| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
H. pylori, a Gram-negative, spiral, flagellated bacterium that colonizes the mucosal lining of the stomach and proximal duodenum, was discovered by Warren and Marshall in 1984 (1). H. pylori has since been recognized as a primary causative agent in the development of gastric and duodenal ulcers (2), chronic active gastritis (3), primary gastric B-cell lymphoma (4), and gastric adenocarcinoma (5). Dixon's metaanalysis of ulcer patients yielded a 93% H. pylori infection rate for those with duodenal ulcers and 80% for those with gastric ulcers (2). Treatment of patients with H. pylori infections with combinations of H2 blockers (i.e., ranitidine) or proton pump inhibitors (i.e., omeprazole), bismuth, and antibiotics (i.e., amoxicillin, metronidazole, tetracycline, clarithromycin) has resulted in cure rates of up to 90% (6)(7)(8). If H. pylori infection is eradicated, the reinfection rate for disease is <0.5%/year (9)(10). Patients treated with bismuth and H2 blockers alone experienced ulcer recurrence rates of 95% for duodenal ulcer and 74% for gastric ulcer (11). In 1994, the NIH Consensus Conference on H. pylori in Peptic Ulcer Disease concluded that H. pylori infection caused peptic ulcer disease and recommended that all patients diagnosed with H. pylori and peptic ulcers be treated for the infection with antimicrobial drugs (12).
Several validated methods are used to diagnose H. pylori infection. Until recently, the preferred method of diagnosis was endoscopy and biopsy coupled with a rapid urease test, culture, and (or) histology. These methods require specialized equipment, trained personnel, hours to days to obtain test results, and cost between $200 and $1200 per patient. With the observation that H. pylori infection is usually associated with a systemic immune response, immunoassays that detect IgG antibodies to H. pylori provide a cost-effective alternative to endoscopic biopsy for the initial diagnosis of H. pylori infection.
In this communication we describe the functional characteristics and clinical performance of FlexSure HP, a new rapid test for the detection of IgG antibodies against H. pylori in serum. FlexSure HP is a simple, qualitative test that can be performed safely and easily in <5 min in the physician's office or at the point of care.
| Materials and Methods |
|---|
|
|
|---|
|
high-molecular-mass cell-associated protein enzyme immunoassay
(hm-cap eia)
HM-CAP EIA (Enteric Products) is a microwell ELISA method for the
detection of IgG serum antibodies to H. pylori in
serum.1
Aliquots (5 µL) of each serum sample were
diluted with 500 µL of wash buffer. Next, 100 µL of each diluted
sample was transferred to the appropriate well of the microtiter plate.
Two wells were run for each sample. In addition, the kit controls were
run according to the manufacturer's instructions; two wells each were
run with the negative, low-positive, and high-positive calibrators. The
plates were incubated for 20 min at room temperature in a 95% humidity
chamber. The plates were emptied, washed three times with 350 µL/well
of wash buffer, and then 100 µL of 1x conjugate was added to each
well and incubated for 20 min in a 95% humidity chamber. The plates
were emptied, washed three times with 350 µL/well of wash buffer, and
100 µL of 1x substrate was placed in each well. The plates were
incubated for 10 min at room temperature, and then 100 µL of stop
solution was added to each well and the absorbance for each well was
measured at 450 nm with a BioTek plate reader. DeltaSoft II software
was used to generate a calibration curve with the serum calibrators
expressed in ELISA value (EV) units. The absorbance values for each
sample were subsequently converted to EVs by using the calibration
curve. Samples with EVs <1.8 were negative, EVs 1.82.2 were
indeterminate (inclusive), and EVs >2.2 were positive.
cross-reactivity
The cross-reactivity test method was adopted from Perez-Perez et
al. (13). Bacterial suspensions grown in standard cultures
were harvested, washed, and centrifuged to a pellet. The pellet of
bacteria was resuspended in an aliquot of previously characterized
undiluted serum. The bacterial suspensions were centrifuged to a pellet
and the serum supernatant was transferred to a tube with a fresh pellet
of bacteria. The procedure was repeated until the aliquot of serum had
been treated with five successive pellets of bacteria. All serum
samples for evaluation were processed in this manner. After each of the
five adsorption cycles for each bacterial strain, a small volume (<100
µL) of serum was removed for testing with FlexSure HP and HM-CAP EIA.
Test results for each sample were compared with the untreated serum.
The following bacterial species were used: Campylobacter
jejuni (four strains), C. fetus (three strains),
C. coli (three strains), Escherichia coli (four
strains), H. mustelae (one strain), or H. pylori
(five positive control strains). If antibodies in a serum sample
cross-reacted with the HM-CAP antigens and proteins in specific
bacterial strains used in the adsorption, the antibody titer would be
reduced in the sample by the adsorption step. This would be indicated
by a lower EV for the HM-CAP EIA test and (or) a negative FlexSure HP
test result (if the amount of anti-H. pylori antibodies had
been reduced below the threshold of the FlexSure HP test). If
antibodies in a serum sample did not cross-react with the HM-CAP
antigen and the specific bacterial strain used in the adsorption, the
antibody titer would remain the same. Within experimental variation, no
change in the EV or the FlexSure HP test would be observed, when
compared with the untreated serum.
user study
A panel of 50 blind-coded serum samples were sent to eight
individuals working in various medical capacities (three physicians,
one registered nurse, one licensed practical nurse, one medical
technologist, and two clerical staff). Each end-user received only the
written product instructions as a basis for performing the FlexSure HP
tests. Test results were compared with those obtained by five trained
scientists who were very familiar with the FlexSure HP test. The serum
samples spanned a range of antibody titers from negative to high
positive as determined by HM-CAP EIA.
within- and between-site reproducibility
Five serum samples with EVs of 0.2, 1.8, 2.5, 3.5, and 6.1 were
prepared in blind-coded sets comprising 10 replicates of each serum for
a total of 50 samples. The within-site study was performed by three
technicians working independently at the same site. The between-site
study was performed by individual technicians working at three
geographically separated laboratories. The frequency of positive and
negative test results were calculated for each study and reported as
percent correct.
clinical samples
Serum samples (n = 551) were obtained as part of an
epidemiological study of the prevalence of H. pylori
infection in the Houston metropolitan area (14) in healthy
volunteers without symptoms referable to the upper gastrointestinal
tract (n = 354) and from patients undergoing upper
gastrointestinal endoscopy for evaluation of dyspepsia (n = 197).
Serum samples were stored frozen in aliquots. H. pylori
infection status was determined in all patients and volunteers by using
the [C]urea breath test (15). In some
patients infection status was confirmed by histology and (or) culture.
All serum samples were coded and forwarded to the clinical trial site (Enteric Products) for testing by FlexSure HP and HM-CAP diagnostic tests. All tests were stored at 4 °C until used. The test kits and clinical trial sera were equilibrated to room temperature before use. The test devices were run according to the manufacturers' instructions. During testing, laboratory personnel remained blinded with regard to sample identification, H. pylori infection status, clinical group, and the results of competing tests. After completion of all testing, the code for H. pylori infection status was broken and the data analyzed.
| Results |
|---|
|
|
|---|
|
As expected, serum samples adsorbed with different strains of H. pylori had lower titers (EVs) with each successive adsorption and resulted in negative FlexSure HP test results. Of the five strains of H. pylori used in the study, the serum pool was negative by FlexSure HP in two strains following the first adsorption, in two other strains following the second adsorption, and in the fifth strain following the fourth. The final EVs for the serum samples adsorbed with H. pylori ranged between 0.7 and 3.5. The apparent false-negative FlexSure HP test run on the sample with an EV of 3.5 may be attributed to the presence of antibodyantigen complexes remaining in the serum after centrifugation that were detected by ELISA but not the rapid test. H. mustalae showed only slight cross-reactivity, which resulted in a decline in serum titer from 6.5 EV to 4.9 EV
clinical samples
The sample population of patients and healthy volunteers was
selected to provide an approximately equal distribution between
[C]urea breath test-positive and -negative
samples. The patients and volunteers included African Americans,
Caucasians, Hispanics, and Asians between the ages of 4 and 82 (mean
age 44 ± 17 years), 36% female and 64% male. All patients
(n = 197) had undergone upper gastrointestinal clinical evaluation
for dyspepsia with primary diagnosis by endoscopy. In addition to
endoscopy, biopsies were obtained from 75 of the patients to confirm
diagnosis. Diagnoses included 44 patients with gastric ulcers, 140 with
duodenal ulcers, 6 with pernicious anemia, 3 with nonulcer dyspepsia, 3
with ulcers from nonsteroidal antiinflammatory drugs, and 1 with
esophagitis. None of the volunteers (n = 354) reported symptoms
associated with upper gastrointestinal disease and received no other
follow-up. All subjects provided serum samples and consented to the
[C]urea breath test (15). The sensitivity,
specificity, and overall agreement of FlexSure HP with
[C]urea breath test or HM-CAP EIA were determined.
Analysis of 551 serum samples comparing FlexSure HP test with
[C]urea breath test (reference method) yielded a
sensitivity of 94.4%, specificity of 87.6%, and an overall agreement
of 91.1% (Table 2
). The predictive value for a positive test in this population
of patients and volunteers was 89.1%; the predictive value of a
negative FlexSure HP test result was 93.5%. There were 49 instances
where the FlexSure HP test did not agree with the
[C]urea breath test: 16 FlexSure HP false-negative and
33 FlexSure HP false-positive tests. The discordant samples were run on
HM-CAP EIA to determine if antibodies to H. pylori were
detectable by another immunoassay.
|
Of the 33 FlexSure HP-positive/[C]urea breath test-negative patients, 18 serum samples were positive by HM-CAP EIA. Although those serum samples contained antibodies to H. pylori, the patients may not have had active infections or bacterial loads sufficient to yield positive [C]urea breath tests. Three of the 16 FlexSure HP-negative/[C]urea breath test-positive samples were negative by HM-CAP EIA, indicating false serologic test results for FlexSure HP.
As stated above, the subjects used in this study consisted of both patients and healthy volunteers. In the population of symptomatic patients (n = 197), FlexSure HP yielded positive serologic test results in 169 of 178 patients positive with the [C]urea breath test for a sensitivity of 94.9%. There were too few symptomatic patients to determine the specificity of FlexSure HP. Asymptomatic volunteers (n = 354) were recruited for the study to provide sufficient numbers of [C]urea breath test-negative subjects to estimate the specificity of FlexSure HP. FlexSure HP and [C]urea breath tests yielded concordant positive test results on 98 of 106 breath test-positive and concordant negative tests on 224 of 249 breath test-negative volunteers. The sensitivity, specificity, and overall agreement of FlexSure HP were 93.3%, 90.0%, and 91.0%, respectively.
FlexSure HP was compared directly with HM-CAP with 551 frozen serum samples, of which 299 were positive (EV >2.2), 237 were negative (EV <1.8), and 15 were indeterminate (EV 1.82.2) by EIA. The indeterminate samples were removed from further data analysis. FlexSure HP test results agreed with HM-CAP EIA on 285 of 299 EIA positive and 222 of 237 EIA negative samples to yield a sensitivity of 95.3% and a specificity of 93.7%. The overall agreement between the two tests was 94.6%.
ROC curve analysis (16) was used to determine the
analytical cutoff for FlexSure HP by using the HM-CAP EIA as the
standard (Fig. 2
). Theoretical EIA cutoffs for positive and negative tests were
established at 0.5 EV intervals from 0.0 to 8.5 with the upper limit of
the interval set as the cutoff for a "negative" ELISA. The fraction
of true-positive and true-negative FlexSure HP tests was calculated for
each theoretical EV cutoff. For example, if 0.0 EV cutoff was set as a
lower limit for a positive test result, all positive FlexSure HP tests
would be considered true positive and all negative FlexSure HP tests
would be false positive. As the cutoff is increased, the fraction of
true positive tests will decrease to zero and the fraction of negative
tests will increase to 1.0. The optimal performance for FlexSure HP
(e.g., best combination of sensitivity and specificity) will be on a
45-degree line bisecting the graph.
|
The EV at which optimal FlexSure HP sensitivity and specificity was observed was 2.0 EV. This coincided with the positive/negative test cutoff previously established for HM-CAP EIA with [C]urea breath test as the gold standard (17).
reproducibility
In both the within- and between-site reproducibility studies,
tests results reported by the technicians agreed with the expected
result for serum samples with EVs 0.2 (negative), 3.5 (positive), and
6.1 (positive). In addition, all test results with the 1.8-EV serum
sample agreed with the expected result (negative) for the within-site
technicians. Technicians in the between site had one false-positive
test result of 60 for the 1.8-EV sample. The within- and between-site
technicians each reported 3 of 60 false-negative test results for the
2.5-EV sample. No procedural problems were reported, indicating that
the test yielded good reproducibility, particularly around the
analytical cutoff.
user study
The ability of personnel in the clinical or physician office
laboratory setting to accurately perform FlexSure HP was evaluated by
providing a group of eight inexperienced users with education and
experience backgrounds ranging from office clerks to physicians with 50
different serum samples each, spanning the range of EVs from 0.2 to
9.1. The accuracy of the inexperienced users was compared with that
obtained by four experienced scientists. Each group reported one
false-positive test result. The experienced scientists reported no
false-negative test results, whereas the inexperienced group reported
six false-negative tests of 200. Four of the inexperienced users
reported no false-positive or false-negative tests. The results from
two groups were independently pooled and a correlation coefficient
calculated (r = 0.9891; 95% confidence interval =
0.98610.9921). The results of the user study indicate that
individuals inexperienced in performing FlexSure HP, regardless of
their education and professional training, were able to perform the
test with ease and accuracy equal to that of trained, experienced
technicians.
| Discussion |
|---|
|
|
|---|
Most invasive methods for diagnosis involve endoscopic biopsy followed by histology, rapid urease test, and (or) culture. These tests are nearly 100% specific for diagnosis of H. pylori infection, since a positive diagnosis is determined by direct observation of H. pylori in tissue sections, detection of urease activity, or culture of H. pylori from biopsies.
The most widely used noninvasive tests are immunochemical tests designed to detect the presence of IgG antibodies to H. pylori in serum involving either ELISA, immunoblot, or immunochromatographic formats. The other principal noninvasive method is the urea breath test, which requires the patient to ingest C- or C-labeled urea (15)(18). In the presence of H. pylori, bacterial urease decomposes the urea to ammonia and CO2 or CO2. The labeled CO2 is subsequently detected in the breath 10 to 60 min after ingestion of the substrate. The [C]- and [C]urea breath tests were recently cleared by the FDA.
The simplest and lowest-cost approach for the initial diagnosis of H. pylori infection is an accurate and rapid office-based test for the presence of serum IgG antibodies against H. pylori. Such tests permit the physicians to begin immediate anti-H. pylori treatment for patients with positive test results or prescribe other appropriate medical follow-up for subjects with negative test results. To be effective, physician office tests must be simple to use and interpret, exhibit no cross-reactivity with other bacteria, have no interference with normal blood constituents, have a medically relevant reactive cutoff, and provide high diagnostic specificity and sensitivity.
A key requirement to developing a serology test with high sensitivity and specificity is the antigen preparation used in the test. Soon after Warren and Marshall described this organism in 1984 (1), the correlation between gastritis and serum anti-H. pylori antibody was documented (19)(20)(21)(22) in complement fixation, bacterial agglutination, and EIAs with crude antigen preparations. First-generation immunoassays had low specificity because they involved crude antigen preparations containing proteins that cross-reacted with antibodies directed against other bacteria (e.g., C. jejuni, C. fetus, and E. coli) (22)(23). The evolution of antigen preparations used in serum-based tests and the performance of those tests was recently reviewed by Graham et al. (24).
A second-generation antigen described by Evans et al. (17) was used to develop an ELISA (HM-CAP EIA) with reduced cross-reactivity, 100% specificity, and 99% sensitivity, relative to the urea breath test. They described the antigen preparation as a HM-CAP complex extracted from H. pylori cells with a molecular mass range from 450 000 to 700 000 kDa containing urease activity. We have incorporated this same antigen preparation into FlexSure HP.
FlexSure HP was designed to perform comparably with commercially available ELISAs. The concentrations of HM-CAP antigen and conjugate in FlexSure HP were balanced to obtain the highest clinical utility (i.e., test accuracy) and to show performance characteristics comparable with those of the HM-CAP EIA. ROC curves, which describe the overall performance of a given method independent of a predefined cutoff, were constructed for FlexSure HP with HM-CAP EIA as the standard. The optimum combined sensitivity and specificity was achieved at an EV of 2.0, the same analytical cutoff previously established for the HM-CAP EIA with the [C]urea breath test as the standard.
FlexSure HP is an easy-to-use, rapid, point-of-care test for the detection of antibodies to H. pylori with performance characteristics essentially equivalent to other immunoassays and the urea breath test, but with results in a fraction of the time. In a clinical evaluation of 569 patients and healthy volunteers, FlexSure HP yielded a relative sensitivity of 95% and relative specificity of 93% when compared directly with HM-CAP EIA, but provided results in 5 min as opposed to 60 min for the EIA. The availability of a rapid, accurate, in-office test like FlexSure HP for the diagnosis of H. pylori infection should simplify the physician's task of deciding on appropriate treatment strategies for patients presenting with symptoms suggestive of gastritis, peptic ulcer disease, gastric cancer, or mucosa-associated lymphoid tissue lymphoma. Following this practice could reduce the need for invasive endoscopy except for those in whom recurrent or recrudescent infection or cancer is suspected.
| Footnotes |
|---|
1 Nonstandard abbreviations: HM-CAP, high-molecular-mass cell-associated protein; EIA, enzyme immunoassay; and EV, ELISA value. ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. H. Peruski and L. F. Peruski Jr. Immunological Methods for Detection and Identification of Infectious Disease and Biological Warfare Agents Clin. Vaccine Immunol., July 1, 2003; 10(4): 506 - 513. [Full Text] [PDF] |
||||
![]() |
W. K. Leung, F. K. L. Chan, M. S. Falk, R. Suen, and J. J. Y. Sung Comparison of Two Rapid Whole-Blood Tests for Helicobacter pylori Infection in Chinese Patients J. Clin. Microbiol., November 1, 1998; 36(11): 3441 - 3442. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |