|
|
||||||||
Molecular Diagnostics and Genetics |
1
Neo Gen Screening, 110 Roessler Road, Pittsburgh, PA 15220.
2
California Department of Health Services, Genetic
Disease Branch, 2151 Berkeley Way, Annex 4, Berkeley, CA 94704.
3
Division of Medical Genetics, Department of Pediatrics,
Duke University Medical Center, Research Triangle Park, NC 27709.
a Author for correspondence. Fax 412-341-8926; e-mail dhchace{at}neogenscreening.com.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Few studies have been described that have systematically examined the quantification of Phe in blood spots collected before 24 h. To reduce or eliminate false-negative results with these early specimens, Cunningham et al. (3) reduced the diagnostic cutoff to 258 µmol/L (4.3 mg/dL) for a fluorometric technique. However, they observed the relatively high rate of 4% for false-positive results. An initial study using HPLC and tandem mass spectrometry (MS/MS) demonstrated the ability to quantify Phe and Tyr simultaneously (4); it has also been suggested that calculation of the ratio of Phe to Tyr (Phe/Tyr) will allow detection of affected newborns under 24 h of age without increasing the rate of false positives (5). This study was designed to test this hypothesis by comparing the results of previous analyses by fluorometry with those by MS/MS, in which Phe and Tyr are quantified, in samples collected from newborns <24 h of age.
| Materials and Methods |
|---|
|
|
|---|
258 µmol/L (4.3 mg/dL) were
classified as positive. An additional specimen was collected from each
infant with a positive result. If the subsequent specimen was negative,
the result was classified as a false positive. Metabolic specialists
made the final diagnoses of cases of classical PKU or variant PKU. The
group classified as variant PKU had a mean Phe concentration of 430
µmol/L (7.1 mg/dL) in the second specimen. The group classified as
classical PKU had a mean Phe in the second specimen of 1188 µmol/L
(19.6 mg/dL). These specimens were assigned study numbers and sent
blinded to the Mass Spectrometry Facility at Duke University Medical
Center for analysis by MS/MS. The specimens are categorized in Table 1
|
Quantitative amino acid profiles were obtained by isotope dilution liquid secondary ion MS/MS, using methods described previously (4). A cutoff of 180 µmol/L (3.0 mg/dL) was used to identify increased Phe, and a Phe/Tyr molar ratio of 2.5 was used to provide additional evidence of a metabolic defect in the conversion of Phe to Tyr, indicative of a primary hyperphenylalaninemia. These cutoffs were based on values used in a previous study (4) for negative samples collected >24 h post delivery. Results were sent to the California Genetic Disease Branch where the study was unblinded.
| Results |
|---|
|
|
|---|
|
The results of Phe quantification (µmol/L) by MS/MS are summarized in
Table 2
. Fig. 2
is a scatter plot of the data set. It should be noted that all
but three specimens originally classified as false positives were
correctly categorized by MS/MS as negative. Comparison of results
obtained by fluorometry with those obtained by MS/MS reveals a high
correlation, with a Pearson correlation coefficient of 0.817. In Fig. 2
, the cutoff value for MS/MS is 180 µmol/L (3.0 mg/dL), in
comparison with the fluorometric cutoff value of 258 µmol/L (4.3
mg/dL). The results of the Tyr quantification (µmol/L) by MS/MS are
also summarized in Table 2
. A scatter plot of each data point for Tyr
(data not shown) showed no partitioning of data points by sample
groups, i.e., negative, positive for PKU, or positive for
hyperphenylalaninemia, in contrast to that observed for Phe, although
the Tyr value for infants with PKU was distinctly lower than those for
the other groups. The values for the Phe/Tyr ratio are summarized in
Table 2
, and a scatter plot of each data point is shown in Fig. 3
, using a Phe/Tyr molar ratio of
2.5 to confirm the diagnosis
of hyperphenylalaninemia. The Phe/Tyr ratio eliminated the
false-positive designation from two of the three specimens that had
remained in the false-positive category when only the Phe value
obtained by MS/MS was used. Thus, with values for both Phe and Phe/Tyr,
only one specimen that was originally false positive by fluorometric
analysis remained false positive by MS/MS, whereas the other 90
specimens originally classified as false positives were negative. The
relationships between Phe concentration (µmol/L) and Phe/Tyr ratio
determined by MS/MS and time of collection (hours) among the groups and
in one infant with PKU from whom serial samples were collected during
the first 24 h of life are presented in Figs. 4
and
5, respectively. In this infant, both the Phe value and the
Phe/Tyr ratio rose consistently during the first 24 h of life. In
other infants with PKU or variant hyperphenylalaninemia, the Phe value
also tended to be higher at increasing hours of age within the first
24 h (Fig. 4
). This was not true for the Phe/Tyr ratio; however,
the value at every hour of age sampled was above the cutoff value (Fig. 5
).
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The addition of the Phe/Tyr molar ratio determined by MS/MS with a cutoff of 2.5 allowed the accurate detection of PKU and variant hyperphenylalaninemia in all infants originally determined to be positive, with only one false-positive result. Thus, although both fluorometry and MS/MS detected classical PKU or hyperphenylalaninemia in specimens collected <24 h post delivery, including one collected as early as 4 h post delivery, MS/MS greatly reduced the number of false-positive identifications.
The differences in quantitative cutoff values between MS/MS and fluorometry may be attributed to predictable fluorescent interference, which raises the concentration of Phe. However, we have found that MS/MS results are slightly lower when compared with other methods as well, specifically HPLC and the bacterial inhibition assay. Additional work is in progress to compare MS/MS with these methods. However, previous work has shown a good correlation for HPLC and MS/MS for the measurement of Phe in plasma samples (4). Additional comparative studies are in progress.
The measurements of Phe concentration in unaffected newborns by MS/MS
and fluorometry are tightly clustered around the median of 68 µmol/L
(Fig. 2
). The phenylalanine hydroxylase deficiency that produces
classical PKU or variant hyperphenylalaninemia dictates a high Phe/Tyr
ratio in association with the increased Phe. As shown in Fig. 3
, the
initial negative (normal) samples are tightly clustered around a median
molar Phe/Tyr ratio of 0.73. There is one negative sample (
) that is
clearly an outlier and near the cutoff value of 2.5. This sample had
increased Met as well as increased Phe and decreased Tyr. Review of the
case identified this specimen as being from a newborn receiving
intravenous feeding. The abnormal Leu/Phe and Met/Phe ratios indicate
that the sample did not have either classical PKU or variant
hyperphenylalaninemia. The range of the Phe/Tyr ratio for initial
false-positive samples is 0.911.07, which is only slightly increased
compared with initial negative samples. Only one false-positive sample
was greater than the cutoff of 2.5. All of the classical PKU or variant
hyperphenylalaninemia patient specimens had Phe/Tyr ratios >2.5.
It is noteworthy that 88 of 91 false-positive results had a Phe concentration within the reference interval by MS/MS, suggesting that most false-positive results previously detected by fluorometry were attributable to interference from other amino acids or from other compounds present in the blood, rather than an actual increase of Phe. This result also supports previous work that indicated that MS/MS is a more accurate method for measuring Phe concentration. Moreover, the Phe/Tyr ratio determined by MS/MS in this study is a molar ratio, rather than the signal-intensity ratio used by many other methods. Therefore, data using a Phe/Tyr molar ratio may not be directly comparable to the ratios of signal intensities published in other articles.
In specimens collected at various times during the first 24 h post
delivery from infants with PKU or variant hyperphenylalaninemia, as
depicted in Figs. 4
and 5
, the concentration of Phe increases over
time, becoming >180 µmol/L at 6 h. The Tyr concentration does
not show a trend toward decreasing concentration with time (data not
shown). In the one infant studied serially, the Phe/Tyr ratio became
greater than the cutoff value by 6 h and then plateaued. The rate
of increase of the Phe/Tyr ratio over time is not as noticeable as the
rate of change for Phe.
When MS/MS was used, the newborn detection of infants confirmed to be positive for PKU or variant hyperphenylalaninemia was equally valid using either Phe or the Phe/Tyr ratio. The precision of the method, however, was greater for the Phe/Tyr ratio than for Phe alone (data not shown). These results demonstrate the utility of MS/MS in the detection of PKU in newborns discharged early and the reduction of false-positive results achieved through the higher accuracy of the Phe measurements and the simultaneous determination of the Phe/Tyr ratio.
MS/MS can diagnose PKU with an extremely low false-positive rate, only ~1/100 that of fluorometry, with excellent accuracy and precision, as described previously (4). With the cutoff values that we have used, it is anticipated that few if any false-negative results will occur. Because the expected false-positive rate is so dramatically lower, the cutoff concentration could be set to <180 µmol/L with only a small increase in the false-positive rate. However, this is a decision that must be made in individual laboratories and through experience. At this time, with the experience of testing >550 000 blood specimens from newborns by MS/MS at Neo Gen Screening, no known false-negative results have occurred.
In addition to detection of PKU, MS/MS can also detect other aminoacidopathies, including maple syrup disease (6) and homocystinuria (7). Because a single analysis measures not only Phe and the Phe/Tyr ratio but also amino acids other than Phe and Tyr, the method is very cost-effective. In addition to other aminoacidopathies, MS/MS adds newborn screening for the organic acid disorders and disorders of fat metabolism, such as medium chain acyl-CoA dehydrogenase deficiency (8)(9). The technique is cost-effective and meets the current need to expand methods of preventive medicine. Because of its low false-positive rates, this method is an efficient way to screen for large numbers of disorders in individuals while minimizing the cost of follow up because of low false-positive rates. The cost-benefits of MS/MS and its applications in newborn screening seem to be solidly based (10).
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
E. Altmaier, S. L. Ramsay, A. Graber, H.-W. Mewes, K. M. Weinberger, and K. Suhre Bioinformatics Analysis of Targeted Metabolomics--Uncovering Old and New Tales of Diabetic Mice under Medication Endocrinology, July 1, 2008; 149(7): 3478 - 3489. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ho, Z. Lukacs, G. F. Hoffmann, M. Lindner, and T. Wetter Feature Construction Can Improve Diagnostic Criteria for High-Dimensional Metabolic Data in Newborn Screening for Medium-Chain Acyl-CoA Dehydrogenase Deficiency Clin. Chem., July 1, 2007; 53(7): 1330 - 1337. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. I. Kaye and and the Committee on Genetics Newborn Screening Fact Sheets Pediatrics, September 1, 2006; 118(3): e934 - e963. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. H. Chace, T. A. Kalas, and E. W. Naylor Use of Tandem Mass Spectrometry for Multianalyte Screening of Dried Blood Specimens from Newborns Clin. Chem., November 1, 2003; 49(11): 1797 - 1817. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Schulze, M. Lindner, D. Kohlmuller, K. Olgemoller, E. Mayatepek, and G. F. Hoffmann Expanded Newborn Screening for Inborn Errors of Metabolism by Electrospray Ionization-Tandem Mass Spectrometry: Results, Outcome, and Implications Pediatrics, June 1, 2003; 111(6): 1399 - 1406. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Sweetman Newborn Screening by Tandem Mass Spectrometry: Gaining Experience Clin. Chem., November 1, 2001; 47(11): 1937 - 1938. [Full Text] [PDF] |
||||
![]() |
T. H. Zytkovicz, E. F. Fitzgerald, D. Marsden, C. A. Larson, V. E. Shih, D. M. Johnson, A. W. Strauss, A. M. Comeau, R. B. Eaton, and G. F. Grady Tandem Mass Spectrometric Analysis for Amino, Organic, and Fatty Acid Disorders in Newborn Dried Blood Spots: A Two-Year Summary from the New England Newborn Screening Program Clin. Chem., November 1, 2001; 47(11): 1945 - 1955. [Abstract] [Full Text] [PDF] |
||||
![]() |
National Institutes of Health Consensus Developmen National Institutes of Health Consensus Development Conference Statement: Phenylketonuria: Screening and Management, October 16-18, 2000 Pediatrics, October 1, 2001; 108(4): 972 - 982. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Enns Newborn Screening by Tandem Mass Spectrometry NeoReviews, August 1, 2001; 2(8): e201 - 207. [Full Text] [PDF] |
||||
![]() |
D. Matern and M. J. Magera Mass Spectrometry Methods for Metabolic and Health Assessment J. Nutr., May 1, 2001; 131(5): 1615S - 1620. [Abstract] [Full Text] |
||||
![]() |
Serving the Family From Birth to the Medical Home. Newborn Screening: A Blueprint for the Future - A Call for a National Agenda on State Newborn Screening Programs Pediatrics, August 1, 2000; 106(2): 389 - 422. [Full Text] |
||||
![]() |
E. W. Naylor and D. H. Chace Automated Tandem Mass Spectrometry for Mass Newborn Screening for Disorders in Fatty Acid, Organic Acid, and Amino Acid Metabolism J Child Neurol, November 1, 1999; 14(1_suppl): S4 - S8. [Abstract] [PDF] |
||||
![]() |
D. H. Chace, B. W. Adam, S. J. Smith, J. R. Alexander, S. L. Hillman, and W. H. Hannon Validation of Accuracy-based Amino Acid Reference Materials in Dried-Blood Spots by Tandem Mass Spectrometry for Newborn Screening Assays Clin. Chem., August 1, 1999; 45(8): 1269 - 1277. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. L. Levy Newborn Screening by Tandem Mass Spectrometry: A New Era Clin. Chem., December 1, 1998; 44(12): 2401 - 2402. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |