|
|
||||||||
Lipids and Lipoproteins |
1
Academic Hospital Rotterdam, 3015 GD Rotterdam, The Netherlands.
2
Istituto Scientifico H.S. Raffaele, Milan, Italy.
3
Klinikum Grosshadern, Munich, Germany.
4
Klinikum der Universität des Saarlandes, Hamburg,
Germany.
5
Boehringer Mannheim, Mannheim, Germany.
6
Klinikum der Universität, Freiburg, Germany.
a Address correspondence to this author at: Department of Clinical Chemistry, Lipid Reference Laboratory, Academic Hospital Rotterdam, Dr. Molewaterplein 40, 3015 GD Rotterdam, The Netherlands. Fax 31-10 436 7894; e-mail: boersma{at}ckcl.azr.nl.
| Abstract |
|---|
|
|
|---|
-cyclodextrin, was assessed for
precision, accuracy, and cholesterol and triglyceride interference. In
addition, its analytical performance was compared with that of a
phosphotungstic acid (PTA)/MgCl2 precipitation method
(HDL-P). Within-run CVs were
1.87%; total CVs were
3.08%.
Accuracy was evaluated in fresh normotriglyceridemic sera using the
Designated Comparison Method (HDL-H = 1.037 Designated Comparison
Method + 4 mg/L; n = 63) and in moderately hypertriglyceridemic
sera by using the Reference Method (HDL-H = 1.068 Reference
Method - 17 mg/L; n = 41). Mean biases were 4.5% and 2.2%,
respectively. In hypertriglyceridemic sera (n = 85), HDL-H
concentrations were increasingly positively biased with increasing
triglyceride concentrations. The method comparison between HDL-H and
HDL-P yielded the following equation: HDL-H = 1.037 HDL-P + 15
mg/L; n = 478. We conclude that HDL-H amply meets the 1998 NCEP
recommendations for total error; its precision is superior compared
with that of HDL-P, and its average bias remains below ±5% as long as
triglyceride concentrations are
10 g/L and in case of moderate
hypercholesterolemia. | Introduction |
|---|
|
|
|---|
Accordingly, HDL-c determinations are included in most national coronary heart disease prevention programs to predict an individual's risk and to guide treatment (1)(9). As an example, the US National Cholesterol Education Program (NCEP) Adult Treatment Panel II has identified an HDL-c concentration <0.35 g/L as a major risk factor for coronary heart disease and considers HDL-c concentrations >0.60 g/L as a negative risk factor (9). Because of the enhanced role of HDL-c in medical practice, reliable and easy-to-perform assays are warranted. Several techniques to determine blood HDL-c have been described, including ultracentrifugation, high-performance liquid chromatography, electrophoresis, and precipitation-based methods (10). In routine clinical chemistry laboratories, HDL-c is frequently measured by means of chemical precipitation of the apolipoprotein B-containing lipoproteins with either polyethylene glycol (PEG), dextran sulfate, or PTA/MgCl2, followed by quantitation of the cholesterol content in the HDL-containing supernate (10). Standardization of these HDL-c assays is challenging because, in addition to matrix effects, substantial variability exists between the commonly used precipitation reagents (11)(12)(13). In addition, in certain laboratories, precipitation methods do not meet the imprecision goal for usefulness of a medical test (14). Finally, precipitation methods are time-consuming, require relatively large vol-umes of sample, and are not amenable to full automation.
Recently, a direct assay for HDL-c was developed in Japan
(15) and commercialized by Boehringer. The assay is based on
specific enzymatic hydrolysis of cholesterol esters and oxidation of
cholesterol in HDL particles. The specificity is achieved by the use of
sulfated
-cyclodextrin, PEG-modified cholesterol esterases and
oxidases, and by optimization of pH and Mg2 concentration
(15). Determinations can be made directly from serum or
heparin plasma, without any pretreatment of samples. In this study, we
aimed to evaluate this homogeneous HDL-c assay with respect to accuracy
and traceability to the Reference Methods in two CDC Network Labs. A
major focus was put on the accuracy evaluation in case of moderate and
severe dyslipidemia in comparison with a conventional chemical
precipitation method (PTA/MgCl2). Furthermore, an extensive
precision study was conducted in one CDC Network Lab. Finally, a field
method comparison between the PTA/MgCl2 precipitation
method and the new homogeneous HDL-c method was performed in five
university hospital laboratories.
| Materials and Methods |
|---|
|
|
|---|
Five European university laboratories participated in the method comparison part of the study (Rotterdam, Milan, Munich, Freiburg, and Hamburg). In all centers, uniform protocols and identical lots of reagents, calibrators, and controls were used. Moreover, all centers were equipped with Hitachi 911 or Hitachi 717 analyzers (Boehringer Mannheim), and official Boehringer applications were used.
specimens
Because matrix limitations exist for HDL-c, the mechanism for
transferring the accepted accuracy base involves comparisons using
fresh human specimens. To this end, 4 x 6 mL venous whole blood
was drawn from fasting blood donors (n = 41) and from dyslipidemic
outpatients from the lipid clinic (n = 85), of whom informed
consent was obtained to establish traceability of the homogeneous HDL-c
assay to the HDL-c Reference Method, whereas 2 x 6 mL whole blood
was sufficient for the reference standardization vs the Designated
Comparison Method (n = 98 and 35, i.e., from blood donors and
outpatients from the lipid clinic, respectively)
(10)(14). Blood was allowed to clot at room
temperature, and serum was obtained by centrifugation at
1500g during 15 min. Subsequently, specimens were split and
handled according to a standard CDC protocol; one aliquot was stored at
4 °C and analyzed with the direct HDL-c and
PTA/MgCl2-based method at the day of sample collection, and
the other aliquot was stored at -70 °C for a maximum of 4 weeks
until analysis with the Reference Methods took place.
In the cholesterol interference study (n = 35), the bias of both normocholesterolemic (n = 19) and hypercholesterolemic (n = 16) specimens was checked vs the HDL-c Designated Comparison Method. All specimens had serum triglycerides <2 g/L. In the triglyceride interference study (n = 85), sera with isolated hypertriglyceridemia (n = 32), with isolated hypercholesterolemia (n = 1), and with mixed hyperlipidemia (n = 52) were evaluated vs the HDL-c Reference Method.
For the method comparison between homogeneous and precipitation-based HDL-c assays, fresh native sera from 478 inpatients and outpatients, ~100 sera per center, that were neither icteric or hemolytic on visual inspection and had triglycerides <4 g/L, were collected from the hospital routine and analyzed. Direct and precipitation-based HDL-c determinations were performed in each participating center at the day of blood collection. HDL-c measurements were performed in singlicate in five independent assays.
HDL-C REFERENCE METHOD AND HDL-C DESIGNATED
COMPARISON METHOD
Reference points for HDL-c measurements, as recommended by
the NCEP Lipoprotein Measurement Working group, are:
(a) the CDC Reference Method, a three-step procedure
involving ultracentrifugation, precipitation, and cholesterol analysis;
and (b) the CDC Designated Comparison Method (DCM)
(10)(14)(16). The Reference Method
combines removal of VLDL by a ß-quantification ultracentrifugation
procedure, isolation of HDL by precipitation of LDL from the
ß-quantification bottom fraction (d = 1.006 kg/L)
using 46 mmol/L heparin-Mn2, and cholesterol analysis of
the HDL-c supernate by the CDC-modified Abell-Kendall Reference Method
(17). The DCM involves isolation of HDL by dextran
sulfate-Mg2 precipitation, followed by cholesterol
analysis with the Abell-Kendall method, and requires
normotriglyceridemic sera because of its limited precipitation
efficiency (16).
In the multicenter part of the study, HDL-c measurements were performed in five analytical assays; the sera were analyzed in duplicate with the CDC Reference Method (n = 41) and the DCM (n = 98) and in singlicate with the homogeneous and precipitation-based HDL-c method. In the triglyceride interference study (n = 85), reference HDL-c analyses were done in duplicate and routine HDL-c analyses were done in singlicate.
homogeneous HDL-c assay
The reagents for the homogeneous HDL-c assay were obtained
from Boehringer Mannheim. Reagent 1 was composed of 0.5 mmol/L
-cyclodextrin sulfate, 0.5 g/L dextran sulfate, 2 mmol/L
MgCl2, and 0.3 g/L
N-ethyl-N-(3-methyl-phenyl)-N'-succinyl-ethylene
diamine in 30 mmol/L 3-(N-morpholino)propane sulfonic acid
buffer, pH 7.0. Reagent 2 contained PEG-coupled cholesterol esterase
(
1 kU/L), PEG-coupled cholesterol oxidase (
5.6 kU/L), peroxidase
(30 kU/L), and 4-aminophenazone (0.5 g/L) in 30 mmol/L
3-(N-morpholino)propane sulfonic acid buffer, pH 7.0. In the
first step, 4 µL of sample was mixed with 300 µL of reagent 1 and
incubated for 5 min at 37 °C. In a second step, 100 µL of reagent
2 was added and incubated for an additional 5 min. The resulting color
was measured bichromatically at 600 nm (main wavelength) and 700 nm
(subsidiary wavelength).
The first reagent (HDL homogeneous R1, lot 131AEJ, 75 mL) was ready for use. The second reagent was composed of a lyophilized component (HDL homogeneous R2a Lyo, lot 120 AEJ) to be reconstituted with an aliquot of R2 buffer (HDL homogeneous R2, lot 120 AEJ, 20 mL). After reconstitution, the reagents were kept on board of the Hitachi analyzers. According to the manufacturer, the stability of the reconstituted R2 reagent and calibration stability is up to 4 weeks at 48 °C. For this study, the reagent was calibrated at the time each assay was run. For calibration, a human-based calibrator (HDL/LDL calibrator high, lot ML01 0058) was used, which was preliminarily "targeted" by the manufacturer using the PTA/MgCl2 precipitation method (assigned value, 0.704 g/L). The reconstituted calibration material is claimed to be stable up to 1 day at 48 °C.
To assess traceability to the Reference Methods, complete analytical systems should be evaluated (including instrument model, application, reagent lot, and calibrator lot). In this reference standardization study, the lot numbers displayed above were evaluated; the analyzers used were a Hitachi 911 in Rotterdam and a Hitachi 717 in Milan. For precision evaluation, two additional lots of reagent were investigated on a Hitachi 911 (kit lots were 135AEK and 67241101, respectively). For the bias survey in dyslipidemic specimens, calibrator lot 19079901 and reagent lot 68079401 were used.
HDL-C DETERMINATION BY APTA/MgCl2 PRECIPITATION METHOD
The analytical performance of the homogeneous HDL-c method was
compared with the performance of a conventional
PTA/MgCl2 precipitation method from the same
manufacturer (Boehringer Mannheim, cat. no. 543003, lot 660393-01). The
PTA/MgCl2-based assay was calibrated with 3.5-fold diluted
Calibrator for Automated Systems (CFAS, Boehringer Mannheim, cat. no.
759350, lot 184927) in all clinical laboratories, according to the
manufacturer's instruction (denoted as PTA/MgCl2[CFAS]).
To precipitate apolipoprotein B-containing lipoproteins, 200 µL of
serum was mixed with 500 µL of precipitating reagent, composed of
0.44 mmol/L PTA and 20 mmol/L MgCl2. The mixture was
allowed to stand 5 min and was subsequently centrifuged
(3000g, 10 minutes, 20 °C). Cholesterol in HDL was
determined with the Boehringer Mannheim
cholesteroloxidase-phenol-aminophenazone (CHOD-PAP) reagent (cat. no.
1489437, lot 664358-01).
HDL-C DETERMINATION BY A COMBINED ULTRACENTRIFUGATION
AND PTA/MgCl2 PRECIPITATION METHOD
Laboratories 3 and 4 performed a combined ultracentrifugation and
precipitation assay as an additional comparison method. Hitherto, the
protocol of the Lipid Research Clinics Program was followed, with
modifications published previously (18).
triglyceride assay
Triglycerides were determined with a
glycerolphosphateoxidase-phenol-aminophenazone method (GPO-PAP)
(Boehringer Mannheim, cat. no. 1058550) on Hitachi 911 or Hitachi 717
analyzers. The application with Selective Mode Solution (i.e., 200
mmol/L HCl; Boehringer Mannheim, cat. no. 1360922) as R2 was used,
according to the manufacturer's protocol.
analyzers
Homogeneous and precipitation-based HDL-c determinations were
performed on Hitachi 911 (n = 4) or Hitachi 717 (n = 1)
analyzers (Boehringer Mannheim) in each participating center. Daily
maintenance and operation was performed according to the instructions
of the manufacturer.
control sera
Precision studies were performed by using processed quality
control materials from Boehringer Mannheim: PrecinormR
L (lots 185597 and 186586) and PrecipathR L neu (lots ML01
0062 and ML02 2716). Besides, several batches of fresh frozen human
sera at three concentrations (low, medium, and high) were examined.
data analysis
Before starting the multicenter data analysis, it was verified
whether control sera were within the preset control limits (mean
± 3 SD). In the multicenter part of the study, 1 of 25 homogeneous
HDL-c assays had to be omitted (lab 2, first run, n = 19).
In case of up to 200 paired data points, regression analyses were
performed according to the method of Passing and Bablok
(19); in the case that more than 200 data points had to be
compared, orthogonal regression was used. Total error (%) of the HDL-c
assays was calculated as follows: [(1.96 x analytical
imprecision (%)) absolute mean bias (%)] (14).
Calculation of mean bias and total error vs the DCM was,
according to CDC recommendations, based solely on specimens with
triglycerides
2 g/L (16). In case of the
PTA/MgCl2 precipitation assay, only data from clear and
undiluted supernates, reflecting complete precipitation of
apolipoprotein B-containing lipoproteins, were included in the final
data analysis.
| Results |
|---|
|
|
|---|
|
|
bias vs the cdc ultracentrifugation reference method
The regression equations of homogeneous precipitation-based HDL-c
method vs the CDC Reference Method are presented in Table 5
(HDL-c range, 0.070.799 g/L; triglyceride range, 0.443.94
g/L). For the homogeneous HDL-c method (Y) vs the CDC
Reference Method (X) the slope of the regression equation is
significantly higher than one at
= 0.05, whereas the intercept is
not significantly different from zero. Fig. 1
, A and B, demonstrates a mean bias of 2.2% for the homogeneous
HDL-c method and -3.9% for the PTA/MgCl2 assay,
respectively. Assuming that the overall imprecision of the homogeneous
HDL-c assay is maximally 3.08% (Table 2
), the total error of the assay
is 8.2%.
|
|
bias vs the designated comparison method
The regression equations of homogeneous precipitation-based
HDL-c assays vs the DCM are presented in Table 5
(HDL-c range,
0.1861.052 g/L; triglycerides,
2 g/L; n = 63). Analogously,
the slope of the regression equation for the homogeneous HDL-c method
(Y) vs the CDC DCM (X) is significantly higher
than one at
= 0.05. Fig. 1
, C and D, demonstrates a mean bias of
4.5% for the homogeneous HDL-c method and -5.0% for the
PTA/MgCl2 assay, respectively. Assuming an overall
imprecision of maximally 3.08% (Table 2
) for the homogeneous assay,
the total error of the method is 10.5%.
method comparison with the PTA/MgCl2PRECIPITATION METHOD
The results of the method comparison between homogeneous and
precipitation-based HDL-c methods are presented in Table 6
(HDL-c range, 0.071.25 g/L; triglycerides,
4 g/L). From the
individual and pooled regression equations, it can be seen that the
homogeneous HDL-c assay produced results that were significantly higher
(P <0.05) than those produced with the
PTA/MgCl2 assay. The same holds for laboratories 3 and 4,
which used the combined ultracentrifugation and precipitation
method.
|
cholesterol interference
In Fig. 2
, A and C, it is illustrated that the homogeneous HDL-c assay
does not become biased in the case of hypercholesterolemia. All
specimens examined here (n = 35) were normotriglyceridemic
(triglycerides <2 g/L), whereas 16 specimens were hypercholesterolemic
(cholesterol >2.4 g/L). The cholesterol range tested varied between
1.63 and 3.91 g/L, whereas average and median biases were 2.5% and
3.0%, respectively (range, -3.0% to 10.5%). In each of the 35
samples examined, the bias of the homogeneous HDL-c method was less
than ±13%.
|
In Fig. 2
, B and D, the biases of the PTA/MgCl2
precipitation method are displayed as a function of HDL-c and
cholesterol concentration, respectively. Average and median biases were
-1.9% and -2.0%, respectively (range, -9.4% to 4.5%).
Analogously, no bias could be demonstrated across the cholesterol range
tested.
triglyceride interference
In Fig. 3
, A and C, it is illustrated that the homogeneous HDL-c assay
becomes positively biased in case of severe hypertriglyceridemia, a
condition frequently encountered in conjunction with low HDL-c
concentrations. The triglyceride range tested varied between 2 and 59
g/L. In case of triglyceride concentrations
20 g/L (n = 78),
average and median biases were 4.3% and 3.7%, respectively (range,
-26.3% to 47.2%). In all specimens with triglycerides >20 g/L, huge
positive biases were demonstrated for the homogeneous HDL-c method,
ranging between 51% and 578%. At triglyceride concentrations
10
g/L, the bias of the homogeneous HDL-c method was less than ±13% in
60 of 66 samples; from 10 g/L on, unacceptable triglyceride
interference could eventually appear.
|
In Fig. 3
, B and D, the biases of the PTA/MgCl2
precipitation method are displayed as a function of HDL-c and
triglyceride concentration, respectively. Overall, a tendency to an
increasingly negative bias with increasing triglyceride concentrations
existed, especially >10 g/L of triglycerides. If triglycerides were
20 g/L (n = 78), average and median biases were -7.3% and
-7.5%, respectively (range, -31.1% to 8.5%). In the case of
triglycerides >30 g/L (n = 4), it was no longer possible to get a
clear and homogeneous supernatant after precipitation of apolipoprotein
B-containing lipoproteins, not even after making a twofold sample
predilution with physiological saline solution. Of the 81 remaining
samples in which a chemical precipitation was performed, five samples
demanded predilution to get a clear supernatant.
| Discussion |
|---|
|
|
|---|
In 1995, the NCEP Working Group on Lipoprotein Measurement issued
performance guidelines for HDL-c measurements that clinical
laboratories should achieve by 1998 (14). In the case of
HDL-c, the percentage of total error of routine HDL-c determinations
should be reduced to
13%. Per definition, total error can be
interpreted as an error budget one can divide between analytical
imprecision (CVa) and average bias. One set of
conditions that is consistent with the total error goal is that at an
HDL-c concentration of
0.42 g/L the CVa should be
4%
and the average bias less than or equal to ±5%.
In this study, it is illustrated that the homogeneous HDL-c assay,
applied on Hitachi type analyzers, has an excellent reproducibility;
total CVs ranged between 1.88% and 3.08% and were similar among
different reagent lots (Table 2
). From the field method comparison
study, it became obvious that total CVs of the
PTA/MgCl2 assay differed substantially among
laboratories in both lyophilized and frozen control materials (Tables 3
and 4
). For example, CVs ranged between 1.36% and 6.59% in
processed controls and between 0.58% and 6.60% in frozen sera,
signifying that not all laboratories that use the PTA/MgCl2
assay are able to reach the 1998 imprecision goal
(14). In contrast, total CVs of the homogeneous HDL-c method
were less by one-half and much more consistent among participating
centers, ranging between 0.70% and 2.26% in lyophilized controls and
between 1.00% and 2.66% in fresh frozen sera. From the precision
data, it is evident that the precision of the homogeneous HDL-c assay
amply meets the 1998 NCEP requirements. In addition, the random error
of the homogeneous HDL-c assay fulfills the generally accepted
criterion for usefulness of a medical test, stating that the
CVa should be no greater than one-half the
average biological variation (the biological CV of HDL-c is usually
considered to be 7.5%) (14).
|
|
The bias survey in normo- and moderately hypertriglyceridemic sera
displayed a positive mean bias of similar magnitude in either CDC
Network Laboratory (Table 5
; Fig. 1
, A and C). Notwithstanding the
presence of a significant positive bias, the homogeneous assay met the
1998 total error goal, suggesting acceptable calibration of the method
and adequate specificity for HDL, at least in normo- and moderately
hypertriglyceridemic specimens. Analogously, results of the field
method comparison (Table 6
) demonstrated a positive bias of the
homogeneous HDL-c assay compared with the PTA/MgCl2
assay. The systematic difference is explained, first, by the arbitrary
initial value setting of the homogeneous HDL-c calibrator, and second,
because the CFAS calibrator of the PTA/MgCl2 assay is value
assigned by the Definitive Method, which produces results that are
approximately 1.5% lower compared with those produced with the
Reference Method (16).
As CDC Network Laboratories strive to assist manufacturers in
documenting and improving their method accuracy
(14)(16), a calibrator reassignment was
recommended for the homogeneous assay. Accordingly, Boehringer Mannheim
adjusted the calibrator value to 97% of its preliminary target value.
Assuming a maximum CVa of 3.08% (Table 2
), the
calibrator reassignment implies that total error of future homogeneous
HDL-c measurements will be about 5.2% to 7.5% instead of 8.2% to
10.5%.
By means of the reference standardization part of the study, traceability of the Boehringer homogeneous HDL-c assay to the CDC Reference Methods has been established. Of course, the results presented here only apply to the specific analytical system evaluated (instrument model, reagent lot, and calibrator lot) and do not guarantee accuracy of future reagent and calibrator lots. Notwithstanding, conventional in-house quality control procedures at the manufacturer's site should be adequate to monitor the system in the future. If shifts are observed or suspected, another direct comparison with the HDL-c Reference Methods should be undertaken to reset the system for optimal accuracy.
The comparative bias surveys of conventional and direct HDL-c assays
using fresh, dyslipidemic sera illustrate that the new Boehringer
homogeneous HDL-c assay is as robust as the PTA/MgCl2
precipitation method up to at least 3.91 g/L serum cholesterol (Fig. 2
)
and up to 10 g/L of serum triglycerides (Fig. 3
). Above 10 g/L of serum
triglycerides, the homogeneous HDL-c assay suffers from serious
nonspecificity, leading to overestimation of the HDL-c concentration,
whereas the PTA/MgCl2 precipitation method becomes
increasingly negatively biased. The observed percentage of bias largely
varied among individuals having similar triglyceride concentrations and
likely illustrates an effect of lipoprotein composition (Fig. 3
, C and
D). Whether sample predilution or the use of a reduced sample volume (3
µL in stead of 4 µL) could reduce bias in grossly dyslipidemic sera
by using the Boehringer homogeneous HDL-c assay warrants further
investigation.
We conclude that the new homogeneous HDL-c assay amply meets the 1998
NCEP recommendations for total error; its precision is superior
compared with the PTA/MgCl2 precipitation method, and
its average bias remains well below 5% in case of moderate, isolated
hypercholesterolemia or as long as serum triglyceride concentrations
are
10 g/L.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
-cyclodextrin. Clin Chem 1995;41:717-723.
The following articles in journals at HighWire Press have cited this article:
![]() |
W. E. Connor, P. B. Duell, R. Kean, and Y. Wang The Prime Role of HDL to Transport Lutein into the Retina: Evidence from HDL-Deficient WHAM Chicks Having a Mutant ABCA1 Transporter Invest. Ophthalmol. Vis. Sci., September 1, 2007; 48(9): 4226 - 4231. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. R. Warnick, M. Nauck, and N. Rifai Evolution of Methods for Measurement of HDL-Cholesterol: From Ultracentrifugation to Homogeneous Assays Clin. Chem., September 1, 2001; 47(9): 1579 - 1596. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Wagner, J. L. Sanchez-Quesada, A. Perez, M. Rigla, M. Cortes, F. Blanco-Vaca, and J. Ordonez-Llanos Inaccuracy of Calculated LDL-Cholesterol in Type 2 Diabetes: Consequences for Patient Risk Classification and Therapeutic Decisions Clin. Chem., November 1, 2000; 46(11): 1830 - 1832. [Full Text] [PDF] |
||||
![]() |
M. M. Kimberly, E. T. Leary, T. G. Cole, and P. P. Waymack Selection, Validation, Standardization, and Performance of a Designated Comparison Method for HDL-Cholesterol for Use in the Cholesterol Reference Method Laboratory Network Clin. Chem., October 1, 1999; 45(10): 1803 - 1812. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Cobbaert, P. G.H. Mulder, H. Baadenhuijsen, L. Zwang, C. W. Weykamp, and P. N.M. Demacker Survey of Total Error of Precipitation and Homogeneous HDL-Cholesterol Methods and Simultaneous Evaluation of Lyophilized Saccharose-containing Candidate Reference Materials for HDL-Cholesterol Clin. Chem., March 1, 1999; 45(3): 360 - 370. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. J. Lackner and G. Schmitz ß-VLDL of Patients with Type III Hyperlipoproteinemia Interferes with Homogeneous Determination of HDL-Cholesterol Based on Polyethylene Glycol-modified Enzymes Clin. Chem., December 1, 1998; 44(12): 2546 - 2548. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |