(Clinical Chemistry. 1998;44:674-675.)
© 1998 American Association for Clinical Chemistry, Inc.
Urine Total Protein Measurement with the Vitros Dry Reagent Technology: Modification of Diluent to Resolve Positive Bias of Diluted Samples
Patrick Lionel Mark Lynch1,a,
John Savory1,2,
and Doris Mather Haverstick1
1
Depts. of Pathol. and
2
Biochem., Univ. of Virginia, Charlottesville, VA 22908;
a author for correspondence: fax 804-924-2574, e-mail pll8n{at}avery.med.virginia.edu
The measurement of urinary total protein continues to
have a place in the repertoire of clinical laboratory tests,
particularly in the clinical assessment of patients with renal disease.
Although the test volume is often modest, there exists the need to
perform the assay on the major automated chemistry analyzers currently
in use in most hospital laboratories. We have evaluated the performance
of a new urine total protein procedure developed for dry reagent slide
technology for use on the Vitros 950 analyzer (Johnson & Johnson). In
this method, protein in the sample binds to a pyrocatechol
violetmolybdate complex, resulting in an absorption shift. The method
is calibrated with aqueous solutions of bovine serum albumin supplied
by Johnson & Johnson. We compared the new method with that in current
use on the Cobas Fara analyzer (HoffmannLa Roche), which involves
pyrogallol red dye binding of urine total protein (1)
calibrated with human serum albumin (Biotrol). For this comparison we
analyzed 53 patient samples (median 0.65 g/L, range 0.025.44 g/L,
pyrogallol red results) and obtained the following Deming linear
regression analysis: r = 0.807, slope = 3.01,
intercept = -1.19 g/L, Sy
x =
0.72 g/L (x = Cobas Fara pyrogallol red,
y = Vitros method). These results showed a strong
positive bias with the Vitros method with respect to the pyrogallol red
method. The differences between the procedures appeared greatest in
samples that required dilution, i.e., those >2.00 g/L, the linear
range of both methods. Dilutions, when required, were carried out with
deionized water in accordance with both manufacturers' instructions.
Deming linear regression analysis of the samples that did not require
dilution (n = 36, median 0.23 g/L, range 0.021.51 g/L,
pyrogallol red results) gave the following: r = 0.980,
slope = 1.16, intercept = 0.06 g/L,
Sy
x = 0.06 g/L. Thus the
discrepancies between the two methods were markedly increased when
dilutions were involved.
At this point, three specimens were analyzed with Vitros methodology
with various dilutions in deionized water; the results are given in
Table 1
. Clearly, urine total protein measurements with Vitros
methodology are dilution dependent, showing an increasing positive bias
the greater the dilution. Similar results were obtained with human
serum albumin/deionized water solutions (data not shown). We
hypothesize that the denaturing conditions of a deionized water
dilution enhances the reaction of the proteins.
View this table:
[in this window]
[in a new window]
|
Table 1. Urine total protein results (g/L) with Vitros methodology
with deionized water and 75 and 150 mmol/L KCl solutions as
diluents.
|
|
Because urine contains a relatively high concentration of potassium,
the three specimens were then diluted with 150 and 75 mmol/L KCl
solutions and analyzed as before. These results are also shown in Table 1
. The KCl concentrations were chosen to reflect the normal range of
potassium excretion in urine (25125 mmol per day) (2).
They show that the use of a KCl solution effectively reduces the large
positive bias observed when deionized water alone is used as the
diluent. The same response to KCl was observed when human serum albumin
solutions were analyzed (data not shown).
To validate further the modified dilution protocol we compared the
pyrogallol red and Vitros methods to a manual biuret reference
procedure. This method provides essentially equal sensitivity for
different proteins and is claimed to be the most accurate procedure for
urine total protein (3)(4). The biuret method
used was based upon the ethanolic HCl phosphotungstic protein
precipitation of Savory et al. (5) with modifications
described by Rice (6). Diluted human serum total protein
calibrator was used for calibration. All the reagents and the
calibrator were obtained from Sigma Chemical Co. Deionized water and
150 and 75 mmol/L KCl were used as diluents for the Vitros method when
dilutions were indicated. A total of 23 patient urine samples were
analyzed with a median concentration of 1.44 g/L and range 0.288.93
g/L by the biuret method. Twelve of these specimens required dilution
for analysis on the Vitros analyzer. The results are given in Table 2
%It is clear from the Deming linear regression analyses in Table 2
that
the Vitros method, with dilution with 75 mmol/L KCl for samples >2.00
g/L, agrees well with the biuret reference method, which strongly
suggests that this provides an accurate measurement of urine total
protein concentration. These results, combined with the multiple
dilutions shown in Table 1
, suggest that for samples exceeding the
linear range of the Vitros method, dilution with 75 mmol/L KCl allows
for more accurate urine total protein determinations.
View this table:
[in this window]
[in a new window]
|
Table 2. Deming linear regression analysis data for pyrogallol red
and Vitros methods with different diluents compared with the reference
biuret procedure.
|
|
References
-
Orsonneau JL, Douet P, Massoubre C, Lustenberger P, Bernard S. An improved pyrogallol redmolybdate method for determining total urinary protein. Clin Chem 1989;35:2233-2236.
[Abstract/Free Full Text]
-
Painter PC, Cope JY, Smith JL. Appendix. Burtis CA Ashwood ER eds. Tietz textbook of clinical chemistry 2nd ed. 1994:2203 WB Saunders Philadelphia. .
-
Hohnadel DC, Koller A. Urine protein, total. Pesce AJ Kaplan LA eds. Methods in clinical chemistry 1987:35-45 CV Mosby St. Louis. .
-
Silverman LM, Christenson RH. Amino acids and proteins. Burtis CA Ashwood ER eds. Tietz textbook of clinical chemistry 2nd ed. 1994:717-723 WB Saunders Philadelphia. .
-
Savory J, Pu PH, Sunderman FW, Jr. A biuret method for the determination of protein in normal urine. Clin Chem 1968;14:1160-1171.
[Abstract]
-
Rice EW. Improved biuret procedure for routine determination of urinary total proteins in clinical proteinuria. Clin Chem 1975;21:398-401.
[Abstract]