| Abstract |
|---|
|
|
|---|
12%. We evaluated the
analytical and clinical performance of the direct N-geneous LDL-C assay
(Equal Diagnostics). The N-geneous method correlated highly with the
modified beta-quantification assay (r = 0.95;
y = 0.91x + 70.6 mg/L; n = 199),
showed no significant effect of increased triglyceride or other common
interferants, and performed adequately in serum samples from nonfasting
individuals. This assay demonstrated a mean total error of 6.75% over
a wide range of LDL-C concentrations. In addition, at the medical
decision cutoff points, this LDL-C assay showed positive predictive
values of 7895% and negative predictive values of 8499%. We
conclude that the N-geneous LDL-C meets the currently established
analytical performance goals and appears to have a role in the
diagnosis and management of hypercholesterolemic patients. | Introduction |
|---|
|
|
|---|
In most clinical laboratories, LDL-C is currently derived by the Friedewald equation (2). Although the estimation method correlates highly with the beta-quantification, it has several shortcomings (3)(4). This calculation is invalid when a specimen is collected in the nonfasting state or from a patient with Type III hyperlipoproteinemia, or in the presence of increased triglycerides (>4000 mg/L). Furthermore, according to the NCEP Laboratory Standardization Panel (LSP), the LDL-C concentration must be determined with a total analytical error not exceeding ± 12% (<±4% imprecision and <±4%bias) for a patient to be correctly classified into the NCEP risk categories (4)(5). The Friedewald calculation may not meet these relatively rigid criteria because it combines the inaccuracy and imprecision of three independent measurements: total cholesterol (TC), triglycerides (TGs), and HDL-cholesterol (HDL-C) (3).
Chemical precipitation and immunoseparation methods for the quantification of LDL-C concentration have been reported (4). These methods are affected, to different degrees, by TG concentration, do not measure all "LDL components" as determined by beta-quantification, marginally meet the LSP analytical performance criteria, and require a pretreatment step (6)(7)(8)(9)(10)(11). Here we describe the analytical and clinical performance of a new homogeneous method, N-geneous LDL-C (Equal Diagnostics). We also compare this method with the ultracentrifugation/dextran sulfate-Mg2 (UC-DS) method (the modified Lipid Research Clinics beta-quantification method for LDL-C) and the Friedewald calculation.
| Materials and Methods |
|---|
|
|
|---|
TC, TGs, and HDL-C.
Serum cholesterol and TG
concentrations were determined enzymatically on the Hitachi 911
analyzer (Boehringer Mannheim). The measurement of TGs was corrected
for the presence of endogenous glycerol. The HDL fraction was
separated, as previously described (13), using
DS-Mg2 precipitation reagents (same method as described
above for the beta-quantification), and the cholesterol component was
measured enzymatically on the Hitachi 911 analyzer. TC was determined
with a day-to-day variation (CV) of 1.3% at concentrations of 1400 and
2000 mg/L; TG CVs were 2% and 1.6% for concentrations of 1000 and
2000 mg/L, respectively; HDL-C CVs were 2.7% and 2.0% at
concentrations of 240 and 480 mg/L, respectively. TC was determined
with an average ± SD bias from the Centers for Disease Control
and Prevention (CDC) target values of 0.62% ± 1.1% for
concentrations ranging from 1278 to 2360 mg/L; the TG average
± SD bias of -0.89% ± 3.1% for concentrations ranging from 536
to 2340 mg/L; and the HDL-C average ± SD bias of -1.0%
± 2.4% for concentrations ranging from 270 to 794 mg/L. Our laboratory
is certified by the National Heart, Lung, and Blood Institute and CDC
Lipid Standardization Program.
N-geneous LDL-C.
This assay was performed according to the
manufacturer's specifications on the Hitachi 911 analyzer. The assay
is available from Equal Diagnostics, Exton, PA (cat. no. 7120). After
Reagent 1 is mixed with the serum specimen, Detergent 1 specifically
disrupts non-LDL lipoproteins and causes the release of cholesterol.
The formed free cholesterol, a result of hydrolysis by cholesterol
esterase, reacts with cholesterol oxidase, generating hydrogen
peroxide. The latter is consumed by a peroxidase in the presence of
4-aminoantipyrine to generate a colorless product. On the addition of
Reagent 2, Detergent 2 specifically releases cholesterol from LDL
particles. A similar enzymatic reaction to that described above occurs,
except for hydrogen peroxide, which reacts with
N,N'-bis-(4-sulfobutyl)-m-toluidine
disodium salt, to generate a colored product. The intensity of the
generated color is proportional to the concentration of LDL-C. The
assay was calibrated daily, using the calibrators provided (cat. no.
7272; LDL-C concentrations of 0 and 1130 mg/L).
Friedewald calculation.
LDL-C was estimated by the Friedewald
calculation [LDL-C = TC - (HDL-C TG/5)], only when
fasting TGs are <4000 mg/L, where TG/5 is an estimate of
VLDL-cholesterol (VLDL-C), and all concentrations are expressed in
mg/L. When values are expressed in mmol/L, VLDL-C is estimated as
TG/2.22.
Triglyceride-rich lipoproteins isolation.
The
triglyceride-rich lipoproteins fraction (>1.006 kg/L) was isolated
from serum samples by UC as described above. The isolated fractions of
several samples were mixed, and the TG content of the pool was
determined.
samples
One hundred and ninety-nine fasting serum samples were obtained
over a 1-month period for the comparison study from the daily pool of
received specimens at several clinical chemistry laboratories,
including Beth Israel-Deaconess Medical Center, the Veteran
Administration Medical Center in Washington, D.C. and Seattle, and
Quest Labs.
To determine the postprandial effect on the determination of LDL-C by the N-geneous method, paired 9- to 12-h fasting and nonfasting (3 h after high fat meal) blood samples were obtained from 36 individuals on the same day. The high fat meal consisted of a McDonald's Sausage McMuffin and hash browns (total fat, 31 g; calories from fat, 180). In addition, to establish the effect of different anticoagulants on the determination of LDL-C by the N-geneous assay, blood samples from 19 individuals were simultaneously collected from a single venipuncture in tubes containing (in random order) EDTA (final concentration 35 g/L), heparin, or no anticoagulant. Serum samples were allowed to sit for 2030 min at room temperature before centrifugation. EDTA- and heparinized-plasma were separated from blood cells within 15 min of collection.
analytical performance evaluation
Precision.
Within-run and day-to-day precision studies were
performed using fresh-frozen human serum with LDL-C concentrations
ranging from 832 to 1566 mg/L. Aliquots from the different pools were
prepared and stored at -80 °C.
Ascorbate, bilirubin, hemoglobin, and TG interference.
Five
pools were prepared by combining sera from ~30 individuals. The TC,
TG, HDL-C, and LDL-C concentrations of the pools were as follows: pool
1, 1730 mg/L TC, 1190 mg/L TG, 447 mg/L HDL-C, and 910 mg/L LDL-C; pool
2, 1720 mg/L TC, 1670 mg/L TG, 359 mg/L HDL-C, and 970 mg/L LDL-C; pool
3, 2100 mg/L TC, 990 mg/L TG, 521 mg/L HDL-C, and 1305 mg/L LDL-C; pool
4, 1850 mg/L TC, 1400 mg/L TG, 490 mg/L HDL-C, and 955 mg/L LDL-C; and
pool 5, 2260 mg/L TC, 1010 mg/L TG, 669 mg/L HDL-C, and 900 mg/L LDL-C.
For the ascorbate interference study, various amounts of the stock ascorbic acid solution (1000 mg/L in isotonic saline; Sigma Chemicals) were added to the different pools to give final concentrations of 0, 50, 100, and 150 mg/L. The dilutions were done such that the same amount of saline was delivered to each sample. The concentration of ascorbic acid was confirmed by HPLC (14).
For the bilirubin interference study, various amounts of the stock bilirubin solution (2000 mg/L in 0.1 mol/L sodium hydroxide; Pfanstiehl) were added to the different pools to give final concentrations of 3.4, 41, 108, and 178 mg/L. The dilutions were done such that the same amount of sodium hydroxide was delivered to each sample. The final concentration of total bilirubin was established on the Hitachi 911 with the use of 2,5-dichlorophenyldiazonium tetrafluoroborate as the diazonium salt, and the absorbance was monitored at 570 nm.
For the hemoglobin interference study, hemolysate was prepared by (a) isolating red blood cells, (b) washing the cells with isotonic saline five times, (c) bursting the cells by freezing, and (d) removing sedimented red cell ghosts and membrane debris by centrifugation at 2000 rpm for 15 min. The hemoglobin concentration in the hemolysate was determined using a Technicon H3(TM) analyzer. Various amounts of the prepared hemolysate (100 g/L) were added to the different pools to give final hemoglobin concentrations of 0.48, 2.12, 5.20, and 10.30 g/L. The final concentration of hemoglobin was confirmed on the Hitachi 911, using the hemolysis index function.
For the TG interference study, TG-rich lipoprotein fractions were prepared as described above, and various amounts of the isolate (TGs, 45 000 mg/L) were added to the different pools to give mean final TG concentrations of 972, 3084, 6088, and 10 780 mg/L. The concentrations of TGs were determined enzymatically as described above.
Stability study.
Four serum pools were prepared and stored in
small aliquots at either 4, -20, or -80 °C up to 1 month. LDL-C
concentrations were measured daily, using the N-geneous assay, over a
period of 4 weeks (n = 20).
statistical analysis
The means, medians, and SDs were calculated with Microsoft Excel,
Ver. 5.0 (Microsoft). Student's t-test and least-squares
linear regression analysis were performed using SigmaPlot statistics
program (Jandel Scientific). The t-test was considered
significant at P <0.05. Biases were calculated as test
procedure result (N-geneous assay or Friedewald calculation) minus
reference procedure result.
Total error was calculated as the summation of the systematic and random error [total error (%) = bias (imprecision x 1.96)] (15)(16). Systematic error, including the constant and the proportional error, was derived from the linear regression equation, y = bx a, where b was the slope of the regression equation and represented the proportional error and a was the y-axis intercept and represented the constant error. Systematic error at a particular LDL-C concentration (xc) was defined as the absolute value of yc - xc, where yc = bxc a. Random error was calculated as the day-to-day precision multiplied by 1.96.
The positive predictive value (PPV) of an LDL-C assay at each specified cutoff point was calculated as [true positive/(true positive false positive)] x 100, where true positive meant that LDL-C results of both the reference procedure and the test method (N-geneous assay or Friedewald calculation) were greater than or equal to the cutoff concentration, and false positive meant that the test method LDL-C result was greater than the cutoff point when the reference procedure LDL-C value was less than the cutoff point. The negative predictive (NPV) value of an LDL-C assay at each specified cutoff point was calculated as [true negative/(true negative false negative)] x 100, where true negative meant that LDL-C results of both the reference procedure and the test method were less than the cutoff point concentration, and false negative meant that the test method LDL-C result was less than the cutoff point when the reference procedure LDL-C value was greater than or equal the cutoff point concentration. The sensitivity of an assay was calculated as [true positive/(true positive false negative)] x 100, and the specificity as [true negative/(true negative false positive)] x 100.
| Results and Discussion |
|---|
|
|
|---|
The reproducibility of the assay was examined using fresh-frozen serum pools with different LDL-C concentrations. The within-run imprecision values (n = 20) at LDL-C concentrations of 832, 1156, and 1566 mg/L were 1.63%, 1.71%, and 2.68%, respectively. Furthermore, the day-to-day imprecision values (n = 21) at LDL-C concentrations of 900, 965, 1060, and 1291 mg/L were 3.1%, 3.2%, 3.2%, and 3.1%, respectively. According to the NCEP-LSP performance goals, LDL-C must be determined with an imprecision of <±4%. The N-geneous assay has met this performance criterion in serum pools with different LDL-C concentrations.
For any new assay to be accepted clinically, it must demonstrate
comparable performance to the beta-quantification method, which is the
accuracy base on which the NCEP cutoff points are determined. The LDL-C
concentrations of the serum samples collected from fasting
normolipidemic and hyperlipidemic individuals were simultaneously
determined, over a 1-month period and 15 different analytical runs, by
the modified beta-quantification method (UC-DS), the N-geneous assay,
and the currently recommended routine method, the Friedewald
calculation. The LDL-C concentrations were calculated by the latter
method only for those samples with TG concentrations <4000 mg/L. The
lipid and lipoprotein concentrations of the study population are
presented in Table 1
. The comparison-of-methods plot [UC-DS (x) vs test
method (y)] showed a least-squares linear regression
equation of y = 0.91x 70.6 mg/L (n =
199; r = 0.95) for the N-geneous assay and
y = 0.96x 43.4 mg/L (n = 164;
r = 0.94) for the Friedewald calculation (Fig. 1
). To better illustrate the performance of the N-geneous assay
in hypertriglyceridemic samples, specimens with TGs <4000 mg/L and
those with TGs
4000 mg/L were compared separately with the
UC-DS assay. Least-squares linear regression equations of
y = 0.89x 89.8 mg/L (n = 164;
r = 0.94) for samples with TGs <4000 mg/L and
y = 1.02x - 29 mg/L (n = 35;
r = 0.96) for samples with TG
4000 mg/L were
obtained. In addition, Table 1
indicates a mean LDL-C bias ± SD,
as determined by the N-geneous assay of -48.4 ± 119 mg/L for the
total population (n = 199), -58.4 ± 118 mg/L for those with
TGs <4000 mg/L (n = 164), and -1.6 ± 110 mg/L for those
with TGs
4000 mg/L (n = 35). This study demonstrates that LDL-C
measured by the N-geneous assay is comparable with that determined by
the modified beta-quantification method. Our findings also suggest that
this method measures the cholesterol of the "LDL components", as
determined by beta-quantification, which includes LDL,
intermediate-density lipoproteins, and lipoprotein(a). However, the
exact reason for the negative bias seen in samples with TGs <4000 mg/L
is not known at the present. Perhaps the cholesterol component of
either LDL, intermediate-density lipoproteins, or lipoprotein(a) is not
totally measured by the N-geneous assay in the absence of increased
TGs. This observation could represent a limitation of the N-geneous
assay because the great majority of the population has TGs <4000 mg/L.
|
|
No significant bias from LDL-C measured by the reference procedure was
seen over a wide range of TG concentrations (37011 320 mg/L; Fig. 2
A), and the addition of various amounts of TG-rich lipoproteins
(up to 10 780 mg/L) to five different pools revealed no statistically
significant change in measured mean LDL-C values by the N-geneous
assay. The mean and SD of LDL-C concentrations at baseline were
699 ± 181 mg/L (mean TGs, 972 mg/L); after the addition of
TG-rich lipoproteins, the mean and SD were 696 ± 177, 694
± 173, and 685 ± 176 mg/L at mean TG concentrations of 3084, 6088,
and 10 780 mg/L, respectively. The experiments clearly indicate that,
unlike previously described methods (6)(7)(8)(9)(10)(11), the N-geneous
assay for LDL-C is not affected by the presence of increased
concentrations of TG. In addition, no significant bias in calculated
LDL-C by the Friedewald equation from the reference procedure (Fig. 2B
)
was seen with increasing TG concentrations up to 4000 mg/L.
|
The NCEP performance goals require that LDL-C be measured with an analytical bias of <±4% (5). The substitution of LDL-C values used for medical decisions (10001900 mg/L) into the least-squares linear regression equation for the N-geneous assay yielded systematic errors that ranged from 0.4% to 1.6%. According to the NCEP-LSP, the total allowable analytical error, which is the combination of random error and systematic error, for LDL-C must not exceed ±12% (5). When calculated at the different LDL-C concentrations used in the day-to-day precision study (900, 965, 1060, and 1291 mg/L), the average total analytical error of the N-geneous assay was 6.75% (SD = 0.24%; range 6.57.0%). Therefore, the N-geneous assay appears to analytically meet the performance goals established by the NCEP. For the sake of comparison, the same parameters were estimated for the Friedewald calculation. The bias and imprecision values for TC, HDL-C, and TGs were combined. The average systematic error, random error (imprecision multiply by 1.96), and total error for the calculated LDL-C were 0.8%, 10.8%, and 11.6%, respectively.
Bilirubin and ascorbic acid are reducing compounds that are known to
interfere with peroxidase-dependent cholesterol measurement. In
addition, hemoglobin is a common interferant in routine clinical
laboratory testing. In this study, the potential interference from
these three analytes was evaluated by the addition of increasing and
known concentrations of hemoglobin, bilirubin, or ascorbate to
different serum pools (Fig. 3
). A slight positive trend was seen in LDL-C values as
hemoglobin concentration increased in the test pools (Fig. 3A
). These
differences, however, fell short of statistical significance. In
contrast, a slight negative trend was observed in LDL-C values as
bilirubin concentrations increased in the various pools (Fig. 3B
). The
addition of up to 258 mg/L bilirubin caused LDL-C concentrations to
decrease by no more than 5% (data not shown). This difference was not
statistically significant. Concurrently, the addition of up to 150 mg/L
of ascorbic acid did not result in a significant change in LDL-C, as
measured by the N-geneous assay (Fig. 3C
). Our findings indicate the
N-geneous assay is apparently unaffected by these common interferants.
|
In this study, the effect of feeding on the determination of LDL-C concentration by the N-geneous assay was examined in paired samples from 36 subjects. As expected, TGs increased significantly after feeding (1090 ± 635 mg/L vs 1830 ± 1090 mg/L; P <0.0001). However, no statistically significant difference was seen in TC [fasting (mean ± SD) 1943 ± 385 mg/L vs nonfasting 1952 ± 368 mg/L], HDL-C (560 ± 156 mg/L vs 550 ± 167 mg/L) or LDL-C (1122 ± 299 mg/L vs 1113 ± 273 mg/L)] concentration in samples obtained in the fasting state and postprandially. It is important to note, however, that 2 of the 36 subjects who participated in this study would have been misclassified into NCEP risk categories if postprandial serum samples were used in the determination of their LDL-C (subject 1, 1580 mg/L fasting vs 1200 mg/L nonfasting; subject 2, 1480 mg/L fasting vs 1120 mg/L nonfasting). Parallel decreases in TC postprandially have also been noted in the same two subjects and would have caused them to be misclassified into NCEP risk categories as well (subject 1, 2530 mg/L fasting vs 2000 mg/L nonfasting; subject 2, 2340 mg/L fasting vs 1920 mg/L nonfasting). Compared with values in the fasting state, LDL-C determined postprandially had a PPV of 100% and an NPV of 93%. This assay appears to perform better than the LDL-C immunoprecipitation method in postprandial samples (7). To further characterize the performance of the assay, the fasting and nonfasting LDL-C values were examined separately in the 11 hypercholesterolemic (LDL-C >1300 mg/L) and the 25 normocholesterolemic subjects in this group. In the hypercholesterolemic group, the fasting and nonfasting mean ± SD of LDL-C concentrations were 1475 ± 117 mg/L and 1404 ± 142 mg/L (P <0.23), respectively. The two discrepant subjects described above fell into this category. In the normocholesterolemic group, the fasting and nonfasting LDL-C values were 967 ± 207 mg/L and 984 ± 211 mg/L (P <0.78), respectively. Physiological changes in lipoprotein concentrations are known to occur postprandially. An average decrease of 22% in LDL-C concentration as determined by beta-quantification 3 h after fat feeding has been reported (19). It is not known at present why the N-geneous LDL-C assay is unaffected postprandially. Perhaps some VLDL remnants, which are increased postfeeding, are measured as part of LDL-C, thus compensating for the physiologically decreased LDL.
A stability study was conducted to determine a possible change in LDL-C
concentrations of four different freshly prepared serum pools stored at
4, -20, or -80 °C for up to 4 weeks. The mean (SD) of LDL-C
concentrations of the various pools at baseline and after 1-week,
2-week, 3-week, and 4-week storage periods at 4, -20, or -80 °C
are presented in Table 2
. None of these changes were statistically significant. The
short-term stability of this assay at various temperatures provides the
clinical laboratory with an added practical advantage and suggests a
role for this method in clinical and epidemiological studies. However,
a long-term stability study must demonstrate the lack of effect from
storage at -20 or -80 °C on LDL-C values before the N-geneous
method is considered for longitudinal trials. Such a study is ongoing
in our laboratory at present.
|
Total cholesterol has been used as a surrogate test for LDL-C in
screening for hypercholesterolemia. We evaluated the N-geneous LDL-C
assay as a potential screening test for high cholesterol in this study
population (n = 199). Hypercholesterolemia was defined as LDL-C
1300 mg/L, as determined by beta-quantification. The sensitivity and
specificity of TC (at concentrations
2000 mg/L) as the screening test
were 98% and 60%, respectively (Table 3
). The high incidence of false positives when TC is used usually
diminishes by the measurement of HDL-C along with TC at the initial
screening stage, as has been recommended by the NCEP-Adult Treatment
Panel II (1). In this study population, when TC -
(HDL-C 300 mg/L, which represents the mean VLDL-C in American adults)
was used (at concentrations
1300 mg/L), the specificity was increased
by over 10% (Table 3
). The sensitivity and specificity of LDL-C (at
concentrations
1300 mg/L), as determined by the N-geneous assay, were
77% and 92%, respectively. The sensitivity of the N-geneous LDL-C
assay, however, improved when a lower LDL-C cutoff point of
1200 mg/L
was used to detect hypercholesterolemic subjects (LDL-C
1300 mg/L by
beta-quantification; Table 3
). Our data indicate that the direct
measurement of LDL-C by this assay, using a lower cutoff point, has
better specificity (81% vs 71%) and sensitivity (91% vs 87%) than
the TC - (HDL-C 300 mg/L) currently used in screening for
hypercholesterolemia. Furthermore, we evaluated the suitability of the
N-geneous LDL-C method in the management of patients with
hypercholesterolemia. According to the NCEP-Adult Treatment Panel II
guidelines (1), the management of hyperlipidemic patients,
using either dietary or drug therapy, is based on four LDL-C cutoff
points (
1000,
1300,
1600, or
1900 mg/L). The ability of this
assay to correctly classify subjects at the medical decision cutoff
points was evaluated in this study population (n = 199). For the
interest of comparison, the PPV and NPV of LDL-C estimated by the
Friedewald calculation were also examined, using only those subjects
with TGs <4000 mg/L (n = 164). LDL-C concentrations determined by
either the Friedewald calculation or the N-geneous assay classified
over 80% of the subjects into the above mentioned cutoff points,
correctly. The PPV of LDL-C estimated by either method decreased as
LDL-C concentrations increased (PPV range 9578% for N-geneous assay
and Friedewald calculation; Fig. 4
A). In contrast, the NPV of LDL-C estimated by either method
increased as LDL-C concentrations increased (NPV range 8499% for
N-geneous assay and 8899% for Friedewald calculation; Fig. 4B
).
About 65% of patients with LDL-C ranging from 1300 to 1590 mg/L were
classified into this category by either the N-geneous assay or the
Friedewald calculation (N-geneous PPV, 64%; NPV, 84%; Friedewald PPV,
69%; NPV, 89%). The N-geneous assay appears to be capable of
classifying patients into NCEP cutoff points, regardless of their TG
concentrations.
|
|
In conclusion, the N-geneous LDL-C assay meets the established NCEP analytical performance goals, is unaffected by major interferants, including increased TGs, and appears to be a potential screening test for hypercholesterolemia and a possible useful tool in the management of hyperlipidemic patients.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
S. Mora, N. Rifai, J. E. Buring, and P. M Ridker Comparison of LDL Cholesterol Concentrations by Friedewald Calculation and Direct Measurement in Relation to Cardiovascular Events in 27 331 Women Clin. Chem., May 1, 2009; 55(5): 888 - 894. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Giovannucci, Y. Liu, B. W. Hollis, and E. B. Rimm 25-Hydroxyvitamin D and Risk of Myocardial Infarction in Men: A Prospective Study Arch Intern Med, June 9, 2008; 168(11): 1174 - 1180. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Wu, W. C. Willett, N. Rifai, and E. B. Rimm Plasma Fluorescent Oxidation Products as Potential Markers of Oxidative Stress for Epidemiologic Studies Am. J. Epidemiol., September 1, 2007; 166(5): 552 - 560. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. de Ferranti, D. Shapiro, R. Markowitz, E. Neufeld, N. Rifai, and H. Bernstein Nonfasting Low-Density Lipoprotein Testing: Utility for Cholesterol Screening in Pediatric Primary Care Clinical Pediatrics, June 1, 2007; 46(5): 441 - 445. [Abstract] [PDF] |
||||
![]() |
S. Santosa, I. Demonty, A. H. Lichtenstein, K. Cianflone, and P. J.H. Jones An Investigation of Hormone and Lipid Associations after Weight Loss in Women J. Am. Coll. Nutr., June 1, 2007; 26(3): 250 - 258. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. K. Welty, K. S. Lee, N. S. Lew, and J.-R. Zhou Effect of Soy Nuts on Blood Pressure and Lipid Levels in Hypertensive, Prehypertensive, and Normotensive Postmenopausal Women Arch Intern Med, May 28, 2007; 167(10): 1060 - 1067. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. B. Ebbeling, M. M. Leidig, H. A. Feldman, M. M. Lovesky, and D. S. Ludwig Effects of a Low-Glycemic Load vs Low-Fat Diet in Obese Young Adults: A Randomized Trial JAMA, May 16, 2007; 297(19): 2092 - 2102. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Pischon, C. J. Girman, F. M. Sacks, N. Rifai, M. J. Stampfer, and E. B. Rimm Non-High-Density Lipoprotein Cholesterol and Apolipoprotein B in the Prediction of Coronary Heart Disease in Men Circulation, November 29, 2005; 112(22): 3375 - 3383. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. Peeters, A. W. van den Beld, H. Attalki, H. v. Toor, Y. B. de Rijke, G. G. J. M. Kuiper, S. W. J. Lamberts, J. A. M. J. L. Janssen, A. G. Uitterlinden, and T. J. Visser A new polymorphism in the type II deiodinase gene is associated with circulating thyroid hormone parameters Am J Physiol Endocrinol Metab, July 1, 2005; 289(1): E75 - E81. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. B Ebbeling, M. M Leidig, K. B Sinclair, L. G Seger-Shippee, H. A Feldman, and D. S Ludwig Effects of an ad libitum low-glycemic load diet on cardiovascular disease risk factors in obese young adults Am. J. Clinical Nutrition, May 1, 2005; 81(5): 976 - 982. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. T. Bairaktari, K. I. Seferiadis, and M. S. Elisaf Evaluation of Methods for the Measurement of Low-Density Lipoprotein Cholesterol Journal of Cardiovascular Pharmacology and Therapeutics, January 1, 2005; 10(1): 45 - 54. [Abstract] [PDF] |
||||
![]() |
I. Shai, E. B. Rimm, S. E. Hankinson, G. Curhan, J. E. Manson, N. Rifai, M. J. Stampfer, and J. Ma Multivariate Assessment of Lipid Parameters as Predictors of Coronary Heart Disease Among Postmenopausal Women: Potential Implications for Clinical Guidelines Circulation, November 2, 2004; 110(18): 2824 - 2830. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Pischon, C. J. Girman, G. S. Hotamisligil, N. Rifai, F. B. Hu, and E. B. Rimm Plasma Adiponectin Levels and Risk of Myocardial Infarction in Men JAMA, April 14, 2004; 291(14): 1730 - 1737. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Riddler, E. Smit, S. R. Cole, R. Li, J. S. Chmiel, A. Dobs, F. Palella, B. Visscher, R. Evans, and L. A. Kingsley Impact of HIV Infection and HAART on Serum Lipids in Men JAMA, June 11, 2003; 289(22): 2978 - 2982. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Usui, H. Kakuuchi, M. Okamoto, Y. Mizukami, and M. Okazaki Differential Reactivity of Two Homogeneous LDL-Cholesterol Methods to LDL and VLDL Subfractions, as Demonstrated by Ultracentrifugation and HPLC Clin. Chem., November 1, 2002; 48(11): 1946 - 1954. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Miller, P. P. Waymack, F. P. Anderson, S. F. Ethridge, and E. C. Jayne Performance of Four Homogeneous Direct Methods for LDL-Cholesterol Clin. Chem., March 1, 2002; 48(3): 489 - 498. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Nauck, G. R. Warnick, and N. Rifai Methods for Measurement of LDL-Cholesterol: A Critical Assessment of Direct Measurement by Homogeneous Assays versus Calculation Clin. Chem., February 1, 2002; 48(2): 236 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Willett, M. Stampfer, N.-F. Chu, D. Spiegelman, M. Holmes, and E. Rimm Assessment of Questionnaire Validity for Measuring Total Fat Intake using Plasma Lipid Levels as Criteria Am. J. Epidemiol., December 15, 2001; 154(12): 1107 - 1112. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. D. Ragland, R. J. Konrad, C. Chaffin, C. A. Robinson, and R. W. Hardy Evaluation of a Homogeneous Direct LDL-Cholesterol Assay in Diabetic Patients: Effect of Glycemic Control Clin. Chem., November 1, 2000; 46(11): 1848 - 1851. [Full Text] [PDF] |
||||
![]() |
H. Fei, S. Maeda, H. Kirii, S. Fujigaki, N. Maekawa, H. Fujii, H. Wada, K. Saito, and M. Seishima Evaluation of Two Different Homogeneous Assays for LDL-Cholesterol in Lipoprotein-X-positive Serum Clin. Chem., September 1, 2000; 46(9): 1351 - 1356. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Esteban-Salan, A. Guimon-Bardesi, J. M. de la Viuda-Unzueta, M. N. Azcarate-Ania, P. Pascual-Usandizaga, and E. Amoroto-Del-Rio Analytical and Clinical Evaluation of Two Homogeneous Assays for LDL-Cholesterol in Hyperlipidemic Patients Clin. Chem., August 1, 2000; 46(8): 1121 - 1131. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Gomez, J. Camps, J. M. Simo, N. Ferre, and J. Joven Agreement Study of Methods Based on the Elimination Principle for the Measurement of LDL- and HDL-Cholesterol Compared with Ultracentrifugation in Patients with Liver Cirrhosis Clin. Chem., August 1, 2000; 46(8): 1188 - 1191. [Full Text] [PDF] |
||||
![]() |
P. Benlian, C. Cansier, G. Hennache, O. Khallouf, P. Bayer, F. Duron, F. Carrat, R. Couderc, O. Chazouilleres, J. Bardet, et al. Comparison of a New Method for the Direct and Simultaneous Assessment of LDL- and HDL-Cholesterol with Ultracentrifugation and Established Methods Clin. Chem., April 1, 2000; 46(4): 493 - 505. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Nauck, M. S. Graziani, D. Bruton, C. Cobbaert, T. G. Cole, F. Lefevre, W. Riesen, P. S. Bachorik, and N. Rifai Analytical and Clinical Performance of a Detergent-based Homogeneous LDL-Cholesterol Assay: A Multicenter Evaluation Clin. Chem., April 1, 2000; 46(4): 506 - 514. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Kondo, Y. Muranaka, I. Ohta, and T. Kanno Dynamic Reaction in a Homogeneous HDL-Cholesterol Assay Visualized by Electron Microscopy Clin. Chem., November 1, 1999; 45(11): 1974 - 1980. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |