|
|
||||||||
Drug Monitoring and Toxicology |
1
Abteilung Klinische Chemie, Georg-August-Universität Göttingen, D-37075 Göttingen, Germany.
2
Department of Pathology & Laboratory Medicine,
University of Pennsylvania Medical Center, Philadelphia, PA 19104.
a Address correspondence to this author at: Abteilung Klinische Chemie, Zentrum Innere Medizin, Georg-August-Universität, Robert Koch Strasse 40, D-37075 Göttingen, Germany. Fax 49-551-398551; e-mail ewieland{at}med.uni-goettingen.de.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Results from controlled clinical trials with renal transplant patients have shown that the addition of MMF to immunosuppression protocols with cyclosporine and steroids substantially reduces the early acute rejection rate and decreases the need for antirejection therapy while showing a favorable safety profile (6)(7)(8). However, the full clinical potential of the drug has not yet been realized. The use of MMF in heart, liver, and lung transplantation as well as in autoimmune diseases is not yet sufficiently established (3)(4)(9). Limited data are available regarding its pharmacokinetic properties in children or after multiple dosage in patients with renal or liver dysfunction. The role of therapeutic drug monitoring to improve the therapeutic efficacy and to minimize adverse side effects is still under investigation (10)(11).
To address these questions, accurate and sensitive methods are needed to determine the free and total MPA concentrations, as well as the concentration of its glucuronide metabolite (MPAG) in patient blood specimens. Because of the high binding to plasma albumin (>97%) (12), free MPA concentration (MPAf) measurement may be a better indicator of total drug exposure. Therefore, a suitably sensitive method with a very low detection limit is required.
Several gas chromatographic (13)(14) and HPLC methods (15)(16)(17) have been developed for the determination of MPA. Previously described methods suffered from several disadvantages, such as a lack of sensitivity and specificity or the need for a large sample volume (13)(14). To date, only two methods allow the simultaneous measurement of MPA and MPAG (15)(17). The analysis of MPA and MPAG as described by Tsina et al. (17) is relatively complex. MPA and MPAG are analyzed either sequentially in two separate analytical runs using one HPLC system or in parallel by splitting the extract into two portions and injecting them onto two separate HPLC systems. Alternatively, MPAG has been cleaved enzymatically and MPA determined with and without enzymatic pretreatment of the sample (15). Compared with the extraction procedure described here, both methods use time-consuming and laborious extraction protocols. The approach described by Tsina et al. (17) can be automated by use of a robotic system, thereby making it more convenient. However, this robotic system is not readily available in every laboratory.
Here, we describe a rapid and reliable reversed-phase HPLC method for the simultaneous determination of MPA and MPAG in plasma, as well as for the determination of the MPAf concentration in plasma water, based on simple sample preparation and gradient elution chromatography. The method has a lower detection limit than previously published procedures. This renders the method particularly suitable for measurement of the MPAf in plasma samples.
| Materials and Methods |
|---|
|
|
|---|
sample preparation
EDTA-blood was centrifuged at 3000g; plasma was
collected and either immediately analyzed or stored at -20 °C until
measurement. For determination of the plasma concentrations of MPAG and
total MPA, 200 µL of the sample (calibration, control, or plasma) and
100 µL of acetonitrile containing the internal standard MPAC (15
mg/L) were vortex-mixed in a 1.5-mL polypropylene tube for 5 s.
This was followed by the sequential addition of 20 µL of 150 g/L
perchloric acid and 20 µL of 250 g/L sodium tungstate to the tube,
with vortex-mixing for 15 s after each addition. The sample was
then centrifuged for 5 min at 10 000g, and 50 µL of the
supernatant was removed for chromatography. The
Centrifree® Micropartition System (Amicon) described by
Nowak and Shaw (12) was used to obtain an ultrafiltrate for
MPAf determination. Because impurities were found in some batches of
the Centrifree filters, all filters were routinely washed immediately
before use. The micropartition tubes were soaked in a mixture of 500
mL/L methanol and 500 mL/L water and sonicated for 10 min. They were
subsequently flushed with a fresh mixture of 500 mL/L methanol and 500
mL/L water and centrifuged for 20 min at 2000g, leading to
the complete accumulation of the flush solution in the collection cups.
The flush solution was then discarded, and the cups were carefully
dried with a cotton tip. For the ultrafiltration procedure, 300 µL of
either calibration or control solution or plasma sample was added to
the sample reservoir, and the tube was centrifuged for 40 min at
2000g at 20 °C, yielding ~150 µL of ultrafiltrate.
The internal standard MPAC (2.5 mg/L) was mixed with the ultrafiltrate
at a ratio of 1:10 (by volume), and a 100 µL aliquot was then
injected directly onto the column.
chromatographic conditions
The HPLC system consisted of a chromatographic pump (M480), an
automatic injector (GINA 50), a diode array detector (UVD 340S), a
computer interface system controller linked to a PC (Gyncotek) and a
250 mm x 4.6 mm Symmetry-C18 reversed-phase column
(Waters Associates). The column was maintained at 38 °C in a column
temperature compartment (Du Pont Instruments) to improve separation.
The mobile phase consisted of solution A (250 mL of acetonitrile and
750 mL of 20 mmol/L phosphate buffer, pH 3.0) and solution B (700 mL of
acetonitrile and 300 mL of 20 mmol/L phosphate buffer, pH 6.5) that
formed the following gradient: 04.5 min, 3% B; 4.55 min, 30% B;
512 min, 30% B; 1212.5 min, 100% B; 12.514.5 min, 100% B;
14.515 min, 3% B. The flow rate was 1.2 mL/min, and the compounds
were quantified in parallel by absorbance at 254 and 215 nm. The
Gyncosoft program, Ver. 5.42 (Gyncotek), was used for recording and
calculating the data and also for recording the UV spectra when
required. Calculations were made in the internal standard mode, using
peak area ratios. Calibration and quality control were performed by use
of in-house prepared solutions that were analyzed with each run.
Calibration and control materials for total MPA and MPAG were prepared
by adding the analytes to drug-free plasma. MPA was calibrated at a
concentration of 3 mg/L; MPAG was calibrated at 200 mg/L. Solutions
containing 0.2, 1.0, and 25.0 mg/L MPA and 10, 50, and 250 mg/L MPAG
were used for control of those analytes. A protein-free NaCl (9 g/L)
solution adjusted to pH 7.4 with phosphate buffer (67 mmol/L) and to
which 0.05 mg/L MPA was added was used for calibration of the MPAf
determination. For accuracy control of MPAf measurements, 0.025 and 0.1
mg/L MPA were added to the NaCl solution. The precision of MPAf
measurement was established by use of drug-free plasma to which 1 and 5
mg/L MPA were added.
assessment of performance characteristics
The detection limit of the method was calculated using a
signal-to-noise ratio of 3. For this purpose, the noise signal of the
baseline was obtained from a segment of the chromatogram that preceded
each MPA or MPAG peak, respectively. The linearity of the method was
established using drug-free plasma with MPA and MPAG added to yield
concentrations from 0.04 to 50 mg/L for MPA and 1500 mg/L for MPAG.
Within-run and between-run precision, as well as extraction efficiency,
were studied with drug-free plasma to which MPA and MPAG were added to
yield final MPA concentrations of 0.2, 1, 5, and 25 mg/L and final MPAG
concentrations of 10, 50, and 250 mg/L.
Extraction efficiency was calculated by comparing peak areas obtained from the extracted plasma samples with added MPA or MPAG with peak areas obtained with acetonitrile/water (80:20, by volume) containing the same amounts of MPA and MPAG, which were directly injected onto the column without extraction. To obtain information on the accuracy of the method, including the influence of the plasma protein concentration, the recovery was determined by adding known amounts of MPA (0.2, 1, 5, and 25 mg/L) and MPAG (10, 50, and 250 mg/L) to drug-free plasma pools containing 41, 67, and 91 g/L total protein. The percentage of recovery was calculated by comparing the measured concentrations with the expected concentrations.
Potential chromatographic interference by commonly administered drugs was evaluated by analysis of patient specimens received for routine therapeutic drug monitoring, including transplant patients under immunosuppressive therapy without MMF, human therapeutic drug monitoring quality-control sera (Ciba Corning Diagnostics), and drug-containing methanol standards. In addition, the existence of endogenous chromatographic interferences was evaluated by separate analysis of 60 patient specimens, including transplant recipients without MMF therapy, sent to the laboratory for routine clinical chemical tests.
method comparison
Twenty-five plasma specimens derived from pharmacokinetic studies
in renal transplant recipients were used for method comparison with the
HPLC procedure that has been used for the acquisition of extensive
pharmacokinetic/pharmacodynamic data on MPA and MPAG
(17)(18). Frozen plasma specimens in which MPA
and MPAG concentrations had been determined by the method of Tsina et
al. (17) at the Philadelphia laboratory were shipped to
Göttingen. After the samples were thawed, they were analyzed
using the new HPLC procedure without prior knowledge of the results
from Philadelphia.
statistics
The nonparametric regression procedure of Passing and Bablok
(19) was used for comparison of the two HPLC methods
(EVAPAK, Ver. 2.08, Boehringer Mannheim). The regression equations are
given in Results, together with the 95% confidence
intervals for the estimates of slope and intercept in parentheses. The
dispersion of the residuals are documented as the 68% median distance.
For comparison, the standard deviation of the residuals,
Sy
x, calculated using the standard principal
component procedure, is also given.
| Results |
|---|
|
|
|---|
|
|
The detection limit (signal-to-noise ratio, 3) at 215 nm for plasma samples was 0.01 mg/L for total MPA and 0.03 mg/L for MPAG. The corresponding values for detection at 254 nm were 0.03 mg/L for total MPA and 0.1 mg/L for MPAG. In the case of the MPAf concentration, even lower detection limits of 0.005 mg/L at 215 nm and 0.015 mg/L at 254 nm could be achieved, because of the lower background noise in the chromatograms associated with the use of ultrafiltrates.
The linearity of the assay was verified up to 50 mg/L MPA and up to 500
mg/L MPAG (correlation coefficient >0.999). Performance
characteristics of this method were tested at different concentrations
of MPA and MPAG (Table 2
). In the working ranges for MPA (0.225 mg/L) and MPAG
(10250 mg/L), the extraction efficiency was 5659% for MPA and
7377% for MPAG (n = 3). The extraction efficiency for MPAC was
65% (n = 3). The analytical recovery of MPA was in the range of
95106%; for MPAG, analytical recovery was 96106% (Table 3
). There were no major differences in the recovery of MPA and
MPAG, using plasma specimens with low, normal, and increased total
protein concentrations (Table 3
).
|
|
The within-day CVs for measurement of MPAf were 11.8% (at 17.9 µg/L) and 6.8% (at 210.1 µg/L) at 254 nm and 11.6% (at 18.4 µg/L) and 6.5% (at 209.2 µg/L) at 215 nm (n = 12). The between-day CVs were 15.8% (at 22.8 µg/L) and 7.5% (at 197.2 µg/L) at 254 nm and 14.6% (at 23.2 µg/L) and 7.2% (200.6 µg/L) at 215 nm (n = 12).
Very good agreement (Fig. 2
A) was observed between the plasma MPA concentrations determined
by our new HPLC procedure [y] and those obtained with the
method of Tsina et al. [x] (17):
y = 1.086 (1.0421.119)x 0.224
(0.0660.323), 68% median distance = 0.136 mg/L,
Sy
x = 0.157 mg/L. This was also true for the
comparison of plasma MPAG concentrations (Fig. 2B
) determined with the
two procedures: y = 1.033 (0.9731.075)x
1.016 (-5.9358.896), 68% median distance = 6.278 mg/L,
Sy
x = 11.387 mg/L.
|
The results obtained using the new HPLC procedure for samples from an
international MPA proficiency testing scheme are presented in Table 4
. Excellent agreement was found between the measured MPA
concentrations in the samples and the target concentrations.
|
The applicability of the method was proven by analyzing MPA and MPAG
concentrations in >1500 plasma samples and MPAf in >600 samples
collected from liver and kidney transplant recipients. The
concentration-time profile of MPA and MPAG in a pediatric kidney
recipient who was treated with 0.6 g/m twice per day
MMF and who is participating in an ongoing pharmacokinetic study is
shown in Fig. 3
. The determination of MPAf in 305 plasma samples drawn for
pharmacokinetic profiles in children after kidney transplantation
revealed a median for protein-bound MPA of 98.1% (range, 92.9% to
99.3%). MPAf concentrations below the respective detection limits were
observed in 27 of 305 (8.7%) samples at 215 nm and in 38 of 305
(12.6%) at 254 nm.
|
| Discussion |
|---|
|
|
|---|
The sensitivity of this procedure allows the quantification of MPAf concentration after separation of the protein-bound MPA, using the ultrafiltration procedure described by Nowak and Shaw (12). The observation of these authors that binding of MPA to the ultrafiltration system is insignificant was confirmed in the present investigation (data not shown). The sample preparation described with acetonitrile/perchloric acid/sodium tungstate (5:1:1, by volume) leads to an extraction efficiency of >56% and >72% with MPA and MPAG, respectively. The efficiency for MPA can be improved by increasing the volume of acetonitrile, with a concomitant decrease in the recovery of MPAG. The conditions described here represent a compromise to extract both MPA and MPAG simultaneously in one extraction step. With a detection limit of 0.03 mg/L at 254 nm and 0.01 mg/L at 215 nm for MPA, the extraction efficiency was sufficient for all plasma specimens investigated thus far. There was no need, therefore, to modify the protocol. Our equipment allowed us to measure the peaks at both 254 and 215 nm in parallel. Analysis of the spectra under our assay conditions revealed three absorption peaks at 215 nm, 251 nm, and 305 nm for MPA, whereas MPAG had three absorption peaks at 215 nm, 251 nm, and 294 nm. Both compounds displayed their absorption maximum at 215 nm. Therefore, it is also possible to follow MPA and MPAG at either 254 nm or 215 nm. The detection at 215 nm was superior to that at 254 nm, which has been used exclusively before. In our experience, measurement at 215 nm not only improved the detection limit and the sensitivity of the method but also the specificity. Because of the lower detection limit, detection at 215 nm is also recommended for determination of MPAf.
By changing the gradient conditions to 04 min, 3% B; 44.5 min, 26% B; 4.57.5 min, 26% B; 7.59 min, 35% B; 914.5 min, 35% B; 14.515.5 min, 100% B; 15.518 min, 100% B; 1818.5 min, 3% B, MMF can also be measured using this method (retention time ~10 min; extraction efficiency, 60%; data not shown). In addition, when no MPAG determination is required, MPA alone can be separated using isocratic conditions (mobile phase: 450 mL acetonitrile/550 mL 20 mmol/L phosphate buffer, pH 4.5), a flow rate of 1.2 mL/min, and a chromatography run time of 13 min.
The results show that the presented method is a reliable and convenient procedure for the quantification of plasma concentrations of total and free MPA, as well as of MPAG.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. Jain, R. Venkataramanan, R. Sharma, T. Kwong, P. Abt, M. Orloff, R. Kashyap, G. Tsoulfas, and A. Bozorgzadeh Pharmacokinetics of Mycophenolic Acid in Live Donor Liver Transplant Patients vs Deceased Donor Liver Transplant Patients J. Clin. Pharmacol., May 1, 2008; 48(5): 547 - 552. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Brandhorst, F. Streit, S. Goetze, M. Oellerich, and V. W. Armstrong Quantification by Liquid Chromatography Tandem Mass Spectrometry of Mycophenolic Acid and Its Phenol and Acyl Glucuronide Metabolites Clin. Chem., October 1, 2006; 52(10): 1962 - 1964. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Cussonneau, M. Bolon-Larger, and R. Boulieu Adsorption of Mycophenolic Acid and Its Phenolic Glucuronide to Glass, Polystyrene, and Polypropylene Containers. Clin. Chem., May 1, 2006; 52(5): 904 - 906. [Full Text] [PDF] |
||||
![]() |
H. Tian, J. Ou, S. C. Strom, and R. Venkataramanan PHARMACOKINETICS OF TACROLIMUS AND MYCOPHENOLIC ACID ARE ALTERED, BUT RECOVER AT DIFFERENT TIMES DURING HEPATIC REGENERATION IN RATS Drug Metab. Dispos., March 1, 2005; 33(3): 329 - 335. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Streit, M. Shipkova, V. W. Armstrong, and M. Oellerich Validation of a Rapid and Sensitive Liquid Chromatography-Tandem Mass Spectrometry Method for Free and Total Mycophenolic Acid Clin. Chem., January 1, 2004; 50(1): 152 - 159. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. T. Weber, M. Shipkova, V. W. Armstrong, N. Wagner, E. Schutz, O. Mehls, L. B. Zimmerhackl, M. Oellerich, and B. Tonshoff Comparison of the Emit Immunoassay with HPLC for Therapeutic Drug Monitoring of Mycophenolic Acid in Pediatric Renal-Transplant Recipients on Mycophenolate Mofetil Therapy Clin. Chem., March 1, 2002; 48(3): 517 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. T. Weber, M. Shipkova, V. W. Armstrong, N. Wagner, E. Schuutz, O. Mehls, L. B. Zimmerhackl, M. Oellerich, and B. Tonshoff The Pharmacokinetic-Pharmacodynamic Relationship for Total and Free Mycophenolic Acid in Pediatric Renal Transplant Recipients: A Report of the German Study Group on Mycophenolate Mofetil Therapy J. Am. Soc. Nephrol., March 1, 2002; 13(3): 759 - 768. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Shipkova, V. W. Armstrong, M. G. Kiehl, P. D. Niedmann, E. Schutz, M. Oellerich, and E. Wieland Quantification of Mycophenolic Acid in Plasma Samples Collected during and Immediately after Intravenous Administration of Mycophenolate Mofetil Clin. Chem., August 1, 2001; 47(8): 1485 - 1488. [Full Text] [PDF] |
||||
![]() |
P. Morissette, M. Beaulieu, and B. Vinet A Specific Artificial Antibody toward Mycophenolic Acid Prepared by Molecular Imprinting Clin. Chem., September 1, 2000; 46(9): 1516 - 1518. [Full Text] [PDF] |
||||
![]() |
M. Shipkova, E. Schutz, V. W. Armstrong, P. D. Niedmann, M. Oellerich, and E. Wieland Determination of the Acyl Glucuronide Metabolite of Mycophenolic Acid in Human Plasma by HPLC and Emit Clin. Chem., March 1, 2000; 46(3): 365 - 372. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Schutz, M. Shipkova, V. W. Armstrong, E. Wieland, and M. Oellerich Identification of a Pharmacologically Active Metabolite of Mycophenolic Acid in Plasma of Transplant Recipients Treated with Mycophenolate Mofetil Clin. Chem., March 1, 1999; 45(3): 419 - 422. [Full Text] [PDF] |
||||
![]() |
M. Shipkova, V. W. Armstrong, T. Schneider, P. D. Niedmann, E. Schutz, E. Wieland, and M. Oellerich Stability of Mycophenolic Acid and Mycophenolic Acid Glucuronide in Human Plasma Clin. Chem., January 1, 1999; 45(1): 127 - 129. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |