(Clinical Chemistry. 1998;44:1302-1308.)
© 1998 American Association for Clinical Chemistry, Inc.
|
Endocrinology and Metabolism |
Non-separation assay for glycohemoglobin
Stuart Blinckoa,
and Raymond Edwards
a Author for correspondence. Fax 44-171-606-5937.
 |
Abstract
|
|---|
The determination of glycohemoglobin [HbA1c,
HbA1, or total glycohemoglobin (GHb)] has become an
established procedure in the management of diabetes mellitus. Here, we
describe the development of a simple, fluorescence, non-separation
assay for the percentage of GHb (%GHb). The fluorescence of an
eosin-boronic acid derivative when it was mixed with hemolysates of
unwashed erythrocytes was quenched in proportion to the percentage of
glycohemoglobin. Measurement of the fluorescence intensity gave an
estimate of GHb in the sample, and measurement of light absorbance gave
an estimate of total hemoglobin. A combination of the two measurements
gave the assay response. Comparison with HPLC (Menarini-Arkray HA-8140
fully automated analyzer) for the percentage of HbA1
(%HbA1) gave %GHb(NETRIA) = 1.1(SD
±0.03)%HbA1 +0.6(SD ±0.3), Sy
x
= 0.821, r = 0.972, n = 80; comparison for
HbA1c gave %GHb(NETRIA) = 1.3(SD
±0.04)%HbA1c + 1.8(SD ±0.3),
Sy
x = 0.813, r = 0.973,
n = 80. Precision, estimated as the percentage of the CV of the
%GHb assay results, was <2% (intraassay, range 522% GHb) and
<4.2% (interassay, range 416% GHb). Dilution of a high-percentage
GHb sample lysate showed that the assay was linear, and addition of
glucose (60 mmol/L), bilirubin (250 µmol/L), and triglycerides (14
mmol/L) to low, medium, and high %GHb samples showed no clinical
interference in assay results.
 |
Introduction
|
|---|
Glycohemoglobin has become the established method for
assessing long-term glycemic control for diabetic patients
(1). A percentage of hemoglobin becomes glycated during the
lifetime of erythrocytes (120 days), reflecting the integrated mean
blood concentration of glucose over the preceding 812 weeks
(2)(3).
There is a wide range of methods for the determination of
glycohemoglobin [HbA1c, HbA1, or total
glycohemoglobin (GHb)], and they have been reviewed recently
(1)(4). Glycohemoglobin is made up of a number
of chemically distinct species, which are produced by the binding of
glucose and other carbohydrates to various sites on hemoglobin.
HbA1c is glucose bound via Amadori rearrangement to the
N-terminal valine of the ß chain. HbA1 is composed of all
the carbohydrate species, including glucose, derived from adducts bound
to the N-terminal valine. GHb is the total of all carbohydrate
derivatives bound anywhere on the hemoglobin molecule. To date, no
clinical advantage has been shown for measuring HbA1c as
opposed to HbA1 or GHb (1).
Here, we present the development of a non-separation method for the
determination of GHb. The principle on which the method is based is
that the emissions of certain fluorescent derivatives of boronic acids
are quenched when bound to glycohemoglobin. In previous publications,
the quenching of a fluorescein-boronic acid derivative was described
(5)(6). The fluorophore showed fluorescence
quenching when mixed with purified glycohemoglobin but not with
non-glycohemoglobin at similar concentrations. Addition of the
fluorophore to erythrocyte lysates gave quenching values that
correlated well with the electroendosmosis method for HbA1
(Corning Glytrac; r = 0.91, n = 25), and gave
valid results with three heterozygous HbA/S samples. It was considered
that the fluorescence quenching method could be improved by the
selection of a fluorophore whose emission gave a high degree of
spectral overlap with an absorption band of hemoglobin. This overlap
would, it was predicted, lead to enhanced quenching of the fluorophore
when bound to glycohemoglobin. The emission spectra of an eosin-boronic
acid derivative gave the required spectral characteristics, with
emission coincident with the 540 nm absorption band from hemoglobin
(Fig. 1
). We describe results using the eosin-boronic acid derivative
and the development of an assay for GHb.

View larger version (16K):
[in this window]
[in a new window]
|
Figure 1. Absorption spectrum of hemoglobin and emission spectra of
a fluorescein-boronic acid (FBA)
(5)(6) and the eosin-boronic acid derivative
(EBA).
Fluorescence intensity scale is in arbitrary units.
|
|
 |
Materials and Methods
|
|---|
eosin-boronic acid derivative
An eosin-boronic acid derivative of eosin-5-isothiocyanate and
3-aminophenyl boronic acid was prepared by the method described
previously (6). An aqueous solution was made and
stored at 4 °C. Assays were performed over an 11-month period from
one batch of reagent, which showed no signs of degradation. Other
batches have also been prepared and gave close agreement in tests.
assay buffer
The buffer was composed of 100 mmol/L glycine (AnalaR; BDH) and 50
mmol/L magnesium chloride hexahydrate (AnalaR; BDH), pH 8.5 (pH
adjusted to 8.5 by addition of saturated sodium hydroxide).
samples
Aliquots were taken from clinical samples (EDTA whole blood) taken
for routine clinical analysis. All measurements were performed within 7
days of the sample being taken. None of the samples gave evidence of
abnormal hemoglobins by the HA-8140 HPLC method.
calibrants
Lyophilized red blood cell lysates (Menarini) were reconstituted
in distilled water to give an absorbance (405 nm) within the correction
range of the assay (Fig. 4
).

View larger version (8K):
[in this window]
[in a new window]
|
Figure 4. Relationship between absorbance (405 nm) and correction
applied to fluorescence-quenching results ( QI).
|
|
fluorometer and spectrophotometers
The fluorometer was a PerkinElmer LS50B adjusted to the
following settings: excitation, 523 nm; emission, 542 nm; slits, 5 nm/5
nm; and integration time, 5 s. All intensity values were on an
arbitrary scale. The spectrophotometers were a Cecil Instruments 5233
and Anthos HT II, equipped with a 405 nm filter.
hplc method
The HPLC was an Menarini-Arkray HA-8140 fully automated analyzer
(Arkray, formerly KDK) (7). Results were given as
percentages of HbA1c, HbA1, and HbF.
assay development
Concentration of reagents.
The concentrations of the
eosin-boronic acid derivative and red cell lysates were adjusted to
give optimum discrimination between low and high percentage GHb (%GHb)
samples.
Time of reaction.
A sample lysate-fluorophore mixture was
prepared according to the assay method in a square 1-cm glass cuvette
(four clear sides) sited in the fluorometer. On addition of the
eosin-boronic acid solution to the lysate, continuous measurement of
the fluorescence intensity was begun and was monitored for 800 s.
A graph of fluorescence intensity against time was plotted (Fig. 2
).

View larger version (9K):
[in this window]
[in a new window]
|
Figure 2. Time course for reaction between sample lysate and
eosin-boronic acid.
Fluorescence intensity plotted vs time.
|
|
Fluorescence quenching results.
The fluorescence intensity of
the eosin-boronic acid solution with and without sample lysates allowed
calculation of the percentage of quenching to be made. Subtraction of
the background gave the quenching index (QI). Comparison of
fluorescence QI results with the percentage of HbA1
(%HbA1) showed some agreement (r = 0.81,
n = 34, Fig. 3
). The samples showing positive deviation from the
%HbA1 result were all found to have higher absorbances
than those that showed negative deviation.

View larger version (11K):
[in this window]
[in a new window]
|
Figure 3. Quenching results (QI), before correction for hemoglobin
concentration, plotted vs %HbA1 (HA-8140).
r = 0.81, n = 34; QI(NETRIA) = 0.99(%HbA1)
+ 0.19.
|
|
Correction of quenching results estimated by absorbance.
The
relationship between absorbance (405 nm) and quenching correction
(
QI) was obtained from experiments where the difference between QI
results and %HbA1 results were plotted vs absorbance.
Tests were carried out for individual samples at different
concentrations and for different samples at different concentrations.
Investigations of the relationship between absorbance (405 nm) and
QI for samples at different concentrations (absorbances 0.370.52)
gave a correlation coefficient of 0.98 (n = 20) and a slope of
55.7 (linear regression). Samples at different concentrations were
found to lie close to this slope; therefore, the slope of the
regression relationship, shown in Fig. 4
, was adopted with minor modification to the intercept for use
in assays. Application of
QI to the quenching data gave the assay
response as the corrected quenching index (QIcor, Fig. 5
).

View larger version (11K):
[in this window]
[in a new window]
|
Figure 5. Quenching results corrected for hemoglobin concentration
to give an assay response, QIcor, plotted vs
%HbA1 (HA-8140).
r = 0.97, n = 34; QIcor(NETRIA) =
0.99(%HbA1) + 1.6.
|
|
Calibration.
Assay responses (QIcor) gave a
significant correlation with %HbA1 (r =
0.972). A Bland-Altman plot (Fig. 5
) (8) showed that a
linear method of calibration would be appropriate. The lyophilized red
cell lysates described above allowed linear calibration of assays with
%GHb values assigned (5.0% and 14.5%). These values were assigned to
the lyophilized pools to give a range of %GHb values that gave good
agreement with the HPLC method.
Assay method.
A dilute solution of the eosin-boronic acid
derivative was prepared from stock in the assay buffer containing
detergent. This was prepared immediately before use in each assay.
One milliliter of whole blood (EDTA) was pipetted into a 5-mL test tube
and centrifuged at 173g or allowed to settle overnight.
Fifty microliters of the packed erythrocyte layer were added to 400
µL of lysing buffer (detergent solution in assay buffer), gently
mixed by hand until homogeneous, and allowed to stand for a minimum of
10 min.
Twenty microliters of lysate or reconstituted calibrant were pipetted
into glass vials (in duplicate) and 4 mL of the eosin-boronic acid
solution was added by repeater pipette. In addition to the sample and
calibrant vials, 4 mL of the eosin-boronic acid solution was added to
20 µL of lysing buffer in separate vials to estimate the total
fluorescence signal. The sample-fluorophore mixtures were allowed to
stand for 20 min out of direct light, and the fluorescence intensity
was read (
ex523 nm/
em542 nm, details
above). The absorbance of each sample was determined by pipetting 200
µL of each sample-fluorophore mixture into microtiter wells and
measuring the absorbance at 405 nm on a spectrophotometer (see below).
Assay method summary.
Following is a summary of the assay:
(a) One milliliter of EDTA whole-blood sample was
centrifuged (173g) or allowed to settle overnight.
(b) Fifty microliters of packed red cells were added to 400
µL of lysing buffer, mixed gently, and allowed to sit for 10 min.
(c) Twenty microliters of lysate or reconstituted calibrant
were mixed with 4000 µL eosin-boronic acid solution and left in the
dark for 20 min.
(d) Fluorescence intensity (
ex523
nm/
em542 nm) and absorbance (405 nm) were read.
calculation of results
Data from assay.
The data from the assay were calculated using
the following variables: T, fluorescence intensity of the
total solution (eosin-boronic acid solution with no sample or
calibrant); F, fluorescence intensity of the eosin-boronic
acid solution with calibrants or samples; and absorbance 405
nm, the absorbance of a sample at 405 nm, measured with a 5-mm
pathlength.
The percentage of quenching was calculated as [(T
- F)/T] x 100%. Subtraction of background
gives the QI. The results were corrected for hemoglobin concentration
(using the absorbance 405 nm) to give an assay response.
Values for %GHb were obtained for samples by linear interpolation from
calibrant responses.
Intraassay precision.
All sample lysates were measured in
duplicate, and a precision profile was plotted (CV values vs mean
result; Fig. 9
). The curve fit (quadratic) was plotted.

View larger version (11K):
[in this window]
[in a new window]
|
Figure 9. Intraassay precision profile.
CV duplicate results (%) plotted vs mean %GHb result, n = 80.
Quadratic curve plotted (solid line).
|
|
Interassay precision.
Four assays were performed using
separate assay reagents on frozen lysates from patient samples (n
= 6) stored at -40 °C, and a precision profile was plotted (CV
values vs mean result; Fig. 10
). The curve fit (quadratic) was plotted.

View larger version (10K):
[in this window]
[in a new window]
|
Figure 10. Interassay precision profile for frozen lysates (n =
6, in four assays).
CV results (%) from four assays plotted vs mean %GHb result.
Quadratic curve plotted (solid line).
|
|
Linearity.
Lysate from a high-percentage GHb sample (20.5%
GHb) was diluted in lysate from a low-percentage GHb sample (8.7%
GHb). The lysate mixtures (four dilutions and the high and low lysates)
were measured according to the assay protocol. The assay results for
each dilution (duplicates) were plotted vs the calculated %GHb for the
dilutions (Fig. 11
).

View larger version (9K):
[in this window]
[in a new window]
|
Figure 11. Dilution of lysate from a high glycohemoglobin sample
(20.5%GHb) in lysate from a low glycohemoglobin sample (8.7%GHb).
Assay results (%GHb found) for each dilution (x 4) and
high and low GHb sample plotted vs %GHb calculated. Dotted
line is the line of equivalence.
|
|
effect of glucose
Three whole-blood samples (low, medium, and high %GHb) were each
divided into two aliquots. To one aliquot, 9 g/L NaCl was added, and to
the other a high concentration of D-()-glucose
(Sigma) was added to give a final concentration of ~60 mmol/L.
Glucose was measured for both aliquots of blood by the hexokinase
method (on an automated analyzer, ILab 900, Instrumentation
Laboratories). The two aliquots were also assayed for %GHb, according
to the protocol. Results from the samples with added glucose were
compared with those for the sample without added glucose and expressed
as a percentage (Table 1
).
effect of bilirubin
The method for bilirubin was similar to the method for glucose,
with the final concentration of added bilirubin (mixed isomers, Sigma)
measured by the serum total bilirubin method (modified Jendrassik-Grof
assay on automated analyzer ILab 900). The results are shown in Table 2
.
effect of triglycerides
The method for triglycerides was similar to the method for
glucose, with the final concentration of triglycerides (Intralipid,
Pharmacia) measured by the serum triglyceride method (automated assay
on an automated analyzer ILab 900). The results are shown in Table 3
.
correlations and regressions
The correlations and regressions were calculated according to the
method of Pearson.
 |
Results
|
|---|
Optimization of the concentrations of sample and the eosin-boronic
acid derivative gave a quenching-percentage range (after subtraction of
background) between 5% and 20% for samples with similar
%HbA1 values.
Fluorescence intensity with time measurements of a sample lysate and
eosin-boronic acid mixture showed the reaction reached steady-state
after 7 min (Fig. 2
).
Data from the assay with 34 samples were used to show the fluorescence
quenching results before and after correction for hemoglobin
concentration (absorbance 405 nm). The uncorrected quenching results
for 34 samples (Fig. 3
) showed some agreement with the
%HbA1 results (r = 0.81). The samples
showing positive deviation from the %HbA1 result were all
found to have higher absorbances than those that showed negative
deviation. Application of the correction calculated from absorbance
(
QI, Fig. 4
; see the section concerning assay development in
Materials and Methods) to the quenching data gave the assay
response (Fig. 5
). These results gave a significant correlation with
%HbA1 (r = 0.97). A Bland-Altman plot
(8) (Fig. 6
) showed that a linear method of calibration would be
appropriate. Lyophilized pools of red blood cell lysates with high and
low %GHb were found have stable absorbances (405 nm) in the assay and
were used as calibrants. %GHb values of 5.0% and 14.5% were assigned
to the pools (see description of assay development in Materials
and Methods). Sample results calibrated as %GHb gave a slope of
1.1 (intercept 0.6) compared with %HbA1 results.

View larger version (10K):
[in this window]
[in a new window]
|
Figure 6. Bland-Altman plot (8), (QIcor -
%HbA1) plotted vs the mean of the two results.
Mean difference (± SD) = 1.52 (± 0.70), n = 34.
|
|
The performance of the developed method was evaluated by comparison
with the Menarini-Arkray (HA-8140) HPLC (7).
A total of 80 samples were measured in duplicate: 34 in one analytical
run and 46 in another. The values for %GHb were plotted vs
%HbA1 (Fig. 7
) and %HbA1c (Fig. 8
). Regression and correlation results were as follows:
%GHb(NETRIA) = 1.1(SD ±0.03) %HbA1 0.6(SD ±0.3),
Sy
x = 0.821, r = 0.972, n =
80 (34 46); %GHb(NETRIA) = 1.3(SD ±0.04) %HbA1c
1.8(SD ±0.3), Sy
x = 0.813, r =
0.973, n = 80 (34 46).

View larger version (13K):
[in this window]
[in a new window]
|
Figure 7. %GHb results from two assays plotted vs
%HbA1 (HA-8140).
%GHb(NETRIA) = 1.1(SD ± 0.03)%HbA1 +
0.6(SD ± 0.3), Sy x = 0.821,
r = 0.972, n = 80 (34 + 46).
|
|

View larger version (13K):
[in this window]
[in a new window]
|
Figure 8. %GHb results from two assays plotted vs
%HbA1c (HA-8140).
%GHb(NETRIA) = 1.3(SD ± 0.04)%HbA1c +
1.8(SD ± 0.3). Sy x = 0.813,
r = 0.973, n = 80 (34 + 46).
|
|
The intraassay precision (CV, duplicate results plotted vs mean %GHb
result, n = 80) for sample values of 522% GHb was <2% (Fig. 9
). The interassay precision for six samples assayed four times
was <4.2% for sample values of 416% GHb (Fig. 10
).
The assay showed linearity of response with a high %GHb sample diluted
in a low %GHb sample (Fig. 11
).
Glucose up to a concentration of ~60 mmol/L was found to have no
clinical interference on the assay for low, medium, and high %GHb
samples (Table 1
). Bilirubin up to a concentration of 250 µmol/L
showed no clinical interference on the assay for low, medium, and high
%GHb samples (Table 2
). Triglycerides up to a concentration of 14
mmol/L showed no clinical interference for low, medium, and high %GHb
samples (Table 3
).
 |
Discussion
|
|---|
The results from measuring the quenching of the eosin-boronic acid
derivative by hemolysates and correcting for the hemoglobin
concentration of the samples demonstrate a novel assay for
glycohemoglobin.
Hemoglobin is known to quench the fluorescence of associated and
covalently bound fluorophores (9)(10)(11)(12). The eosin-boronic
acid derivative has a fluorescence emission spectrum that peaks at 541
nm. This peak is coincident with one of the absorption peaks of
hemoglobin (540 nm, Fig. 1
). It may be postulated that the covalently
bound, excited eosin-boronic acid loses energy by resonance transfer to
glycohemoglobin in a manner similar to other fluorophores bound to the
ß93 Cys residue on hemoglobin (10). The resonance transfer
mechanism only occurs for donor-acceptor pairs in close proximity
(<1012 nm) with a high degree of spectral overlap (13).
Hemoglobin is 6.4 x 5.5 x 5.0 nm in size; therefore,
covalently bound eosin will be in close proximity to one of the four
heme units (14). This quenching process forms the specific
component of the quenching observed and depends on the amount of
glycohemoglobin in the sample covalently bound to the eosin derivative.
In addition to this quenching, the presence of hemoglobin in solution
with an eosin derivative (not bound) quenches the signal by "inner
filter effect" (i.e., the emitted light from the eosin derivative is
reabsorbed by hemoglobin) (10). This component of the
quenching depends on the absorbance of the sample. Therefore, it can be
estimated (Fig. 4
) and used to correct the quenching results (Fig. 3
, uncorrected; Fig. 5
, corrected for hemoglobin concentration). The value
of the corrected results (QIcor) was the assay response,
and linear interpolation from calibrant responses gave %GHb results.
In contrast to this non-separation method, fluorescence quenching has
been used to estimate glycohemoglobin in a separation assay developed
for the Abbott IMx(TM) automated analyzer
(9). The method separates glycated and nonglycated
hemoglobin and estimates the concentrations by the fluorescence
quenching of 4-methylumbelliferone. The interaction between the
fluorophore and hemoglobin is thought to be a noncovalent association.
The use of calibrants was considered necessary to ensure maximum
consistency between assays. The validity of linear calibration for this
method is demonstrated by the Bland-Altman plot (Fig. 6
). Calibration
for %GHb is the most appropriate for this method because boronic acids
bind all cis-diol-containing adducts bound to hemoglobin. The values
assigned to the calibrants gave an appropriate range of values for
%GHb. Final calibration of this method (as for others) is not yet
possible until an international standard becomes available. The method
could, if desired, be calibrated with %HbA1c or
%HbA1. The assay gave a linear response with dilutions of
a high-%GHb sample lysate, demonstrating the the validity of the
calibration and method of calculating assay response (Fig. 11
). The
close correlations with the Menarini-Arkray HA-8140 HPLC method
demonstrate that this method can accurately determine glycohemoglobin
concentrations for diabetic management (Figs. 7
and 8
).
The precision results of <2% (intraassay, Fig. 9
) and 4.2%
(interassay, Fig. 10
) are well below the specified target of 5%
(intra- and interassay) for glycohemoglobin assays (1).
These precision results are an improvement on the manual boronic acid
column method (within-batch CV, 35%; between-day CV, 38%)
(15).
Glucose up to a concentration of 60 mmol/L showed no clinical
interference in the assay. This finding is similar to those with other
methods using boronic acids (9)(16). Bilirubin,
a fluorescent compound, may be expected to interfere in a fluorescence
non-separation method. However, bilirubin gives negligible fluorescence
emission at 542 nm when excited at 523 nm (the wavelengths used in the
assay), and results show that it does not clinically interfere with the
assay. Triglyceride concentrations up to 14 mmol/L did not clinically
interfere with the assay.
Additional studies are required to assess whether abnormal hemoglobin
samples are measured equipotently with the HbA samples and whether
anemia affects the clinical interpretation of results from this assay.
The majority of methods for measuring glycohemoglobin involve a
separation step, which increases the technical complexity
(9)(16)(17)(18)(19)(20)(21)(22)(23)(24). This non-separation method appears
to have the advantage of simplicity without compromising precision and
accuracy. Furthermore, this non-separation methodology demonstrates a
lack of interference from glucose, bilirubin, and triglycerides. Future
developments include the manufacture of a purpose-built combined
fluorescence and absorption spectrophotometer and the substitution of
packed red cell lysates with whole-blood lysates. The method has the
potential to be inexpensive and capable of high sample throughput.
 |
Acknowledgments
|
|---|
We thank Graham Hitman, Garry John, and Alan Williams at
the Royal London Hospital, Whitechapel, London, for help and advice.
Aspects of the method and principles described in this paper are the
subject of an international patent application (6).
 |
Footnotes
|
|---|
NETRIA, St Bartholomew's Hospital, The Royal Hospitals NHS Trust, London EC1A 7BE, UK.
 |
References
|
|---|
-
John WG. Glycated haemoglobin analysis. Ann Clin Biochem 1997;34:17-31.
-
Koenig RJ, Peterson CM, Jones RL, Saudek C, Lehrman M, Cerami A. Correlation of glucose regulation and hemoglobin A1c in diabetes mellitus. N Engl J Med 1976;295:417-420.
[Abstract]
-
Bunn JF, Gabbay KJ, Gallop MP. The glycosylation of haemoglobin: relevance to diabetes mellitus. Science 1978;200:21-27.
[Abstract/Free Full Text]
-
Standing SJ, Taylor RP. Glycated haemoglobin: an assessment of high capacity liquid chromatographic and immunoassay methods. Ann Clin Biochem 1992;29:494-505.
-
Blincko S, Edwards R. Quenching of fluorescence by glycated haemoglobin: early development of a novel assay. Ann Clin Biochem 1998;35:140-142.
-
Edwards R, Blincko S, applicants. Glycated proteins
assay. Int Pat Appl PCT/GB95/01766, US Pat Appl 08/765,461..
-
John WG, Braconnier F, Miedema K, Anlesa C, Piras G. Evaluation of the Menarini-Arkray HA-8140 hemoglobin A1c analyser. Clin Chem 1997;43:968-975.
[Abstract/Free Full Text]
-
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;i:307-310.
-
Wilson DH, Bogacz JP, Forsythe CM, Turk PJ, Lane TL, Gates RC, Brandt DR. Fully automated of assay of glycohemoglobin with the Abbott IMx(TM) analyser. Novel approaches for separation and detection. Clin Chem 1993;39:2090-2097.
[Abstract]
-
Hirsch RE. Front face fluorescence spectroscopy of haemoglobins. Methods Enzymol 1994;232:231-246.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Kuwajima T, Asai H. Synthesis of organic phosphates and their equilibrium binding to bovine oxyhemoglobin. Biochemistry 1975;14:492-497.
[Medline]
[Order article via Infotrieve]
-
Horiuchi K, Asai H. A reaction cell for simultaneous measurements of fluorescence and absorption in a hemoglobin solution at various oxygen pressures. Biochem Biophys Res Commun 1980;97:811-818.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Campbell AK. Chemiluminescence. Principles and
applications in biology and medicine. Chichester: Ellis Horwood,
1988:499510..
-
Perutz MF, Rossmann MG, Cullis AF, Muirhead H, Will G, North ACT. Structure of haemoglobin. Nature 1960;158:416-422.
-
John WG, Bullock DG, MacKenzie F. Methods for the analysis of glycated haemoglobins: What is being measured?. Diabetic Med 1992;9:15-19.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Klenk DC, Hermanson GT, Krohn RI, Fujimoto EK, Mallia AK, Smith PK, et al. Determination of glycosylated hemoglobin by affinity chromatography: comparison with colorimetric and ion-exchange methods, and effects of common interferences. Clin Chem 1982;28:2088-2094.
[Abstract/Free Full Text]
-
Trivelli LA, Ranney HM, Lai HT. Hemoglobin components in patients with diabetes mellitus. N Engl J Med 1971;284:353-357.
-
Baumgart R, France MW, UaConaill D, Murnagham S, Perrira M, Drurg MI. An assessment of a microcolumn chromatographic method for the measurement of haemoglobin A1. Ann Clin Biochem 1980;17:101-104.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Gruber CA, Koets MD. Quantitation of hemoglobin A1a+b and hemoglobin A1c by automated high pressure liquid chromatography. Clin Chem 1979;25:1970-1971.
[Abstract/Free Full Text]
-
Menard C, Dempsey ME, Blankstein LA, Aleyasine H, Wacks M, Soeldner JS. Quantitative determination of glycosylated hemoglobin A1 by agar gel electrophoresis. Clin Chem 1980;26:1598-1602.
[Free Full Text]
-
John WG, Gray MR, Bates DL, Beacham J. Enzyme immunoassay: a new method for the estimation of HbA1c. Clin Chem 1993;39:663-666.
[Abstract/Free Full Text]
-
Mallia AK, Hermanson GT, Krohn RI, Fujimoto EK, Smith PK. Preparation and use of a boronic acid affinity support for the separation and quantitation of glycosylated hemoglobins. Anal Lett 1981;14:649-661.
-
Noffsinger J. Rapid glycohemoglobin measurements. Clin Chem 1993;39:1556.[Web of Science][Medline]
[Order article via Infotrieve]
-
Doelman CJA, Siebelder CWM, Nijof WA, Weykamp CW, Janssens J, Penders T. Capillary electrophoresis system for glycated haemoglobin evaluated. Clin Chem 1997;43:644-648.
[Abstract/Free Full Text]