(Clinical Chemistry. 1998;44:944-949.)
© 1998 American Association for Clinical Chemistry, Inc.
|
Enzymes and Protein Markers |
Serum protein electrophoresis and immunofixation by a semiautomated electrophoresis system
Xavier Bossuyta,
Ann Bogaerts,
Gilberte Schiettekatte,
and Norbert Blanckaert
a Address correspondence to this author at: Central Clinical Laboratory, Clinical Pathology Department, University Hospital Leuven, Kapucijnenvoer 33, B-3000 Leuven, Belgium. Fax 32 16 332896; e-mail xavier.bossuyt{at}uz.kuleuven.ac.be.
 |
Abstract
|
|---|
Semiautomated agarose electrophoresis and immunofixation performed with
Hydrasys-Hyrys(TM) (Sebia) were compared with conventional,
manually performed methods, including cellulose acetate
electrophoresis, immunoelectrophoresis, and immunofixation. Reference
intervals for agarose electrophoresis with Hydrasys-Hyrys were
determined. Within-run imprecision (CV) for fraction quantitation with
the semiautomated system was between 1% (albumin) and 4.5%
(ß-globulin). Total imprecision (CV) was between 2.7% (albumin) and
7.3% (ß-globulin). Semiautomated agarose electrophoresis showed
linear correlation with cellulose acetate electrophoresis. Thirty-four
specimens with monoclonal components were analyzed by manual
immunoelectrophoresis and immunofixation and by Hydrasys. In one case,
a light-chain disease was missed with Hydrasys when the sample was
diluted 1:3 (the routine dilution) but not when the sample was assayed
undiluted. In another case, the Hydrasys system revealed a small IgG
monoclonal component in addition to the IgA monoclonal component
detected by the manual methods. In the other cases, no differences
between the manual methods and the semiautomated method were seen with
respect to paraprotein identification.
 |
Introduction
|
|---|
Serum protein electrophoresis is widely used in clinical
laboratories, especially for the detection and identification of
paraproteins. Traditional clinical electrophoretic procedures are
manual methods that use agarose gels or cellulose acetate membranes as
the separation bed. Quantitation of the five major serum fractions is
done by densitometric scanning of the gel or the membrane. Clinical
interpretation is based on the alteration of the content of one or more
of the five fractions. Agarose as the supporting medium for protein
electrophoresis is reported to give better resolution than cellulose
acetate, with increased ability to detect paraproteins
(1)(2).
Classification of paraproteins is accomplished by technically demanding
immunoelectrophoresis or immunofixation. Immunoelectrophoresis is
lengthy in terms of obtaining the results and is less sensitive than
immunofixation (3)(4)(5)(6).
Several automated protein electrophoresis systems, such as REP and its
variants manufactured by Helena, have been released during the last few
decades. Recently, a new semiautomated system for agarose
electrophoresis of human serum proteins has become commercially
available (Hydrasys-Hyrys(TM), Sebia). The system automatically carries
out the different phases of electrophoresis: sample application,
migration, incubation, staining, destaining, and drying. The system is
also adapted for immunofixation.
We compared the semiautomated Hydrasys-Hyrys system and conventional
manual serum protein separations that use cellulose acetate membrane
electrophoresis. In addition, we also compared immunofixation performed
by Hydrasys-Hyrys with manually performed immunofixation and
immunoelectrophoresis.
 |
Materials and Methods
|
|---|
electrophoretic methods
Semiautomated agarose electrophoresis and immunofixation was
performed with the Hydrasys automate according to the manufacturer's
instructions, using Hydragel 15 and 30 protein(e) and Hydragel 2 and 4
immunofixation gels (Sebia). For protein electrophoresis, 10 µL of
sample was applied manually to the sample template. The sample was
allowed to diffuse for 5 min in the template. The application (30 s),
electrophoresis (pH 8.6, 20 W, and 20 °C for 7 min), and drying
(65 °C for 10 min) were then performed automatically in the
migration compartment of the instrument. The gel was manually
transferred to the staining compartment, in which staining (4 min with
amido black), destaining (three times, for 3 min, 2 min, and 1 min,
respectively), and drying (75 °C for 8 min) were done automatically.
The gels were scanned with the Hyrys densitometer. The immunofixation
procedure on the Hydrasys automate consisted of three steps. In the
first step, the sample was applied by use of a template to 6 different
positions on an agarose gel (Hydragel 2 and 4 immunofixation, Sebia),
and electrophoretic separation (20 W and 20 °C for 9 min) was done
automatically. Either fixative or monospecific antisera (IgG, IgA, IgM,
, or
) was then applied to the electrophoresis lanes. This was
followed by a 5 min incubation step to allow for fixation and
immunoprecipitation. The final step consisted of staining (4 min with
violet blue), destaining (3, 2, and 6 min, respectively), and drying (8
min at 75 °C). This was done automatically in the staining
compartment of the instrument. Detection of monoclonal bands was by
visual inspection of stained gels. The system simultaneously processes
four immunofixation samples. The manufacturer's recommendation for
serum dilution before immunofixation was 1:6 for the IgG lane and 1:3
for the other lanes. The dilution was adapted depending on the
concentration of the paraprotein. Generally, if the immunochemical
measurement of the paraprotein exceeded 25 g/L, the dilution was 1:10
for the IgG lane and 1:5 for the other lanes. On the other hand, if the
immunochemical measurement of the paraprotein was <10 g/L, the
dilution was 1:4 for the IgG lane and 1:2 for the other lanes. Bence
Jones detection and characterization with Hydrasys was performed with
nonconcentrated urine and Hydragel 2 and 4 Bence Jones gels (Sebia).
The antibodies used for immunofixation were from Sebia.
Cellulose acetate electrophoresis was done manually with Sepharose
cellulose polyacetate electrophoresis strips for the Microzone System
(Gelman Sciences). The gels were equilibrated in diethylbarbital
buffer, pH 8.6. We applied ~1 µL of each sample and carried out the
electrophoresis at 220 V for 25 min in diethylbarbital buffer (pH 8.6).
Proteins were stained by incubating the gels in Ponceau S (Analis) for
10 min. The individual fractions were quantified by densitometry
(Beckman Appraise, Beckman Instruments). For conventional
immunofixation, the Paragon kit (Beckman Instruments) was used
according to the manufacturer's instructions. For small paraproteins
(<10 g/L by immunochemical measurement), a 1:5 dilution was used.
Immunoelectrophoresis was performed as described by Grabar and Williams
(7), using barbital buffer, pH 8.6, and antibodies from
Sanofi Diagnostics Pasteur. The gels were from Kallestad Laboratories.
Detection and identification of Bence Jones proteins with the manual
techniques were done after concentration of the urine using Minicon
(Amicon).
evaluation details
Reference intervals were established in accordance with NCCLS
guideline C28-A (8). Cellulose acetate protein
electrophoresis and nephelometric determination (IFCC-standardized) of
IgG, IgA, IgM, haptoglobin, C3c, albumin, transferrin, and
1-acid glycoprotein (see below) were performed on all
candidate samples. We excluded samples that were hemolytic, icteric,
lipemic, that displayed an M-component, or that had an abnormal value
for one of the specific proteins determined.
Within-assay and total imprecision were calculated as described in
NCCLS EP5-T2 (9) from 50 determinations of a serum pool
analyzed in duplicate over 13 days. Each day, two assays with two
aliquots of the pool were analyzed. The assays were separated by a
minimum of 2 h. The serum pool was prepared from patient specimens
filtered through 8 µm (pore size) filters (Elkay Products). The pool
was aliquoted and stored at -20 °C until assay. The pool
concentration was 77 g/L.
Method comparisons were carried out with linear regression analysis on
50 patient specimens analyzed over 5 days, as described in NCCLS
guideline EP9-A (10).
other analyses
Quantification of serum proteins (IgG, IgA, IgM, kappa light
chains, lambda light chains, haptoglobin, C3c, albumin, transferrin,
and
1-acid glycoprotein) was done by endpoint
nephelometry using a Behring BN 100 instrument (Behringwerke). All
reagents and calibrators were from Behring (Behringwerke). Total
protein was determined using the Boehringer Mannheim reagent kit and
application on a Hitachi 747 analyzer (Boehringer Mannheim).
statistical analysis
Statistical analysis was performed with the use of the Statistical
Analysis System (SAS Institute).
 |
Results and Discussion
|
|---|
protein electrophoresis and immunofixation by hydrasys
Agarose electrophoresis performed by Hydrasys on serum samples
from hospitalized patients is shown in Fig. 1
. Batchwise, 30 serum protein samples could be electrophoresed
simultaneously every 20 min with the Hydrasys system. The processing of
the first series took 40 min. In comparison, Fig. 2
shows cellulose acetate electrophoresis of 16 of the same
samples processed by the Hydrasys system. Samples 18 of the upper
cellulose acetate gel correspond to samples 310 of Fig. 1
. Samples
18 of the lower gel correspond to samples 1825 of Fig. 1
. The
resolution is higher and the clarity of the electrophoresis bands is
better with agarose (Hydragel) than with cellulose acetate (Microzone).
For example, the oligoclonal pattern that is obvious in lane 8 on
agarose gel electrophoresis (Fig. 1
) is less clear on cellulose acetate
electrophoresis (Fig. 2
, upper gel, lane 6). Fig. 3
shows immunofixation performed by Hydrasys on four samples. The
immunofixation revealed a IgG
paraprotein (serum 1, lower lefthand
panel), an IgG
monoclonal protein (serum 2, lower righthand panel),
absence of a monoclonal protein (serum 3, upper lefthand panel), and an
IgG
monoclonal protein (serum 4, upper righthand panel). Fig. 4
shows the manual immunofixation (Paragon) of the lower
righthand sample of Fig. 3
. The immunoelectrophoresis of the same
sample is shown in Fig. 5
. Both manual techniques reveal a IgG
paraprotein.

View larger version (104K):
[in this window]
[in a new window]
|
Figure 1. Agarose protein electrophoresis using Hydrasys was
performed on 30 samples from hospitalized patients as described in
Materials and Methods.
|
|

View larger version (84K):
[in this window]
[in a new window]
|
Figure 2. Cellulose acetate electrophoresis was performed as
described in Materials and Methods.
Samples 18 (top to bottom) of the
upper gel correspond to samples 310 of Fig. 1
. Samples
18 (top to bottom) of the lower gel
correspond to samples 1825 of Fig. 1
.
|
|
reference intervals
Reference intervals for the five electrophoretic serum protein
fractions were determined with Hydrasys-Hyrys from a population of 76
healthy men and 76 healthy women. Median values and the central 95% of
the distribution, expressed as a fraction percentage and in absolute
values (g/L), are presented in Table 1
. All distributions were gaussian. No statistically significant
(P <0.05) sex-related differences were observed (Student's
t-test).
imprecision
In a preliminary precision test, 20 aliquots of a serum were
assayed by Hydrasys-Hyrys in sequence, as proposed by NCCLS EP5-T2
(3). The CV values were 0.90%, 4.61%, 1.95%, 1.86%, and
2.4% for the albumin,
1-globulin,
2-globulin, ß-globulin, and
-globulin fractions,
respectively. The total imprecision and within-run imprecision for
protein electrophoresis by Hydrasys-Hyrys are listed in Table 2
. These CV values are clearly better than previously published
CV values for manually performed cellulose acetate electrophoresis
(11)(12) and CV values for cellulose acetate
electrophoresis obtained in our laboratory. The latter were 1.6%,
8.8%, 6.3%, 7.9%, and 6.2% (n = 49; over 25 days) for the
albumin,
1-globulin,
2-globulin,
ß-globulin, and
-globulin fractions, respectively. The excellent
CV values that can be obtained with Hydrasys-Hyrys probably are related
to the automation of most steps of the procedure, including sample
application, migration, and staining.
method comparison studies
To compare Hydrasys-Hyrys with cellulose acetate electrophoresis,
50 samples were analyzed with both systems. The five major serum
fractions were quantified. The linear regression plots are shown in
Fig. 6
. For each of the five fractions, good linear correlation was
found between both methods, as reflected by the good coefficients of
correlation (>0.98).

View larger version (22K):
[in this window]
[in a new window]
|
Figure 6. Linear correlation between cellulose acetate
electrophoresis (x-axis) and Hydrasys-Hyrys
(y-axis) for each of the five serum protein fractions.
Linear regression analysis was performed on 50 patient specimens
analyzed over 5 days. The parameter estimates of the regression
analysis for, respectively, albumin, 1-globulin,
2-globulin, ß-globulin, and -globulin are as
follows: slope (1.01 ± 0.02, 0.81 ± 0.05, 0.89 ±
0.04, 1.00 ± 0.08, and 0.92 ± 0.03); intercept (-0.01
± 0.08, -0.01 ± 0.01, 0.16 ± 0.03, 0.01 ± 0.05, and
-0.01 ± 0.03); and correlation coefficient (0.99, 0.91, 0.95,
0.89, and 0.98).
|
|
monoclonal immunoglobulin and light chain disease
Thirty-four specimens with a monoclonal component were analyzed
with Hydrasys and with the two reference methods, immunoelectrophoresis
and immunofixation. We investigated 5 cases of light chain disease
(4
and 1
), 2 of IgA (1 IgA
and 1 IgA
), 23 of IgG (14 IgG
and 9 IgG
), and 4 of IgM (2 IgM
and 2 IgM
) monoclonal
paraproteins. In addition, two samples with biclonal paraproteins were
also evaluated.
The results of the comparison study are summarized in Table 3
. Generally, no clinically significant differences between the
manual methods and Hydrasys were seen with respect to characterization
of monoclonal paraproteins. In one case, the Hydrasys system revealed a
small IgG
monoclonal component in addition to the IgA monoclonal
component detected by the manual comparison methods. In another case,
the heavy chain but not the light chain could be identified by Hydrasys
using the standard dilution. The light chain type was revealed by
immunoelectrophoresis.
The group of paraproteins that we studied also included five cases of
light chain disease. Of these five cases, one case was missed with
Hydrasys when the usual 1:3 serum dilution was used. However, when the
serum sample was assayed undiluted, free light chains could be detected
by this system. Free light chains in serum are more easily detected by
immunoelectrophoresis than by immunofixation. This is because
immunoelectrophoresis includes a diffusion step that guarantees
antibody-antigen equivalence. With immunofixation, great care in the
control of the antigen concentration to maximize sensitivity is
necessary. It should be pointed out that the usual dilution recommended
by the manufacturer is not appropriate for all samples. Each sample
should be evaluated individually and the serum dilution adapted
according to the amount of paraprotein. Therefore, high-resolution
agarose electrophoresis should be available before immunofixation is
performed.
In addition to monoclonal paraproteins, we have also investigated two
cases of biclonal (oligoclonal) disease. In one case, Hydrasys revealed
an IgG
in combination with an IgM
and free
, whereas
immunoelectrophoresis and manual immunofixation disclosed IgG
and
free
. In another case, the manual techniques revealed an IgG
in
combination with IgG
, whereas the same paraproteins with an
additional IgG
was found with Hydrasys.
We also evaluated 20 randomly selected samples that did not reveal a
monoclonal component by the manual techniques. None of them revealed a
monoclonal protein by Hydrasys.
urinalysis
Identification of Bence Jones proteins in urine was evaluated on
16 samples. The samples included 10
and 6
Bence Jones proteins.
An IgG monoclonal protein was present in 4 of the 16 cases. No
discordances were found between Hydrasys and the two manual methods. In
four additional cases, a polyclonal increase of the light chains was
equally detected by all three techniques. The manual method used
concentrated urine, whereas the Hydrasys system used nonconcentrated
urine, which is a time-saving feature. Concentrated urine can also be
used by Hydrasys with the Hydragel protein(e) gel and might be the more
sensitive method.
In conclusion, Hydrasys provides for reproducible and rapid serum
protein electrophoresis. The system also provides for accurate
identification of paraproteins, which ascertains the quality and
appropriateness of the specific antibodies.
 |
Acknowledgments
|
|---|
We acknowledge H. Raveschot and M. Artoos for their expert
technical assistance. We wish to thank G. Marien and E. Stevens for
their contribution in generating the immunoelectrophoresis data.
 |
Footnotes
|
|---|
Central Clinical Laboratory, Department of Clinical Pathology, University Hospital of Leuven, Kapucijnenvoer 33, B-3000 Leuven, Belgium.
 |
References
|
|---|
-
Jeppson JO, Laurell CB, Franzén B. Agarose gel electrophoresis. Clin Chem 1979;25:629-638.
[Free Full Text]
-
Aguzzi F, Jayakar SD, Merlini G, Petrini C. Electrophoresis: cellulose acetate vs agarose, visual inspection vs densitometry. Clin Chem 1981;27:1944-1945.
[Free Full Text]
-
Gerard S, Chen K, Khayam-Bashi H. Immunofixation compared with immunoelectrophoresis for the routine characterization of paraprotein disorders. Am J Clin Pathol 1987;88:198-203.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Duc J, Morel B, Peitrequin R, Frei PC. Identification of monoclonal gammopathies: a comparison of immunofixation, immunoelectrophoresis and measurements of kappa- and lambda-immunoglobulin levels. J Clin Lab Immunol 1988;26:141-146.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Guinan JEC, Kenny DF, Gatenby P. Detection and typing of paraproteins: comparison of different methods in a routine diagnostic laboratory. Pathology 1989;21:35-41.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Chu SY, MacLeod JE, Bocci L, Monteith M. Characterization of small monoclonal protein bands with Beckman's "Paragon" immunofixation system. Clin Chem 1987;:617.
-
Grabar P, Williams CA. Méthode permettant l'étude conjugée des propriétés électrophoreétique et immuno-chimiques d'un mélange de protéines. Application en serum sanguin. Biochim Biophys Acta 1953;10:193-194.
[Medline]
[Order article via Infotrieve]
-
National Committee for Clinical Laboratory Standards.
How to define and determine reference intervals in the clinical
laboratory: approved guideline. NCCLS Document C28-A. Villanova, PA:
NCCLS, 1992;15..
-
National Committee for Clinical Laboratory Standards.
Evaluation of precision performance of clinical chemistry devices, 2nd
ed. Tentative guideline. NCCLS Document EP5T2. Villanova, PA: NCCLS
1992;12..
-
. National Committee for Clinical Laboratory Standards. Method comparison and bias estimation using patient samples: approved guideline. NCCLS Document EP9-A 1995:15 NCCLS Villanova, PA. .
-
Baars JD, Lombarts AJ. Imprecision of protein electrophoresis. Clin Chem 1986;32:1425-1426.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Kahn SN, Strony LP. Imprecision of quantification of serum protein fractions by electrophoresis on cellulose acetate. Clin Chem 1986;32:356-357.
[Abstract/Free Full Text]