequency of
unacceptable runs is expected to be very low when this ELISA is in
routine use. Thus, QC procedures with 1 positive control and having
Ped slightly <0.90 would be satisfactory.
The HBsAg ELISA is a single-analyte assay. Assays for other infectious
disease markers, such as human immunodeficiency virus, human
T-lymphocyte virus, and hepatitis C virus, detect multiple analytes and
thus may not give a direct linear response of controls to assay
sensitivity. Further study is necessary to determine if the methods
used here will apply to other ELISAs.
Because this experiment was designed to study the performance
characteristics of positive controls, we did not run negative controls
other than the negative materials used to calculate the cutoff
absorbance. In ELISAs for HBsAg, absorbance values for negative
controls are very close to 0 (typically ranging from <0 to 0.005
A). Because reactive specimens are repeated in duplicate in
a second run, a negative sample must test falsely positive 3 times in 2
runs to be reported as reactive for HBsAg. Thus, negative controls
function mostly to detect gross contamination and the presence of
splashing or poor washing during the assay.
The assay protocol for Ortho Antibody to HBsAg ELISA Test System 3
specifies running 3 negative calibrators and uses a test of their
range, as well as a test of their mean absorbance, to judge the
acceptability of their absorbance values before their mean absorbance
is used to calculate the cutoff absorbance. In this case, the negative
calibrators provide the same information as negative controls. In our
experience, nonrepeatable false positives usually occur from assignable
causes, such as splashing from an adjacent well. Reference sample QC
has limited ability to detect random conditions that favor these types
of false positives. We did not study the ability of reference sample QC
to detect an systematic increase in sensitivity (e.g., causing negative
samples to test positive). Because the mean of HBsAg-negative results
is >6 s from the cutoff, such an increase in sensitivity will only
rarely cause a false-positive result. This may not hold true for other
ELISAs for hepatitis markers, in which the absorbance of negative
controls has considerable between-run variance.
These studies demonstrate that QC procedures that guarantee a
stated amount of sensitivity can be developed for qualitative HBsAg
ELISAs. These rules can be selected to give a high probability of error
detection while maintaining low probabilities of false rejection. In
practice, it will be necessary for manufacturers to continue to provide
the general QC limits that would apply to all laboratories in initially
establishing an assay. User laboratories can improve the performance of
the assay by determining their own quality requirements, evaluating
assay precision in their own environments, and developing custom
QC rules, as is the practice with most clinical chemistry
assays.
 |
Acknowledgments
|
|---|
We appreciate the technical assistance provided by Marla Fosdick,
Michelle Anson, and Susan Merkel of the Blood Bank of Delaware. R.N.C.
is a consultant to Ortho Diagnostic Systems, and J.O.W. is a principal
in WesTgard QC.
 |
Footnotes
|
|---|
1 Nonstandard abbreviations: HBsAg, hepatitis B surface antigen; Ped, probability of error detection; Pfr, probability of false rejection; S/C, ratio of sample absorbance to cutoff absorbance; and
SEcrit, critical systematic error. 
 |
References
|
|---|
-
Clinical Laboratory Improvement Act of 1988; final
rule. Fed Reg 1992;57:7164..
-
Quality control standards. Albany, NY: New York State
Department of Health, July 1,1994:911..
-
Westgard JO. Planning and validating QC procedures
[Workshop Manual]. Ogunquit, ME: WesTgard QC, 1994:109pp..
-
Courouce A-M, Drouet J, LeMarrec N, Drouet A, Soulier JP. Blood donors positive for HBsAg and negative for anti-HBC antibody. Vox Sang 1985;49:26-33.
[Web of Science][Medline]
[Order article via Infotrieve]
-
National Committee for Clinical Laboratory Standards.
Evaluation of precision performance of clinical chemistry devices, 2nd
ed. Tentative guideline. NCCLS document EP5T2 (ISBN 1-56238-145-8).
Villanova, PA: NCCLS, 1988;5:1992..
-
Cembrowski GS, Carey RN. Laboratory quality management: QC & QA 1989:110-111 ASCP Press Chicago. .
-
Westgard JO, Groth T, Aronsson T, Falk H, de Verdier C-H. Performance characteristics of control rules for internal quality control: probabilities for false rejection and error detection. Clin Chem 1977;23:1857-1867.
[Abstract/Free Full Text]
-
Westgard JO, Barry PL. Cost-effective quality control: managing the quality and productivity of analytical processes 1986:47-49 AACC Press Washington, DC. .