|
|
||||||||
Drug Monitoring and Toxicology |
1
Division of Forensic Toxicology, Office of the Armed Forces Medical Examiner, Armed Forces Institute of Pathology, Rockville, MD 20850.
a Address correspondence to this author at: Forensic Toxicology, AFIP Annex, 1413 Research Blvd., Rockville, MD 20850. Fax 301-319-0628; e-mail paul{at}email.afip.osd.mil.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
LSD is a semisynthetic compound derived from naturally occurring ergot alkaloids. The major source of these alkaloids is a parasitic fungus, Claviceps purpuria, that grows on rye and other grains (6). These alkaloids are also found in the seeds of morning glory, Rivea corymbosa (7). In illicit preparation of LSD, these compounds are hydrolyzed to produce a common compound, lysergic acid, which is then reacted with diethylamine to produce the LSD.
Theoretically, four optical isomers are possible from the two
asymmetric carbon atoms in the LSD molecule (C-5 and C-8 in Fig. 1
). Many of the ergot alkaloids are isomeric at the C-8 position
(8)(9)(10), but none of these alkaloids is
isomeric at C-5. Therefore, only two compounds, lysergic acid and
isolysergic acid, are formed after hydrolysis of the ergot alkaloids.
In lysergic acid, the C-5 hydrogen atom and the C-8 carboxylic acid are
in cis configuration, and in isolysergic acid, these two
functions are in trans configuration
(11)(12). If these two isomers are not
separated before the illicit preparation of LSD, both LSD and iso-LSD
are formed. Moreover, a strong base and long reaction time in the
hydrolytic step may transform some of the lysergic acid to isolysergic
acid (9).
|
The metabolic disposition of LSD in animals and humans has been reviewed (13). Soon after administration, LSD metabolizes extensively, with only <1% excreted in urine unchanged (13)(14)(15). The metabolism primarily involves aromatic hydroxylation of LSD to 13- and 14-hydroxy-LSD. The compounds, however, could not be conclusively identified because no reference compounds were available to compare the results. Two other urinary metabolites, 2-oxo-LSD and N-desmethyl-LSD, were also detected in animals (14)(15). The identity of N-desmethyl-LSD in human urine was confirmed by tandem mass spectrometry and negative chemical ionization mass spectrometry (16).
For the past 10 years, we have been testing LSD in urine to curb LSD abuse in the Department of Defense. In routine analysis, a RIA screening method is used to identify the presumptive positive specimens in a large number of test samples. Presence of LSD in urine is then confirmed by a gas chromatographic and electron impact mass spectrometric technique (17). In confirmation, we observed that several specimens contained iso-LSD. The compound is also present in some urine specimens, when the LSD concentration is below the limit of detection of the procedure (50 ng/L). Although iso-LSD can be detected by comparing retention time and two ion ratios with that of a reference calibrator, procedures to quantify iso-LSD in urine failed because calibration solutions with definite iso-LSD concentrations cannot be prepared. Some of the iso-LSD was changed to LSD during standard laboratory storage conditions.
Iso-LSD is not psychoactive (18), but like LSD, the compound is classified as a schedule I drug under the Controlled Substances Act of 1970. Because LSD is prepared from ergot alkaloids with isomeric configuration at C-8 position, both LSD and iso-LSD are present in most illicit preparations. To control the use of LSD and iso-LSD, establishing a procedure to detect and quantify these compounds in urine is prudent. In this report, we describe a method that can be used routinely to detect the iso-LSD in large numbers of urine specimens.
| Materials and Methods |
|---|
|
|
|---|
instrument
A Hewlett-Packard (HP) gas chromatographymass spectrometry
(GC-MS) system consisted of an HP 5890 series II Plus GC and 5972
quadrupole mass selective detector (MSD), and the Vectra XM2 4/100i
computer workstation was used.
instrument conditions
The flow rate of helium through a DB-5MS capillary column [5:95
phenyl:methyl siloxane, 15 m x 0.25 mm (i.d.)] (J&W Scientific)
was 1.4 mL/min. The instrument was operated in splitless and
temperature program modes. Oven temperature was increased from 160 to
300 °C at 30 °C/min. The oven was held at 160 °C and 300 °C
for 1 and 3 min, respectively. Injection and detector temperatures were
285 °C and 300 °C, respectively. Glass wool was not used in the
GC liner. Active sites in the glass wool decreased the instrument
sensitivity. The MSD was operated in the electron ionization mode. The
electron multiplier voltage of the detector was 8001000 V above the
autotune value. The selected ion monitoring window and dwell time of
the ions were 0.2 amu and 50 ms, respectively.
preparation of sodium ethoxide in ethanol (0.5 mol/l)
The reagent was prepared by dissolving 575 mg of sodium (38 mm
spherical) in portions in 50 mL of anhydrous ethanol. The reaction was
conducted in a round-bottom flask protected with a drying tube and away
from any flame and spark. When the evolution of hydrogen gas subsided
(~30 min), the solution was transferred into a screw-capped bottle.
The bottle was tightly closed and stored in a refrigerator. The reagent
was found suitable for isomerization of iso-LSD to LSD for at least 6
months.
initial extraction of lsd and iso-lsd from urine
Internal calibrator, LAMPA (0.1 mL, 80 µg/L), was added to 10-mL
urine specimens and standard urine solutions containing known amounts
of iso-LSD. The solutions were made basic with 0.5 mL of
NH4OH (14.8 mol/L) and saturated with approximately
5 g of NaCl. The drugs were then extracted with 5 mL of
1-chlorobutane by shaking the mixtures for 15 min on an orbital shaker
at 60 rpm. The mixtures were then centrifuged. The clear supernatants
were transferred to other tubes and evaporated to complete dryness.
isomerization of iso-lsd to lsd
The dried extract containing iso-LSD was dissolved in 100 µL of
0.5 mol/L ethanolic sodium ethoxide. The tubes were capped and heated
in a metal block at 50 °C for 10 min. The reaction was terminated by
adding 2 mL of distilled water to the uncapped tubes. Approximately
1 g of NaCl and 3 mL of 1-chlorobutane were added to the
solutions. The mixtures were vortex-mixed for 1 min and centrifuged.
The clear supernatants were transferred to other tubes and evaporated
to dryness.
solid-phase purification
The isomerized compound and the internal calibrator were dissolved
in 2 mL of isooctane:CH2Cl2:triethylamine
(TEA) (50:50:0.1). The solutions were poured into SPE tubes prewashed
with 3 mL each of MeOH:TEA (100:0.1),
CH2Cl2:TEA (100:0.1), and
isooctane:CH2Cl2:TEA (50:50:0.1). The extracts
were allowed to pass through the sorbent by its gravity flow. The
columns were washed with 3 mL of CH2Cl2:TEA
(100:0.1) to remove low polar impurities and then dried under a vacuum
(300 mmHg) for approximately 10 s. The drugs were eluted from the
columns with 3 mL of MeOH:CH2Cl2:TEA
(0.2:10:0.01). The solutions were evaporated to dryness.
acidbase purification
After the solid-phase purification, the extracts were dissolved in
3 mL of 1-chlorobutane. The alkaloids were then extracted with 3 mL of
phosphate buffer (0.1 mol/L, pH 4.5) by vortex-mixing the solutions for
1 min. The solvents were discarded. An additional 3 mL of
1-chlorobutane was used to wash the aqueous phases and then discarded.
This process removed most of the acidic and neutral impurities from the
acidic solutions. The solutions were made basic with 0.1 mL of
NH4OH (14.8 mol/L) and then saturated with
approximately 1 g of NaCl. The basic drugs were extracted with 3
mL of 1-chlorobutane by vortex-mixing the mixtures for 1 min and
centrifugation. The clear organic solutions were transferred into 5-mL
Reacti-vials and evaporated to dryness. The extracts were dissolved in
0.1 mL of ethanol containing 1 mL/L TEA and the solutions were again
evaporated to dryness. The TEA efficiently dissolved the LSD bound to
the silicic acid on the glass tube.
derivatization
The purified extracts were dissolved in 20 µL of BSTFA. The
vials were capped and vortex-mixed for 1 min. The solutions were heated
in a metal block at 70 °C for 15 min. Approximately 3 µL was used
for GC-MS analysis. The product was sensitive to moisture but was
stable for at least 5 days when protected from moisture and stored at
26 °C.
| Results and Discussion |
|---|
|
|
|---|
Freshly prepared solutions of iso-LSD (2000 ng/L) in methanol or acetonitrile were evaporated to dryness. The compound was derivatized with BSTFA at 70 °C for 15 min and analyzed by the GC-MS method. On the basis of the total ion chromatogram, approximately 57% LSD was detected in the iso-LSD samples. Although heating was required to ensure complete derivatization, a substantial amount of compound was found reacted with BSTFA even at room temperature. Derivatization at different times and temperatures (room temperature/15 min, 50 °C/10 min, and 70 °C/15 min) showed no significant difference in isomerization, indicating iso-LSD was not isomerized to LSD at the derivatization step. The isomerization was also unaffected by the GC-MS conditions. The percent of LSD remained almost the same when the derivatized iso-LSD was tested at different injection (245285 °C) and oven temperatures (285305 °C). Therefore, either the commercial source of solid iso-LSD contained 57% LSD or the compound isomerized almost immediately when dissolved in organic solvents.
Iso-LSD rapidly isomerized to LSD when dissolved in organic solvents
and stored at 26 °C. When a freshly prepared iso-LSD solution was
tested, the amount of LSD was 57%. In 3 days, 1618% of iso-LSD
was changed to LSD (Table 1
). The isomerization continued for approximately 10 months until
it reached an equilibrium (K = 0.5). No further
conversion was observed after this period.
|
Because of the instability of iso-LSD, the compound cannot be used to quantify iso-LSD in urine with a GC-MS method of analysis. Initially, we attempted to isomerize iso-LSD to LSD by aqueous sodium hydroxide and analyze the product by the procedure described (17). When iso-LSD in 0.1 mol/L aqueous sodium hydroxide was heated at 50 °C for 60 min, a considerable amount of iso-LSD remained unchanged. Moreover, >50% of iso-LSD was lost during the incubation period. The loss was apparently due to the amide hydrolysis of iso-LSD.
Conversion of isolysergic acid to lysergic acid was extensively investigated (9). When isolysergic acid was dissolved in 100 g/L aqueous potassium hydroxide and heated on a steam bath for 1 h, only 60% of the compound was isomerized to lysergic acid. Similar base-catalyzed conversion of lysergic acid amide produced only 45% isolysergic acid amide. Approximately 33% of the compound remained unchanged. The conditions used to catalyze the isomerization were also used to hydrolyze lysergic acid amide or other ergot alkaloids to lysergic acid. Therefore, the remaining 22% of lysergic acid amide was lost because of hydrolysis of lysergic acid amide to lysergic acid. None of the methods effectively converted one isomer to the other.
In iso-LSD, the C-8 hydrogen atom is activated by the adjacent
carbonylamide group and the C-9 double bond (Fig. 1
). The hydrogen atom
can be easily removed by a nonaqueous base. When treated with sodium
ethoxide in ethanol, iso-LSD formed a C-8 carbanion. Addition of water
selectively protonated the carbanion to LSD. The conversion to LSD was
found to be almost quantitative. The selective protonation is likely
due to steric effects of the N,N-diethylamide
group on the C-8 carbanion. The reaction was almost immediate (72.3%)
after the addition of sodium ethoxide followed by water, but the
optimum yield was found by heating the iso-LSD in 0.5 mol/L ethanolic
sodium ethoxide at 50 °C for 10 min (98%) (Table 2
). When the same condition was used to isomerize LSD to iso-LSD,
no detectable peak was observed at the retention time of iso-LSD.
|
The effect of sodium ethoxide on the internal calibrator LAMPA was also investigated. A known amount of drug and internal calibrator was treated with sodium ethoxide at 50 °C over a period of 60 min. The ion area ratios of LSD and LAMPA remained almost the same during the incubation period, indicating LAMPA as a suitable internal calibrator for quantification of iso-LSD. Deuterated LSD (d3-LSD) cannot be used as internal calibrator because the trimethylsilyl derivatives of d3-LSD and LSD appear at the same retention time and produce common ions (17).
In extraction, the urine was made basic and saturated with salt. The drugs were then extracted with a solvent. The organic solution was evaporated to dryness. The dried extract was treated with sodium ethoxide to convert iso-LSD to LSD. The product was purified by using two principles of purification. In the initial solid-phase purification technique, the compounds were fractionated to remove low and high polar compounds. After the initial adsorption of the compound in the solid-phase material, the sorbent was washed with a mixture of solvents to remove the low polar impurities. The drugs were then eluted with a suitable mixture of solvents, leaving behind the polar impurities in the solid-phase column. The isomerization of iso-LSD to LSD before the solid-phase purification was essential, because it allowed a narrow solid-phase chromatographic band that contained the least amount of low polar impurities.
Solid-phase purification alone was insufficient to remove all undesirable compounds from the extract. Although it removed many urinary impurities, it introduced other impurities present in the column materials. Extensive solvent wash of the column materials did not remove all the impurities. The acidbase separation after the solid-phase purification successfully removed most of the remaining impurities. Sometime variation in polarity of the column material required adjustment in the methanol concentration at the final elution step. Concentrations in the range of 1.52.0% were found to be most suitable. Excess methanol was avoided because it extracted polar impurities that interfered with the detection process.
LSD bound irreversibly with silicic acid on the glass surface. In extraction, when glass tubes were used and the solution was not basic, TEA was added to the solution. The strongly basic nitrogen in the TEA selectively binds to the active sites of the glass surface and protects LSD from breakdown. Similar loss of LSD derivative was observed when the compound was analyzed in the GC column. Loss of sensitivity of samples injected at the beginning of the run was a major problem. Pure LSD derivatized with BSTFA was injected 23 times to condition the column. The trimethylsilyl groups of LSD molecules effectively blocked the active sites of silica inside the GC column. A considerable improvement in instrument sensitivity was observed after the column conditioning.
The recovery of the procedure was determined by adding internal calibrator at the beginning and end of the extraction and purification steps. The overall recovery was found to be 69% ± 4%. A major problem in LSD and iso-LSD analysis is the coeluting impurities. The amount of drugs is so low that minor impurities may seriously interfere in the detection process. A clean extract was considered more important than the overall yield of the drugs. Iso-LSD in urine at concentrations ranging from 25 to 5000 ng/L was analyzed with this procedure. Excellent linearity was observed over the concentration range 502000 ng/L. The slope, intercept, and correlation coefficient were 1.1, -8.2, and 1.0000, respectively. On the basis of the curve, when the iso-LSD concentrations at 50 and 2000 ng/L were recalculated, the amounts were found to be 53 and 1992 ng/L, respectively. Interrun quantifications were also within ±10% of the theoretical values. Limit of detection was 50 ng/L. Below this concentration, qualifying ion ratios exceed ±20% of that of reference calibrator at concentration 200 ng/L.
The procedure was successfully applied to analyze iso-LSD in a large
number of urine specimens. When specimens contained both LSD and
iso-LSD and were analyzed by this process, the LSD concentration
represented the total amount of LSD (LSD iso-LSD). The amount of
iso-LSD was calculated from the difference of total LSD and the LSD
tested without isomerization. Fig. 2
shows a specimen analyzed with and without the isomerization
step. Both iso-LSD and LSD were detected in the initial detection of
LSD (Fig. 2A
). The concentration of LSD was 220 ng/L. After
isomerization, the concentration of the total LSD increased to 416 ng/L
(Fig. 2B
). The iso-LSD in the specimen was then calculated from
the difference in concentrations of total LSD and LSD (196
ng/L).
|
Iso-LSD is a byproduct of illicit preparation of LSD. They are stereoisomers and classified as schedule I drugs. Therefore, detection of any of the compounds or their combination is an indication of LSD use. Although two GC-MS procedures may be required to determine iso-LSD in urine, in forensic investigation only one GC-MS confirmation procedure that detects the iso-LSD and LSD together is recommended because it saves time and is cost effective.
| Acknowledgments |
|---|
| Footnotes |
|---|
1 Nonstandard abbreviations: LSD, lysergic acid
diethylamide; LAMPA, lysergic acid methylpropylamide; SPE, solid-phase
extraction; BSTFA, bis(trimethylsilyl)trifluoroacetamide; GC-MS, gas
chromatographymass spectrometry; MSD, mass selective detector; and
TEA, triethylamine. ![]()
The opinions expressed herein are those of the authors and are not to be construed as official or as reflecting the views of the Department of the Army or the Department of Defense.
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
F.-L. Sauvage, J.-M. Gaulier, G. Lachatre, and P. Marquet Pitfalls and Prevention Strategies for Liquid Chromatography-Tandem Mass Spectrometry in the Selected Reaction- Monitoring Mode for Drug Analysis Clin. Chem., September 1, 2008; 54(9): 1519 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Skopp, L. Potsch, R. Mattern, and R. Aderjan Short-Term Stability of Lysergic Acid Diethylamide (LSD), N-Desmethyl-LSD, and 2-Oxo-3-hydroxy-LSD in Urine, Assessed by Liquid Chromatography-Tandem Mass Spectrometry Clin. Chem., September 1, 2002; 48(9): 1615 - 1618. [Full Text] [PDF] |
||||
![]() |
P. D. Anderson, K. Naik, C. Kinemond, and A. ImObersteg Forensic Testing for Drugs of Abuse Journal of Pharmacy Practice, June 1, 2000; 13(3): 226 - 235. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |