(Clinical Chemistry. 1998;44:1942-1946.)
© 1998 American Association for Clinical Chemistry, Inc.
|
Drug Monitoring and Toxicology |
Evaluation of the tacrolimus II microparticle enzyme immunoassay (MEIA II) in liver and renal transplant recipients
Jan L. Cogill1,
Paul J. Taylor1,a,
Ian S. Westley2,
Raymond G. Morris2,
Stephen V. Lynch3,
and Anthony G. Johnson1
1
The University of Queensland Department of Medicine, Princess Alexandra Hospital, Brisbane, Queensland, Australia 4102.
2
Department of Clinical Pharmacology, The Queen Elizabeth
Hospital, Adelaide, South Australia, Australia 5011.
3
The University of Queensland Department of Surgery,
Princess Alexandra Hospital, Brisbane, Queensland, Australia 4102.
a Address correspondence to this author at: Centre for Clinical and Experimental Therapeutics, First Floor, Lions Clinical Research Building, Princess Alexandra Hospital, Ipswich Road, Brisbane, QLD, Australia 4102. Fax 61 7 3240 5031; e-mail Ptaylor{at}gpo.pa.uq.edu.au.
 |
Abstract
|
|---|
We evaluated the MEIA II with blood samples with added tacrolimus (3.0,
5.0, 11.0, and 22.0 µg/L). The assay had acceptable recoveries
(99103%) and intraday imprecision (<16.0%) across the range of
concentrations studied, except for the recoveries at 3.0 µg/L
(86.3%) and 5.0 µg/L (80.7%). Comparison of liver (n = 116)
and renal (n = 113) patient samples measured by MEIA II against
HPLC-tandem mass spectrometry (HPLC-MS/MS) found a mean overestimation
of 15.6%. From these comparison data it can be calculated that at
values of 5 and 20 µg/L in liver or renal transplant patient samples,
measured by HPLC-MS/MS, MEIA II will have the corresponding range
estimates of 3.67.9 µg/L and 20.925.4 µg/L, respectively. No
clinically significant difference in results, in terms of
overestimation or correlation, was observed between the two transplant
groups studied. The MEIA II is an improvement on the previous MEIA I
and is suitable for the therapeutic drug monitoring of tacrolimus where
HPLC-MS/MS is unavailable.
Key Words: TDM, therapeutic drug monitoring MEIA, microparticle enzyme immunoassay MS, mass spectrometry MS/MS, tandem mass spectrometry.
 |
Introduction
|
|---|
Tacrolimus (FK506, Prograf®), a potent
immunosuppressive agent, was introduced into the clinic in the late
1980s for use as both primary and rescue therapy in patients receiving
solid organ transplants (1) . The dose-related efficacy and
toxicity, narrow therapeutic index, cytochrome P450-mediated drug
interactions, and considerable interpatient variability in
pharmacokinetics of tacrolimus (2) has led to the consensus
that therapeutic drug monitoring
(TDM)4 of steady-state trough tacrolimus concentrations is required
(3) . To date, the most frequently used methodologies for the
TDM of tacrolimus have been immunoassay-based, either ELISA or
microparticle enzyme immunoassay (MEIA) (4)(5)(6)(7) . A recent
survey reported by Holt and Johnston (8) on the European
Tacrolimus Quality Assessment Scheme showed that all but one of the 74
participating members were using immunoassay-based methods. These
immunoassays use an anti-tacrolimus monoclonal antibody that recognizes
not only the parent drug but also several of its metabolites
(9)(10) . Several studies have reported that, in
different patient groups, these immunoassays show overestimation in
tacrolimus concentrations when compared with methods specific for the
parent drug (11)(12)(13)(14)(15) .
Tacrolimus has been shown to exhibit a potency of 100-fold greater than
cyclosporin (16) , and thus tacrolimus dosage is reduced
correspondingly. The resulting blood concentrations (low µg/L) make
specific measurement of tacrolimus difficult. However, Gonschior et al.
(17) utilized HPLC-mass spectrometry (HPLC-MS) to measure
tacrolimus and three of its metabolites simultaneously in blood and
urine. Recently, we developed a HPLC-tandem mass spectrometry
(HPLC-MS/MS) assay for the analysis of tacrolimus in blood that is
specific, sensitive, and rapid (18) . Under the guidelines
recommended by Büttner (19) , both HPLC-MS and
HPLC-MS/MS are considered reference procedures for the TDM of
tacrolimus. The use of HPLC-MS, however, is limited by its lack of
widespread availability and initial capital cost.
Recently, a tacrolimus II MEIA (MEIA II) has become available. The
lower analytical range of this method, compared with the previous
version, MEIA I (7) , may provide an alternative to HPLC-MS
in the TDM of tacrolimus. In this study, we determined the accuracy and
imprecision of the MEIA II, using blood containing weighed-in amounts
of tacrolimus. In addition, we evaluated the performance of the MEIA II
in the clinical setting against a HPLC-MS/MS assay, using blood samples
obtained from two patient groups, liver and renal transplant
recipients.
 |
Materials and Methods
|
|---|
meia ii
The MEIA II was performed on an IMx analyzer®
according to the manufacturer's instructions (Abbott Laboratories).
Tacrolimus calibrators and controls were used as supplied by Abbott.
The IMx analyzer was calibrated (0, 3.0, 6.0, 12.0, 20.0, and 30.0
µg/L) before sample analysis. A series of controls (5.0, 11.0, and
22.0 µg/L) were run with each batch to confirm assay integrity
throughout the study. Sample preparation involved blood samples (150
µL) pretreated with a methanol solution containing zinc sulfate and
ethylene glycol (150 µL) in 1-mL polypropylene centrifuge tubes.
These samples were vortex-mixed and centrifuged for 5 min at
10 600g. The supernatant was dispensed into the respective
reaction cells, and the automated analysis was performed.
hplc-ms/ms procedure
HPLC-MS/MS analysis was performed according to a modified method
of our previously reported assay (20) . Briefly, sample
preparation consisted of an initial organic solvent precipitation
(containing an internal standard: FR900520, a tacrolimus analog,
Fujisawa Pharmaceutical Co.) of blood samples (500 µL), followed by a
C18 solid phase cartridge extraction. The extracted samples
were chromatographed using a C8 column, 2 x 100 mm
(Brownlee Laboratories) at ambient temperature. The mobile phase
consisted of 200 mL of 40 mmol/L ammonium acetate buffer (pH 5.1) and
800 mL of methanol per liter and was pumped at a flow rate of 400
µL/min. The flow was split postcolumn at a ratio of 1:12 into the
mass spectrometer. Mass spectrometric detection was via an electrospray
interface using multiple reactant ion monitoring (tacrolimus
m/z 821.5
768.4; internal standard m/z
809.5
756.4). For each batch of samples analyzed, a six-point
calibration curve (0, 3.0, 6.0, 12.0, 20.0, and 30.0 µg/L) was
constructed. A weighted 1/X2 regression analysis was used
to fit the calibration curves.
assessment of meia ii against hplc-ms/ms
As part of patients' routine clinical care, blood samples were
collected by venipuncture 12 h after the last oral dose of
tacrolimus, just before the next dose, into tubes containing EDTA. The
total number of samples collected from 50 liver and 43 renal transplant
patients was 116 and 113, respectively. A maximum of five samples was
collected from any one patient. All samples were stored at -18 °C
and assayed within 3 months of collection.
statistical analysis
To assess the accuracy and imprecision of MEIA II compared with
HPLC-MS/MS, three controls (5.0, 11.0, and 22.0 µg/L) were assayed in
quadruplicate over 4 days by both methods. Accuracy was calculated as
the ratio of the mean concentration (n = 12) over the known
"weighed-in" concentration, expressed as a percentage. Imprecision
was determined from the analysis of variance, using the methods of
Krouwer and Rabinowitz (21) . To investigate assay
performance at the low end of the therapeutic range, the
manufacturer's calibrator B (3.0 µg/L) was analyzed in replicate on
1 day by MEIA II (n = 10) and HPLC-MS/MS (n = 5). The
acceptable limit of accuracy was defined as ±15% of the
weighed-in concentration and ±20% at the limit of quantitation.
Similarly, the acceptable limit of imprecision is defined as CV <15%,
except for the limit of quantitation, where it should be <20%
(22) . Regression analysis and the methods described by Bland
and Altman (23) were used to compare the results obtained by
the two methods for the liver and renal transplant groups.
 |
Results
|
|---|
A summary of the analytical performances of the MEIA II and
HPLC-MS/MS methods at three concentrations (5.0, 11.0, and 22.0 µg/L)
is shown in Table 1
. The total imprecision of MEIA II was acceptable (<13.0%)
across the range of controls studied. The recovery of the MEIA II was
acceptable at 11.0 µg/L (102.5%) and 22.0 µg/L (99.7%); however,
the recovery at 5.0 µg/L (80.7%) was unacceptable. The intraday
imprecision and recovery determined at 3.0 µg/L, using the
manufacturer's calibrator B, for MEIA II were 15.3% and 86.3%,
respectively. These data compare with a total imprecision of <4.0%
and a recovery range of 99.5100.6% for the HPLC-MS/MS method over
the range of controls studied. The imprecision and recovery at 3.0
µg/L for the HPLC-MS/MS were 2.5% and 102.2%, respectively.
The tacrolimus concentrations measured by both methods for all the
transplant patient samples studied are shown in Fig. 1
. The equation for the line of best fit is described by the
equation MEIA II = 1.16 (± 0.0184) x HPLC-MS/MS - 0.00561 (±
0.163) µg/L (Sy
x = 1.12, n = 229). Fig. 2
shows the above data as the difference in tacrolimus
concentrations (MEIA II minus HPLC-MS/MS) against the mean tacrolimus
concentrations of both methods. The line of best fit is described by
the equation MEIA II minus HPLC-MS/MS = 0.179 (± 0.0155) x Mean
Concentration - 0.261 (± 0.149) µg/L
(Sy
x = 1.03, mean bias = 1.26 µg/L). From
the 95% confidence intervals it can be calculated that concentrations
of 5 and 20 µg/L, measured by HPLC-MS/MS, will give corresponding
estimates for MEIA II of 3.67.9 µg/L and 20.925.4 µg/L,
respectively. When the data for the two transplant groups were examined
independently, little difference, in terms of correlation, was
observed. The mean overestimations for the liver and renal transplant
groups were 13.1% and 18.2%, respectively.

View larger version (18K):
[in this window]
[in a new window]
|
Figure 1. Tacrolimus blood concentrations in liver ( , n =
116) and renal ( , n = 113) transplant recipients, measured by
MEIA II and HPLC-MS/MS).
The solid line represents the line of identity (i.e.,
slope = 1).
|
|
 |
Discussion
|
|---|
Although the recommendations of the Lake Louise International
Consensus Conference on Immunosuppressive Drugs called for the use of
specific methods in the TDM of tacrolimus ((3)), the majority
of centers use ELISA and MEIA (8) . This is partially because
accurate measurement of tacrolimus in blood using traditional HPLC
detection is not possible due to the drug's lack of intrinsic
fluorescence, end absorption in the ultraviolet spectrum
(24) , and circulating concentrations are low compared with
cyclosporin. The development of reliable HPLC-MS methods has provided
one means for the specific measurement of tacrolimus
(17)(18) . The major problem to overcome for the
routine use of HPLC-MS is the specialized and expensive
instrumentation. The instrumentation required for ELISA and MEIA are
less expensive and are found routinely in clinical laboratories.
However, the consumable costs of running HPLC-MS analysis is the same
or less than traditional HPLC and also less than the cost of the
commercial kits for the MEIA II. Finally, the simple sample preparation
and rapid turnaround time make the MEIA II very attractive to the
clinical laboratory.
The current therapeutic ranges of tacrolimus for liver and renal
transplant recipients are in the range of 520 µg/L (25) .
The original MEIA I assay was found to have insufficient sensitivity to
monitor patients with tacrolimus concentrations <5 µg/L
(13) . Therefore, the development of a MEIA II method with a
lower analytical range is of considerable interest to transplantation
centers measuring tacrolimus. Our investigation of the analytical
performance of the MEIA II, using known weighed-in controls, found the
assay to be accurate and precise except for the recovery at 5.0 µg/L
(80.7%), although the mean value of 4.0 µg/L obtained for the 5.0
µg/L control was considered acceptable under the manufacturer's
guidelines (3.07.0 µg/L) (26) . Furthermore, the MEIA II
assessed using the manufacturer's calibrator B (3.0 µg/L) gave a
recovery and imprecision within the guidelines of Shah et al.
(22) for the limit of quantitation. Overall, the analytical
performance of MEIA II, determined using weighed-in controls, was
superior to our previously reported evaluation for MEIA I
(15) .
Several groups have reported increased tacrolimus concentrations in
patient samples measured by immunoassays compared with results obtained
by specific methods (11)(12)(13)(14)(15) . In this study, we found that
the overestimation was similar for both transplant groups. For the
liver and renal transplant groups, the maximum percentages of
overestimation (mean) were 47.8% (13.1%) and 117.9% (18.2%),
respectively. These data compare with the work of Firdaous et al.
(13) , in which the reported overestimation for the liver and
renal transplant groups ranged from 18% to 48%. The clinical
importance of this overestimation in tacrolimus concentrations is yet
to be determined, although Braun et al. (14) reported a case
study of a renal transplant patient with ongoing rejection, in which
the difference in concentrations measured by MEIA I (median, 10.5
µg/L), ELISA (median, 7.92 µg/L), and HPLC-MS (median, 2.93 µg/L)
could be considered clinically important.
In conclusion, the MEIA II is an improvement on the previously
available MEIA I and is suitable for the TDM of tacrolimus where
HPLC-MS/MS is not available. The random sample selection for our study
was an attempt to obtain a cross-section of our total population;
however, determining the clinical importance of the observed
overestimation in results will require longitudinal studies with large
numbers of patients in all transplant populations.
 |
Acknowledgments
|
|---|
This project was supported in part by the Princess Alexandra
Hospital Research and Development Foundation. The MEIA II reagents for
this study were a kind gift from Janssen-Cilag Australia. We thank all
the hospital staff who cared for the patients.
 |
References
|
|---|
-
Steinmuller DR. FK506 and organ transplantation 1994:19-74 CRC Press Boca Raton, FL. .
-
Peters DH, Fitton A, Plosker GL, Faulds D. Tacrolimus. A review of its pharmacology, and therapeutic potential in hepatic and renal transplantation. Drugs 1993;46:746-796.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Jusko WJ, Thomson AW, Fung J, McMaster P, Wong SH, Zybler-Katz E, et al. Consensus document: therapeutic monitoring of tacrolimus (FK-506). Ther Drug Monit 1995;17:606-614.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Tamura K, Kobayashi M, Hashimoto K, Kojima K, Nagase K, Iwasaki K, et al. A highly sensitive method to assay FK-506 levels in plasma. Transplant Proc 1987;5(Suppl 6):23-29.
-
Wallemacq PE, Firdaous I, Hassoun A. Improvement and assessment of enzyme-linked immunosorbent assay to detect low FK506 concentrations in plasma or whole blood within 6 hours. Clin Chem 1993;39:1045-1049.
[Abstract/Free Full Text]
-
D'ambrosio R, Jusko WJ, Stabler T, Kobayashi M. Improved procedures for enzyme immunoassay for tacrolimus (FK506) in whole blood [Letter]. Clin Chem 1994;40:159-160.
[Free Full Text]
-
Grenier FC, Luczkiw J, Bergmann M, Lunetta S, Morrison M, Blonski D, et al. A whole blood FK 506 assay for the IMx analyzer. Transplant Proc 1991;23:2748-2749.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Holt DW, Johnston A. Tacrolimus quality assessment [Letter]. Ther Drug Monit 1997;19:243.[Web of Science][Medline]
[Order article via Infotrieve]
-
Iwasaki K, Shiraga T, Nagase Z, Tozuka Z, Noda K, Sakuma S, et al. Isolation, identification, and biological activities of oxidative metabolites of FK506, a potent immunosuppressive macrolide lactone. Drug Metab Dispos 1993;21:971-977.
[Abstract]
-
Iwasaki K, Shiraga T, Matsuda H, Nagase K, Tokuma Y, Hata T, et al. Further metabolism of FK506 (tacrolimus). Identification and biological activities of the metabolites oxidised at multiple sites of FK506. Drug Metab Dispos 1995;23:28-34.
[Abstract]
-
Gonschior A-K, Christians U, Braun F, Winkler M, Linck A, Baumann J, Sewing K-F. Measurement of blood concentrations of FK506 (tacrolimus) and its metabolites in seven liver graft patients after the first dose by h.p.l.c.-MS and microparticle enzyme immunoassay (MEIA). Br J Clin Pharmacol 1994;38:567-571.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Warty V, Zuckerman S, Venkataramanan R, Lever J, Chao J, Mckaveney T, et al. Tacrolimus analysis: a comparison of different methods and matrices. Ther Drug Monit 1995;17:159-167.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Firdaous I, Hassoun A, Otte JB, Reding R, Squifflet JP, Besse T, Wallemacq PE. HPLC-microparticle enzyme immunoassay specific for tacrolimus in whole blood of hepatic and renal transplant patients. Clin Chem 1995;41:1292-1296.
[Abstract/Free Full Text]
-
Braun F, Lorf T, Schütz E, Christians U, Grupp C, Sattler B, et al. Clinical relevance of monitoring tacrolimus: comparison of microparticle enzyme immunoassay, enzyme-linked immunosorbent assay, and liquid chromatography mass spectrometry in renal transplant recipients converted from cyclosporine to tacrolimus. Transplant Proc 1996;28:3175-3176.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Salm P, Taylor PJ, Clark A, Balderson GA, Grygotis A, Norris RLG, et al. High-performance liquid chromatography-tandem mass spectrometry as a reference for analysis of tacrolimus (FK506) to assess two immunoassays in patients with liver and renal transplants. Ther Drug Monit 1997;19:694-700.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Kino T, Inamura F, Sakai K, Nakahara K, Goto T, Okuhara M, et al. Effect of FK-506 on human mixed lymphocyte reaction in vitro. Transplant Proc 1987;19(Suppl 6):36-39.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Gonschior A-K, Christians U, Winkler M, Schiebel HM, Linck A, Sewing K-F. Simplified high-performance liquid chromatography-mass spectrometry assay for measurement of tacrolimus and its metabolites and cross-validation with microparticle enzyme immunoassay. Ther Drug Monit 1995;17:504-510.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Taylor PJ, Jones A, Balderson GA, Lynch SV, Norris RLG, Pond SM. Sensitive, specific quantitative analysis of tacrolimus (FK506) in blood by liquid chromatography-electrospray tandem mass spectrometry. Clin Chem 1996;42:279-285.
[Abstract/Free Full Text]
-
Büttner J. Reference materials and reference methods in laboratory medicine: a challenge to international cooperation. Eur J Clin Chem Clin Biochem 1994;32:571-577.
[Web of Science][Medline]
[Order article via Infotrieve]
-
Taylor PJ, Hogan NS, Lynch SV, Johnson AG, Pond SM. An improved method for the therapeutic drug monitoring of tacrolimus (FK506) using tandem mass spectrometry [Technical Brief]. Clin Chem 1997;43:2189-2190.
[Free Full Text]
-
Krouwer JS, Rabinowitz R. How to improve estimates of imprecision. Clin Chem 1984;30:290-292.
[Abstract]
-
Shah VP, Midha KK, Dighe S, McGilveray IJ, Skelly JP, Yacobi A, et al. Analytical methods validation: bioavailability, bioequivalence, and pharmacokinetic studies. J Pharm Sci 1992;81:309-312.
-
Bland JM, Altman DG. Statistical methods for assessing agree-ment between two methods of clinical measurement. Lancet 1986;i:307-310.
-
Tanaka H, Kuroda A, Marusawa H, Hashimoto M, Hatanaka H, Kino T, et al. Physiochemical properties of FK-506, a novel immunosuppressant isolated from Streptomyces tsukubaensis. Transplant Proc 1987;5(Suppl 6):11-16.
-
McMaster P, Mirza DF, Ismail T, Vennarecci G, Patapis P, Mayer AD. Therapeutic drug monitoring of tacrolimus in clinical transplantation. Ther Drug Monit 1995;17:602-605.
[Web of Science][Medline]
[Order article via Infotrieve]
-
. Abbott Laboratories. Reagent pack insert for Tacrolimus II 1996:4 Abbott Laboratories Abbott Park, IL. .