Clinical Chemistry 43: 1697-1702, 1997;
This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (2)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Oberhardt, B. J.
Right arrow Articles by Pritchard, C. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Oberhardt, B. J.
Right arrow Articles by Pritchard, C. G.
Related Collections
Right arrow Laboratory Management
Right arrow Hemostasis and Thrombosis
Right arrow Oak Ridge Conference
(Clinical Chemistry. 1997;43:1697-1702.)
© 1997 American Association for Clinical Chemistry, Inc.


Oak Ridge Conference

Point-of-care fibrinolytic tests: the other side of blood coagulation

Bruce J. Oberhardt1,a, Patrick D. Mize2 and Cynthia G. Pritchard2

1 Sanguinex, P.O. Box 98385, Raleigh, NC 27624-8685.

2 Cardiovascular Diagnostics, Inc., 5301 Departure Dr., Raleigh, NC 27616.
a Author for correspondence. Fax 919-676-5436; e-mail bjo{at}ntwrks.com


Abstract

Point-of-care (POC) coagulation tests with paramagnetic iron oxide particles have provided alternatives to testing previously done only in the laboratory. With this technology, POC fibrinolytic tests have followed quietly the trail blazed by POC clotting tests and have found specific applications. These include rapid verification of in vivo thrombolytic drug action by in vitro testing with subsequent quantitative drug monitoring of the systemic lytic state, and also the determination of in vitro thrombolytic drug response before the drug is actually administered, to individualize therapy by selection of the most appropriate drug. Other applications include POC coagulation factor assays associated with fibrinolysis, and most recently the POC screening of patients with fibrinolytic defects. In this latter application, plasma from cardiac catheterization (n = 19) and venous thrombosis (n = 47) patient groups were tested. Controls consisted of two independent donor pools (n = 10, n = 21) as negatives and two plasma samples with known genetic defects in the fibrinogen molecule (A{alpha}554 Arg -> Cys) as positives.

Blood coagulation testing at the point of care (POC) now exists in many hospitals.3 POC testing has been driven by economic factors and by improved technology (1). In particular, dry chemistry coagulation testing at the POC with paramagnetic iron oxide particles (PIOP) has provided alternatives to testing previously done only in the laboratory and has opened the door to convenient fibrinolytic testing (2)(3)(4). With this technology, POC fibrinolytic tests have followed quietly the trail blazed by POC dry chemistry clotting tests and have found specific applications. For example, during and after thrombolytic drug administration, a quantifiable measure of a patient's systemic lytic response can be determined rapidly at the POC to verify drug action (5). Furthermore, the response of a patient's in vitro-generated clot to a thrombolytic drug can be determined before the drug is actually given, allowing therapy to be individualized by selecting the most appropriate drug (6). Both of these fibrinolytic applications are being pursued in multicenter clinical trials in several countries. Two additional potential clinical applications of POC fibrinolytic testing are the ability to perform fibrinolytic factor assays4 and to screen for fibrinolytic defects, which will be discussed further in this report.

The PIOP/dry chemistry technology may be used with whole blood or citrated plasma, yielding similar results for most tests. After placement of the test card that contains appropriate dry reagents and PIOP in the instrument, the card is warmed by a 37 °C heat source. The sample is then added to the test card. The liquid sample dissolves the reagents and frees the PIOP. The PIOP then organize into ensembles or "stacks" under the influence of orthogonal magnetic fields. The field perpendicular to the test card surface oscillates, causing the particles to oscillate (3)(7). The motion of PIOP is monitored in the near infrared throughout the test. When the sample clots, the PIOP become entrapped in the gel clot, and their motion diminishes gradually. The peak PIOP oscillation signal before the decay in amplitude is the typical end point for clotting time measurements that correlates well with conventional laboratory methods. A variety of conventional and new clotting tests are possible with this technology. For example, a PIOP/dry chemistry test [ecarin clotting time (ECT)] was developed for antithrombin drugs. Fig. 1 shows a plot of clotting time vs the concentration of the thrombin inhibitor polyethylene glycol (PEG)–hirudin, a therapeutic drug being developed as a replacement for heparin in patients with conditions such as heparin-induced thrombocytopenia. This test is based on ECT (8)(9). Ecarin is a snake venom product from Echis carinatus (EC 3.4.99.27) that provides a test specific for thrombin inhibitors.



View larger version (16K):
[in this window]
[in a new window]
 
Figure 1. Heparin effects on ECT test cards.

Solid line denotes response of ECT test to PEG–hirudin in the absence of heparin. Dotted line denotes the response of the ECT test to PEG–hirudin in the presence of 500 U/L porcine heparin.

Ecarin, a protein prothrombin activator, causes coagulation of citrated whole blood or citrated plasma by the calcium-independent activation of prothrombin. Ecarin catalyzes the hydrolytic cleavage of the 323Arg–324Ile bond in the human prothrombin molecule, whereby thrombin activity is generated without the release of a zymogen fragment. This form of active prothrombin has been termed meizothrombin and is inhibited by r-hirudin or PEG–hirudin, but surprisingly, not efficiently by the heparin–ATIII complex. Ecarin has been used as the basis to develop a sensitive analytical method for the determination of thrombin inhibition by antithrombin drugs such as r-hirudin and PEG–hirudin and is based on the method of Nowak and Bucha (8)(9). The PIOP/dry chemistry ECT test card is intended to determine the concentration of an antithrombin drug in citrated whole blood and can be used to monitor patients receiving antithrombin drugs such as r-hirudin or PEG–hirudin.

The data shown in Fig. 1Up are for citrated whole-blood samples from healthy donors, supplemented with increasing concentrations of PEG–hirudin (Knoll) and (or) porcine heparin (Sigma) and tested in the dry chemistry format. The ECT test cards were prepared from ecarin (Centerchem), calcium, PIOP, buffers, and stabilizers. As the in vivo concentration of antithrombin drug increases, the clotting time increases in a linear dose-dependent manner with essentially no effect of heparin on the clotting time end point, a potentially important clinical consideration when determination of only the anticoagulant effects of the direct thrombin inhibitor is desired.

For fibrinolysis (thrombolysis) measurements, where the clot subsequently dissolves, a new end point known as lysis onset time (LOT) was established (10).5 LOT is the point at which the PIOP oscillation amplitude starts to increase again (Fig. 2 ), eventually resulting in full restoration of movement and is detected well before the clot dissolves, which provides the earliest indication of fibrinolysis. LOT can be precisely measured and is inversely proportional to the concentration of plasminogen activator present in the system (3). A mathematical model for LOT was developed that is supportable by experimental data (7). Fig. 2 shows a PIOP oscillation curve vs time for a lytic test. The peak event corresponding to the induced clotting time is observed along with the subsequent rise event corresponding to LOT. Increased concentrations of thrombotic agent incorporated in the test card dry chemistry decrease the clotting time in the oscillation curve. For high concentrations of a thrombotic agent such as thrombin or batroxobin, clotting time can be reduced to a few seconds. Increased concentrations of thrombolytic agent in the test card or in the sample being tested shorten the LOT in the oscillation curve. For high concentrations of thrombolytic agent, LOT can be <90 s. In this report, data are shown to demonstrate that two new potential applications of LOT may be achievable: fibrinolytic factor assays and fibrinolytic defect screening.



View larger version (15K):
[in this window]
[in a new window]
 
Figure 2. Typical signal amplitude difference curve.

Curve was generated with a test card containing 1500 NIH units/L human thrombin with a normal citrated blood sample with added t-PA (Genetech) at 500 000 U/L. Clotting time is the point at which clotting of the sample starts to occur, and LOT is the point at which the formed clot starts to dissolve.


Materials and Methods

plasminogen assay
Plasminogen-free citrated plasma was prepared by passing pooled normal citrated plasma (Specialty Assayed Reference Plasma, George King) through a lysine–Sepharose column (Pharmacia). This treatment selectively removes plasminogen without affecting other coagulation proteins. Plasminogen (Sigma) was then added to the plasminogen-depleted citrated plasma to yield the different concentrations used in the assay. Dry chemistry plasminogen test cards were prepared by assembling the test card device and injecting 24 µL of a solution [containing 1500 U/L thrombin, 500 000 U/L tissue plasminogen activator (t-PA), 500 000 U/L urokinase, PIOP, salts, buffers, and stabilizers] into the card chamber, freezing the card/reagent mixture, and lyophilizing the test card to yield the dry chemistry formulation. Test cards were placed in a Thrombolytic Assessment System (TAS) analyzer (Cardiovascular Diagnostics), and 30 µL of the various plasminogen-supplemented citrated plasmas were added to the test card. The TAS analyzer was programmed to detect the LOT and display this value to the user.

screening for fibrinolytic defects
Citrated plasma samples were obtained from stockpiled venous thrombosis patients and patients undergoing cardiac catherization. CHIII and Dusart are samples from well-studied thrombosis-prone families with abnormal fibrinogens known to exhibit impaired fibrinolysis (12)(13)(14)(15)(16). The defect in both families consists of the same amino acid substitution of cysteine at the A{alpha}554 arginine site (15). To provide "normals," samples were obtained from local donors at Cardiovascular Diagnostics, and from donors at a major blood center. All citrated plasma samples were maintained frozen at -85 °C before analysis and thawed at 37 °C before use. Dry chemistry test cards suitable for screening patient and normal samples were prepared as follows: 24 µL of a solution containing 1500 U/L thrombin, t-PA at a concentration of 200 000, 400 000, 800 000, or 1 000 000 U/L, PIOP, salts, buffers, and stabilizers were injected into an assembled test card device, frozen, and lyophilized to yield the dry chemistry formulation. The formulation included heparin neutralizers, which negated the effects of heparin, if present. Results shown are for a t-PA concentration of 200 000 U/L. The test cards were placed in the TAS analyzer, and 30 µL of patient or normal donor citrated plasma were added to the test card. The TAS analyzer was programmed to detect and display the LOT value.


Results

plasminogen assay
A set of increasing responses forming a curve resulted when LOT values were plotted vs plasminogen concentration in the sample (Fig. 3 ). The resulting LOT values ranged from 90 s at the higher plasminogen concentrations (120 to 160 mg/L) to 250 s at 45 mg/L. Plotting the LOT results of the plasmin(ogen) assay vs the inverse of the plasminogen concentration results in a linear LOT response (Fig. 4 ).



View larger version (12K):
[in this window]
[in a new window]
 
Figure 3. Plasmin(ogen) calibration curve.

Plasmin(ogen) assay calibration curve generated by addition of plasminogen to pooled plasminogen-depleted citrated plasma added to PIOP/dry chemistry test cards.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 4. Plasmin(ogen) calibration curve; inverse response.

Plasmin(ogen) assay calibration curve generated by addition of plasminogen to pooled plasminogen-depleted citrated plasma added to PIOP/dry chemistry test cards plotting LOT vs inverse plasminogen concentration and showing the resulting linear relation.

screening for fibrinolytic defects
When the low-concentration t-PA (200 000 U/L) test data were obtained, we observed that the venous thrombosis, cardiac catheterization, and two normal groups exhibited a wide range of individual responses. The mean LOT values and their standard deviations (in s) were: 432 ± 102 for the venous thrombosis group (n = 47), 329 ± 88 for the cardiac catheterization group (n = 19), 283 ± 76 for the local "normal" donors (n = 10), and 319 ± 54 for the blood center "normal" donors (n = 21). The LOT values obtained for the Dusart and CHIII samples were 875 and 670 s, respectively. All measurements are means of duplicates. LOT measurements, with low-concentration t-PA, are generally within ± 10%. Precision approaches ± 5% as the amount of t-PA reaches 500 000 U/L, the concentration used to generate the data in Fig. 2Up .


Discussion

The calibration curve for plasminogen indicates that it should be possible to provide a POC test for low-range plasminogen in patient samples. This test would actually measure total available plasmin and provide the composite fibrinolytic effect of plasminogen and free plasmin, if any. Typically, plasminogen would be measured. In the case of a patient undergoing extended thrombolytic therapy, e.g., for deep vein thrombosis, it may be important to determine periodically that the patient has sufficient plasmin(ogen) reserves for the therapy to continue to be effective. A low-range test for total available plasmin is well suited to meet this objective. Normal- and higher-range plasminogen (concentrations >200 mg/L) is not a critical care issue and can be determined with a laboratory test in a routine fashion.

A linear LOT response is obtained by plotting LOT vs the reciprocal of the plasminogen concentration (Fig. 4Up ). This linearization had previously been established for plasminogen activators (3). This depiction of the LOT response allows easier conversion of LOT into plasmin(ogen) concentration. Our test may be suitable for use with thrombolytic therapy patients to assess plasmin(ogen) reserves at the POC.

In screening for fibrinolytic defects, it is important to establish a normal range. Extensive testing was not undertaken to do this, but selected random samples were used from two different donor pools. Whether all of the samples in each pool are "normal" was not addressed by our study. Similarly, some of the cardiac catheterization patients are probably normals. For example, patients with heart valve problems but without vessel disease might have normal LOT values. However, when the means of each "normal" pool and their standard deviations are different, a general range of probable normals may be inferred from the data, as depicted graphically in Fig. 5 . Similarly, the known abnormal samples, Dusart and CHIII, fall well outside of this general range of probable normal values. In fact, LOT values for both Dusart and CHIII are more than 5 SD from the mean of either normal group. Therefore, other samples falling well beyond the normal ranges would be expected to be abnormal as well. It is remarkable, therefore, that the venous thrombosis group contains two or four individuals whose LOT values are >5 SD from the mean value of the "normals," depending upon which "normal" group is used. Therefore, we have been able to demonstrate impaired fibrinolysis in vitro for some of the venous thrombosis patients. Specific fibrinolytic defects were not detected and correlated to clinical outcomes, as that was beyond the scope of the present feasibility study.



View larger version (14K):
[in this window]
[in a new window]
 
Figure 5. Screening for fibrinolytic defects.

Comparison of low-concentration t-PA responses of individual citrated plasma samples in patient and normal groups with PIOP/dry chemistry test cards.

Venous thrombosis is caused by a variety of defects, at least one subset of which are fibrinolytic defects. Finding or identifying this subset of patients is not easily accomplished. Two methods that have been used are euglobulin lysis time and whole-blood clot lysis time. Both methods are labor intensive and difficult to standardize. For the individual samples tested here, the origin of the fibrinolytic defects is unknown. However, the simple, rapid test methodology, as described, should enable convenient screening and identification of abnormal individuals who could undergo further testing to determine the exact cause of the fibrinolytic defect.

Table 1 tabulates POC clotting time measurements and LOT measurements for different end uses. For the end use of lytic state monitoring, clotting time end points, such as in fibrinogen and thrombin time tests, provide an indirect measure of the systemic lytic state. LOT provides a direct measure (3)(5). For drug response, clotting time end points provide a measure of in vitro anticoagulant effects after administration of an anticoagulant drug. With appropriate test chemistry, LOT provides a measure of in vitro fibrinolytic effects before the administration of a thrombolytic drug (6). In factor assays, clotting time and LOT end points (as discussed previously) may be used to determine concentrations of clotting and fibrinolytic factors, respectively. Finally, clotting time end points in tests such as prothrombin time and activated partial thromboplastin time enable screening for clotting defects, and fibrinolytic tests such as the low-concentration t-PA response test described in this report enable screening for fibrinolytic defects.


View this table:
[in this window]
[in a new window]
 
Table 1. Comparison of POC clotting and fibrinolytic end point tests.


Acknowledgments

The contributions of the following individuals are also acknowledged: Omar Iqbal, Zaid Y. Alkadi, Stuart Lind, William H. Bingham, Susan T. Lord, Gregory J. Dehmer, Sharon C. Dermott, David R. Musselman, Jawed Fareed, David C. Sane, Gregory Goslen, David A. Tate, Nancy J. Gresalfi, and Aldo A. Tabares. Support of NIH of this research, in part, through the grants R43 HL 47223, R44 HL 47223, R43 HL 52412, and R44 HL 52412 is gratefully acknowledged.


Footnotes

3 Nonstandard abbreviations: POC, point of care; PIOP, paramagnetic iron oxide particles; ECT, ecarin clotting time; PEG, polyethylene glycol; LOT, lysis onset time; t-PA, tissue plasminogen activator; and TAS, Thrombolytic Assessment System.

4 Normally the term "coagulation factor" refers to protein factors that, when activated, cause both clotting and lysis. We prefer to separate the coagulation factor, plasminogen, as a "fibrinolytic factor" that, when activated, causes lysis of the clot.

5 LOT differs from clot lysis time, an older term found in hematology textbooks. Clot lysis time is related to the time required to dissolve an in vitro clot in tests such as whole-blood clot lysis time. Whole-blood clot lysis time is used as a simple functional screening test to determine lytic tendency and consists of allowing a tube of whole blood to clot and examining the tube periodically to determine how long it takes for the clot to dissolve. In healthy individuals, the clot remains undissolved, even after 24 h. In individuals with abnormal antiplasmin activity or in some fibrinolytic states, the clot is observed to dissolve in a few hours [11]. It should be recognized that the visual determination of clot lysis time cannot be made until a substantial amount of lysis of the thrombus occurs. Euglobulin clot lysis time is a variation of this test.


References

  1. Oberhardt BJ. Thrombosis and hemostasis testing at the point of care. Am J Clin Pathol 1995;104(Suppl1):S72-S78. [ISI][Medline] [Order article via Infotrieve]
  2. Oberhardt BJ, Taylor M, Alkadi ZY, Dermott SC. Diagnostic assay system for convenient monitoring of oral anticoagulant therapy. Presented at the XIIth Congress of the International Society on Thrombosis and Haemostasis, Tokyo, Japan. Thromb Haemost 1989;62:327a.
  3. Oberhardt BJ, Dermott SC, Taylor M, Alkadi ZY, Abruzzini AF, Gresalfi NJ. Dry reagent technology for rapid, convenient measurements of blood coagulation and fibrinolysis. Clin Chem 1991;37:520-526. [Abstract/Free Full Text]
  4. Rose VL, Dermott SC, Murray BF, McIver MM, High KA, Oberhardt BJ. Decentralized testing for PT and activated PTT using a dry chemistry portable analyzer. Arch Pathol Lab Med 1993;117:611-617. [ISI][Medline] [Order article via Infotrieve]
  5. Sane DC, Gresalfi NJ, Enney-O'Mara LA, Califf RM, Greenberg CS, Bovill EG, Oberhardt BJ. Exploration of rapid bedside monitoring of coagulation and fibrinolysis parameters during thrombolytic therapy. Blood Coagul Fibrinolysis 1992;3:47-54. [ISI][Medline] [Order article via Infotrieve]
  6. Musselman DR, Tate DA, Oberhardt BJ, Abruzzini AF, Blauwet MB, Koch G, Dehmer GJ. Differences in clot lysis among patients demonstrated in vitro with three thrombolytic agents (tissue-type plasminogen activator, streptokinase and urokinase). Am J Cardiol 1994;73:544-549. [ISI][Medline] [Order article via Infotrieve]
  7. Oberhardt BJ. Advancing blood coagulation testing to the point of care in response to molecular biology-driven development of pharmaceuticals. Clin Chem 1993;39:1982-1984. [ISI][Medline] [Order article via Infotrieve]
  8. Nowak G, Bucha E. A new method for the therapeutical monitoring of hirudin. Thromb Haemost 1993;69:1306a.
  9. Nowak G, Bucha E. Quantitative determination of hirudin in blood and body fluids. Semin Thromb Hemost 1996;22:197-202. [ISI][Medline] [Order article via Infotrieve]
  10. Oberhardt BJ. Method for performing coagulation assays accurately, rapidly, and simply, using chemical reagents and paramagnetic particles. US Patent 5,110,727 (May 5, 1992)..
  11. Clinical diagnosis and management by laboratory methods. Henry JB, ed. Philadelphia: W.B. Saunders Co., 1991:742..
  12. Carrell N, Gabriel DA, Blatt PM, Carr ME, McDonagh J. Hereditary dysfibrinogenemia in a patient with thrombotic disease. Blood 1983;62:439-447. [Abstract/Free Full Text]
  13. Soria J, Soria C, Caen JP. A new type of congenital dysfibrinogenemia with defective fibrin lysis—Dusard syndrome: possible relation to thrombosis. Br J Haemotol 1983;53:575-586. [ISI][Medline] [Order article via Infotrieve]
  14. Lijnen HR, Soria J, Soria C, Collen D, Caen JP. Dysfibrinogenemia (fibrinogen Dusard) associated with impaired fibrin-enhanced plasminogen activation. Thromb Haemost 1984;51:108-109. [ISI][Medline] [Order article via Infotrieve]
  15. Yasuo W, Lord ST. A correlation between thrombotic disease and a specific fibrinogen abnormality (A alpha 554 Arg->Cys) in two unrelated kindred, Dusart and Chapel Hill III. Blood 1994;84:3709-3714. [Abstract/Free Full Text]
  16. Oberhardt BJ, Gresalfi NJ, Tabares A, Pritchard C, Lord ST. Detecting impaired fibrinolysis in plasma samples with abnormal fibrinogens using a rapid, dry chemistry method. Poster presented at the XVth Congress of the International Society on Thrombosis and Haemostasis, Jerusalem, Israel. Thromb Haemost 1995;73:1145a.



The following articles in journals at HighWire Press have cited this article:


Home page
CLIN APPL THROMB HEMOSTHome page
T. W. Stief and J. Fareed
Point of Care: Diagnostics in Hemostasisthe Wrong Direction?
Clinical and Applied Thrombosis/Hemostasis, July 1, 2003; 9(3): 191 - 195.
[Abstract] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via ISI Web of Science (2)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Oberhardt, B. J.
Right arrow Articles by Pritchard, C. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Oberhardt, B. J.
Right arrow Articles by Pritchard, C. G.
Related Collections
Right arrow Laboratory Management
Right arrow Hemostasis and Thrombosis
Right arrow Oak Ridge Conference


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS