|
|
||||||||
Hematology |
Department of Pathology, Virginia Commonwealth University/Medical College of Virginia, Richmond, VA.
a Address correspondence to this author at: Department of Pathology, Medical College of Virginia, P.O. Box 980250, Richmond, VA 23298-0250. Fax 804-828-2812; e-mail JBENEZRA{at}HSC.VCU.EDU.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
materials and methods
We studied 252 freshly collected urine samples submitted for
diagnostic urinalysis to our laboratory in a tertiary-care university
hospital. After the routine diagnostic microscopic urinalysis was
performed, the samples were analyzed the same day on the UF-100
analyzer; the use of such discarded material has been approved by our
institution's IRB panel. To reduce interobserver variability, the same
technologist performed all the microscopic urinalyses with the same
microscope, using the KOVA® system (Hycor Biomedical)
(8). The areas of a low-power field (LPF) and high-power
field (HPF) were determined with a measured scale to be able to
correlate the cells or particles seen in a microscopic field (usually
measured as cells or particles/LPF or HPF) with the quantitative
results of the UF-100, which are provided in units of cells (or
particles)/µL.1
Between-run quality-control samples consisted of suspensions of particles, provided by the manufacturer, in the sizes of bacteria, erythrocytes (RBCs), leukocytes (WBCs), and squamous epithelial cells (SECs). Within-run imprecision was determined by analyzing specimens with various concentrations of RBCs, WBCs, SECs, and bacteria 1220 times (usually 20) each on the UF-100 analyzer. Linearity was determined by analyzing in triplicate specimens and their dilutions of 1:2, 1:4, 1:5, 1:10, 1:20, and 1:40; the slope, intercept, and proportion of measured value to expected value were determined with EP-Evaluator (Rhoads). Carryover analysis was performed by analyzing the specimen in triplicate, followed by three blank specimens; the percent carryover was determined by the formula:
Carryover = (blank 1 - blank 3)/(specimen 3 - blank 3)
Statistical analysis, including the gamma statistic
(9) for measuring correlation and the McNemar test for
measuring change in the distribution of two dichotomous variables by
the
test, was performed by using SPSS®
for Windows. Gamma is a measure of association between two variables
measured on an ordinal level, and can be thought of as the probability
that a random pair of observations is concordant minus the probability
that the pair is discordant, assuming the absence of ties. Gamma is
symmetric and ranges between 0 and 1.
| Results |
|---|
|
|
|---|
|
For the WBC data (Table 2
), the gamma was 0.970. Analyzing the data by clinical positive
vs negative results (5 WBC/HPF = 26 WBC/µL), we again found
significant differences between methods (P <0.00001).
|
For the SEC data (Table 3
), the gamma was 0.880. With results categorized by clinical
positive vs negative (5 SECs/LPF = 1.95 SECs/µL), the methods
differed (McNemar test, P <0.00001). Samples in which the
UF-100 "overcalled" epithelial cells contained casts and
Trichomonas organisms. The protein in these samples most
likely caused WBCs to aggregate, resulting in a possibly falsely high
SEC count.
|
The UF-100 detected many types of casts, including hyaline, granular,
and cellular casts. For these data (Table 4
), the gamma was 0.880. Analyzing the data by clinical positive
vs negative results (1 cast/LPF = 0.39 casts/µL), we found
significant differences between the two methods (McNemar test,
P <0.00001).
|
precision
Within-run CVs for the RBC analysis are shown in Table 5
. The CV ranged from 33% for low numbers of RBCs (5 RBCs/µL)
to <3% at RBC concentrations >660/µL. Between-run precision was
6% at a measured mean of 50 RBC/µL and 2.7% at a measured mean RBC
of 256/µL.
|
Within-run precision for the WBC analysis is shown in Table 5
. The CV
ranged from 24% for low numbers of WBCs (4 WBCs/µL) to 6% at a WBC
count of 853/µL. Between-run precision was 27% at 54 WBC/µL and
4.9% at a WBC of 228/µL.
Within-run precision ranged from 24% at 3.42 SECs/µL to 2.9% at 180 SECs/LPF. Between-run precision was 9.1% at 22 SECs/LPF and 7.5% at 88 SECs/LPF.
Within-run precision for casts varied from 75% for samples with <0.42 casts/µL to 17% for samples with 4.6 casts/µL.
The within-run CVs for bacterial counts (Table 5
) was 212% over the
concentration of bacteria tested. Between-run CV was 29% at 54
bacteria/µL, 37% at 60 bacteria/µL, and 26% at 146 bacteria/µL.
linearity
Dilutions of urine samples were made as described in
Materials and Methods. For casts, dilutions of 1:2 and 1:4
were made from a sample with 18 casts/µL, and the specimens were
analyzed in triplicate.
For samples with RBC concentrations of 3864 to 15 956 cells/µL, the measured values were 87105% of expected. For three samples with WBC concentrations of 5698 to 5735 cells/µL, values were 92112% of expected. For bacteria (two samples), results were 111% (at 1809 cells/µL) and 133% (4433 cells/µL) of expected, and for casts (one sample) 106%.
carryover
Carryover studies were performed on four urine samples of various
RBC, WBC, SEC, cast, and bacterial counts. Carryover ranged from
-0.13% to 0.04% for RBC of 13 to 2404 cells/µL, -0.62% to 0.03%
for WBC counts of 12 to 855 cells/µL, -1.6% to 0.14% for SEC of
3.0 to 72 cells/µL, 0.00% to 0.44% for 0.26 to 28 casts/µL, and
-0.23% to 0.29% for bacterial counts of 277 to 2744 cells/µL.
These results show that no substantial carryover was detected in any of
the samples, demonstrating that the carryover was not systematic.
| Discussion |
|---|
|
|
|---|
The Yellow IRIS® urinalysis instrument has automated the microscopic evaluation of urine sediment. Several studies have demonstrated that it has increased precision over routine manual urinalysis, and that it can detect more abnormalities than conventional microscopy (1)(3)(4)(6)(7). However, a technologist must be physically present at the instrument to characterize events, and some technologist interpretation of images is required, thus leading to imprecision and inaccuracy. The UF-100 instrument analyzes urine cells and particles on the basis of flow cytometric principles, analogous to the manner by which modern hematology analyzers perform complete blood and differential cell counts.
The UF-100 analyzer in this study detected more RBCs, WBCs, and SECs than did manual microscopy; similar findings have also been reported for the Yellow IRIS (1). This probably represents true detection of cells not identified by manual microscopy, since routine urinalysis has centrifugation, decantation, and resuspension steps that are either incomplete or lead to cellular loss and lysis. Nonetheless, agreement between the manual and UF-100 methods for these cellular elements was excellent (gamma = 0.880.97). Samples in which the UF-100 "overcalled" epithelial cells contained casts and Trichomonas organisms. The protein in these samples possibly caused WBCs to clump, resulting in a possibly falsely high SEC count.
Detection of casts with the UF-100 was less robust than was the detection of cellular elements; this has also been seen with the Yellow IRIS (6), and may be due to differences in the volume of urine sample analyzed. We chose not to evaluate accuracy of bacteria detection, since the "gold standard," quantifying bacterial counts in culture, measures something qualitatively different, live bacteria, from what the UF-100 urine analyzer measures, live and dead bacterial particles. This is a limitation of all automated urinalysis analyzers (10). Nonetheless, the UF-100 flags the presence of casts, crystals, and bacteria, alerting the laboratory to their presence and the need to characterize them further under the microscope.
Automated urinalysis performed by the UF-100 displayed good precision for analyzing cellular elements, bacteria, and casts; the CVs were much less than those reported historically for manual urinalysis (1)(3)(4), and are comparable with those seen with other automated instruments. Moreover, the assays were linear over clinically useful ranges, and no carryover was found with the UF-100 analyzer for the analytes measured.
Comparison of the UF-100 with manual microscopy, the gold standard, is somewhat difficult, since this standard is somewhat tarnished by several methodological steps that lead to imprecision and inaccuracy (1)(2). Our studies show that the UF-100 urine analyzer's results are comparable with those of manual microscopy and published reports of the performance of the Yellow IRIS urine analyzer. In addition, counts of cellular elements with the UF-100 are comparable with those obtained on unspun urines analyzed under the microscope with a hemocytometer (manuscript in preparation). We conclude that the UF-100 can perform accurate and precise quantification of microscopic elements in urine with little or no operator interaction.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
P. Froom, R. Etzion, and M. Barak What Is an Abnormal Test Strip Urinary Erythrocyte Concentration? Clin. Chem., March 1, 2004; 50(3): 673 - 675. [Full Text] [PDF] |
||||
![]() |
J. Penders, T. Fiers, A. M. Dhondt, G. Claeys, and J. R. Delanghe Automated flow cytometry analysis of peritoneal dialysis fluid Nephrol. Dial. Transplant., February 1, 2004; 19(2): 463 - 468. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Gai, G. B. Piccoli, G. P. Segoloni, and G. Lanfranco Microscopic Urinalysis and Automated Flow Cytometry in a Nephrology Laboratory Clin. Chem., September 1, 2003; 49(9): 1559 - 1560. [Full Text] [PDF] |
||||
![]() |
C. Ottiger and A. R. Huber Quantitative Urine Particle Analysis: Integrative Approach for the Optimal Combination of Automation with UF-100 and Microscopic Review with KOVA Cell Chamber Clin. Chem., April 1, 2003; 49(4): 617 - 623. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Penders, T. Fiers, and J. R. Delanghe Quantitative Evaluation of Urinalysis Test Strips Clin. Chem., December 1, 2002; 48(12): 2236 - 2241. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kouri, L. Vuotari, S. Pohjavaara, and P. Laippala Preservation of Urine for Flow Cytometric and Visual Microscopic Testing Clin. Chem., June 1, 2002; 48(6): 900 - 905. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Zaman, S. Roggeman, and J. Verhaegen Unsatisfactory Performance of Flow Cytometer UF-100 and Urine Strips in Predicting Outcome of Urine Cultures J. Clin. Microbiol., November 1, 2001; 39(11): 4169 - 4171. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Apeland, O. Mestad, and O. Hetland Assessment of haematuria: automated urine flowmetry vs microscopy Nephrol. Dial. Transplant., August 1, 2001; 16(8): 1615 - 1619. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. T. Van Acker, J. R. Delanghe, M. R. Langlois, Y. E. Taes, M. L. De Buyzere, and A. G. Verstraete Automated Flow Cytometric Analysis of Cerebrospinal Fluid Clin. Chem., March 1, 2001; 47(3): 556 - 560. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Ziebig, A. Lun, and P. Sinha Leukocyte Counts in Cerebrospinal Fluid with the Automated Hematology Analyzer CellDyn 3500 and the Urine Flow Cytometer UF-100 Clin. Chem., February 1, 2000; 46(2): 242 - 247. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Lun, R. Ziebig, H. Hammer, U. Otting, G. Filler, and P. Sinha Reference Values for Neonates and Children for the UF-100 Urine Flow Cytometer Clin. Chem., October 1, 1999; 45(10): 1879 - 1880. [Full Text] [PDF] |
||||
![]() |
A. Lun, R. Ziebig, F. Priem, G. Filler, and P. Sinha Routine Workflow for Use of Urine Strips and Urine Flow Cytometer UF-100 in the Hospital Laboratory,1 Clin. Chem., August 1, 1999; 45(8): 1305 - 1307. [Full Text] [PDF] |
||||
![]() |
M. R. Langlois, J. R. Delanghe, S. R. Steyaert, K. C. Everaert, and M. L. De Buyzere Automated Flow Cytometry Compared with an Automated Dipstick Reader for Urinalysis Clin. Chem., January 1, 1999; 45(1): 118 - 122. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |