(Clinical Chemistry. 1998;44:606-613.)
© 1998 American Association for Clinical Chemistry, Inc.
|
Automation and Analytical Techniques |
Improved protamine-sensitive membrane electrode for monitoring heparin concentrations in whole blood via protamine titration§
Narayanan Ramamurthy1,
Narayan Baliga2,
Joyce A. Wahr2,
Ulrich Schaller1,
Victor C. Yang3,
and Mark E. Meyerhoff1,a
1
Departments of Chemistry and
2
Anesthesiology, and
3
College of Pharmacy, The University of Michigan, Ann Arbor, MI 48109.
a Address correspondence to this author at: Department of Chemistry, The University of Michigan, 930 N. University Ave., Ann Arbor, MI 48109-1055. Fax 313-647-4865; e-mail mmeyerho{at}umich.edu.
 |
Abstract
|
|---|
An improved protamine-sensitive electrode based on a polymeric membrane
doped with the charged ion exchanger dinonylnaphthalenesulfonate (DNNS)
is used for monitoring heparin concentrations in whole blood. The
electrode exhibits significant nonequilibrium potentiometric response
to polycationic protamine over the concentration range of 0.520 mg/L
in undiluted whole-blood samples. The sensor can serve as a simple end
point detector for the determination of heparin via potentiometric
titrations with protamine. Whole-blood heparin concentrations
determined by the electrode method (n
157) correlate well with other
protamine titration-based methods, including the commercial Hepcon HMS
assay (r = 0.934) and a previously reported
potentiometric heparin sensor-based method (r =
0.973). Reasonable correlation was also found with a commercial
chromogenic anti-Xa heparin assay (r = 0.891) with
corresponding plasma samples and appropriate correction for whole-blood
hematocrit levels. Whereas a significant positive bias (0.62 kU/L;
P <0.001) is observed between the anti-Xa assay and the
protamine sensor methods, insignificant bias is observed between the
protamine sensor and the Hepcon HMS tests (0.08 kU/L;
P = 0.02). The possibility of fully automating these
titrations offers a potentially simple, inexpensive, and accurate
method for monitoring heparin concentrations in whole blood.
 |
Introduction
|
|---|
Heparin, a highly sulfated polysaccharide, is the anticoagulant of
choice in extracorporeal procedures such as cardiopulmonary bypass
surgery (CPB).2
Accurate measurement of heparin
concentrations during such a procedure is important because high
heparin doses can lead to bleeding complications in patients. Accurate
monitoring of blood heparin concentrations can also aid in heparin
therapy by providing a means to determine the minimal protamine dose
required to neutralize the anticoagulant activity of heparin at the
conclusion of CPB. This is of clinical significance since protamine
overdose is known to cause complement activation and other toxic
effects in patients. Furthermore, Despotis et al. recently demonstrated
that closely regulated heparin concentrations at somewhat higher doses
than traditionally used in CPB could significantly decrease
postoperative complications and the use of blood products
(1).
At present, the use of heparin during CPB is typically monitored by
clotting-time measurements, the most common and widely used one being
the activated clotting time (ACT). This method, however, is indirect
and its results can be misleading because ACT values are affected by
many variables such as hemodilution and hypothermia that do not
correlate directly with plasma heparin concentrations
(2)(3). Chromogenic heparin assays based on
factor Xa inhibition, although widely used in clinical laboratories,
cannot be performed with whole-blood samples (4). The only
method currently available to determine heparin concentrations in
whole-blood samples is the Hepcon HMS assay system (Medtronic Blood
Management), which is based on a protamine titration and the use of
clot formation for end point detection. The Hepcon assay is performed
on a cartridge consisting of four to six channels that contain
different amounts of protamine as well as dilute thromboplastin. The
end point of the test is the detection of clot formation, which is
achieved by measuring the movement of a plunger mechanism in each
cartridge. The channel containing the smallest quantity of protamine
that completely neutralizes the heparin in the sample exhibits the
shortest clotting time. The heparin concentration is determined from
the quantity of protamine in that channel (on the basis of
stoichiometry of the protamineheparin complex). Each Hepcon cartridge
thus tests a limited range of blood heparin concentrations. Often,
multiple cartridges must be run, resulting in an expensive and
time-consuming procedure. Moreover, the Hepcon determines heparin in a
"discontinuous" manner (i.e., results are displayed in intervals of
0.7 kU/L), and its accuracy is typically within ± 0.34 kU/L
(Hepcon HMS insert). Nevertheless, the Hepcon assay is the most
reliable and rapid point-of-care test method available. Whole-blood
heparin concentrations determined by the Hepcon assay system have been
shown to correlate well with laboratory-determined plasma anti-Xa
concentrations after correcting for blood hematocrit levels
(5).
Recent research in our laboratories has demonstrated that appropriately
formulated polyvinyl chloride (PVC) films doped with
tridodecylmethylammonium chloride (TDMAC) or potassium
tetrakis(4-chlorophenyl)borate (KTpClPB) exhibit large and reproducible
potentiometric responses toward low concentrations of heparin
(6)(7) and protamine (8),
respectively. Response of these polyion-sensitive electrodes has been
ascribed to the favorable extraction of the polyions into the membrane
phase via cooperative ion-pairing interactions with lipophilic ion
exchanger sites doped within the polymeric membrane phase. This
extraction process yields a nonequilibrium steady-state change in the
phase boundary potential (EMF) at the membrane/sample interface
(9).
Herein we describe the clinical utility of an improved
polycation-sensitive electrode based on a polymeric membrane doped with
the cation exchanger dinonylnaphthalenesulfonate (DNNS) for determining
heparin concentrations in whole blood. Such a DNNS-based polymeric
membrane electrode has been reported previously by our group
(10) for use in developing assays for specific proteases
involving polycationic substrates. Although we have shown earlier that
whole-blood heparin concentrations can be monitored with the
TDMAC-based heparin-sensitive electrode (11),
pseudotitrations with this electrode are very labor intensive, owing to
the irreversibility in the sensor's heparin response. Consequently
such pseudotitrations could only be carried out by using multiple tubes
containing equal aliquots of whole blood but different amounts of
protamine in each tube. In contrast, titrations with the DNNS-based
protamine-sensitive membrane electrode are direct and can be performed
more easily by monitoring the EMF change after the addition of small
aliquots of protamine to a single heparinized sample of whole blood.
Because the sensor does not respond to the heparinprotamine complex,
it can be used to monitor the titration end point (the presence of
excess protamine). The heparin concentration is then determined by
using a predetermined binding stoichiometry between heparin and
protamine.
 |
Materials and Methods
|
|---|
reagents
High-molecular-weight PVC, TDMAC, KTpClPB, calcium
bis[4-(1,1,2,3-tetramethylbutyl)phenyl] phosphate (CaTMBPP),
bis(2-ethylhexyl) sebacate (DOS), 2-nitrophenyloctyl ether (NPOE),
tris(2-ethylhexyl) phosphate (TOP), dioctyl phthalate (DOPth), dioctyl
adipate (DOA), and tetrahydrofuran (THF) were obtained from Fluka
Chemika Biochemika. Protamine sulfate (from herring), beef lung
heparin, and tris[(hydroxy methyl)amino methane] (Tris) were from
Sigma Chemical Co. DNNS was a kind gift from King Industries.
Polyurethane (M48) was kindly supplied by Medtronic Inc. Tecoflex
SG-80A was a gift from Thermedics Inc. and Pellethane 236380AE was
from Dow. Polyurethane PU-2060 was a gift from Prof. G. S. Cha
(Kwangwoon University, Seoul, S. Korea). Injectable heparin (from
porcine intestinal mucosa, 1000 USP kilounits/L) was from Elkins-Sinn
Inc. Fresh frozen human plasma was obtained from the American Red
Cross. All other reagents were of analytical grade. All solutions were
prepared with distilled deionized water. Unless otherwise stated, the
primary buffer solution used in all experiments was 50 mmol/L Tris-HCl,
pH 7.4, containing 120 mmol/L NaCl.
preparation of protamine-sensitive cylindrical membrane electrodes
Protamine-sensitive cylindrical membrane electrodes (Fig. 1
) were prepared as described previously (10). The
membrane casting solution was formulated by dissolving 200 mg of the
components (DNNS:polymer:plasticizer = 1.0:49.5:49.5 by weight) in
2 mL of THF. This casting solution was then dip coated (12 times at
30-min intervals) over rounded glass rods protruding from a narrow-bore
Tygon tube (i.d.
1.31.5 mm) and dried overnight. After soaking in
15 mmol/L NaCl for about 6 h, the glass rods were carefully
removed, and the tube was then internally filled with 50 mmol/L
Tris-HCl, pH 7.4, containing 120 mmol/L NaCl. An Ag/AgCl wire was then
inserted into the inner bore of the tube to serve as the internal
reference electrode. Before use, all cylindrical electrodes were
presoaked in a 15 mmol/L NaCl solution for at least 6 h. Unless
otherwise specified, these cylindrical electrodes were used in all
experiments and disposed of after each use.

View larger version (30K):
[in this window]
[in a new window]
|
Figure 1. Schematic diagram of a protamine-sensitive membrane
electrode used in this study and an expanded view of the extraction and
ion-pairing of the polycation with DNNS within the membrane phase.
|
|
measurement of emf response of dnns-doped cylindrical electrodes
The EMF responses of the cylindrical DNNS-based electrodes were
measured vs a miniature external Ag/AgCl reference electrode
(Bioanalytical Systems Inc.), via a Macintosh IIcx computer coupled
with an NB-MIO analog/digital input/output board (National Instruments)
and a VF-4 electrode interface module (World Precision Instruments)
controlled by LabView 2 software (National Instruments) as described in
our previous work (12).
potentiometric titrations with protamine-sensitive membrane
electrodes
Potentiometric titrations of various heparin preparations were
performed by adding small aliquots of a protamine solution (1 g/L) to
10 mL of a well-stirred buffer solution containing a fixed
concentration of either porcine or beef heparin. The resulting
potential change was recorded 3 min after each addition. Titration
curves were constructed by plotting the potential change (from the
initial baseline value) vs the concentration of added protamine.
Titration end points were computed by the Kolthoff method
(13), followed by applying a subtractive correction
factor. The latter was done to offset the effect of the small protamine
concentration required to elicit sensor response. For blood
measurements, the correction factor was 4 mg/L protamine (which
corresponds to 0.4 kU/L heparin). The heparin/protamine binding ratios
determined from the corrected end points were then used to determine
heparin concentrations in unknown samples.
heparin determination in whole-blood samples
Twenty-two cardiac surgery patients were enrolled in this study
after providing informed consent. The protocol used in these studies
was preapproved by the Human Subjects Internal Review Board at the
University of Michigan. All patients received a bolus dose of 300
units/kg of porcine heparin to achieve systemic anticoagulation.
Additional doses of heparin were administered as needed to maintain a
kaolin ACT of 480 s or above. Whole-blood samples were drawn from
patients before the initial heparin infusion, every 30 min after
heparin infusion, and after neutralization with protamine. Blood
samples were drawn into EDTA-coated Vacutainer Tubes (Becton Dickinson)
and stored at 4 °C until analysis. Hepcon assays were performed
immediately on each blood sample (non-EDTA) in the operating room. A
kaolin-activated ACT (Hemochron; International Technidyne Corp.) value
was obtained at each of the time points as well. Plasma samples for the
anti-Xa (Coatest®, Pharmacia Hepar) heparin assay were
obtained by centrifuging a portion of the respective blood samples at
670g for 20 min. The plasma samples were frozen until
subsequent analysis. Plasma heparin concentrations thus obtained by the
Xa assay were corrected for blood hematocrit with the following
formula:
 |
 |
Heparin determinations with the heparin sensor were carried out
via pseudotitrations with protamine as previously reported
(11)(14). In brief, whole blood (250 µL) was
added to each of a series of 11 tubes, each coated with a known but
increasing quantity of dried protamine sulfate. The EMF response of a
cylindrical heparin sensor in each tube vs an external Ag/AgCl
reference electrode was recorded sequentially with the computerized
LabView setup mentioned earlier. A protamine titration curve was
constructed by plotting the EMF readings against the amount of
protamine in each tube. The regression lines for the two linear
portions of the titration curve were extrapolated to yield the break
point, from which the heparin concentration was determined.
Potentiometric titrations with DNNS-based electrodes in whole blood
were performed by using the procedure described in the previous
section, except that a smaller volume (4 mL) of blood was used for each
titration. Heparin concentrations were determined from the corrected
end points by using the heparin/protamine neutralization stoichiometry
determined with the sensor (1 unit of heparin binds to 10 µg of
protamine).
statistical analysis
Bias among the various methods was calculated as recommended by
Bland and Altman (15). Student's paired t-test
was used to compare the mean whole-blood heparin measurements between
the various methods. Least-squares linear regression was used to
estimate a linear relation and generate correlation coefficients for
heparin determinations from the methods, with P values
<0.05 considered statistically significant. Confidence intervals were
computed at the 95% level.
 |
Results
|
|---|
optimization of protamine-sensitive membrane electrodes
The development and analytical utility of polyion-sensitive
membrane electrodes have been demonstrated only recently
(7). Development of these polymer membrane electrodes
requires the identification of appropriate ion-complexing agents and
membrane chemistries that yield significant potentiometric response to
polyionic macromolecules. The choice of the ion exchanger used within
the membrane is thus very crucial for their optimal performance.
Plasticized PVC membranes doped with lipophilic borate (8)
or organophosphate salts were shown previously to elicit significant
response towards protamine in physiological saline solution. However,
their poor response (<5 mV) in complex media such as human plasma
restricted their analytical utility, thereby prompting the need to
identify an improved ion exchanger that would provide a more favorable
and selective extraction of protamine into the membrane phase. Toward
this end, DNNS was investigated as the ion exchanger for use in the
development of improved polycation-sensitive electrodes. The use of
DNNS as an ion exchanger in polymeric membranes has been reported
previously in ion-selective electrodes for lanthanides
(16) and organic cations (17). Only recently
was DNNS suggested as a useful ion exchanger in the design of
polycation-sensitive electrodes for use in monitoring synthetic
polycationic peptides (10). From Fig. 2
it is obvious that the membrane formulation with DNNS offers a
significant improvement in the maximum EMF response as well as
detection limits towards protamine over other ion-complexing agents
tested. This is very likely due to the high ion-pairing affinity of
DNNS towards protamine within the organic membrane phase of the
electrode. It should be noted that detection limits towards protamine
improve when the DNNS concentration in the membrane phase is reduced
below 1 wt% (in accordance with the mechanism of polyion response
(9)); however, such reduction in DNNS concentrations yield
a much lower total EMF response at higher concentrations of protamine
(e.g., 10 mg/L). Hence, 1 wt% was chosen as a compromise between
detection limits and the magnitude of observed EMF signal.
Plasticizers play an important role in polymeric membrane electrodes by
solvating the charged components within the organic phase. In
polyion-sensitive electrodes, they also influence the diffusion
coefficients of the polyions in the membrane. A host of plasticizers
(DOS, DOA, NPOE, TOP, and DOPth) were examined in films containing DNNS
(Fig. 3
). It is noted that although significant protamine response was
observed with most plasticizers, membranes containing the polar
plasticizer NPOE yielded the best detection limits and total EMF
response. It appears that the relatively high polarity of this
plasticizer (dielectric constant,
= 24 vs
= 35 for other
plasticizers) enables a stronger cooperative ion pairing of DNNS with
protamine in the membrane phase, thereby enhancing the extraction of
protamine into the polymer film. The precise mechanism by which NPOE
enhances the ion pair formation in the membrane phase is currently
under investigation.
The nature of the polymer also plays a significant role in a
membrane's polyion response, especially for applications in undiluted
plasma/whole blood. The EMF response in undiluted plasma is usually
much lower compared with that in a physiological buffer solution,
presumably because of nonspecific adsorption of various plasma proteins
on the surface of the membrane. Because polyurethanes have been
preferred over PVC for use as biomaterials (18), they were
investigated as the polymeric matrix component in the development of an
improved protamine-sensitive membrane electrode. The effect of varying
the polymer matrix on the protamine response of DNNS-doped membranes in
undiluted human plasma is shown in Fig. 4
. The use of polyurethanes as the polymeric membrane component
appears to offer very significant advantages for preparing
protamine-sensitive electrodes. Notably, membranes containing Tecoflex
as the polymeric membrane component exhibited the greatest total EMF
response; however, this was at the expense of detection limits. This is
probably due to the lower rigidity of Tecoflex compared with PVC, which
results in a greater diffusion coefficient in the membrane phase that
affects the accumulation of polyions at the membrane/sample interface
(9) and hence the magnitude of the EMF response at low
protamine concentrations. Alternatively, membranes prepared with the
polyurethanes M48 and PU-2060, which contain a higher proportion of
hard segments, showed improved detection limits. The lower detection
limits and high potentiometric response of these electrodes render them
the optimal choice for titrations of blood heparin concentrations with
protamine. Thus, optimized membranes containing 1.0 wt% DNNS, 49.5
wt% o-NPOE, and 49.5 wt% M48 were used in all subsequent
studies.
heparin determinations via potentiometric titrations
The DNNS-based membrane electrodes can be used to detect the end
point of the titrations of heparin with protamine. Indeed, as shown in
Fig. 5
, heparin concentrations as low as 0.5 kU/L can be readily
determined. Mass neutralization ratios of protamine with bovine and
porcine heparin (powdered forms) as measured with the
protamine-sensitive membrane electrode agree well with values
determined with a heparin-responsive membrane electrode
(19). Titrations in whole-blood samples obtained at
different time intervals during the course of a typical bypass surgery
are shown in Fig. 6
. Heparin concentrations determined with DNNS-based electrodes
correlate well with those determined with the previously reported
heparin sensor-based method (11)(14), the
Hepcon assay, and the chromogenic anti-Xa assay (Table 1
). The correlations for heparin measurements by both sensors
and the Hepcon assay with the ACT, however, were poor. This
finding is expected, given that ACT values are affected by various
factors such as hemodilution and temperature, which yield no effect on
whole-blood heparin concentrations.

View larger version (28K):
[in this window]
[in a new window]
|
Figure 6. Typical changes in whole-blood heparin concentrations over
the duration of heart surgery as monitored by titrations with protamine
with the protamine-sensitive membrane electrode.
Blood samples S1 to S5 were heparinized; S0 was the preheparinized
sample and S6 was obtained after neutralization with protamine. The
average response of three electrodes containing DNNS:NPOE:M48,
1:49.5:49.5 by weight is shown.
|
|
The reproducibility of the sensor-based titration method was assessed
by performing multiple titrations on whole-blood samples supplemented
with 1 and 2 kU/L porcine heparin. Heparin concentrations of 1.03
± 0.15 kU/L (±SD) (n = 5) and 1.93 ± 0.25 kU/L (±SD)
(n = 5) respectively were determined by the sensor by performing
manual titrations on these samples.
 |
Discussion
|
|---|
Currently, the Hepcon assay is the only available method for
determining heparin concentrations in whole blood. The speed of the
assay (30 to 250 s) and a wide measurable range of heparin
concentrations (from 0.0 to 8.2 kU/L) make it particularly attractive
as a point-of-care test method. However, the appropriate cartridge must
be chosen for a given assay, such as to encompass the heparin
concentration in the sample to be tested. Thus, if the heparin
concentration is determined as a result of clot formation in the first
(or last) channel within the cartridge, then the assay must be repeated
with a cartridge capable of measuring a lower (or higher) heparin
concentration. Hence, multiple assays may be required in such cases.
Further, erroneous positive deviations were noted in the Hepcon assay
in five samples within this study, where the Hepcon showed an increase
(
1 kU/L) in heparin concentration even when no additional heparin was
administered to the patient. Such deviations were likely due to
systematic errors associated with the Hepcon (e.g., plunger problems).
The protamine sensor-based method, however, enables determination of
heparin over the entire range of concentrations that are likely to be
encountered during bypass surgery. Further, the sensor can be used to
follow the titration and thus concentrations of several
low-molecular-weight heparin preparations (e.g., Fragmin by Pharmacia
Hepar) (data not shown). Such low-molecular-weight heparins are
currently used clinically for patients with deep venous thromboembolism
and cannot be monitored by clotting time-based assays such as ACT
(20) or activated partial thromboplastin time
(21), since they do not increase clotting times
significantly.
Upon infusion into the patient, the functional blood concentration of
heparin is dependent on the binding of heparin to various plasma
proteins as well as heparin metabolism. Whole-blood heparin
concentrations therefore continually decrease over time during CPB (see
Fig. 6
), often requiring the administration of an additional dose of
heparin to the patients to maintain adequate systemic anticoagulation
on the basis of ACT times. Titrations with the protamine sensor can
follow these blood heparin concentration changes over the duration of
the surgical procedure. Absence of heparin in preheparinized (S0) or
postprotamine (S6) samples is reflected by a lack of a clear break
point in the resulting titration curves.
As shown in Table 1
, the Hepcon and potentiometric sensor-based methods
show good correlation between each other, and reasonable correlation
with heparin measurements with the chromogenic anti-Xa assay. Minimal
bias was found in heparin measurements between the heparin and
protamine sensors (0.17 kU/L; P = 0.001), and between
the protamine sensor and the Hepcon (0.08 kU/L; P =
0.02). However, a significant positive bias was observed with the
anti-Xa assay compared with the other methods. This may be due to the
continuous metabolism of heparin during surgery that yields
lower-molecular-weight heparin fragments. These fragments bind only
weakly to protamine and therefore may not be detected in titrations
with protamine. However, such fragments may still possess significant
anti-Xa activity and thus be detected via the chromogenic assay.
The ACT has long been used in clinical practice to assess the protamine
dose required for heparin neutralization after CPB. Typically, the
desired protamine doses are calculated from the infused heparin
concentrations on the basis of a dose of 1 mg of protamine for every
100 units of infused heparin. However, because of the aforementioned
metabolism of heparin during surgery and binding of heparin to various
plasma proteins, a reduced protamine dose should normally be sufficient
to fully neutralize the remaining heparin. Previous reports indicate
that reduced protamine doses after surgery can decrease perioperative
blood losses (22), perhaps by reducing complement
activation. The excellent detection limits of the DNNS-based
polycation-sensitive electrode towards protamine (< mg/L) makes it
possible to detect a protamine overdose, which is reflected by an
increased baseline EMF value when the electrode is placed into
postprotamine blood samples. Indeed, in this study, the postprotamine
samples showed an average increase of about 10 mV in the absolute
starting potentials when compared with those of the heparinized
samples, indicating a protamine overdose in all cases examined. As
suggested by others (1)(23), to maintain an
appropriate state of hemostasis in patients undergoing open heart
surgery, the actual blood heparin concentrations, and not just the ACT
values, should be closely monitored.
Although the titrations performed in this study involved manual
addition of protamine to the blood samples, resulting in a rather
lengthy assay time (3 min/titration point; 3045 min for entire manual
titration, depending on the number of points), preliminary attempts
were made also to automate such titrations by using a syringe pump for
continuous protamine infusion into heparinized samples. Fig. 7
presents a typical set of automated titration curves obtained
for three different heparin concentrations in buffer solution. Heparin
concentrations can be easily and rapidly determined from the end point
of these curves, providing the concentration and infusion rate of the
protamine solution are known. Preliminary results with whole-blood
samples indicate that the heparin concentrations determined in this
manner are in excellent agreement with those found by the manual
titrations (data not shown). Multiple automated titrations performed on
blood samples supplemented with heparin (1 and 2 kU/L) yielded
concentrations of 1.00 ± 0.17 (±SD) (n = 8) and 1.97
± 0.25 kU/L (±SD) (n = 8) respectively. Indeed, a full clinical
evaluation of this new automated methodology is currently in progress.
In summary, an improved membrane electrode that shows significant
potentiometric response to protamine has been optimized and used to
determine heparin concentrations in whole-blood samples via protamine
titration. Such heparin determinations show good correlation with other
currently available methods, including the Hepcon and the anti-factor
Xa assays. Because this method is specific for heparin and does not
require clot formation for the detection of titration end points, it
can be used in samples devoid of clotting factors (e.g., serum) or
blood samples containing other anticoagulants (e.g., EDTA, citrate). In
addition, the electrode allows for the determination of heparin on a
continuous scale (unlike the Hepcon method), rendering it suitable for
the measurements of a wider range of heparin concentrations. By using
the mode of continuous protamine infusion via a syringe pump, we
envision that a simple portable system equipped with disposable
DNNS-based electrodes could be adapted for bedside heparin monitoring.
 |
Acknowledgments
|
|---|
We thank Theresa Ambrose for her helpful review of this manuscript.
This work was supported in part by NIH grants GM 28882, HL 38353, and
HL 55461 and a research grant from Medtronics Inc.
 |
Footnotes
|
|---|
1 Dedicated to the memory of Dr. Jong Hoon Yun. 
2 Nonstandard abbreviations: CPB, cardiopulmonary
bypass surgery; ACT, activated clotting time; PVC, polyvinyl chloride;
TDMAC, tridodecylmethylammonium chloride; KTpCIPB, potassium
tetrakis(4-chlorophenyl)borate; EMF, electromotive force (phase
boundary potential); DNNS, dinonylnaphthalenesulfonate; CaTMBPP,
calcium bis[4-(1,1,2,3-tetramethylbutyl)phenyl] phosphate; DOS,
bis(2-ethylhexyl) sebacate; NPOE, 2-nitrophenyloctyl ether; TOP,
tris(2-ethylhexyl) phosphate; DOPth, dioctyl phthalate; DOA, dioctyl
adipate; and THF, tetrahydrofuran. 
 |
References
|
|---|
-
Despotis GJ, Joist JH, Goodnough LT. Monitoring of hemostasis in cardiac surgical patients: impact of point-of-care testing on blood loss and transfusion outcomes. Clin Chem 1997;43:1684-1696.
[Abstract/Free Full Text]
-
Cohen EJ, Camerlengo LJ, Dearing JP. Activated clotting times and cardiopulmonary bypass: I. The effect of hemodilution and hypothermia upon activated clotting time. J Extra Corpor Technol 1980;12:139-141.
-
Culliford AT, Gitel NS, Starr N. Lack of correlation between activated clotting time and plasma heparin level during cardiopulmonary bypass. Ann Surg 1981;193:105-111.
[ISI][Medline]
[Order article via Infotrieve]
-
Teien AN, Lie M, Abildgaard U. Assay of heparin in plasma using a chromogenic substrate for activated Factor X. Thromb Res 1976;8:413-416.
[ISI][Medline]
[Order article via Infotrieve]
-
Despotis GJ, Summerfield AL, Joist JH, Goodnough LT, Santoro SA, Spitznagel E, et al. Comparison of activated coagulation time and whole blood heparin measurements with laboratory plasma anti-Xa heparin concentrations in patients having cardiac operations. J Thorac Cardiovasc Surg 1994;108:1076-1082.
[Abstract/Free Full Text]
-
Ma SC, Yang VC, Fu B, Meyerhoff ME. Electrochemical sensor for heparin: further characterization and bioanalytical applications. Anal Chem 1993;65:2078-2084.
[Medline]
[Order article via Infotrieve]
-
Meyerhoff ME, Fu B, Bakker E, Yun JH, Yang VC. Polyion-sensitive membrane electrodes for biomedical analysis. Anal Chem 1996;68:168A-175A.
[Medline]
[Order article via Infotrieve]
-
Yun JH, Meyerhoff ME, Yang VC. Protamine-sensitive polymer membrane electrode: characterization and bioanalytical applications. Anal Biochem 1995;224:212-220.
[ISI][Medline]
[Order article via Infotrieve]
-
Fu B, Bakker E, Yun JH, Yang VC, Meyerhoff ME. Response mechanism of polymer membrane based potentiometric polyion sensors. Anal Chem 1994;66:2250-2259.
[Medline]
[Order article via Infotrieve]
-
Han IS, Ramamurthy N, Yun JH, Schaller U, Meyerhoff ME, Yang VC. Selective monitoring of peptidase activities with synthetic polypeptide substrates and polyion-sensitive membrane electrode detection. FASEB J 1996;10:1621-1626.
[Abstract]
-
Meyerhoff ME, Yang VC, Wahr JA, Lee LM, Yun JH, Fu B, et al. Potentiometric polyion sensors: new measurement technology for monitoring blood heparin concentrations during open heart surgery. Clin Chem 1995;41:1355-1356.
[Free Full Text]
-
Telting M, Collison ME, Meyerhoff ME. Simplified dual-lumen catheter design for simultaneous potentiometric monitoring of carbon dioxide and pH. Anal Chem 1994;66:576-583.
[Medline]
[Order article via Infotrieve]
-
Sergeant EP. Potentiometry and potentiometric titrations. Kolthoff IM Elwing PJ eds. Chemical analysis 1985;Vol. 69:362-364 John Wiley and Sons New York. .
-
Wahr JA, Yun JH, Yang VC, Lee LM, Fu B, Meyerhoff ME. A new method of measuring heparin levels in whole blood by protamine titration using a heparin-responsive electrochemical sensor. J Cardiothorac Vasc Anesth 1996;10:447-450.
[ISI][Medline]
[Order article via Infotrieve]
-
Bland MJ, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307-310.
-
Harrell JB, Jones AD, Choppin GR. A liquid ion-exchange membrane electrode for polyvalent ions. Anal Chem 1969;41:1459-1462.
-
Martin CR, Freiser H. Response characteristics of ion-selective electrodes based on dinonylnaphthalenesulfonic acid. Anal Chem 1980;52:562-564.
-
Anderson JM, Marchant KK. Platelet interactions with biomaterials and artificial devices. CRC Crit Rev Biocompat 1985;1:111-204.
-
Yun JH, Ma SC, Fu B, Yang VC, Meyerhoff ME. Direct potentiometric membrane electrode measurements of heparin binding to macromolecules. Electroanalysis 1993;5:719-724.
-
Greiber S, Weber U, Galle J, Bramer P, Schollmeyer P. Activated clotting time is not a sensitive parameter to monitor anticoagulation with low molecular weight heparin in hemodialysis. Nephron 1997;76:15-19.
[ISI][Medline]
[Order article via Infotrieve]
-
Carter CJ, Kelton JG, Hirsh J, Gent M. Relationship between the antithrombotic and anticoagulant effects of low molecular weight heparin. Thromb Res 1981;21:169-174.
[ISI][Medline]
[Order article via Infotrieve]
-
Despotis GJ, Joist JH, Hogue CW, Alsoufiev A, Kater K, Goodnough LT, et al. The impact of heparin concentration and activated clotting time monitoring on blood conservation. J Thorac Cardiovasc Surg 1995;110:46-54.
[Abstract/Free Full Text]
-
Despotis GJ, Joist JH, Hogue CW, Alsoufiev A, Joiner-Maier D, Santoro SA, et al. More effective suppression of haemostatic system activation in patients undergoing cardiac surgery by heparin dosing based on heparin blood concentration rather than ACT. Thromb Haemost 1996;76:902-908.
[ISI][Medline]
[Order article via Infotrieve]