|
|
||||||||
Automation and Analytical Techniques |
1
Department of Clinical Biochemistry, St. Bartholomew's and the Royal London School of Medicine & Dentistry, Turner Street, London E1 2AD, UK.
2
Dade International, Glasgow Site, Wilmington, Inc.,
Newark, DE 19898.
a Author for correspondence. Fax (44)71-377-1544; e-mail c.p.price{at}mds.qmw.ac.uk.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
At present there is little general agreement on a single test or panel
of laboratory tests that may serve as a definitive means of monitoring
nutritional status (1)(2)(3)(4)(5)(6)(7)(8)(9)(10)(11)(12)(13)(14)(15)(16)(17)(18)(19). The proposed criteria of an
"ideal" serum marker of nutritional status (8) are, in
summary, a short biological half-life and a specific and fast response
to protein/energy status. Circulating proteins and their half-lives
that reflect protein/energy homeostasis
(5)(7)(12)(20) include
albumin (
20 days), transferrin (
8.5 days), prealbumin (
2
days), fibronectin (
424 h), and retinol-binding protein (
812
h). Within this context, prealbumin is the candidate whose
characteristics best match the requirements because the intermediate
half-life yields the best diagnostic sensitivity and specificity
(2)(4)(9)(14).
Prealbumin is especially sensitive to early phases of decreased
nutrition, particularly caloric intake
(1)(2)(7) and nitrogen balance; in
addition, the concentration of prealbumin rapidly returns to values
within the reference interval once the malnutrition has been corrected
(1)(7)(16). For these reasons, the
measurement of prealbumin has been shown to be clinically useful in
monitoring nutritional status (4)(9)(10)(11) and/or
nutritional support (12)(13), including total
parenteral nutrition (14)(15). Because
prealbumin is also a negative acute phase reactant, the potential
contribution of metabolic stress can be determined by the measurement
of a positive acute phase reactant, such as C-reactive protein
(5)(17)(18). C-reactive protein has
a similar half-life to prealbumin and can be measured using the same
technology (19). The physiological role of prealbumin is
actually that of a transport protein of thyroid hormones and the
vitamin A/retinol-binding protein complex to target tissue, whereas in
the circulation prealbumin exists as a stable tetramer bound to
retinol-binding protein (21).
The broad consensus for an adult reference range for prealbumin is
120500 mg/L, with considerable variation within the quoted ranges
depending on the study, the numbers tested, and the methods used
(19)(20)(21)(22)(23)(24)(25)(26)(27). A small difference between the sexes exists, with
the male range being marginally lower than the female range. The range
in infants is
40150 mg/L (24)(25); however,
during childhood and puberty the concentrations for both sexes rise
gradually, and at different stages of development, there appear to be
some marginal differences between sexes
(26)(27). Various studies
(4)(13) have suggested that a concentration of
<110 mg/L is considered high risk, requiring major nutritional
therapy; 110170 mg/L is moderate risk with less intensive nutritional
therapy requirement; and >170 mg/L is of little or no risk. Reference
range values of prealbumin, therefore, represent the high range of any
assay, whereas a decrease from these concentrations will be of clinical
use. Methods available for the measurement of prealbumin are mainly
based on immunoprecipitation and include electroimmunoassay
(19), immunodiffusion (23), and light-scattering
free solution assays (28). The two former techniques are
time-consuming, and the conventional nonenhanced light-scattering
assays operate at the limit of sensitivity of the assay, requiring very
careful choice of antiserum. The opportunity to use a particle-enhanced
method would provide greater sensitivity and reduce the use of
antibody. To this end, a latex-particle-enhanced turbidimetric
immunoassay
(PETIA)1
was developed to meet the clinical criteria of a wide working
range with good precision and specificity.
| Materials and Methods |
|---|
|
|
|---|
Preparation of antibody particle reagent (PR).
A polyclonal
antibody, IgG fraction, (Binding Site; cat. no. PCO66) was covalently
coupled to 40-nm diameter latex particles consisting of a
polyvinyl-naphthalene core and a chemically reactive chloromethyl
styrene shell (Dade International) (29). Particles were
added in a slow dropwise fashion to the diluted antibody with gentle
shaking. The final concentrations of antibody and particles for
coupling were 2.5 g/L protein and 5 g/L dry solids, respectively, in 15
mmol/L sodium phosphate buffer, pH 7.5, 0.5 mL/L Gafac RE610 detergent
(Gafac, Manchester, UK). The mixture was incubated overnight at
37 °C in an orbital shaker (Infors HT, Crewe, UK) at 240 rpm
followed by ultracentrifugation at 40 000g for 60 min at
20 °C. The supernatant was carefully removed and discarded, whereas
the pellet was resuspended in wash buffer (50 mmol/L glycine, 0.05 mL/L
Gafac, and 0.1 g/L azide, pH 7.5) made up to the volume of the starting
reaction mixture. This was followed by three similar wash steps, after
which the final pellet was resuspended in a volume of storage buffer
(500 mmol/L glycine, 0.5 mL/L Gafac, and 1 g/L azide, pH 7.5)
equivalent to one-half of the original volume of the starting coupling
reaction mixture. The final preparation of PR was placed in an ice bath
and sonicated by two bursts of 30 s duration (20 microns intensity)
with an interval of 1 min between bursts, using an MSE Soniprep 150
(Fisons, UK). The PR was now ready for use in the assay; the working
dilution was adjusted with 5 mmol/L glycine, 0.005 mL/L Gafac, and 0.01
g/L azide, pH 7.5, such that the absorbance at the active measuring
wavelength of 383 nm (termed as "initial absorbance") was 1000
milliabsorbance units before the addition of sample.
Coupling reproducibility.
Identical lots of particles and IgG
fractions and identical scales of synthesis (volumes of reaction
mixture) were used to generate calibration curves from separate
syntheses (n = 3) and to determine the variation (CV) in
signal at each concentration of prealbumin.
Calibrators and matrix.
The purified prealbumin used as the
calibrator material was provided by Scipac (Kent, UK; product code
P1711, >96% purity). A stock solution of calibrator (1000 mg/L),
was prepared in an artificial matrix (HEPES-buffered bovine serum
albumin with preservatives) supplied by Dade International. A zero
calibrator and a series of five working calibrators were prepared by
diluting the stock solution to values assigned by the Beckman
Array® and calibrated according to IFCC reference material
CRM470 (30).
assay development
Within the operation framework of the Dimension analyzer, all
selections of materials for reagents and their optimizations were
performed to meet specifications of a maximum change in signal over the
required assay range, to achieve the best reproducibility of
performance and the lowest nonspecific response.
Nonenhanced assay.
The alternative approach of using free
antibody in place of latex enhancement in the assay produced very poor
signal changes of
510 milliabsorbance units at 500 mg/L
prealbumin. This was observed in the presence of relatively high
amounts of polyethylene glycol (PEG; 40 g/L) and was, therefore,
abandoned.
Antibody loading.
A binding site antibody was chosen from a
variety of commercial and in-house sources, both IgG- and
affinity-purified, which best fulfilled the assay criteria. The working
range of the assay (0550 mg/L), was achieved by minimizing the sample
volume and using the maximum amount of antibody within the constraints
of the system. This was done by using high loadings of antibody and
additions of concentrated PR. A 2.5 g/L loading of antibody onto
particles was chosen because the higher loadings that had been used
previously had produced colloidal instability and steric hindrance.
(31).
PR dilution and wavelength selection.
A system requirement for
the assay was an initial absorbance of
1000 milliabsorbance
units at the active wavelength of measurement. This ensured that the
rise in signal caused by immunoagglutination stayed within the 2000
milliabsorbance-unit upper limit of absorbance linearity (BeerLambert
Law), as well as the measuring range of the spectrophotometer on board
the Dimension system. At any given working dilution of PR, the initial
absorbance decreases with increasing wavelength, according to the
Rayleigh light-scattering theory when particle sizes are
<1/10 of
the wavelength of incident light. This provided the means of extending
the working range of the assay by increasing the amounts of PR (by
altering the volume of added PR) at higher wavelengths while
maintaining the initial absorbance at 1000 milliabsorbance units.
Assay buffer.
Conditions were optimized such that the
enhancing effect of PEG 8000-assisted agglutination, (PEG obtained from
Sigma, UK), was offset by maintaining a negative charge on the antibody
at pH 7.5, (IgGs have a pI
6.0) and by the addition of an anionic
detergent to increase PR stability by decreasing nonspecific binding.
The presence of phosphate has also been shown (28) to aid
complex formation (antichaotropic agent); therefore, the concentration
of phosphate was minimized from the starting conditions of a 150 mmol/L
phosphate buffer to 15 mmol/L while maintaining the ionic strength with
sodium chloride. This also prevented the potential carryover of
phosphate interference into other analytical tests. The final
conditions are described in the assay protocol section.
Assay protocol.
Reagents were introduced into empty, punctured
reagent containers (flexes(TM)) as follows: 4 mL of assay buffer (15
mmol/L sodium phosphate buffer, 135 mmol/L sodium chloride, 14 g/L PEG,
and 0.1 mL/L Gafac) was added to wells 14, and 2.6 mL of PR was added
to wells 5 and 6. The immunoagglutination reaction was started by the
addition of 2 µL of sample, from a total sample volume of 30 µL
present in the sample cup, after a 100-s preincubation of assay buffer
(346 µL) with PR (106 µL). The assay signal was determined from the
captured filter data, incorporating a sample blank (i.e., the
difference between two bichromatic readings at the active wavelength of
383 nm and reference wavelength of 700 nm), taken 30 and 300 s
after sample addition.
Determination of prealbumin result.
A calibration curve was
constructed using an iterative five-parameter logistic (logit)
curve-fitting procedure, as found on the Dimension system, but using a
software package (Deltagraph® software; Delta Point) to
calculate the initial coefficients off-line. The concentrations of
unknown prealbumin samples were determined using the logit algorithm on
a Microsoft Excel spreadsheet, where the concentration was calculated
from the signal and generated coefficients.
Statistical analyses.
Individual aspects of reproducibility,
such as assay imprecision, calibration stability, and reagent stability
(on board the Dimension analyzer), were assessed by the total
reproducibility (interassay variation) and the within-run
reproducibility (intraassay variation), using the ANOVA procedure
recommended by the NCCLS (32). Using Microsoft Excel,
Ver. 5.0, software, a spreadsheet was constructed to perform the ANOVA
calculations. A
test was used to assess the
significance of the calculated reproducibility values relative to
chosen values (e.g., CV 5%) or to identify the significance of factors
contributing to the total imprecision. The agreement between methods
was assessed by Passing and Bablok regression (33), which
accounts for errors in both x and y planes, makes
no assumption on the error distribution, and in which the estimated
line of best fit is not unduly influenced by extreme points or
outliers. A BlandAltman (34) procedure was used to assess
bias. The statistical package used was the Astute(TM) (Diagnostic
Development Unit, University of Leeds, UK). Simple linear regression
was used to assess parallelism.
evaluation and validation
Analytical recoveries and parallelism.
Recoveries were
assessed by the addition of two different amounts of puri- fied
prealbumin (stock 1000 mg/L, weighed into the calibrator) to nine
nondiseased serum samples such that the volume of the supplement was
<10% of the total sample volume. The recovery with endogenous
prealbumin was assessed by experiments where 10 samples containing
relatively high prealbumin concentrations (range, 260400 mg/L range)
were each diluted (by factors of 0.2, 0.4, 0.6, or 0.8) with 10 samples
containing relatively low concentrations of prealbumin (range, 60160
mg/L). Parallelism was determined by analyzing 10 serum samples
containing prealbumin concentrations in the 250450 mg/L range that
had been serially diluted with saline (x0.25, x0.5, x0.75, and
x0.875).
Imprecision.
The within-assay and total reproducibility were
calculated by the ANOVA procedure, using a one-assay-per-day design on
duplicate analyses of four serum pools containing prealbumin
concentrations representative of the working assay range. This was
performed on 20 occasions throughout a period of 40 days, using a new
calibration curve each time. In addition, the assay imprecision was
also estimated by a precision profile obtained from duplicate analyses
of 103 samples used for the method comparison study.
Calibration and reagent stability.
Both calibration and
reagent stability were analyzed by ANOVA, giving within-assay and total
reproducibility values. A re-analysis of the imprecision data
calculated according to the calibration from day 1 was performed to
determine the calibration stability. The reagent stability on board the
analyzer, with already punctured flexes as required by the Dimension
analyzer open channels format, was also assessed from captured filter
data over the same occasions and period of time. The results were
likewise calculated according to day 1 calibration.
Accuracy.
The developed PETIA was compared with the selected
reference method (Beckman Array system, Beckman Instruments), which
used a calibrator assigned with the IFCC reference material CRM 470
(30). Serum specimens that had been kept at 4 °C for <1
week were obtained from the hospital routine clinical laboratory for
simultaneous analysis by both methods. Samples were drawn from the
following patient groups: reference interval albumin (3546 g/L;
n = 41), low albumin (<25 g/L; n = 20), renal patients with
urea >15 mmol/L (n = 18), lipemic and icteric (n = 5), and
quality controls (n = 19), obtained from Behring (cat. no. OWX),
Incstar (cat. no. SPQ), Bio-Rad (cat. no. 1587) and Dako (cat. no. X94)
in the United Kingdom and Fitzgerald (cat. no. 35-PC5) in the United
States.
Interference and cross-reaction with albumin.
Interference was
assessed by the addition of the following interferents at the stated
final concentrations in six serum samples containing measured
prealbumin concentrations of 80, 103, 118, 140, 180, and 250 mg/L:
hemoglobin at 500-7700 mg/L, prepared from washed, hemolyzed
erythrocytes and concentrated by ultrafiltration; bilirubin at 4.4550
µmol/L, prepared in a stock solution in sodium carbonate/dimethyl
sulfoxide; and triglyceride at 1.512.2 mmol/L, obtained in the form
of Intralipid 20% fat emulsion from Pharmacia. Conditions were
arranged such that the volume of added interferents was <5% of the
total sample volume.
The effects of commonly used blood sample collection procedures in a routine clinical laboratory environment were also investigated. Blood was obtained from 10 healthy subjects, using serum (plain tubes), serum with SST gel-clot activator, and plasma from Li-heparin and EDTA tubes. Prealbumin was measured during the same day in all samples.
A series of human serum albumin (Behring/Hoechst, UK, cat no. ORHA) samples were prepared in the range 110 g/L and measured in the prealbumin assay to assess cross-reactivity.
Assay range and sensitivity.
Precision profiles were obtained
by curve-fitting a second-order polynomial function to the relationship
between the SD and the CV of duplicate measurements, with respect to
binned prealbumin concentration from the Dimension analyzer comparison
results. In addition, a duplicate analysis was performed on seven
samples that contained low concentrations of prealbumin to allow a
better definition of the binned range, 50100 mg/L. Because of
insufficient volume, these samples were not available for the method
comparison. The assay range was defined as the prealbumin concentration
range within which the CV was acceptable to the analyst, generally
accepted to be at the CV cutoff of 10% (35). The
sensitivity (detection limit) as defined by the error of measurement at
zero dose (35) was obtained from the intercept of the SD
precision profile.
Sample and calibrator stability.
Three serum sample pools and
two freshly prepared calibrators were measured after storage under the
following conditions: five freeze-thaw cycles during a single day,
based on an initial calibration curve, and during days 1, 2, 3, 5, and
10 of incubations at room temperature (
20 °C), 4 °C, and
-20 °C.
| Results |
|---|
|
|
|---|
C2)1)]C0.
|
pr dilution and wavelength selection
An active wavelength of 383 nm was selected; Fig. 2
shows the respective effects, at two different amounts of
PR, of increasing the wavelength on the standard curves of the
assay and the initial absorbances. The 340 nm wavelength gave working
ranges of prealbumin at prohibitively high initial absorbances of
>1360 milliabsorbance units, whereas the higher wavelength (405 nm)
was also avoided because of potential interference from hemolyzed
samples (hemoglobin strongly absorbs at 410 nmreferred to as the
"Soret" band).
|
analytical recovery and parallelism
Mean percentage recoveries (± 1SD) for samples supplemented with
90 and 50 mg/L purified prealbumin were 104.1% ± 4.3% and
100.1% ± 2.3%, respectively; the complete data are shown in
Table 1
. The mean recoveries (± 1SD) of endogenous prealbumin for each
of the fixed proportions were 100% (± 3.1%), 101% (± 2.3%), 101%
(± 2.1%), and 102% (± 2.1%). The results obtained for the
parallelism study gave a linear regression of [Observed] = -1.5
1.02[Expected], with a coefficient of determination of 0.999.
|
imprecision
There were no CV values significantly higher than the 5%
determined by the
analysis for either
within-assay or total reproducibility in the four control serum pools
(Table 2
). The mean CV values (and ranges) obtained from duplicate
analyses from the method comparisons were 1.59% (08.2%) for the
Dimension analyzer assay and 1.22% (08.6%) for the Beckman Array
assay.
|
stabilities of calibration and on-board reagents
Table 2
also provides the estimates of total and within-assay
reproducibility for both assessments of stability in each of the four
control pools; CVs were 15%, except for a CV of 7.6% for the total
reproducibility in the low pool for calibration stability. The daily
pool means of the calibration stability (Fig. 3
) showed no discernible trends over the duration of the study.
|
method comparison
A scattergram and a Passing and Bablok regression for the method
comparison are shown in Fig. 4
. The Dimension analyzer assay showed close agreement of both
slope (1.00 ± 0.06) and intercept (7.0 ± 13) with the
target values of 1.00 and zero within the ±95% confidence limits. An
additional analysis (not shown), by the BlandAltman method,
demonstrated a random scatter in the relationship between method
differences and averages, thus confirming the absence of bias.
|
interference
Samples with added hemoglobin, bilirubin, and intralipid showed no
significant change of results from those of the same samples in the
absence of interferents (Table 3
), with measured values all within 100% ± 5%. There also
were no significant differences (P >0.5 by MannWhitney)
between the various sample collection conditions compared with serum
obtained in plain tubes. At a concentration of 10 g/L albumin, the
prealbumin assay showed a signal equivalent to the zero calibrator,
thus demonstrating the negligible cross-reaction of albumin (data not
shown).
|
assay working range and sensitivity
Using the CV cutoff of 10%, the precision profile of the form
y = 6.37 - 4.3-2x
8.4-5x (Fig. 5
) gave an assay range which covered the range of calibration
(0550 mg/L). The detection limit (sensitivity), given by the
intercept of the SD precision profile curve fit, (of the form
y = 7.92 - 7.0-2x
1.9-4x), was 8 mg/L
prealbumin. The assay working range was taken as the values lying
between the sensitivity and the uppermost standard, or 8550 mg/L.
|
sample and calibrator stability
There were no discernible changes in the sample pools or
calibrators during the freeze-thaw cycles or during days 1 to 10 of
incubations at 4° and -20 °C; recovery results all lay
within ± 5% of the initial readings (data not shown). The
incubations at room temperature began to show appreciable variation
of ± 20% from the running mean after 3 days in one of the
samples and calibrators (Fig. 6
).
|
coupling reproducibility
The variation (CV) in signal over the three syntheses, at each
concentration of prealbumin standard, was <5%.
| Discussion |
|---|
|
|
|---|
The covalent coupling of proteins such as IgG has been shown in other
established immunoassays to promote reagent stability
(31)(36). This was reflected by the behavior of
on-board reagent stability and calibration stability, where acceptable
performance was observed for up to 40 days. The imprecision of the
assay (CV, 15%) was similar to other clinical methods and direct
immunoassays, such as PETIAs, that generally are more precise compared
with their immuno-inhibition counterparts (35). Nutritional
studies and/or reviews, composed of individual management of patients
within defined groups (1)(2)(3)(4)(5)(6)(7)(8)(9)(10)(11)(12)(13)(14)(15)(16)(17)(18)(19), all indicate that this level
of total precision (CV, <5%) is more than sufficient to be of
diagnostic use, particularly in the critical area of interest
(
40170 mg/L) as well as including the higher reference range
concentrations of 170500 mg/L. The upper values of all reported
reference ranges have been shown not to exceed 500 mg/L. Latex
enhancement of immunoagglutination was observed to be an absolute
requirement, because very poor signals were seen when free antibody was
used, even in the presence of high PEG concentrations.
When the sample blank approach was used, the assay was unaffected by any of the added interferents: hemoglobin, bilirubin, or triglyceride (in the form of a lipid emulsion) at concentrations that might be expected in a routine clinical laboratory setting. Corresponding experiments that used baseline readings before sample addition demonstrated unacceptable photometric interferences, despite the relatively small proportion of sample (2 µL) present in a total reaction volume of 500 µL. In addition, the assay was observed to be valid for both serum and plasma collected by the methods most commonly used in any clinical practice. Close agreement of the Dimension analyzer PETIA with the reference method was demonstrated using calibrator values assigned by the Beckman method. Sample and calibrator stability indicated that specimen storage at room temperature for >3 days should be avoided. The small sample volume requirement of 2 µL (in a dead volume of 30 µL), the 10 mg/L detection limit, and the lack of significant interference from bilirubin or lipid makes the method to be suitable for pediatric use. In keeping with the performance of the Dimension AR or ES analyzers for other analytes, a throughput rate of 200 results/h may be achieved for this prealbumin assay. Quoted insert sheet details for commercially available methods, e.g., Beckman Array and Behring Nephelometer Analyzer®, obtained from the manufacturers at the time of this study, indicate sample volumes and lower reportable limits of 150 µL and 70 mg/L for the Beckman Array and 50 µL and 20 mg/L for the Behring Nephelometer Analyzer.
In conclusion, an accurate, precise, reliable method for prealbumin has been developed and automated for use on the Dimension Clinical Chemistry Analyzer. The recommended format is defined in the open channels software, available for implementation onto the main instrument panel of diagnostic tests. Measurements of prealbumin and C-reactive protein can thus be integrated into a routine nutritional profile where the results for both may be obtained within 7 min. The method covers the entire concentration range of clinical interest with no predilution of sample.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
H. Wang, Y. Zhang, B. Yan, L. Liu, S. Wang, G. Shen, and R. Yu Rapid, Simple, and Sensitive Immunoagglutination Assay with SiO2 Particles and Quartz Crystal Microbalance for Quantifying Schistosoma japonicum Antibodies Clin. Chem., November 1, 2006; 52(11): 2065 - 2071. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Englebienne, A. Van Hoonacker, and J. Valsamis Rapid Homogeneous Immunoassay for Human Ferritin in the Cobas Mira Using Colloidal Gold as the Reporter Reagent Clin. Chem., December 1, 2000; 46(12): 2000 - 2003. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |