|
|
||||||||
Lipids and Lipoproteins |
a Address correspondence to this author at: Laboratorio de Análisis Clínicos, Sección de Bioquímica, Hospital Universitario "Del Río Hortega", calle Cardenal Torquemada, s/n, 47010 Valladolid, Spain. Fax 34-983-331566; e-mail marisa.arranz{at}mx2.redestb.es.
| Abstract |
|---|
|
|
|---|
-cyclodextrin sulfate (PEGME; Kyowa); a homogeneous
method based on polyanionpolymer/detergent (PPD; Daiichi); the usual
precipitation method with phosphotungstic acid/MgCl2 (PTA);
and an ultracentrifugation (UC) procedure. The homogeneous
HDL-cholesterol assays (performed with automated analyzers) were
precise and correlated well with the PTA and UC assays. The specificity
and accuracy of the PEGME method were better than those of the PPD
method. | Introduction |
|---|
|
|
|---|
In laboratories, HDL-C has been determined by techniques such as ultracentrifugation (UC) (6), electrophoresis (7), HPLC (8), and precipitation-based methods (9). Of these, the precipitation-based methods have been the most widely used in clinical laboratories. However, they are time-consuming, require large serum volumes, and are not suited to automated analysis; these restrictions have encouraged researchers to focus on homogeneous assays.
Here we report our assessment of the analytical performance of two
homogeneous methods designed for use in routine chemical laboratory
work: a polyethylene glycol-modified enzymes/
-cyclodextrin sulfate
(PEGME) assay from Kyowa (10) and a
polyanionpolymer/detergent (PPD) method from Daiichi
(11)(12). We compared the HDL-C
concentrations obtained by these procedures with the phosphotungstic
acid/MgCl2 precipitation (PTA) method (13)(14)(15)(16)
and a UC method that uses centrifugation to remove triglyceride-rich
lipoproteins from plasma (17). The PTA method has been
chosen because most laboratories use it as a routine method. The UC
comparison method has been also included because it is routinely used
as the comparison method for determining accuracy.
| Materials and Methods |
|---|
|
|
|---|
hdl-c concentrate isolation
HDL-C was isolated and concentrated by precipitating the VLDL and
LDL particles from a serum pool with a mixture of 0.44 mmol/L
phosphotungstic acid and 20 mmol/L MgCl2 at a ratio of
1:3.5. After agitation for 30 s (to assure proper homogeneization)
the mixture was left at room temperature for 10 min and then
centrifuged at 4000g for 10 min. The resulting supernatant
was dialyzed through a semipermeable membrane with phosphate buffer (pH
7.2) to remove phosphotungstic acid and MgCl2, and 290 g/L
(290 mg/mL) KBr was added to the processed dialyzate. The mixture was
then centrifuged at 488 000g for 9 h at 16 °C in a
TLA100.3 ultracentrifuge rotor (Beckman). The HDL-C concentrate thus
obtained was dialyzed with phosphate buffer (pH 7.2) to remove the KBr.
Finally, an assay to verify the absence of lipoproteins other than HDL
was performed using the Paragon system (Beckman) for electrophoresis
and lipoprotein staining.
hdl-c assays
Homogeneous assays.
The PEGME method (Kyowa) involved two
reagents in solution form. Reagent 1 contained dextran sulfate,
MgCl2, sodium
N-(2-hydroxy-3-sulfopropyl)-3,5-dimethoxyaniline, and
-cyclodextrin sulfate. Reagent 2 was reconstituted before use and
consisted of a lyophilized enzyme mixture containing PEG-modified
cholesterol esterase, PEG-modified cholesterol oxidase, peroxidase, and
4-aminoantipyrine (10).
The PPD method (Daiichi) also involved two reagents, both in solution. Reagent 1 contained a polyanion and a synthetic polymer. Reagent 2 consisted of a mixture of cholesterol esterase, cholesterol oxidase, peroxidase, 4-aminoantipyrine, and solvent containing disodium N,N-bis(4-sulfobutyl)-m-toluidine, phthalic acid buffer (pH 5.7), detergent, and other constituents, as listed in the manufacturer's instructions (11).
These two homogeneous assays for HDL-C were performed with a Hitachi 917 automated analyzer (Boehringer Mannheim) according to the manufacturer's specifications.
PTA assay.
In the PTA assay, the samples were mixed
with a precipitating solution containing phosphotungstic acid and
MgCl2 (Boehringer Mannheim). After incubation and
precipitation at room temperature, samples were centrifuged, and the
supernatant was collected for assays. HDL-C was determined with an
enzymatic endpoint assay using cholesterol oxidase, peroxidase, and a
chromogenic reaction with 4-aminophenazone (CHOD-PAP) on a Hitachi 917
analyzer. The reagents were purchased from Boehringer Mannheim, and the
assay was carried out according to the application protocol provided by
the manufacturer.
UC method.
In the UC method, 1.5 mL of NaCl solution
(density, 1.006 kg/L) was added to 1.5-mL serum samples and then
centrifuged at 488 000g for 3 h at 16 °C in 13
x 51 mm polycarbonate tubes in a TLA100.3 rotor. A 1.5-mL aliquot of
the supernatant (the fraction containing the VLDL and chylomicrons) was
isolated. The HDL-C in the infranatant (the fraction containing the LDL
and HDL) was determined by the PTA method described above.
cholesterol and triglyceride measurements
Cholesterol and triglycerides were determined enzymatically with a
Hitachi 917 automated analyzer, using the CHOD-PAP and glycerophosphate
oxidase-peroxidase-4-aminophenazone (GPO-PAP) methods, respectively,
according to manufacturers' recommendations and with reagents
purchased from Boehringer Mannheim.
performance evaluation
Precision.
Four commercial control sera with low and medium
HDL-C concentrations from Boehringer Mannheim (Precipath®
HDL/LDL-C and Precinorm® L) and ITC Diagnostics (cat. no.
767660, levels 1 and 2) were used to compare the precision of the two
homogeneous HDL-C assays.
The interassay imprecision (CV) was calculated from 20 different analyses of aliquots of the four controls performed on 20 different days. The intraassay CV was determined as the average of the variances obtained with 20 assays in the same analytical run in 1 day.
Total error.
Total error is the sum of systematic error plus
random error (18). The systematic error of each homogeneous
assay was calculated by linear regression analysis comparison with the
UC "reference" method (22). Random error was 1.96 x the
total CV, which includes both intraassay and interassay CVs
(18).
Linearity.
Linearity was tested by adding various amounts of
isotonic 9 g/L NaCl to the previously described HDL-C serum pool
concentrate. In addition, another dilution series was obtained with the
Lin-Trol® linearity set for HDL-C from Sigma Diagnostics;
the HDL-C Lin-Trol was mixed with HDL-C Lin-Trol diluent at
concentrations from 0% to 100%.
Interferences.
Interference from hypertriglyceridemia was
analyzed according to Glick et al. (19). Sera from
hypertriglyceridemic subjects were centrifuged in solutions with the
same densities as the sera, which allowed triglyceride-rich
lipoproteins to rise to the top of the sample. After isolation, these
lipoproteins were added to pooled serum at various concentrations
(20) to obtain triglyceride concentrations of 250015 000
mg/L .
Sample storage.
HDL-C was determined both in two pools of
fresh serum samples (at low and medium concentrations) and two plasma
pools (heparin- and EDTA-treated samples). The pools were divided into
aliquots; of these, some were stored for 40 days at 4 °C; others
were stored at -20 °C for the same period. The samples were then
reanalyzed. The HDL-C concentration in each pool was measured 10 times
during the study with the three field methods (the homogeneous PEGME
and PPD assays and the PTA precipitation assay).
Comparison of serum, lithium heparinate, and EDTA plasma.
Lithium heparinate and EDTA plasma samples were compared with serum for
HDL-C in 26 patients. The samples were obtained from a single stick in
no particular order. EDTA plasma was obtained with blood samples (3 mL)
collected in tubes containing 0.06 mL of 0.19 mol/L K3EDTA.
Each sample was measured once.
Intermethod comparison.
We compared the two homogeneous HDL-C
assays with the conventional PTA method, analyzing 199 samples in
parallel (group I). In addition, for patients with sufficient blood
sample volume (88 samples), the homogeneous assays were compared with
the UC method (group II). In both groups I and II we used subjects with
triglyceride concentrations <4000 mg/L (<400 mg/dL). In addition, we
compared the two direct HDL-C assays with the UC method, using 27 sera
from subjects with triglycerides >4000 mg/L (>400 mg/dL; group III).
statistical analysis
Descriptive statistics (means, SDs, and CVs) were calculated with
Microsoft Excel, Ver. 5.0® (Microsoft). Regression
analysis was calculated according the Passing and Bablok method by the
Astute Statistics Add-in program for Microsoft Excel (19931995 DDU
software, University of Leeds, UK).
| Results |
|---|
|
|
|---|
|
Total error.
The linear regression equations for systematic
error calculations were: y = 0.962x + 23.45
for the PEGME method; and y = 0.956x + 38.82
for the PPD method. Using the PEGME method, we obtained total error
values for HDL-C in low (331 mg/L) and high (428 mg/L) concentrations
of 10.90% and 8.03%, respectively. With the PPD method, the total
error values at low (304 mg/L) and high (598 mg/L) HDL-C concentrations
were 16.19% and 7.75%, respectively.
Linearity.
To define the linearity of the homogeneous HDL-C
methods, we plotted the measured HDL-C values resulting from each
method as a function of the expected HDL-C concentrations for a
dilution series obtained with 9 mg/L NaCl. Statistical analysis
revealed that the slopes of the linear regression equations
(y = 1.004x - 0.56 for the PEGME
method and y = 1.019x - 4.6 for the
PPD method) did not differ significantly from the lines of identity of
the scattergrams up to at least 1500 mg/L for the PEGME method and up
to at least 1200 mg/L for the PPD method. These results suggest an
almost complete analytical recovery for both methods.
Data from both the PEGME and PPD homogeneous HDL-C assays from the Lin-Trol dilution experiments indicated that the PPD assay was not as good as desired (recovery, 31.5%), whereas the PEGME method was very superior in this aspect (91%). The linearity for the PEGME method followed the linear regression equation: y = 1.009x - 10.7.
Interferences.
The assay bias data (overestimated
triglyceride-rich lipoprotein result minus the true result, as a
percentage) plotted as a function of triglyceride concentration are
shown in Fig. 1
. Triglyceride concentrations between 2500 and 15 000 mg/L led
to a 32% increase in the HDL-C result in the PEGME method and a 15%
increase in the PPD method.
|
Storage.
The stability of HDL-C as determined by the three
methods in serial measurements of four pools stored for 40 days at
4 °C and -20 °C is shown in Fig. 2
. The initial HDL-C concentration was taken as 100%, and the
plots show the percentage of change in the HDL-C concentrations of the
four pools over time. Excellent agreement was obtained with the PEGME
method [the largest decrease (<9%) was shown at 31 days in the
low-concentration serum pool at 4 °C]. The PTA method also showed
low deviations compared with the results either in fresh native sera or
in heparin- and EDTA-treated samples (the greatest overall decrease was
12% for the low-concentration serum pool stored at 4 °C). However,
aliquots of the low-concentration serum pool stored at 4 °C showed
decreases of 18% at 21 days when assayed with the PPD method.
|
Comparison of serum, lithium heparinate plasma, and EDTA
plasma.
The PEGME, PPD, and PTA methods showed no significant
differences between HDL-C measured in either serum or heparin plasma.
For determinations by the PEGME and PPD methods, EDTA plasma yielded
lower HDL-C results compared with serum (Table 2
).
|
intermethod comparison
The means and SDs for cholesterol, triglycerides, and HDL-C in the
samples included in the comparison of the homogeneous assays with the
PTA method (group I) and the UC method (groups II and III) are shown in
Table 3
. The data in Table 4
show that all the homogeneous HDL-C assays and the PTA and UC
methods were strongly correlated. When we estimated the parameters of
the regression lines according to Passing and Bablok, the slopes and
intercepts ranged from 0.974 to 1.038 mg/L and -10.87 to 40.00 mg/L,
respectively.
|
|
| Discussion |
|---|
|
|
|---|
In the present study, the results of total error with the PEGME method are in agreement with established NCEP criteria (21) and were better than those reported by Harris et al. (22) and Huang et al. (12). The PPD method also met the 1998 NCEP performance goal (<13%) for samples with 598 mg/L HDL-C; however, at low concentrations (304 mg/L) it failed because of large systematic errors. Analogous results were also reported by Harris et al. (11) and Huang et al. (12) for this method.
The linearity results for both the PEGME and PPD homogeneous HDL-C assays were as good as those reported in the literature (12)(17). We observed that the PEGME method yields more ideal results than the PPD assay in Lin-Trol dilution experiments. The poor analytical recovery of the PPD method with the latter materials could be attributable to our use of a different reagent system than the one recommended by the manufacturer (who specifies that only cholesterol-precipitating reagents containing phosphotungstate may be used with Lin-Trol). Nevertheless, this lack of linearity could be nothing but the result of matrix effects.
Analysis of triglyceride interference showed a positive bias in both direct assays being compared. The positive error for our results is similar to that seen by Harris et al. (11). Notwithstanding the presence of a significant positive bias, both homogeneous assays met the NCEP total error goal (bias <5%), suggesting acceptable calibration of the methods and adequate specificity for HDL-C, at least in hypertriglyceridemic specimens with concentrations <10 000 mg/L.
In storage studies, our data showed that the PEGME method can be appropriately applied in samples stored for 40 days either at 4 °C or -20 °C. Nevertheless, with the PPD method, the serum samples stored at 4 °C should be either analyzed before 21 days or frozen; the heparin- and EDTA-treated samples stored at 4 °C should be analyzed before the 16th day; and the heparin- and EDTA-treated samples stored at -20 °C should be analyzed before the 40th day. All these conclusions should be interpreted cautiously because of the inherent random error.
Highly significant differences in HDL-C were obtained when EDTA plasma results were compared with serum for the two homogeneous methods. These differences in HDL-C results are in agreement with analogous average bias results recently reported by Nauck et al. (17). Thus, with EDTA-treated samples there would be disadvantages in analyzing HDL-C.
The statistical study on comparison methods showed that the PEGME method yields the same results as both the PTA and UC methods (i.e., the slopes did not differ significantly from 1, and the intercepts not were significantly different from 0). These findings are better than those of earlier work that report results handicapped by slope or intercept values significantly different from 1 and 0, respectively (12)(17)(22)(24). We are convinced that the PEGME method allows a reliable determination of HDL-C. Concerning the PPD assay, we found that this method gives results similar to those obtained with the UC method. Nevertheless, readings obtained by the PPD were 20 mg/L higher than those obtained by the PTA method because of a constant systematic error.
In conclusion, the two HDL-C homogeneous assays studied represent a significant improvement in our methodology to quantify HDL-C: both use small sample volumes, are easy to handle, are suitable for totally on-line automation, and produce results in 10 min. As shown in this and earlier studies, both homogeneous assays correlated highly and compared very well with the UC method used as a reference procedure. Therefore, we found that the PEGME and PPD methods can play an important role in routine HDL-C testing.
| Acknowledgments |
|---|
| Footnotes |
|---|
1 Nonstandard abbreviations: HDL-C, HDL-cholesterol; NCEP,
National Cholesterol Education Program; UC, ultracentrifugation; PEGME,
polyethylene glycol-modified enzymes/cyclodextrin sulfate; PPD,
polyanion polymer/detergent; PTA, phosphotungstic
acid/MgCl2; CHOD-PAP, cholesterol
oxidase-peroxidase-4-aminophenazone; and GPO-PAP, glycerophosphate
oxidase-peroxidase-4-aminophenazone. ![]()
| References |
|---|
|
|
|---|
-cyclodextrin. Clin Chem 1995;41:717-723.
The following articles in journals at HighWire Press have cited this article:
![]() |
Z. Li, R. Yang, G. Xu, and T. Xia Serum Lipid Concentrations and Prevalence of Dyslipidemia in a Large Professional Population in Beijing Clin. Chem., January 1, 2005; 51(1): 144 - 150. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. R. Warnick, M. Nauck, and N. Rifai Evolution of Methods for Measurement of HDL-Cholesterol: From Ultracentrifugation to Homogeneous Assays Clin. Chem., September 1, 2001; 47(9): 1579 - 1596. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |