(Clinical Chemistry. 1998;44:124-133.)
© 1998 American Association for Clinical Chemistry, Inc.
Proficiency testing of creatine kinase and creatine kinase-2: the experience of the Ontario Laboratory Proficiency Testing Program
A. Ralph Henderson1,2,a,
Suba Krishnan3,
Sharon Webb1,
C. Mark Cheung4,5,
Darius J. Nazir6,7,
and Harold Richardson1,7,1
1
Laboratory Proficiency Testing Program, 250 Bloor St. East, Suite 501, Toronto, ON, Canada M4W 1E6.
2
Department of Biochemistry, University of Western
Ontario, London, ON, Canada.
3
Department of Laboratory Medicine, The Credit Valley
Hospital, Mississauga, ON, Canada.
4
Department of Laboratory Medicine, Sunnybrook Health
Science Centre, North York, ON, Canada.
5
Department of Laboratory Medicine, University of
Toronto, Toronto, ON, Canada.
6
Department of Clinical Chemistry, Hamilton General
Campus, Hamilton Health Sciences Corporation, Hamilton, ON, Canada.
7
Department of Pathology, McMaster University, Hamilton,
ON, Canada.
8
Analyzer types and manufacturers are: aca, Dimension, Paramax, and Stratus systems (Dade International); Access (Sanofi Diagnostics Pasteur); ACS 180 (Chiron); Beckman CX systems (Beckman Instruments); Cobas Mira systems (Roche Diagnostic Systems); Hitachi systems (Boehringer Mannheim); ICON (Hybritech); RA systems and Immuno I (Bayer Corp.); Spectrum, IMx, and AxSYM (Abbott Labs.); Vitros DT, 250, 500, and 700 (Johnson and Johnson Clinical Diagnostics).
9
LPTP requires that proficiency testing challenges are processed as patient specimens. Accordingly, outliers reported in these surveys have not been subjected to statistical trimming.
a Address correspondence to this author at: 19 Linksgate Rd., London, ON, Canada N6G 2A6. E-mail ahenders{at}julian.uwo.ca.
 |
Abstract
|
|---|
The Ontario Laboratory Proficiency Testing Program has regularly
monitored the analytical performance of total creatine kinase (CK)
(
230 participants) and CK isoenzyme-2 (CK-MB) (
160 participants)
throughout the entire province. Consistently, a wide dispersion of
results has been observed not only between different analyzer systems
but also among identical analyzers. Accordingly, the results of the
last three proficiency surveys for these analytes were examined
statistically to establish both the extent of these variations and the
range of values reported for the male upper reference ranges. The
results of many of the analyzer systems were significantly different
from each other, as were many of the reference ranges. This
unsatisfactory situation may only be remedied by the use of reference
materials as shown by others. The consequences of these findings also
effect the reliability of epidemiological surveys such as the WHO
MONICA Project (Circulation 1994;90:583612), which
monitors deaths due to heart disease and includes cardiac enzyme
results in its criteria.
 |
Introduction
|
|---|
The healthcare system in Ontario served, in 1996, a population of
10.8 million people (
38% of Canada's population), rather similar
to that of Michigan (9.5 million), Ohio (11.2 million), or Illinois
(11.8 million), and results of proficiency testing obtained from such a
system are therefore likely to be of interest to other North American
healthcare organizations.
Mandatory external quality assurance for Ontario laboratories comprises
licensing and inspection by the Ontario Ministry of Health and, since
1974, mandatory proficiency testing by the Ontario Medical Association
under a contract from the Ontario Ministry of Health. The Laboratory
Proficiency Testing Program (LPTP), a unit of the Ontario Medical
Association (1), has as its primary functions
(a) the external quality assessment of the performance of
all Ontario patient-testing laboratories, whether owned by hospitals or
commercial enterprises, and (b) the provision of educational
assistance to all such laboratories when there are proven deficiencies
in their preanalytical, analytical, and postanalytical performance.
Testing is assigned to several working committees, one of which is the
Enzymes, Cardiac Markers, and Lipids Committee. This Committee has
tested all Ontario laboratories licensed to determine creatine kinase
(CK; EC 2.7.3.2) and CK isoenzyme-2 (CK-2 or CK-MB) since 1989
(2) with lyophilized human-based serum, and, from 1991
onwards, with fresh frozen human serum supplemented with purified human
CK-3 (CK-MM) and CK-2 (3).
Of all enzyme activity assays, the determination of CK is one of the
most investigated (4) and has been the subject of IFCC
standardization (5). This assay, known as the NAC-EDTA
formulation, is used by
90% of the
230 Ontario laboratories so
licensed, and
99% of all these laboratories can regularly achieve
replications of
6.5% at all levels of CK activity (6).
Although intralaboratory performance is satisfactory, the Committee has
noted consistent differences in results between the same type of
analyzer when used by different laboratories even though reference
ranges are similar. Also, there appear to be marked differences in the
results produced by different analyzers. Similarly the intralaboratory
assay performance for CK-2 challenges is satisfactory, with >90% of
the
160 participants achieving replication of
10.0%.
Interlaboratory performance, however, often shows marked discrepancies
in both activity and mass assay results for this isoenzyme.
With the consolidation of tertiary care services throughout Ontario,
enzyme estimations made at a peripheral hospital may diverge from those
made at the receiving hospital, with obvious consequences for patient
care. Furthermore, the cardiac enzyme criteria of the WHO MONICA
Project (7) for the diagnosis of myocardial infarction
code as abnormal a result more than twice the "upper limit of
normal" for that test in each (our italics) laboratory. If
there are biases in analytical performance or in the reference range,
the resulting diagnostic classification may be erroneous. Accordingly,
the Committee statistically assessed these analytical differences in
three recent surveys (June 1996, Z-9606; October 1996, Z-9610; and
January 1997, Z-9701) to ascertain the extent of analyzer differences
and the effect of the reported male upper reference limit on them.
 |
Materials and Methods
|
|---|
Male upper reference ranges
were obtained from the LPTP
archives for each laboratory for total CK and CK-2 for an adult male.
Every laboratory is required to report its methods, analyzers, and
reference ranges to LPTP.
Survey materials.
Fresh frozen human serum was obtained
from the London branch of the Canadian Red Cross Society. Several
individual bags from different individuals were pooled to provide
sufficient material for each survey. Endogenous CK and CK-2 activities
were uniformly at the lower ends of their respective reference ranges.
Purified human CK-3 (from skeletal muscle) and CK-2 (from heart) were
obtained from Scripps Labs., where they had been prepared by sequential
ammonium sulfate precipitation, ion-exchange chromatography, and
proprietary affinity chromatography. Scripps Labs. established
homogeneity by the appearance of a single protein zone after sodium
dodecyl sulfate electrophoresis and by a single zone of CK activity
after electrophoresis on agarose gel. These enzymes were provided in
liquid form suspended in 500 mL/L glycerol, 5 mmol/L succinate, 1
mmol/L EDTA, 5 mmol/L 2-mercaptoethanol, and 10 mmol/L NaCl. Enzyme
activities were
700 U/L at 37 °C. Survey materials (Table 1
) were prepared by supplementing the fresh frozen serum with
these purified enzymes; the extent of contamination of the survey
materials by the enzyme preservatives was minimal because of the
addition of microliter quantities to liter volumes of the fresh frozen
serum. Survey materials were prepared and distributed as previously
described (3).
Surveys.
The total CK and CK-2 activities and CK-2 mass
assay concentrations used in the three LPTP surveys are listed in Table 1
together with the
analyzers used by participants in each survey (Table 2
). LPTP requests that all vials be analyzed within 24 h of
receipt and that they may also be analyzed in the same test run.
Experience suggests that results are therefore subject only to
within-run variance.
For total CK, results from different analyzers from a single vendor
were aggregated only if there were no statistically significant
differences between them (see below); analyzer groups of less than five
were not assessed and there were insufficient numbers to examine the
effects of different vendor kits on the reported results. For CK-2 the
results of all methods were included in the assessment, although
statistical analyses could not be performed on the results from a
single analyzer group.
Statistical analysis.
Statistical assessment was
conducted with SigmaStat version 2.0 or with SYSTAT version 7.0 (SPSS).
Analytical results and upper reference limits were first tested for
normality (with the KolmogorovSmirnov test with Lilliefors'
correction set at P = 0.05 (8)) and then
for equal variances before testing either parametrically, by an
unpaired t-test, or nonparametrically, by the MannWhitney
rank sum test. A significant difference was defined as P
<0.05 and all such values are reported exactly unless
P <0.001. Linear regression and correlation were
determined for all survey results for individual analyzer groups and
the male upper reference limits reported for these analyzers.
 |
Results
|
|---|
Point plots are shown for the results for survey Z-9606 for total
CK activity (Fig. 1
A), CK-2 activity (Fig. 2
A), and CK-2 mass (Fig. 3
A); for survey Z-9610 for total CK activity (Fig. 1B
); and for
survey Z-9701 for total CK activities (Figs. 1C
, vial 1, and D, vials
25), for CK-2 activities (all vials, Fig. 2B
), and for CK-2 mass (all
vials, Fig. 3B
).

View larger version (26K):
[in this window]
[in a new window]
|
Figure 1. Total CK results by analyzer types from (A)
survey Z-9606 (all vials), (B) survey Z-9610 (all vials),
(C) survey Z-9701 (vial 1), and (D) survey Z-9701
(vials 25).
Results are shown only for groups of five or more instruments (see
Table 2
). Each point represents the mean value of all results reported
for each laboratory.
|
|

View larger version (9K):
[in this window]
[in a new window]
|
Figure 2. CK-2 results determined by activity from (A)
survey Z-9606 (all vials) and (B) survey Z-9701 (vial 1 and
vials 25).
All results are shown for all analytical systems used in each survey
(see Table 2
). Each point represents the mean value of all results
reported for each laboratory.
|
|

View larger version (12K):
[in this window]
[in a new window]
|
Figure 3. CK-2 results determined by mass from (A) survey
Z-9606 (all vials) and (B) survey Z-9701 (vial 1 and vials
25).
All results are shown for all analytical systems used in each survey
(see Table 2
). Each point represents the mean value of all results
reported for each laboratory.
|
|
Statistical assessments of survey results are tabulated in Table 3
(for total CK measurements), in Table 4
(for CK-2 activity measurements), and in Table 5
(for CK-2 mass assays).
Point plots are shown for the adult male upper reference ranges for
total CK activity (Fig. 4
A), CK-2 activity (Fig. 4B
), and CK-2 mass (Fig. 4C
).
Statistical assessment of the total CK reference ranges are tabulated
in Table 6
.

View larger version (15K):
[in this window]
[in a new window]
|
Figure 4. Upper reference limits for adult males, by analyzer type,
reported by all Ontario laboratories licensed to perform total CK or
CK-2 assays.
(A) Total CK activity, (B) CK-2 activity,
(C) CK-2 mass.
|
|
The relation between the male total CK upper reference limits and the
results of each survey for each major analyzer group are shown in Table 7
.
View this table:
[in this window]
[in a new window]
|
Table 7. Coefficients of correlation and determination for the male
upper reference limit and the total CK survey
results.
|
|
 |
Discussion
|
|---|
Total CK activity.
The results obtained for total CK
activity for the major analyzer groups in survey Z-9606 are displayed
in Fig. 1A
. Some instrument groups show a very tight distribution of
resultsnotably the Paramax and RA analyzersalthough others do not.
This has always been a worrisome feature of these survey results. A
statistical analysis of the data is provided in Table 3A. After a
preliminary assessment of the results, we found it necessary to
separate the results of the Vitros analyzers into three instrumental
groups: DT, 250, and the combination of the 500 and 700 analyzers. It
is evident that the Paramax and RA systems provide significantly
different results; many of the other differences are almost certainly
due to the presence of one or two
outliers in several instrument groups. What is of great
concern is that a standard assay system, such as on the Vitros 250, can
produce such a wide range of results.
In survey Z-9610 (Fig. 1B
), the results from the different analyzer
groups, at higher CK activities, have changed. There was no statistical
difference between the individual Vitros analyzers (the DT instrument
was then used by fewer than three laboratories and has been excluded
from these results), so all other Vitros analyzer results have been
consolidated. The tight distributions previously seen for the Paramax
and RA group results are no longer evident, whereas the Dimension and
Hitachi groups display a smaller dispersion of results. Still
noticeable, however, are the significant differences between the
Paramax and all other analyzers (Table 3B).
In survey Z-9701 (Fig. 1C
and 1D
), the two total CK activities were
chosen, respectively, to be below the upper reference limit and to be
midway between the activities used in the previous surveys. Outliers in
the vial 1 results (Beckman, Spectrum, and Paramax) caused an increased
dispersion of the data, leading to significant differences (Table 3C);
the removal of these outliers converted these differences to ones of
nonsignificance. The same outliers in the results from vials 25
(Table 3D) did not alter the statistical assessments.
We therefore conclude that differences between analyzer results exist
across a wide range of total CK activities. Even the use of similar
analyzers does not avoid such variability. As Hyltoft Peterson and
Hørder illustrate in detail, both analytical imprecision and
analytical bias may cause false classification of reported results
(9). As noted above, LPTP CK challenges appear to be
analyzed with acceptable imprecision, so clearly analytical bias is the
major problem. It appears likely, from an inspection of Fig. 1
, that
bias could be both negative and positive, thus further complicating the
effect on patient classifications. A reviewer suggested that the
variability of results could also be due to preanalytical factors such
as delay in analyses or differences in storage conditions. Laboratories
are asked to process LPTP specimens similar to patient specimens, and
indeed each laboratory director is required to certify such handling.
Nonetheless, preanalytical processing is clearly a component
contributing to the reported variability in results.
How are CK measurements processed by each analyzer? The Vitros systems
and the Dimension analyzers use CK calibrators. The remainder use
"self-indicating" reactions based on the reduction of
NADP by the CK assay and its known molar absorptivity
(10). Two possible sources of error suggest themselves:
the CK calibrators are not stable at the time they are used or the
spectrophotometric accuracies of the analyzers are erroneous. Clearly
there is an urgent need for a consensus on the calibration of the CK
assay. Bowers and McComb proposed the concept of an International
Clinical Enzyme Scale and demonstrated the effectiveness of its
application (11)(12). Recent work in the
European Commission has further shown the practicability of using
standard reference materials to remove the variability of enzyme
results in clinical practice (13)(14).
CK-2 activity.
The major techniques used for CK-2
activity determinations are immunoinhibition (Table 2
). The results
obtained in survey Z-9606 (Fig. 2A
) show a very wide variation for
non-J&J immunoinhibition determinations compared with those obtained
with the Vitros analyzers. An analysis of these results (Table 4
) shows
that although there are no differences between these techniques and the
four results obtained by electrophoretic separation of the CK
isoenzymes, there is a highly significant difference between the two
immunoinhibition techniques. In survey Z-9701, no laboratory used
electrophoretic separation (Fig. 2B
) and there continued to be a
significant difference between the immunoinhibition methods. A
comparison between the mass assay results for vial 1 (see below) with
those obtained by the non-J&J analyzers (Fig. 2B
) was made. Serum
containing practically no CK-2 (by mass assay) was reported by non-J&J
immunoinhibition analyzers to have up to 12 U/L, whereas the J&J
analyzers reported values <4 U/L. Users of these non-J&J systems argue
that these high values are matched by higher reference ranges so that
patient misclassifications do not occur. Nonetheless, from an
analytical point of view, the obvious discordance between what CK-2 is
actually present and what is measured must raise doubts about their
accuracy; in addition, there is an adverse impact on the CK-2 fraction
result.
CK-2 mass.
The major technique now used by two-thirds of
all provincial laboratories is the mass assay in seven different
formulations (Table 2
). The results for survey Z-9606 are displayed in
Fig. 3A
. An obvious anomaly are the ACS-180 results. This system is
known to give results about twofold greater than other common
analyzers; Wians et al. (15) derived a regression equation
of ACS 180 (µg/L) = 0.492 2.293 Stratus (µg/L). Results similar
to those shown in Fig. 4A
have occurred regularly in previous LPTP CK-2
surveys. Of concern is that laboratories using the ACS 180 have
identical reference ranges to, for example, the Stratus and IMx
analyzers, although their results are twofold greater. Communication
with users of the ACS 180 brought this data to their attention. They
insisted that their method comparisons justified such reference ranges.
It is also evident that the results produced by the IMx and AxSYM are
significantly different from each other (Table 5
), presumably due to
differences in calibration. Apart from this aspect, however, users of
the IMx analyzer reported results ranging from 25 to 44 µg/L for a
sample containing about 35 mg/L CK-2. Several significant differences
between other analyzers were also observed (Table 5
). In survey Z-9701,
we used lower values for CK-2 to check if the differences observed in
survey Z-9606 occurred across the range of values found in clinical
practice (Fig. 3B
). Vial 1 contained very small quantities of CK-2, but
significant differences were still observed between several
analyzers (Table 5
). Of interest are the ACS 180 results. In this
survey the newly issued recalibrated CKMB II assay (giving results
comparable with other mass assay systems) was used by three of the four
participants to obvious effect. The differences noted between the IMx
and AxSYM results in survey Z-9606 are clearly evident, again, in the
results for vials 25 in survey Z-9701 (Fig. 3B
and Table 5
). Again,
the IMx analyzer reported results ranging from 7 to 28 µg/L on a
specimen containing about 20 µg/L. The Stratus, Access, and the ICON
also show significant differences (Table 5
).
The differences between the results of many mass assays are due, to an
extent, to differences in calibration. Again, the forthcoming
availability of a CK-2 mass standard will reduce this variability
(16)(17). That it will likely not remove it
entirely is evident from the spread of results shown in Fig. 3A
and B.
Reference ranges.
We previously reported on the wide
dispersion of total CK and CK-2 reference ranges in Ontario found in
1991 (18). Subsequent to these findings, the Committee
encouraged many laboratories to revise or modify their reference
values. The considerable effects of preanalytical variables on these
analytes has been extensively documented (19). Such
effects create considerable difficulties in establishing satisfactory
limits. This topic has recently been comprehensively reviewed and
detailed advice on analytical goals for generating reference intervals
is available (20)(21).
Currently, the reported male upper reference limits for total CK
throughout the province form a gaussian curve with a mean value of 199
U/L (SD = 23) with neither kurtosis nor skewing. The individual
values, by instrument group, are shown in Fig. 4A
. A statistical
assessment of these values shows that the upper reference ranges
reported for the Paramax analyzer are significantly different from all
but three analyzer groups (Table 6
). Other analyzer groups also show a
few significant differencesVitros, Mira, CX, Dimension, and the RA
systems. Is there a relation between the individual values for each
analyzer and the results from the three surveys? The Pearson
product-moment coefficients of correlation (r) and
determination (r) have been tabulated (Table 7
). An r value less than about 0.4 can be regarded as
showing little strength of association between the variables, and this
appears to be the case for most analyzers with the notable exceptions
of the Vitros DT, Dimension, and RA analyzers. The influence of the
upper reference limit on the survey CK results is indicated by the
r value, when expressed as a percentage. For
example, for the Vitros 250, only 2% of the variation between the
results from survey Z-9606 is accounted for by the upper reference
limits for these analyzers, whereas for the Dimension analyzers, in the
same survey, nearly 70% of the variation is accounted for by the
reference limits (Table 7
).
The male upper reference limits for CK-2 activity and mass are shown in
Fig. 4B
and C, respectively. Whereas the J&J ranges vary between 5 and
16 U/L, the range for other immunoinhibition assays rise as high as 25
U/L (Fig. 4B
). Mass values for the upper reference range do not exceed
10 µg/L, but in the case of the IMx may be as low as 5 µg/L. In two
recent reports investigating troponin T and troponin I values in the
acute coronary syndromes, both groups used the Stratus analyzer for
CK-2 analyses. One group used 7 µg/L and the other group used 5
µg/L as their upper reference limits
(22)(23). In Ontario, the ranges reported for
the same analyzer were from 3 to 7 µg/L. One Stratus user reported an
upper reference limit of 3 µg/L, whereas values of 4 µg/L were used
by participants with the Immuno 1 and Access analyzers; these values
appear to be far too low.
The Pearson product-moment coefficients of correlation (r)
and determination (r) were also calculated for
CK-2 activity and mass assays. Both the J&J and non-J&J
immunoinhibition assays and the Stratus and IMx assays showed
insignificant r values (and therefore
r values), but the AxSYM assay results showed
high r (0.65) and r (42%) values in
the Z-9606 survey but not in the Z-9701 survey. It is not evident
why this should occur except that the former survey had CK-2 values
twofold the latter.
Do these variations matter? As noted earlier, the WHO MONICA Project
for the diagnosis of myocardial infarction codes an enzyme result as
abnormal if it is more than twice the "upper limit of normal" for
that test in each laboratory (7). It is evident
that the wide range in upper reference limits will undoubtedly alter
the diagnostic classification. A report from the Minnesota Heart Survey
noted that the incorporation of total CK and CK-2 into the Survey's
diagnostic criteria significantly increased the "definite"
myocardial infarction rates (24). But such classification
will be distorted if either or both of the laboratory's upper
reference limits or the analytical value of the determinations are
incorrect. These possibilities are very evident in this recent Ontario
experience. In this regard, the existence of significant interobserver
variation in the interpretation of serum enzyme changes in myocardial
infarction may further complicate patient classification
(25).
We therefore conclude that currently the routine analyses of total
CK and CK-2 in Ontario is unsatisfactory, lacking congruity between
laboratories both as regards analytical results and reference ranges.
 |
Acknowledgments
|
|---|
This work was funded entirely by the Laboratory Proficiency Testing
Program.
 |
Footnotes
|
|---|
1 Address reprint requests to this author at the Laboratory Proficiency
Testing Program. 
 |
References
|
|---|
-
Lohmann RC, Wood DE, Jacobs WI, Elliot ML. Reliability of white blood cell counting. Arch Pathol Lab Med 1989;13:989-994.
-
McQueen MJ, Patten RL, Henderson AR, Krishnan S, Wood DE, Webb S. Proficiency testing for creatine kinase isoenzyme CK-2 (CK-MB) in Ontario. Clin Chem 1990;36:2102-2105.
[Abstract/Free Full Text]
-
Henderson AR, Stark JA, McQueen MJ, Patten RL, Krishnan S, Wood DE, Webb S. Is the determination of creatine kinase-2 after electrophoretic separation accurate?. Clin Chem 1994;40:177-183.
[Abstract/Free Full Text]
-
Gerhardt W, Wulff K. Creatine kinase. In: Bergmeyer HU,
Bergmeyer J, Grassl M, eds. Methods of enzymatic analysis, Vol.
III, 3rd ed.: enzymes 1: oxidoreductases, transferases. Weinheim:
Verlag-Chemie, 1983:50839..
-
Hørder M, Elser RC, Gerhardt W, Mathieu M, Sampson EJ. Approved IFCC recommendation on methods for the measurement of catalytic concentration of enzymes: Part 7. IFCC method for creatine kinase (ATP:creatine N-phosphotransferase, EC 2.7.3.2). Eur J Clin Chem Clin Biochem 1991;29:435-456.
[ISI][Medline]
[Order article via Infotrieve]
-
Henderson AR, Webb S, Richardson H, Wood DE. A comparison of pooled, fresh-frozen, and lyophilized sera as a matrix for enzyme proficiency testing: the experience of the Ontario Laboratory Proficiency Testing Program. Clin Biochem 1996;29:183-185.
[ISI][Medline]
[Order article via Infotrieve]
-
. WHO MONICA Project,Tunstall-Pedeo H, Kuulasmaa K, et al. Myocardial infarction and coronary deaths in the World Health Organization MONICA project. Registration procedures, event rates, and case-fatality rates in 38 populations from 21 countries in four continents. Circulation 1994;90:583-612.
[Abstract/Free Full Text]
-
Conover WJ. Statistics of the KolmogorovSmirnov type. Practical nonparametric statistics 2nd ed. 1980:344-393 John Wiley and Sons New York. .
-
Hyltoft Petersen P H, ørder M. Influence of analytical quality on test results. Scand J Clin Lab Invest 1992;52(Suppl 208):65-88.
[Medline]
[Order article via Infotrieve]
-
Moss DW, Henderson AR. Enzymes. Burtis CA Ashwood ER eds. Tietz textbook of clinical chemistry 2nd ed. 1994:735-896 WB Saunders Co. Philadelphia. .
-
Bowers GN, Jr, McComb RB. A unifying reference system for clinical enzymology: aspartate aminotransferase and the International Clinical Enzyme Scale [Review]. Clin Chem 1984;30:1128-1136.
[Abstract/Free Full Text]
-
McComb RB, Bowers GN, Jr. Alkaline phosphatase and the International Clinical Enzyme Scale. Am J Clin Pathol 1985;84:67-73.
[ISI][Medline]
[Order article via Infotrieve]
-
Moss DW, Maire I, Calam DH, et al. Reference materials in clinical enzymology: preparation, requirements and practical interests. Ann Biol Clin (Paris) 1994;52:189-198.
[Medline]
[Order article via Infotrieve]
-
Lessinger J, Férard G, Grafmeyer D, et al. Usefulness of reference materials in calibration of enzyme activities. Eur J Clin Chem Clin Biochem 1995;33:859-864.
[ISI][Medline]
[Order article via Infotrieve]
-
Wians FH, Baskin LB, Keffer JH, Koch DD. Evaluation and implementation of an immunochemiluminometric assay (ICMA) for quantifying CK-MB concentration. Clin Lab Sci 1995;8:334-338.
-
Vaidya H, Apple F, Boches F, Christensen R, Jones K, Landt Y, et al. Preparation of preliminary standards for creatine kinase-MB (CKMB) mass immunoassays [Abstract]. Clin Chem 1993;39:1256.
-
Green S, Onoroski M, Moore R, Wu A, Lehrer M, Vaidya H. Use of proposed CK-MB standardization material for calibration of mass measurements: clinical correlation of patient sera [Abstract]. Clin Chem 1993;39:1269.
-
Henderson AR, McQueen MJ, Patten RL, Krishnan S, Wood DE, Webb S. Testing for creatine kinase and creatine kinase-2 in Ontario: reference ranges and assay types. Clin Chem 1992;38:1365-1370.
[Abstract/Free Full Text]
-
Young DS. Effects of preanalytical variables on clinical
laboratory tests, 2nd ed. Washington: AACC Press, 1997:1-15-11..
-
Harris EK, Boyd JC. Statistical bases of reference values
in laboratory medicine. New York: Marcel Dekker, 1995:361pp..
-
Gowans EM, Hyltoft Petersen P, Blaabjerg O, Horder M. Analytical goals for the acceptance of common reference intervals for laboratories throughout a geographical area. Scand J Clin Lab Invest 1988;48:757-764.
[ISI][Medline]
[Order article via Infotrieve]
-
Ohman EM, Armstrong PW, Christenson RH, et al. Cardiac troponin T levels for risk stratification in acute myocardial ischemia. N Engl J Med 1996;335:1333-1341.
[Abstract/Free Full Text]
-
Antman EM, Tanasijevic MJ, Thompson B, et al. Cardiac-specific troponin I levels to predict the risk of mortality in patients with acute coronary syndromes. N Engl J Med 1996;335:1342-1349.
[Abstract/Free Full Text]
-
Burke GL, Edlavitch SA, Crow RS. The effects of diagnostic criteria on trends in coronary heart disease morbidity: the Minnesota Heart Survey. J Clin Epidemiol 1989;42:17-24.
[ISI][Medline]
[Order article via Infotrieve]
-
Gjørup T, Kelbæk H, Sørensen F, Stenbygård L. Reliability in the interpretation of the course of serum enzyme concentration changes in suspected acute myocardial infarction. Int J Cardiol 1992;35:43-47.
[ISI][Medline]
[Order article via Infotrieve]