|
|
||||||||
Laboratory Management |
1
Departments of Endocrinology and
2
Clinical Chemistry, Odense University Hospital, 5000 C, Denmark.
a Address correspondence to this author at: Department of Endocrinology, Odense University Hospital, 5000 Odense C, Denmark. Fax 45 65 90 63 27; e-mail MAR.ANDERSEN{at}WINSLOEW.OU.DK.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Plasma contains >100 forms of GH, and what is commonly understood as "plasma GH," i.e., the free monomeric 22-kDa form, represents only 21% of the total immunoreactivity in plasma (1). However, the 22-kDa form of GH is the most prevalent form and is the prototype of pituitary GH. This form is a single-chain, 191-amino acid protein. The 20-kDa form is the second most abundant in the circulation. During peak secretion, the relative amounts of the 22-kDa and 20-kDa forms are fairly stable and independent of secretagogues (2), the relative mean percentage concentrations being 76.4% and 15.8%, respectively. This is in contrast to the basal state, where only a fraction of immunoreactive GH in the blood can be attributed to known GH forms (3). From the above description of the nonhomogeneity of GH, it is clear that the antibodies used in an assay would be of great importance. The use of polyclonal antibodies would yield higher results, because epitopes on different GH forms could bind the antibodies, in contrast to monoclonal antibodies directed only against the 22-kDa form of GH. The degree of cross-reactivity for the monoclonal and polyclonal assays has been quantified and specific differences between the two methods shown (4).
Because we previously had evaluated a new stimulation test, the pyridostigmine-growth-hormone-releasing-hormone (PD-GHRH) test, including a reliable cutoff limit for the test (5), it was with some doubt we started comparisons between the well-established Pharmacia assay and the new Delfia assay.
The problems with hormones in general, e.g., deterioration and lack of transferability, and with GH in particular suggested that it would be difficult to compare results obtained from the Pharmacia method based on polyclonal antibodies and those obtained from the Delfia method using monoclonal antibodies specific to the 22-kDa GH form of the hormone. The difficulties with comparisons between assays were demonstrated in a recent paper (4); these difficulties may contribute to the lack of confidence in GH stimulation tests (6).
To permit comparison between results obtained with the Pharmacia and the Delfia methods, we used both methods to measure GH concentrations in samples from 71 individuals: 40 healthy adults and 31 patients with suspected GHD. Both basal samples and peak GH responses after stimulation with pyridostigmine (PD) in combination with GHRH were studied. We chose to compare the results using ratio plots, a modification of the Altman-Bland difference plot (7), with the application of a 95% prediction interval and with acceptability criteria based on inherent analytical imprecision and on analytical quality specifications (8)(9). We also chose to perform a step-by-step evaluation using well-defined serum samples from groups of healthy adults and patients before and during a stimulation test.
| Materials and Methods |
|---|
|
|
|---|
The PD-GHRH test.
The participants attended the clinic after
fasting overnight; during the experiment they were not permitted to
sleep, smoke, eat, or drink anything but tap water. A cannula was
inserted into an antecubital vein for blood sampling and the
administration of GHRH. At 0900, (time -60), 120 mg of PD (Mestinon,
Hoffmann-La Roche) was administered orally; 60 min later, at 1000 (time
0), 1 µg/kg body weight of GHRH (Groliberin, Pharmacia & Upjohn,),
was administered intravenously as a bolus. Blood samples for the
measurement of GH were taken at -60, 0, 20, 30, 60, and 90 min. The GH
concentrations corresponding to basal and peak GH responses are
presented here.
The samples used in this study were obtained in connection with other investigations (5). The Declaration of Helsinki(II) was observed and the local Ethical Committee has approved the study. All subjects were volunteers, and they signed an informed consent document before taking part in the study.
methods
The two assays are immunometric: Pharmacia hGH RIA (Pharmacia &
Upjohn) and Delfia 22-kDa hGH (Wallac). The Pharmacia hGH assay is a
2-site immunoradiometric method using polyclonal rabbit/sheep
antibodies with I as the reporter molecule and
Sepharose particles as the solid phase. The Delfia hGH assay is a
solid-phase, 2-site fluoroimmunometric assay based on the direct
sandwich technique, in which two murine monoclonal antibodies are
directed against two separate antigenic determinants on the 22-kDa hGH
molecule with europium as the reporter molecule. All assays were
performed according to the manufacturers' protocols. The standards
provided in the kits were calibrated against the 1st International
Standard 80/505 from WHO in which 1 mg = 2.6 IU of GH. The
estimated precision profile for the expected CVs is shown in Table 1
. The in-house cross-reactivity of the Pharmacia and Delfia
assays with biosynthetic recombinant hGH (Genotropin, KabiVitrum AB)
were 134% and 114%, respectively (ratio 1.18). The quality
assessments carried out externally with Ringversuche (Deutsche
Gesellschaft für Klinische Chemie E.V.) showed that the median
ratio (minimum-maximum) between the means of the participants was 1.10
(1.031.14) and close to the ratio estimated in our laboratory on a
single occasion (1.18). In both assays, the GH standard was calibrated
against the 1st International Standard 80/505 from WHO. This 80/505
standard is pituitary-derived and therefore contains a mixture of GH
forms (dimerized and oligomerized) and GH fragments. The detection
limit for the Pharmacia assay was 0.4 mIU/L; the intraassay and
total CVs for the assay at 0.5 mIU/L were 14% and 17%,
respectively, and at 22 mIU/L, the intraassay and total CVs were
4% and 6%, respectively. The detection limit for the Delfia assay was
0.03 mIU/L. The intraassay and total CVs for the assay at 0.5
mIU/L were 5% and 8%, respectively, and at 18 mIU/L, the CVs
were 2% and 4%, respectively.
|
data analysis
The mean of ratios [individual GH results determined by the
Pharmacia assay divided by the results obtained in the Delfia assay,
(GHPharmacia/GHDelfia)] and the expected
distribution of ratios were calculated.
The paired results were evaluated using a modification of the Altman-Bland difference plot (7). A 95% prediction interval was calculated to express the interval within which we would expect 95% of the data points to be found (8)(9). A 95% prediction interval of this kind is in contrast to the 95% limits of agreement described for difference plots by Altman-Bland (7), which are based on calculations of the measured difference between the two methods, and thereby describes the 95% interval for measured differences. During GH stimulation in healthy adults, the percentage of the major GH forms in blood is assumed to be reasonably stable (2). Therefore, we have used the mean of ratios of the peak GH responses to the PD-GHRH test in healthy adults and the known total analytical CVs for the two assays to estimate the 95% prediction interval. The 95% prediction interval is the mean ratio ± 1.96 (CVPharmacia2 CVDelfia2)1/2 x (mean ratio). With this information, it is possible to set up a prediction interval within which a certain fraction (~95%) of the ratio points are expected to be distributed and to test whether the prediction interval can be applied to related sets of less homogeneous data. This is in accordance with the graphical interpretation of analytical data using difference plots (8)(9). Visual inspection of the scatter of points in relation to the prediction interval can be supplemented by calculation of the mean of ratios and CVs for the different subgroups.
A more accurate 95% prediction interval could be based on the formula
(which can be evaluated from Geary (10) and Hinkley
(11)):
![]() |
0.5CVwithin-subject where
(CVwithin-subject) represented the CV of
within-subject variation in accordance with the CotloveHarris concept
(14). This criterion was also applied to the ratio plots
(mean ratio ± 1.96 x
1/2CVwithin-subject x mean ratio) as
an alternative acceptability criterion setting, the
CVwithin-subject, equal to 36%
(5).
|
| Results |
|---|
|
|
|---|
sera from peak gh responses in healthy adults
The mean ratio (r; Pharmacia/Delfia) for the peak
GH responses to the PD-GHRH test in 40 healthy adults was
r = 1.59. The ratio points around this value should be
dispersed in accordance with the expectation from the analytical
imprecisions of the two methods (analytic CV (CVA), 6% and
4%, respectively) for GH concentrations > 20 mIU/L. This
corresponds to an expected CV of the ratios (CVRatio) of
(6 4)1/2 = 7.2%, or a standard
deviation of the ratios (SRatio) of 1.59 x 0.072
= 0.11. Thus, the 95% prediction interval for the ratio points is
1.59 ± 1.96 x 0.11 (1.59 ± 0.22), or from 1.37 to
1.82, assuming constant analytical CV values >20 mIU/L. In Fig. 2
A, the 95% prediction interval is shown together with the 40
individual ratios in a ratio plot. For concentration values <20
mIU/L, the imprecision increases. In Fig. 2A
, this expansion is
performed by interpolation of the limits to the combined imprecision at
0.5 mIU/L, corresponding to (17
8)1/2 = 19%; the SRatio was
1.59 x 0.19 = 0.30. Thus, the 95% prediction interval for
the "lower concentration" ratio points is 1.59 ± 1.96 x
0.30 (1.59 ± 0.59), or from 1.00 to 2.18.
|
Only 3 of the 40 points (7.5%) lie outside the prediction interval,
indicating that at the peak response, the ratios between the measurable
GH variants are fairly constant using the two methods. The measured
SRatio was 0.16, corresponding to the expected
SRatio of 0.12. Accordingly, each group of ratios from
healthy individuals, i.e., men
40 years, men >40 years,
premenopausal women taking no oral contraception, premenopausal women
taking oral contraception, and postmenopausal women, were
(mean ± S) 1.60 ± 0.14, 1.64 ± 0.24, 1.55
± 0.15, 1.64 ± 0.07, 1.54 ± 0.18, respectively. Furthermore,
Fig. 2A
shows that the ratios are independent of the peak GH responses
to the PD-GHRH test and clearly distinct from the ratio of the
biosynthetic recombinant hGH of 1.18. The quality goal (analytic
quality specification) for imprecision based on the biological concept
is CVA
1/2CVwithin-subject, where the
within-subject biological variation for GH is 36% (also shown in Fig. 2A
as the lines 1.59 ± 1.96 x 0.18 x 1.59 ~
1.59 ± 0.56). The use of a common conversion factor of 1.59 is,
therefore, clearly acceptable from the biological standpoint as well.
sera from peak gh responses in patients
Fig. 2B
shows the peak GH responses in patients together with
the 95% prediction interval, based on the conversion factor 1.59, the
analytical goal, and biosynthetic recombinant hGH lines from Figure 2A
.
For values >1 mIU/L, the ratio points are distributed as ratio
points for healthy individuals (mean ± S, 1.50 ± 0.19;
n = 23). Below this point, there is a slight tendency to
decreasing ratios for decreasing values. Of the eight patients with
peak GH responses
1 mIU/L, three produced no detectable GH
response to either method, and the other five produced a GH response of
1 mIU/L to both methods. However, these five all met the
analytical criterion and remained close to the biological criterion for
using a common conversion factor of 1.59. From a clinical point of
view, these low values are far below the cutoff limit for GHD in
adults. The data from the three patients with undetectable GH responses
in both assays are not included.
sera from unstimulated healthy adults
The ratios from measurements of unstimulated healthy
individuals are shown in Fig. 3
A. For concentration values >1 mIU/L, the distribution of
ratios is close to the distribution of stimulated healthy adults
(1.73 ± 0.23; n = 24), but premenopausal women taking oral
contraceptives have a slightly higher mean ratio (1.85 ± 0.20;
n = 8). However, all these points fulfilled the biological
criterion, and only one point was outside the analytical limits. The
results from the other groups as well as the total distribution clearly
satisfied the criterion for values >1 mIU/L. The distribution
below this value is much wider, as was also to be expected from the
analytical imprecisions. The most extreme values are high and low,
respectively, which indicates that the dominating factor is the
analytical imprecision. Our evaluation indicates that the two methods
can be used as substitutes for each other (with the factor) even for
values <1 mIU/L.
|
sera from unstimulated patients
Ratios from measurements of unstimulated patients are shown
in Fig. 3B
. For concentration values >1 mIU/L, the distribution
of these ratios is close to the distribution of ratios in stimulated
healthy adults, 1.60 ± 0.29 (n = 11). Of the 20 patients
with peak GH responses
1 mIU/L, 15 patients had no detectable
GH. Data from these 15 patients are not included.
| Discussion |
|---|
|
|
|---|
In addition to our work, others (4)(20) have shown that the use of the 22-kDa form as a standard cannot alone explain the difference between measured GH concentrations using monoclonal and polyclonal assays. However, it is always preferable to have a standard that is as pure and reproducible as possible. Recently, the WHO Expert Committee on Biological Standardization established the preparation coded 88/624 as the first International Standard for Somatropin (Recombinant DNA-derived Human Growth Hormone) (21).
Some previous studies also examined the relationship between GH measurements in monoclonal and polyclonal assays, because marked discrepancies were noted only after monoclonal assays came into widespread use (4)(22)(23). However, in method comparison, neither the correlation coefficient nor techniques such as regression analysis is appropriate (7). The Pharmacia (polyclonal) and Delfia (monoclonal) assays were compared in a paper by Albertsson-Wikland et al. (22). A high coefficient of correlation was almost guaranteed (7) because the entire range of basal (24-h profiles) and stimulated GH concentrations from young females and males was investigated. They used the same calibrator that we used.
In the same laboratory, Jansson et al. (4) used this
standard (WHO 80/505) and reported the comparison of the same two
assays using basal samples (24-h profiles) from young females. Their
results may be from the same study (22). From the reported
regression coefficient of 0.78 (4), we have calculated the
mean of ratios, 1/0.78 = 1.28 (their Fig. 2D
). This ratio is
considerably lower than our estimation of 1.59, and the scatter of
points about the regression line is much wider than the scatter of our
ratios about the mean ratio. These differences are more distinct than
expected from otherwise comparable studies using patient samples to
compare two analytical methods for measuring the same component.
However, our results are in agreement with the comparison in the same
paper (4) of another monoclonal assay
(Pharmaciamonoclonal) with the polyclonal assay (Pharmacia;
their Fig. 2C
).
In a study of GH assays by Barth et al. (23), no comparisons were made between Pharmacia and Delfia. However, both of these assays were compared with IDS Gamma BCT, which is essentially a monoclonal assay that does not recognize the 20-kDa GH (24). They used difference plots to present data and found that the difference between Pharmacia and IDS Gamma BCT was greater at stimulated GH concentrations >20 mIU/L. They did not characterize this further, but their results do not conflict with ours because we found that the Pharmacia/Delfia ratio was constant throughout the measured concentration range. In their study, the difference plot is not so informative. More information would probably have been obtained using a ratio plot. Moreover, they used very weak stimulation tests, which, in contrast to the PD-GHRH test (5), will not stimulate GH secretion in all healthy individuals (25).
The advantage of using ratio plots rather than difference plots is evident in situations where the analytical CV values are (or are assumed to be) constant over large concentration ranges and when a proportional relationship is expected. When compared with difference plots of ln-values, the advantage is that the ratios allow immediate interpretation of the data (for expected CV values of ratios up to ~0.15 (15%), whereas the information from a ln-difference plot must first be anti-ln-transformed to be interpreted. An additional advantage of the ratio plot is the clear illustration of the absence of concentration dependency. The lack of concentration dependency supports the hypothesis that, except for the conversion factor (1.59), the GH concentration measurements obtained using the Delfia and Pharmacia methods largely remain within the inherent imprecision of the two methods.
For small values of analytical imprecision (CV<0.1 ~ 10% for both the numerator and the denominator) the use of simple calculations for the limits of the 95% prediction interval is justified by the relevant comparison of the calculated CV of measured ratios with the predicted CV based on imprecision alone (CVPharmacia CVDelfia), which makes interpretation easy. For values >20 mIU/L, the effect of the simple calculation method compared with the correct method (10)(11) is that the ratio for the lower limit is 1.365 instead of 1.374 and that 2.0% of the values will be below this limit instead of the ideal of 2.5%. The corresponding results for the upper limit were 1.815 instead of 1.825 and 3% variance instead of 2.5%. For concentrations <1 mIU/L, the percentages for the lower and upper limits were 1.9% and 3.4%, respectively, compared with the correct value, 2.5%.
This study has demonstrated the use of ratio plots in methods comparison. The importance of a step-by-step evaluation was demonstrated by the use of well-characterized samples from healthy volunteers and patients, according to the expected distribution of ratios based on analytical imprecision and on analytical quality specifications. We have performed an appropriate data analysis for the evaluation of comparability between two GH assays: a polyclonal assay (Pharmacia) and a monoclonal assay (Delfia). We found that it was possible to use a conversion factor to alternate between the two methods without losing diagnostic power from the GH measurements.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
R. Marino, E. Chaler, M. Warman, M. Ciaccio, E. Berensztein, M. A. Rivarola, and A. Belgorosky The Serum Growth Hormone (GH) Response to Provocative Tests Is Dependent on Type of Assay in Autosomal Dominant Isolated GH Deficiency because of an ARG183HIS (R183H) GH-I Gene Mutation Clin. Chem., June 1, 2003; 49(6): 1002 - 1005. [Full Text] [PDF] |
||||
![]() |
P. Lovendahl, J. Adamsen, R. Lund, and P. Lind Technical note: Time-resolved immunofluorometric assay for growth hormone in ruminants J Anim Sci, May 1, 2003; 81(5): 1294 - 1299. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Andersen, C. H. Jensen, R. K. Stoving, J. B. Larsen, H. D. Schroder, B. Teisner, and C. Hagen Fetal Antigen 1 in Healthy Adults and Patients with Pituitary Disease: Relation to Physiological, Pathological, and Pharmacological GH Levels J. Clin. Endocrinol. Metab., November 1, 2001; 86(11): 5465 - 5470. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Chaler, A. Belgorosky, M. Maceiras, M. Mendioroz, and M. A. Rivarola Between-Assay Differences in Serum Growth Hormone (GH) Measurements: Importance in the Diagnosis of GH Deficiency in Childhood Clin. Chem., September 1, 2001; 47(9): 1735 - 1738. [Full Text] [PDF] |
||||
![]() |
C. H. Flowers and J. D. Cook Dried Plasma Spot Measurements of Ferritin and Transferrin Receptor for Assessing Iron Status Clin. Chem., October 1, 1999; 45(10): 1826 - 1832. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |