Department of Cardiology, University Hospital Eppendorf, Martinistrasse 52, 20246 Hamburg, Germany.
a Author for correspondence. Fax 49-40 4717 3986; e-mail heeschen{at}uke.uni-hamburg.de.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Measurements of creatine kinase isoenzyme MB (CK-MB) activity are rapid and inexpensive but suffer from a low cardiac specificity and poor sensitivity to detect minor myocardial injury (2)(4)(9) . CK-MB measurements, even when determined immunologically, appear to be inadequate for clinical decision-making in the emergency room (9)(10)(11)(12) .
Cardiac troponins have recently been introduced as cardiac markers (12)(13)(14)(15)(16)(17)(18)(19) . Only cardiac troponin I (cTnI) has not yet been detected in skeletal muscle cells at any stage of development or pathological condition (15)(18)(19) . cTnI can be distinguished by monoclonal antibodies recognizing the amino acid sequence distinct for cardiac muscle cells (17)(18)(19)(20) . cTnI determination promises higher diagnostic efficacy and potency for new diagnostic options (12)(13)(14) :
A recently developed rapid assay for cTnI can be performed at the point of care within 15 min by medical or paramedical staff and might facilitate risk stratification of patients in emergency rooms (12)(21) . Therefore, we compared its analytical performance with two established ELISA systems.
| Materials and Methods |
|---|
|
|
|---|
We added 200 µL of heparinized whole blood or centrifuged plasma to
the device. After separation of cellular blood components from the
plasma fraction by a glass fiber fleece, the migrating plasma dissolves
a buffer and solubilizes the adsorbed antibodies. The antibodies and
the patient's cTnI molecule form sandwich complexes, which migrate to
the signal zone and accumulate in the reading window by means of
biotin-streptavidin interaction. Positive results (cTnI
0.1 µg/L;
manufacturer-defined discriminator) are indicated by a color line that
appears within 15 min.
The unreacted indicator antibodies pass on and bind to the control line
consisting of solid phase anti-mouse IgG antibodies (
0.2 µg). The
appearance of this control line distal to the signal line confirms
impeccable test function, including unimpeded plasma flow.
quantitative cardiac marker determination
The two identical indicator antibodies, but as Fab fragments, were
used on a semiautomated one-step cTnI ELISA system on the basis of
"big bead" technology performed on the Cobas EIA®
(Roche Diagnostic Systems)
(17)(23)(24) . After incubation of
the serum samples with 100 µL of conjugate and the coated bead, the
exceeding conjugate was removed by radial elution. The second
incubation period was performed with 250 µL of solution of
3,3',5,5'-tetramethyl-benzidin and hydrogen peroxide converted into a
colored product by horseradish peroxidase-labeled antibodies. The
reaction was terminated by adding 1000 µL of stopping solution (50
mL/L sulfuric acid) to allow cTnI quantification photometrically. The
lower detection limit of this assay system is 0.03 µg/L
(17) , and a discriminator value of 0.1 µg/L was used. For
calibration, six calibrators containing purified recombinant cTnI
(Spectral Diagnostics), ranging from 0 to 6.0 µg/L, were used.
Interassay precision was 6.2% at 0.2 µg/L and 5.5% at 2.4 µg/L.
In parallel, cTnI and CK-MB mass were quantified by means of a
commercially available analyzer system (Access®
analyzer, Beckmann Diagnostics). It is based on chemiluminescence
technology, and the capture antibody is fixed to magnetic
microparticles. The turnaround time is 18 min. The detection limit for
cTnI is 0.03 µg/L. Values
0.1 µg/L were considered to be positive
(25) , and day-to-day precision (CV) was 8.1% at 0.2 µg/L
and 4.6% at 2.4 µg/L. The detection limit for CK-MB mass is 0.15
µg/L, and the upper reference is 5.0 µg/L (26) . The
interassay CVs were 8.4% at 8.2 µg/L and 7.2% at 14.7
µg/L.
Catalytic activity of total CK was measured colorimetrically in the emergency laboratory at room temperature (Hitachi 717, Boehringer Mannheim) and with different cutoffs for women (70 U/L) and men (80 U/L) according to the manufacturer's specifications. All biochemical analyses were performed by technicians unaware of the patients' histories.
evaluation of test performance
Blood samples were collected from 10 patients with acute
myocardial infarction and 5 patients with minor myocardial injury (high
frequency count ablation, unstable angina, and myocarditis). After
quantitative determination of cTnI two times by Cobas EIA, the samples
were diluted stepwise with human serum from healthy volunteers (cTnI
concentration
0.03 µg/L) down to a cTnI concentration below the
analytical sensitivity of the Cobas EIA system.
Testing of each dilution step was performed five times separately by blinded investigators (n = 6) who were trained in the assessment of the rapid test. The cutoff for the rapid assay was calculated as the cTnI concentration determined with the Cobas EIA (identical antibodies) detected in all trials by each of the blinded observers.
To evaluate the effects of temperature, rapid tests were duplicated on devices heated to 37 °C, cooled to 4 °C, and frozen to -20 °C for 7 days.
clinical evaluation
The study population was recruited from the emergency room of the
University Hospital Hamburg between September 1, 1995 and October 31,
1997. All of these patients had chest pain lasting 12 h or less
and were stratified into four groups of consecutive patients: group 1,
159 patients with acute myocardial infarction, which was defined in the
absence of ST-segment elevations as an increase of total CK activity
within 24 h after onset of chest pain higher than twice the upper
limit of normal associated with increased CK-MB; group 2, 321 patients
with angina at rest associated with ST-T-segment aberrations. Patients
had unstable angina according to the Braunwald classification III B
(53%) + C (47%) (27) ; group 3, 37 patients with cardiac
chest pain of nonischemic origin related to acute congestive heart
failure, perimyocarditis, ventricular arrhythmia, or rejection after
heart transplantation who had no marked coronary stenosis (
70%
diameter stenosis); and group 4, 120 patients with acute chest pain but
no evidence of acute myocardial injury. These patients had no
ST-segment changes and underwent treadmill testing, stress
echocardiography, or angiogram, providing no abnormalities. During 30
days of follow-up, no cardiac event was documented in this group.
Furthermore, we investigated 27 asymptomatic patients (ages, 34 ± 12 years) with end-stage renal failure but no history of coronary heart disease recruited from the department of nephrology at the time of periodic hemodialysis.
statistical Analysis
All results for continuous variables are expressed as means
± SDs (range), and for comparison between two groups, the
MannWhitney U-test was used. Comparison of categorical
variables (positive/negative) were generated by the McNemar test
(two-sided) with appropriate degrees of freedom. The calculations were
done with SPSS 6.1 (SPSS, Inc.), and P values <0.05 were
considered statistically significant.
Analytical results were considered concordant if the rapid assay was interpreted as positive and the quantitative determination by ELISA was above cutoff or if rapid assay was interpreted as negative and the ELISA result was below cutoff. Otherwise, the result was categorized as discordant (false-negative/false-positive).
The diagnostic sensitivity was calculated as the number of positive test results among all patients with acute myocardial infarction observed, and the diagnostic specificity was defined as the number of negative test results among all patients with excluded acute myocardial infarction.
| Results |
|---|
|
|
|---|
|
All cTnI values
0.14 µg/L (Cobas EIA; n = 302) were
categorized as positive rapid test results. For cTnI values between
0.06 and 0.14 µg/L, an increasing rate of positive test results was
found. Only two samples with cTnI values <0.06 µg/L showed a
positive rapid test result (0.03 and 0.04 µg/L) and have to be
considered as false-positive. One test kit indicated malfunction by
absence of a control signal, despite obviously adequate plasma
migration.
Intensity and speed of color development were apparently proportional
to the cTnI concentration. In samples with cTnI amounts
4 µg/L, a
positive test signal developed within 5 min, and samples with cTnI
concentrations
0.4 µg/L could not be rated earlier than 15 min
after serum application (Fig. 2
). To prevent false-positive test results, the time interval
from blood application to test assessment should not exceed 15 min.
|
interobserver variability
The rapid test results of a stepwise dilution series revealed from
patients with myocardial injury were categorized by six blinded
observers who were trained but varied in experience (501000 test
assessments). In the range from 0.06 to 0.14 µg/L, significant,
(P = 0.02) interobserver variability was observed (Fig. 3
). At a cTnI concentration of 0.1 µg/L, four observers rated
the test as positive and two as negative. The discordance correlated
with the observer's experience in assessment of rapid assays.
Experienced observers rated all tests at cTnI concentrations of
0.08 µg/L as positive, whereas trained but unexperienced observers
rated all of these tests as negative (P < 0.01).
Additionally, insufficient artificial illumination and hemolytic or
lipemic samples increased the incidence of discrepant classifications.
All test results with cTnI concentrations above 0.14 µg/L were
considered positive by all observers, independent of experience.
|
temperature stability
Incubation of 15 test devices at 37 °C for 7 days did not
influence the analytical sensitivity of the test system (0.10 vs 0.11
µg/L for incubated devices). Similarly, lowering the temperature of
20 tests devices down to 4 °C was not associated with a decrease of
analytical sensitivity (0.10 vs 0.10 µg/L for cooled devices).
However, the freezing of 20 test devices at -20 °C for 7 days did
affect the sensitivity (0.10 vs 0.45 µg/L for frozen devices;
P < 0.01).
clinical evaluation
Patients with acute myocardial infarction (n = 159).
In
159 patients with acute myocardial infarction, the first determination
of the cardiac markers was performed 3.5 ± 2.7 (0.512) h after
onset of symptoms. A positive cTnI bedside test was found in 60% of
the patients, with increased cTnI concentrations with the Cobas EIA in
58% and with the Access analyzer in 54% of the patients (not
significant for both).
Four hours later, at least one positive bedside test was obtained in 98%. Increased quantitative cTnI concentrations with the Cobas EIA were found in 94% and in 92% of the patients with the Access analyzer (not significant for both).
Immunoassay determination of CK-MB revealed increased values at arrival
in 48% (P = 0.004 compared with cTnI rapid assay) and
for conventional CK activity in 36% of the patients (P
<0.001). Four hours later, in 91% and 61% of the patients, CK-MB
mass (P = 0.03) and CK enzyme activity
(P < 0.001), respectively, were increased, which was
equal to the diagnostic sensitivity of the cTnI rapid device at the
time of arrival (60%) (Fig. 4
).
|
Patients with unstable angina (n = 321).
Among 321 patients
with unstable angina, 122 patients (38%) had at least one positive
cTnI test result within 4 h after arrival and 7.4 ± 2.9
h after onset of symptoms, respectively (Fig. 5
). Positive results were confirmed by quantitative determination
in 92% of the patients for Cobas EIA. Only one patient with a negative
cTnI rapid assay had a cTnI value of 0.12 µg/L (Cobas EIA) and 0.09
µg/L (Access). In 98% of the cTnI-positive patients, a substantial
coronary stenosis could be detected by angiogram. Increased CK-MB mass
was found in 13 patients (4%), and in 7 patients (2%), CK enzyme
activity was increased (81134 U/L).
|
Patients with cardiac chest pain of nonischemic origin (n =
37).
Patients had acute congestive heart failure (n = 15),
perimyocarditis (n = 11), ventricular arrhythmia (n = 8), or
rejection after heart transplantation (n = 3). In patients with
congestive heart failure (73% ischemic), eight positive rapid test
results (53%) were documented, but none of them had marked coronary
stenosis. In seven patients with perimyocarditis (58%), a positive
test result was documented, and all patients with ventricular
arrhythmia and rejection of transplanted heart had increased cTnI
concentrations (Fig. 5
). All positive test results were confirmed by
Cobas EIA and in 95% by the Access analyzer.
Patients with chest pain of noncardiac origin (n = 120).
Patients had acute pulmonary embolism (n = 5), acute
pleuritis (n = 7), pneumonia (n = 12), musculoskeletal
syndrome (n = 51), or traumatic chest pain (n = 45). Positive
cTnI tests were found in two patients with fulminant pulmonary embolism
and echocardiographic signs of right ventricular overloading (Fig. 5
).
Both were confirmed by Cobas EIA and Access analyzer. No cTnI was
detected in the other patients, both by rapid device and ELISA. CK-MB
mass was increased in 32 (71%) patients with traumatic chest pain, in
3 (6%) patients with musculoskeletal syndrome, and in 4 (33%)
patients with pneumonia.
Patients with end-stage renal failure (n = 27).
In
27 patients with end-stage renal failure but no history of coronary
heart disease, 53 blood samples were collected before and after planned
hemodialysis. No positive bedside tests were detected, and likewise no
increased cTnI concentrations were determined by ELISA (Fig. 5
). CK-MB
mass (9.5 ± 7.2 µg/L) was increased in 8 patients (30%), and
CK activity (128 ± 56 U/L) was increased in 12 patients (44%).
| Discussion |
|---|
|
|
|---|
The major advantage of a rapid test system is the applicability of whole blood anticoagulated by heparin so that additional time-consuming preparations such as clotting and centrifugation can be omitted. The performance of the rapid test system at the point of care by medical or paramedical personnel avoids additional time delay and possible sample damage or confusion. But the absence of a quantitative test result represents a disadvantage because positive correlation between the amount of troponin release and cardiac risk has been shown (8)(28) . Although the speed of color development and the intensity of the test signal apparently depends on the cTnI concentration in the blood sample, the rapid device provides at best a semiquantitative test result. However, no different therapeutic strategies according to the troponin concentration have been established thus far (12)(28) , favoring cTnI rapid testing for rapid decision-making and in facilities without access to quantitative troponin measurements.
The analytical sensitivity of the rapid assay is influenced by factors other than the analytical characteristics of the test device, as well. The variability of visual assessment as to the presence or absence of the signal line at low troponin concentrations represents a critical factor in clinical routine. The cutoff defined in our analysis as the cTnI concentration leading to positive test results by all observers in all repetitive samples was settled at 0.14 µg/L. However, we could demonstrate that between 0.10 and 0.14 µg/L, the sensitivity depends on the users' training status. Therefore, a careful acquaintance with training samples is recommended. The highly experienced personnel achieved a detection limit of 0.08 µg/L cTnI without an increase in false-positive test results.
The rapid test device revealed a distinct improvement to CK-MB mass determination with higher diagnostic sensitivity for the detection of acute myocardial infarction (60% vs 48%; 3.5 ± 2.7 h after onset of pain; P <0.01). The cTnI determination is expected to reveal absolute myocardial specificity because cTnI is not expressed in fetal and healthy or diseased adult human skeletal muscle tissue, promising no false-positive test results in patients with substantial skeletal muscle damage or renal failure (29)(30)(31)(32)(33) . The investigated cTnI test systems produced no positive results in patients with end-stage renal failure and acute or chronic skeletal muscle injury, whereas 30% and 71% of the patients, respectively, had increased CK-MB mass concentrations.
Diagnostic specificity, however, for the detection of acute myocardial infarction according to WHO criteria was lower than that of CK-MB because of an marked rate of positive cTnI test results in patients with unstable angina (38% vs 4% for CK-MB mass; P <0.001). The low incidence of patients with increased CK-MB concentrations might be related to strict exclusion of all patients with non-Q-wave acute myocardial infarction. Additional positive cTnI test results were documented in patients with acute congestive heart failure, myocarditis, ventricular arrhythmia, and rejection of transplanted heart. Recent studies indicate that, independent of the diagnosis of myocardial infarction, the detection of troponins is associated with increased subsequent cardiac risk; therefore, these positive test results are not to be rated as false-positive (12)(13)(14) .
In conclusion, the use of this cTnI rapid device could improve efficacy and safety of decision-making in patients with chest pain that might produce more cost-effective use of intensive care facilities. Future prospective studies must define the role of the new diagnostic marker in patient management to improve their adverse prognosis (12)(34) .
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
F. Peng, Z. Wang, S. Zhang, R. Wu, S. Hu, Z. Li, X. Wang, and D. Bi Development of an Immunochromatographic Strip for Rapid Detection of H9 Subtype Avian Influenza Viruses Clin. Vaccine Immunol., March 1, 2008; 15(3): 569 - 574. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Heeschen Biomarkers in acute coronary syndromes and their role in diabetic patients Diabetes and Vascular Disease Research, October 1, 2005; 2(3): 122 - 127. [Abstract] [PDF] |
||||
![]() |
M. T. Sandri, D. Cardinale, L. Zorzino, R. Passerini, P. Lentati, A. Martinoni, G. Martinelli, and C. M. Cipolla Minor Increases in Plasma Troponin I Predict Decreased Left Ventricular Ejection Fraction after High-Dose Chemotherapy Clin. Chem., February 1, 2003; 49(2): 248 - 252. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. W. Hamm and E. Braunwald A Classification of Unstable Angina Revisited Circulation, July 4, 2000; 102(1): 118 - 122. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Heeschen, B. U. Goldmann, L. Langenbrink, G. Matschuck, and C. W. Hamm Evaluation of a Rapid Whole Blood ELISA for Quantification of Troponin I in Patients with Acute Chest Pain Clin. Chem., October 1, 1999; 45(10): 1789 - 1796. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||