|
|
||||||||
Endocrinology and Metabolism |
a Address correspondence to this author at: Washington University School of Medicine, Division of Laboratory Medicine, Box 8118, St. Louis, MO 63110. Fax 314-362-1461; e-mail mscott{at}labmed.wustl.edu.
| Abstract |
|---|
|
|
|---|
alanine substitution at
position 1 of the ß chain. The amino-terminal alanine in this variant
Hb is posttranslationally modified by acetylation, leading to a charge
difference similar to glycation and making the behavior of
HbA1C and Hb Raleigh virtually identical in the
ion-exchange HPLC method. This observation suggests that it is
important to confirm HbA1C values in excess of 15%,
especially if they are not consistent with the clinical picture, by an
independent HbA1C method such as immunoassay or boronic
acid affinity chromatography. However, for this particular variant Hb,
even these latter methods might be misleading, because the acetylated
N-terminal amino acid of the Hb-Raleigh ß chain cannot be glycated. | Introduction |
|---|
|
|
|---|
HPLC-based ion-exchange methods for HbA1C have recently been automated (16), and interferences by Hb species such as HbF and HbS have been minimized. However, several reports have described artificially high HbA1C results with hemoglobin variants such as Hb-Sherwood Forest (9) (Arg 104 Thr of the ß chain) and Hb-South Florida (10) (Val 1 Met of the ß chain) and others (16) when an automated ion-exchange HPLC method is used. We recently observed a HbA1C value of 46% for a sample that was analyzed on the Bio-Rad Variant(TM) automated ion-exchange HPLC method for HbA1C and investigated this, using a series of protein chemistry and molecular biology techniques.
| Materials and Methods |
|---|
|
|
|---|
alkaline and citrate agar gel electrophoresis of Hb
HB electrophoresis was performed using the Ciba Corning Alkaline
and Citrate Systems (Chiron Corporation) according to the
manufacturer's instructions.
Hb and
and ß chain purification
Hb chains were purified by a variation of the procedure published
by Bucci (17). Briefly, the globin fraction from the
patient's erythrocyte lysate was precipitated at -20 °C with 15
volumes of acetone containing 20 mL/L concentrated HCl. The precipitate
was washed four times with acetone and lyophilized. Thirty milligrams
of the globin fraction were dissolved in 5 mL of buffer (20 mmol/L
sodium 2-[N-morpholino]ethanesulfonate, pH 6.6, 8 mol/L
urea, 30 mmol/L 2-mercaptoethanol) and applied to a Mono-S FPLC(TM)
column (Pharmacia) equilibrated with the same buffer. Protein was
eluted with a 0150 mmol/L NaCl linear gradient in the same buffer,
and the absorbance at 280 nm was monitored. Peaks were collected and
dialyzed against 10 mmol/L NH4HCO3, pH 7.8.
isoelectric focusing
Polyacrylamide gel electrophoresis-isoelectric focusing of the
various ß chain preparations was performed using precast IEF 310
MiniPlus SepraGels(TM) according to the manufacturer's instructions
(Integrated Separation Systems).
liquid chromatography-mass spectrometry (lc-ms) and digestion
of ß chain
Purified Hb ß chains were digested with sequencing grade
modified trypsin (Promega, Madison, WI) by adding 5 µL of aqueous 1
mol/L dithiothreitol and 5 µL of a 1 g/L trypsin solution to 100 µL
of a 1 g/L solution of ß chain in 100 mmol/L
NH4HCO3, pH 7.8. The reaction mixture was
incubated for 24 h at 37 °C, after which 7.5 µL of aqueous
100 mL/L trifluoroacetic acid was added to quench the reaction.
Reaction products were stored at 4 °C until use. Preparative HPLC
was performed on Waters hardware, using Millennium 2.10 software
(18). Samples were chromatographed on a Vydac
C18 reversed-phase column (218TP54) purchased from the NEST
Group (Southborough, MA), using a linear trifluoroacetic
acid/acetonitrile gradient as described previously (19).
Amino acid analysis of selected fractions was performed using
6-aminoquinolyl-N-hydroxysuccinimidyl carbamate
derivatization after 6 mol/L HCl hydrolysis for 24 h at 110 °C
as described (20). A Finnigan MAT TSQ-7000(TM) mass
spectrometer (San Jose, CA) with an electrospray ionization source was
used for molecular mass analysis. Samples were either infused at 5
µL/min in a mixture of 500 mL/L methanol, 490 mL/L water, and 10 mL/L
acetic acid or analyzed in an LC-MS mode after microbore chromatography
on a Michrom Bioresources UMA(TM) system on a Reliasil 1 x 150 mm
C18 reversed-phase column with minor modifications
(21)(22)(23). Collision-induced dissociation (CID) experiments
were used to obtain partial sequence information on isolated tryptic
peptides (23). N-terminal Edman sequence analysis was
performed on an ABI Model 477(TM) sequencer (PerkinElmer).
pcr and dna sequencing of [00af]Hbß gene
Genomic DNA was purified from the patient's leukocytes, using
sodium dodecyl sulfate-proteinase K digestion (24). A
1.7-kb DNA fragment containing all three coding exons of the Hb ß
chain gene was amplified by PCR (30 cycles of 30 s at 95 °C,
30 s at 55 °C, 120 s at 72 °C). The primers were as
follows: 5' of the Hb ß gene, TGGGCATAAAAGTCAGGGCA; and 3' of the Hb
ß gene, CAGATGCTCAAGGCCCTTCATA. The entire open reading frame was
sequenced directly from the purified PCR product using the
PerkinElmer Dye Primer Cycle Sequencing system (PerkinElmer) and an
Applied Biosystems 373 DNA sequencer (25). In addition to
the PCR primers, the following oligonucleotides were used in
sequencing: EX23' CCTTCCTATGACATGAA-CTTAACCA; EX35'
ACAACTACAATCCAGCTACCA; EX13' GGTAGA-CCACCAGCAGCCT; and EX25'
GGCATGTGGAGACAGAGAAGACT.
| Results and Discussion |
|---|
|
|
|---|
To rule out HbS or HbF as the cause of the unusually high
HbA1C value from the HPLC method, the patient's blood
sample was analyzed with the Bio-Rad Variant "ß thalassemia short
program," which distinguishes HbF, HbS, and HbA. No HbF or HbS was
identified in the patient's Hb, but a large peak eluting earlier than
A0 was evident (not shown). This peak was not identified by
the ß-thalassemia Short program, but the accompanying package insert
states that peaks in this area are usually HbA1c. These
results clearly demonstrated the presence of a Hb species less
positively charged than HbA0; however, the abundance was
unlikely for HbA1c. Electrophoretic analysis at pH 8.6 in a
cellulose acetate gel failed to show any difference between normal and
patient Hb, whereas electrophoresis at pH 6.2 in a citrate acetate gel
detected an abnormal band with mobility similar to HbF (not shown).
This also suggested the loss of a positively charged group with a
pI < 8.6. Such a positive charge loss would most readily be
ascribed to modification of either an N-terminal amino-terminal group,
a histidine, an arginine, or lysine or to a mutation leading to the
loss of a positively charged residue. For example, a similar pattern of
Hb electrophoresis and falsely positive HbA1c was
previously reported from a patient with the Hb variant, Hb Sherwood
Forest, which has an Arg
Thr mutation in the ß chain
(9).
To identify whether the charge alteration resided in the
or ß
chain, preparative cation-exchange chromatography was performed under
denaturing conditions (6 mol/L urea) to separate the
and ß
chains. This procedure revealed three peaks from the patient sample,
with one ß chain peak (y) and one
chain peak corresponding to the
normal
and ß peaks, but another ß chain peak (x) eluted earlier
than expected (Fig. 1
). The earlier eluting ß chain (x) peak was approximately
equal in size to the normal ß (y) peak, suggesting that the patient
is heterozygotic for a ß chain variant. Isoelectric
focusing-polyacrylamide gel electrophoresis (Fig. 1
, inset) of the two
ß chain peaks confirmed a charge difference, because the earlier
eluting ß (x) peak has a more acidic pI compared with the normal ß
(y) peak.
|
To further isolate the location of the mutation, the molecular masses
of all three fractions were determined by electrospray ionization MS.
The values were 15 126 vs the 15 126.4 expected for the
chain and
15 867 vs the 15 867.2 expected for ß (y), but 15 881 for the ß
(x) peak. Together these results suggested the patient to be
heterozygotic for a ß chain mutation that produces a decreased
positive charge and an increased mass of 14 daltons.
the mutation is the n-terminal valine to alanine
To further identify the site(s) of the mutation, each ß chain
from the patient (peaks x and y) was digested with trypsin, and the
resulting peptide mixture chromatographed on reversed-phase media in
both a preparative and an LC-MS mode. There was little difference in
the preparative chromatogram between the samples except for two peaks,
A and B, that eluted between 36 and 37 min (Fig. 2
). Peak A was much greater for the variant ß (x) chain (Fig. 2
), whereas the opposite was true for the normal ß (y) chain digest
(not shown). LC-MS yielded molecular mass values of 952 Da for tryptic
digest T1 (peak B) and 966 Da for tryptic digest T1 (peak A). Peak B
thus represents the T1 peptide of the wild-type ß (y) chain (residues
18) and peak A is the T1 peptide of the mutant ß (x) chain, which
is 14 Da larger. On the basis of the tryptic peptides, over 94.5% of
the sequence of normal ß globin was found in the map of the ß (x)
chain, virtually eliminating the possibility that other sites in the
protein contain a mutation. Thus, the site of mutation in the
patient's Hb ß chain was localized to the N-terminal eight residues.
|
CID spectra for the normal T1 peptide with m/z ion of 952.4
and the corresponding mutant TI peptide with m/z ion of
966.4 allows immediate assignment of the mutation site to one of the
first two residues (Fig. 3
). The b2 fragment ion [nomenclature from
(26)] for the normal peptide (upper panel) gives the
expected value of 237 Da, whereas the mutant spectra (lower panel)
provides a value of 251 Da, (i.e., 14 Da), for the same N-terminal
fragmentation.
|
Because it was not immediately obvious how a 14-Da change within the
first two amino acids could lead to a loss of a positive charge, we
characterized the mutation at the gene level. The DNA sequence of the
patient's Hb ß gene revealed a single base heterozygous mutation
(T
C) at base 2 of the codon for the first amino acid, producing a
Val
Ala substitution. Amino acid analysis for each T1 peptide
(containing the first eight amino acids of ß chains) confirmed that
the substitution of the first amino acid valine by alanine in the
mutated ß (x) chain was the only difference between the ß (x) and
ß (y) chains. Nevertheless, the substitution of valine by alanine
does not result in either loss of charge or the 14-Da difference
observed between the ß (x) and ß (y) chains. Indeed, the mass
difference for a Val
Ala change is -28 Da.
Interestingly, standard Edman N-terminal sequence analysis of the mutant ß (x) chain gave no sequence information suggesting that the N terminus of the mutant ß (x) chain may be blocked because of a posttranslational modification. In fact, Hb-Raleigh has characteristics similar to those of this patient's Hb. In Hb-Raleigh, the ß chain amino-terminal valine is replaced by acetylalanine (27). Such a mutation is consistent with both the loss of positive charge of ß chain and the 14-Da difference we observed here. Furthermore, the acetylated N-terminal residue would explain why this protein was refractory to Edman degradation. Thus, the combination of protein biochemical studies and DNA sequence analysis determined that this patient is heterozygous for Hb-Raleigh and that HB-Raleigh produces a falsely increased HbA1c in this automated HPLC method.
Ion-exchange HPLC is a widely used methodology (1) in the measurement of normal and abnormal Hb components, and the difference in charge between HbA and HbA1C as a result of glycation of the N-terminal amino acid of Hb ß chain is the basis of the Bio-Rad HPLC HbA1C method. However, this cation-exchange column-based method, as well as other similar methods, can clearly misidentify as HbA1C some Hb variants that have a loss of positive charge (7)(8)(9)(10)(11)(12)(16), as well as carbamylated Hbs (4)(5)(6).
In most cases, the markedly high HbA1C will cause the laboratory to take note and investigate. HPLC and electrophoretic methods are available to clearly identify or rule out HbF or HbS. For variant Hb with decreased charge or carbamylated Hb, alternative glycohemoglobin methods, such as immunoassays that utilize antibodies specific to the glycated N-terminal values of the ß chain, can provide accurate values (13). Similarly, boronate affinity methods (5) can overcome the falsely increased values in ion-exchange methods caused by variant or carbamylated Hbs. Furthermore, although immunoassays, boronate affinity, and ion-exchange methods detect different modifications of the Hb molecule, extensive work towards standardization (15)(28) has made results reasonably comparable with the "gold-standard" ion-exchange method used in the DCCT trials. Thus, when abnormally high values are identified, it is almost always possible to provide useful clinical information by alternative methods.
The rare Hb variant Hb-Raleigh may, however, be an exception
because of the ß chain N-terminal modification. Because this mutation
produces an acetylated alanine at the N terminus, Hb-Raleigh ß chain
cannot be glycosylated at the N terminus and thus will not react with
the immunoassay methods. Furthermore, it would exhibit decreased
reactivity with boronate affinity methods. Therefore, it may not be
possible to utilize Hb glycosylation as a means to monitor glucose
control in patients with Hb-Raleigh. For such individuals,
determination of the glycosylation of other proteins, such as albumin
(28)(29), may be necessary. To our knowledge,
this conundrum is restricted only to Hb-Raleigh, Hb-Long Island, and
HbA2-Niigata, which has a
chain N-terminal acetylated
alanine (30), and which was recently described as having an
interference in HbA1C methods similar to the interference
described here (31).
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
Gln]2; a new hemoglobin variant mistaken for glycosylated hemoglobin. Biochim Biophys Acta 1988;955:214-219.
[Medline]
[Order article via Infotrieve]
1(NA1)VAL-ALA]: a new
chain variant found in the Japanese population. Hemoglobin 1991;15:335-339.
[Web of Science][Medline]
[Order article via Infotrieve]
The following articles in journals at HighWire Press have cited this article:
![]() |
B. I. Dalal and M. L. Brigden Factitious Biochemical Measurements Resulting From Hematologic Conditions Am J Clin Pathol, February 1, 2009; 131(2): 195 - 204. [Abstract] [Full Text] [PDF] |
||||
![]() |
L.-Y. Tsai, S.-M. Tsai, M.-N. Lin, and S.-F. Liu Effect of Hemoglobin Variants (Hb J, Hb G, and Hb E) on HbA1c Values as Measured by Cation-Exchange HPLC (Diamat) Clin. Chem., April 1, 2001; 47(4): 756 - 758. [Full Text] [PDF] |
||||
![]() |
L. Bry, P. C. Chen, and D. B. Sacks Effects of Hemoglobin Variants and Chemically Modified Derivatives on Assays for Glycohemoglobin Clin. Chem., February 1, 2001; 47(2): 153 - 163. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. M. Clarke and T. N. Higgins Laboratory Investigation of Hemoglobinopathies and Thalassemias: Review and Update Clin. Chem., August 1, 2000; 46(8): 1284 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Bisse, N. Zorn, A. Eigel, M. Lizama, P. Huaman-Guillen, W. Marz, A. Van Dorsselaer, and H. Wieland Hemoglobin Rambam (ß69[E13]Gly->Asp), a pitfall in the assessment of diabetic control: characterization by electrospray mass spectrometry and HPLC Clin. Chem., October 1, 1998; 44(10): 2172 - 2177. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |