(Clinical Chemistry. 1998;44:1397-1403.)
© 1998 American Association for Clinical Chemistry, Inc.
|
Molecular Diagnostics and Genetics |
Rapid sequencing of the p53 gene with a new automated DNA sequencer
Bhupinder S. Bharaj,
Katerina Angelopoulou,
and Eleftherios P. Diamandisa
Department of Pathology and Laboratory Medicine, Mount Sinai Hospital, Toronto, Ontario, Canada M5G 1X 5; and Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, Ontario, Canada M5G 1L5.
a Address correspondence to this author at: Department of Pathology and Laboratory Medicine, Mount Sinai Hospital, 600 University Avenue, Toronto, Ontario, Canada M5G 1X 5. Fax 416-586-8628; e-mail ediamandis{at}mtsinai.on.ca.
 |
Abstract
|
|---|
p53 is the most commonly mutated gene in human cancers.
Approximately 90% of the p53 gene mutations are localized
between domains encoding exons 5 to 8. Sequencing methods currently
available are tedious and time-consuming and are not suitable for
routine laboratory testing. In an effort to identify a simple and rapid
sequencing method, we analyzed 16 preselected breast tumors and 18
preselected ovarian tumors, using a newly developed automated DNA
sequencer. p53 gene mutations had been previously
identified in these tumors, using a conventional automated sequencing
procedure. Exons 5 to 8 were amplified by PCR, and the PCR products
were subsequently subjected to cycle sequencing with the Sanger chain
termination method, using Cy5.5-labeled primers. The sequencing mixture
was then resolved on a newly developed automated DNA sequencer that can
sequence ~300 bases of DNA in 30 min. Of these 16 breast tumors, two
had mutations in exon 5, four in exon 6, three in exon 7, and three in
exon 8. Of the 18 ovarian tumors, two had mutations in exon 5, five in
exon 6, two in exon 7, and three in exon 8. In all cases, we identified
the same mutations by both the new and the conventional sequencing
procedures. Most mutations affected an arginine codon. These data
demonstrate that the new method has the capability to provide accurate
sequencing information in a fraction of the time and labor in
comparison with current automated sequencing techniques. When such
procedures are used, DNA sequencing may become a routine tool for
identifying clinically important mutations for diagnosis and prognosis
of patients with genetic, malignant, infectious, and other diseases.
 |
Introduction
|
|---|
p53 is a 53-kDa protein that is frequently mutated in human cancer
(1). Many studies have confirmed that the mutations at the
p53 locus constitute the single most common genetic
alteration observed in human cancer (2). This tumor
suppressor gene is known to control cellular growth after DNA damage
through mechanisms involving cell cycle arrest and programmed cell
death (3). The wild-type p53 functions as a cell cycle check
point, activated by insults against the integrity of the genome.
Cellular stresses such as DNA damage induce p53 protein accumulation,
which transcriptionally modulates the expression of genes, triggering
either growth arrest or apoptosis. p53 inactivated by mutations is
generally nonfunctional, leading to a failure to arrest or delete cells
harboring potentially tumorogenic mutations. Some mutated forms of
p53 gene may even gain the ability to stimulate cell growth
(4) and promote tumor induction and progression
(5).
The human p53 gene is located on chromosome 17p13 and is
composed of 11 exons and 10 intervening introns. The protein consists
of a highly charged acidic amino- terminal domain, which can interact
with components of the transcriptional machinery, and a highly charged
basic carboxy-terminal region, which can mediate oligomerization,
nuclear localization, and possibly the recognition of damaged DNA. The
central portion contains the DNA-binding core, within which the vast
majority of single amino acid substitutions occur when the
p53 gene is mutated. A database assembled from nearly 1500
p53 mutations from cancers of different cell or tissue types indicated
that almost 85% are point missense mutations leading to a functionally
defective protein (6). Approximately 90% of these missense
mutations have been shown to be localized in the DNA-binding domain of
p53, which is encoded by exons 5 to 8 (7). Another
consequence of missense mutation in this domain is the increased
stability of the encoded mutant p53 protein, which accumulates in tumor
cell nuclei. Amino acids R248 and R273, the most frequently mutated
residues in the p53 gene, exhibit direct contact with DNA in
the consensus p53 binding sequence. In total, ~40% of the missense
mutations are localized at the "hotspot" residues R175, G245, R248,
R249, R273, and R282 (8), which play a vital role in the
structural integrity of the DNA-binding domain. Sequencing of the core
region alone allows the detection of the majority of the mutations
(9).
Many techniques have been used to identify p53 abnormalities at either
genetic or protein levels. These procedures, including
immunohistochemistry and single-strand conformation polymorphism, are
plagued with inconsistencies and often are difficult to interpret
(6)(10). Therefore, the use of sensitive and
reproducible methods is imperative if the objective is to gain
definitive and reliable assessment of the mutational status of the
p53 gene. Such information is very useful and has
established prognostic implications for the cancer patient
(11)(12)(13)(14). Direct sequencing is the most accurate technique
to assess the mutational status of the p53 gene, but many
clinical laboratories do not use it because of the complexity, cost,
and labor-intensive nature of the currently available techniques. This
study evaluates a new DNA-sequencing method for assessing the
p53 gene status. The method is based on new instrumentation
that simplifies some steps of the procedure and appears to be suitable
for routine DNA diagnostic applications.
 |
Materials and Methods
|
|---|
tumor extraction
Breast and ovarian tumor tissues that had been previously
sequenced using the ALFexpress automated DNA sequencer (Pharmacia
Biotech) were used in this study (15). Approximately 200 mg
of the tumor tissue, which contained >70% tumor cells as determined
by histologic examination, was pulverized to a fine powder at
-70 °C. DNA was extracted and purified using a conventional
phenol-chloroform-based procedure (16), quantified by
measurements at 260 nm, and stored at 4 °C until analysis.
pcr amplification
The paired primer sequences flanking each of the exons 5 to 8 of
the p53 gene are shown in Table 1
. All oligonucleotide primers were designed using computer
software Oligo 5.0 (National Biosciences), according to the
p53 sequence deposited in GenBank by Chumakov et al.
(accession no. 54156). PCR amplification of each exon was performed in
a final volume of 25 µL, containing ~100 ng of template DNA, 10
mmol/L tromethamine (Tris) buffer, pH 8.3, 50 mmol/L KCl, 1 unit of
AmpliTaq Polymerase (Hoffmann-La Roche), 250 µmol/L deoxynucleoside
triphosphates, 1.5 mmol/L MgCl2, and 1 µmol/L of each
primer. PCR primers and other PCR reagents were part of the
p53 gene sequencing kit that was used in this study and
provided by Visible Genetics Inc., Toronto, Ontario, Canada
(www.visgen.com or info@visgen.com). The thermal cycling profile
consisted of a 30-s denaturation step at 94 °C, a 30-s annealing
step at 60 °C, and a 30-s extension step at 70 °C, for a total of
35 cycles. Each PCR was initiated with a 5-min denaturation at 94 °C
and terminated with a 5-min extension at 70 °C. The success of
the PCR was verified by running a 5-µL aliquot of the PCR
product on a 0.8% agarose gel containing ethidium bromide.
dna sequencing
The primers used for sequencing the PCR-amplified p53
exons 5 to 8 with the Visible Genetics system (see below) were labeled
at the 5' end with the fluorescent dye Cy5.5 (Amersham International).
Their sequences are shown in Table 1
. The same primers, labeled with
Cy5 fluorescent dye, were used to sequence the PCR products on the
Pharmacia ALFexpress system. The detailed procedure for sequencing
these samples with the ALFexpress system has been described elsewhere
(15).
The sequencing primers, labeled with Cy5.5, are also part of the
Visible Genetics p53 gene sequencing kit. The primers were
prepared as follows. During standard phosphoramidite synthesis of the
primers, the
6-(trifluoroacetylamino)propyl-(2-cyanoethyl)-(N,
N-diisopropyl)-phosphoramidite (5'-amino modifier, C3-TFA)
was introduced at the 5' end of each sequencing primer. The base-labile
TFA protecting group of the 5'-amino modifier TFA was removed by
ammonium hydroxide treatment. The free amino group of the primer was
then reacted with an N-hydroxysuccinimide ester of the
fluorescent dye Cy5.5 (Amersham) to obtain the labeled primer, which
was purified by reversed-phase high performance liquid chromatography.
In the sequencing procedure, we used reagents from Visible Genetics.
Briefly, an aliquot of the PCR product (18 µL depending on the
yield, no purification necessary) was mixed with
SequenaseTM buffer, dimethyl sulfoxide, sequencing
primer, water, and ThermosequenaseTM enzyme (Amersham).
This mixture was then distributed into four tubes (5 µL/tube, labeled
A, C, G, or T), and the nucleotide termination mix was added. The
mixture was then cycled for 35 cycles on a PCR machine as follows:
94 °C for 30 s, 55 °C (exon 5) or 60° C (exons 6 and 8)
or 68° C (exon 7) for 30 s, and 70 °C for 60 s. The
thermal cycling was initiated with a 5-min denaturation step at
94 °C and terminated with a 5-min extension step at 70 °C. At the
end, 6.0 µL of the loading dye was added to each of the four tubes to
stop the sequencing reaction. The sequencing samples were heated at
95 °C for 5 min and then placed on ice before 2.0 µL was loaded on
the sequencing gel. Details of the sequencing mixes, including
concentrations of dNTPs and ddNTPs, are described in the insert of the
p53 gene sequencing kit.
The Microgene BlasterTM automated DNA sequencer is a
compact sequencing device containing a visible light laser diode (675
nmol/L) as the excitation source. The laser power is 0.5 mW/lane. The
machine applies 1500V across the sequencing gel and can complete a run
in ~30 min, resolving ~300 bases of sequence. The instrument
accepts a 14 x 14 cm sequencing gel, 50 µm thick. The gel is
cast between two disposable glass plates (MicroCelTM
cassette). In our study, each gel was capable of analyzing four patient
samples (16 lanes; 4 lanes per sample for A, C, G, and T sequencing
reaction tubes). The cassette was filled with 6% acrylamide solution
containing a photo-initiator that was activated by UV light on a
special unit (Gel ToasterTM polymerizing unit). The process
of gel filling and casting took ~5 min.
Once the gel was ready, it was positioned into the sequencer, and the
buffer chambers were filled with 150 mL of a 1x Tris-borate-EDTA
buffer, pH 8.3. The gel temperature and voltage were set at 50 °C
and 1500 V, respectively. A 5-min prerun was initiated to bring the gel
and the temperature to the set values. Each of the 16 lanes was loaded
with 2 µL of the cycle sequencing reactions. The electrophoresis time
was 30 min. Data acquisition and interpretation was achieved with Gene
ObjectsTM software (Visible Genetics).
The Gene Objects software includes modules for both base calling and
fragment analysis. Part of the base calling algorithm is the selection
of thresholds for heterozygote detection, i.e., no threshold, 50%,
20%, and 15% thresholds. When the instrument detects two peaks of
different intensities on the same position, it will call it an
ambiguity (possible heterozygosity) if the intensity of the second
peak, in comparison with the main peak, is more than the specified
threshold. In addition, the software includes a comparator tool that
compares two sequences and automatically lists all the differences
including insertions, deletions, and base changes. Once a difference is
identified, the comparator can realign downstream bases of the
differing allele to produce a clear comparison. This option allows the
operator to compare a generated sequence with a reference sequence
without the need for manual comparison. The Gene Librarian tool is also
part of the software and allows comparison of a newly generated
sequence with a whole database of known genes stored in the library.
 |
Results
|
|---|
We present in detail all the mutations identified by the new
sequencing methodology in Table 2
. Among the total of 34 specimens tested (16 breast and 18
ovarian tumors), 24 (71%) were found to harbor a mutation (12 breast
and 12 ovarian tumors). The rest of the specimens had the wild-type
p53 gene sequence. These data do not reflect mutation rates
in serial samples from breast and ovarian carcinoma, because the
specimens were preselected from a larger series to contain ~6070%
mutant and 3040% wild-type p53 gene sequences. This
optimized the comparison between the new method and the conventional
sequencing technique in terms of ability to detect mutations. Among all
tumors, the new sequencing method agreed in terms of p53
gene status in all cases. Representative examples of sequencing
tracings obtained with the new technique as well as with the ALFexpress
comparative method are shown in Figs. 1
and
2.
To check the ability of the new sequencing method to detect
heterozygotes, we sequenced the exon 6 of the p53 gene in
the breast carcinoma cell lines MCF-7 (wild-type sequence) and T-47D
(homozygous mutation within exon 6). We then mixed DNA from these cell
lines at various proportions and repeated the sequencing. The data
(Fig. 3
) confirm the ability of the new technique to detect the
heterozygous state. However, in Fig. 3B
, the presence of ~25% of
mutant DNA was not detected, presumably because the intensity of the
mutant DNA signal in the final mixture was below the threshold (15%)
of the base-calling algorithm.

View larger version (27K):
[in this window]
[in a new window]
|
Figure 3. Ability of the new sequencing method to detect
heterozygosity.
(A) Wild-type p53 gene sequence of part of exon
6, using DNA from the breast carcinoma cell line MCF-7. (E)
Homozygous mutation in exon 6 of the p53 gene (C to T),
using DNA from the breast carcinoma cell line T-47D. (BD)
Sequencing of a mixture of MCF-7/T-47D DNA at ratios of 75%, 50%, and
25%, respectively. The software efficiently detects heterozygosity in
(C) and (D) but not (B). Y: C or
T.
|
|
 |
Discussion
|
|---|
DNA sequencing is one of a few fundamental techniques that have
revolutionized the generation of new genetic knowledge. This method has
been used in research settings for many years. More recently, the needs
of the Human Genome Project necessitated the introduction of
instrumentation that can generate large amounts of sequencing
information in an automated fashion, without using radioactivity. Such
instruments, which are mainly based on gel electrophoresis and
laser-induced fluorescence detection, now have been introduced in
clinical laboratories for studying genetic, infectious, and malignant
diseases. The number of genes that are cloned every year is expanding
exponentially. Many of these genes are altered by mutations, causing
various human diseases. Examples of genes that are implicated in
carcinogenesis and harbor mutations include p53
(17), ras (18), BRCA1 and
BRCA2 (19)(20)(21), and Rb
(22). To study such mutations, clinical laboratories may
apply an array of genome-screening techniques, recently reviewed in
this journal (10). None of these methods is as reliable as
DNA sequencing. Thus, current and future clinical applications will
require simple, reliable, and cost-effective methods for DNA
sequencing.
The sequencing protocol starts from DNA preparation, which is generally
a manual procedure. The same applies to the PCR amplification step and
the performance of the sequencing reactions. The new protocol described
here does not need a PCR product purification step because the PCR
reactions have been carefully optimized to produce a single PCR band,
and only a fraction of PCR product is used for sequencing. In some
sequencing protocols, PCR product purification is recommended (Fig. 3
).
In addition, the sequencing reaction products do not need precipitation
for the purpose of preconcentration and cleaning from unincorporated
labeled primer or nucleotides. The use of the more sensitive label,
Cy5.5, in comparison with Cy5, allows for loading of only a fraction of
the sequencing reaction product.
The major contributions of this new technology involve the gel casting
procedure and the electrophoresis time. Gel casting using low cost
disposable cassettes can be completed within 5 min and is a major
advantage over conventional reusable plates, which need more time to
prepare and clean. Running time of the gel can be completed within 35
min, an order of magnitude faster than conventional DNA sequencers like
the ALFexpress. This is mainly due to the use of thinner gels. In
general, sequencing data with the new system may be obtained
within 1 working day, whereas with a conventional system, data are
usually generated in 2 working days. Hands-on time with the new
procedure is relatively limited. Importantly, the new system provides
faster data output while maintaining the accuracy of the provided
information, as outlined in Table 2
and Fig. 1
. An outline of the
workflow comparing the new system with the ALFexpress is shown in Fig. 4
.

View larger version (20K):
[in this window]
[in a new window]
|
Figure 4. Flowchart describing approximate times and steps necessary
to complete a sequencing experiment with the new method and the
ALFexpress procedure.
The new method provides sequencing data within 1 working day; the
comparison method within 2 days.
|
|
The sequencing data of Table 2
were generated by selecting specimens
from a series of ~80 tumors (50 ovarian and 30 breast). Because the
type of the mutation was not preselected in this study, it is
remarkable that in 15 of 24 cases with missense mutations, the affected
codon encodes for arginine [CG (ACGT)]. These mutations will probably
affect the biological activity of p53, because many arginine molecules
are necessary for DNA binding of p53 (23). These
observations were also made by others (8).
In conclusion, we here provide evidence that the new
instrumentation efficiently detects mutations in the p53
gene while reducing the complexity of the technique and substantially
decreasing the electrophoresis time. This system has the capability to
be used for detecting mutations in routine diagnostic laboratories.

View larger version (52K):
[in this window]
[in a new window]
|
Figure 2. DNA sequencing of the p53 gene with the
ALFexpress automated DNA sequencer.
Mutation detection of the same patients described in Fig. 1
with this
comparative method.
|
|
 |
Acknowledgments
|
|---|
This work was supported by a grant to E.P. Diamandis from Visible
Genetics, Inc., Toronto, Ontario, Canada.
 |
References
|
|---|
-
Masuda H, Miller C, Koeffler P, Battifora H, Clines MJ. Rearrangement of p53 gene in human osteogenic sarcomas. Proc Natl Acad Sci U S A 1987;84:7716-7719.
[Abstract/Free Full Text]
-
Hollstein M, Sidransky S, Vogelstein B, Harris CC. p53 mutations in human cancers [Review]. Science 1991;253:49-53.
[Abstract/Free Full Text]
-
Yonish RE, Resnitsky D, Lotem J, Sachs K, Kim-Chi A, Oren M. Wild type p53 induces apoptosis of myeloid leukemic cells that are inhibited by interleukin-6. Nature 1991;353:345-347.
-
Singerland JM. The transformation and suppressor functions of p53 alleles: effects of mutations that disrupt phosphorylation, oligomerization and nuclear translocation. EMBO J 1993;2:1029-1037.
[ISI][Medline]
[Order article via Infotrieve]
-
Wynford-Thomas D. p53 guardian of cellular senescence. J Pathol 1987;180:118-121.
-
Hollstein M, Rice K, Greenbelt MS, Soussi T, Fuch R, Sorlie T, et al. Database of p53 gene somatic mutations in human tumors and lines. Nucleic Acids Res 1994;22:3351-3355.
-
Velculesu VE, El Deiry WS. Biological and clinical importance of the p53 tumor suppressor gene [Review]. Clin Chem 1996;42:858-868.
[Abstract/Free Full Text]
-
Cho Y, Gorina S, Jaffe PD, Paveletich NP. Crystal structure of a p53 tumor suppressor-DNA complex: understanding tumorigenic mutations. Science 1994;265:346-355.
[Abstract/Free Full Text]
-
Levine AJ. p53 the cellular gatekeeper for growth and division [Review]. Cell 1997;88:323-331.
[ISI][Medline]
[Order article via Infotrieve]
-
Nollou P, Wagener C. Methods for detection of point mutations: performance and quality assessment [Review]. Clin Chem 1997;43:1114-1128.
[Abstract/Free Full Text]
-
Bergh J, Torbjorn N, Sjogren S, Lindgren A, Holmsberg L. Complete sequencing of the p53 gene provides prognostic information in breast cancer patients, particularly in adjuvant systemic therapy and radiotherapy. Nat Med 1995;1:1029-1034.
[ISI][Medline]
[Order article via Infotrieve]
-
Thor AD, Yandell DW. Prognostic significance of p53 overexpression in node negative breast cancer patients. J Natl Cancer Inst 1992;84:1109-1114.
[Abstract/Free Full Text]
-
Allred DC, Clark GM, Elledge R, Fuqua S, Brown R, Chamness G, et al. Association of p53 protein expression with tumor cell proliferation rate and clinical outcome in node negative breast cancer. J Natl Cancer Inst 1993;85:200-206.
[Abstract/Free Full Text]
-
Levesque MA, Katsaros D, Yu H, Zola P, Sismondi P, Giardina G, Diamandis EP. p53 protein overexpression is associated with poor outcome in patients with well or moderately differentiated ovarian carcinoma. Cancer 1995;75:1327-1338.
[ISI][Medline]
[Order article via Infotrieve]
-
Angelopoulou K. Immune response against the p53 tumor
suppressor gene product: clinical studies and molecular mechanisms
[PhD Thesis]. Toronto, Ontario, CANADA: University of
Toronto, 1997:202 pp..
-
Ausubel FM, Brent R, Kingston RE, Moore DD, Seidman JG, Smith JA, Struhl K. Current protocols in molecular biology 1995:2.1.1-2.3.3 John Wiley & Sons New York. .
-
Elledge RM. Assessing p53 status in breast cancer prognosis: where should you put the thermometer if you think you are p53 sick [Review]. J Natl Cancer Inst 1996;88:141-143.
-
Sidransky D, Tokino T, Hamilton SR, Kinzler K, Levin B, Frost P, et al. Identification of ras oncogene mutations in the stool of patients with curable colorectal tumors. Science 1992;256:102-105.
[Abstract/Free Full Text]
-
Miki Y, Swensen J, Shattuck-Eidens D, Futreal A, Harshman K, Tavtigian S, et al. A strong candidate for the breast and ovarian cancer susceptibility gene BRCA1. Science 1994;266:66-71.
[Abstract/Free Full Text]
-
Futreal PA, Liu Q, Shattuck-Eidens D, Cochran C, Harshman K, Tavtigian S, et al. BRCA1 mutations in primary breast and ovarian carcinomas. Science 1994;266:120-122.
[Abstract/Free Full Text]
-
Wooster R, Bigwell G, Lancaster J, Shaft S, Seal S, Mangion J, et al. Identification of breast cancer susceptibility gene BRCA2. Nature 1995;378:789-792.
[Medline]
[Order article via Infotrieve]
-
Lee EYHP, To H, Shew JY, Bookstein R, Scully P, Lee W. Inactivation of the retinoblastoma susceptibility gene in human breast cancers. Science 1988;241:218-221.
[Abstract/Free Full Text]
-
Vogelstein B, Kinzler KW. X-rays strike p53 again. Nature 1994;370:174-175.
[Medline]
[Order article via Infotrieve]