(Clinical Chemistry. 1998;44:1959-1963.)
© 1998 American Association for Clinical Chemistry, Inc.
Multivariate approach to quality control in clinical chemistry
Jerry Decherta,
and Kenneth E. Case
a Author for correspondence. Fax 405-325-7555; e-mail dechert{at}mailhost.ecn.ou.edu.
 |
Abstract
|
|---|
When monitoring analyzer performance in the clinical setting,
laboratories are required to test multiple concentrations of control
material on a daily basis. Because of the nature of laboratory testing,
there is the potential for correlation between the concentrations of
control material being monitored. Although traditional clinical
quality-control approaches make an underlying assumption of
independence with respect to the control concentrations, this will not
always be the case. The presence of correlation in some circumstances
suggests the use of a new approach for evaluating clinical laboratory
monitoring data: the multivariate control chart. Such a chart (the
2 chart) is evaluated and compared with traditional
quality-control approaches used in the laboratory setting. Results
indicate that the multivariate approach provides an attractive
alternative to many traditional methods of quality assurance when
control concentrations are correlated.
 |
Introduction
|
|---|
Many approaches have been used for the purpose of monitoring and
controlling clinical laboratory testing over the years. In fact, the
original application of statistical monitoring of clinical quality
control dates back to the 1950s (1) . This initial
application by Levey and Jennings involved testing duplicates of each
concentration of control material and plotting the results against
± 3 SD limits. Although a number of alternative methods have been
proposed in the literature, they all focus on the individual
concentrations of control material and fail to acknowledge the
multivariate nature of the monitoring problem. This paper will explore
the use of a multivariate quality-control chart for application in the
clinical laboratory and compare its performance with some traditional
clinical monitoring approaches.
Although the original application of statistical monitoring dates back
to Levey and Jennings, a number of alternative approaches have been
identified over the years (2)(3) . Probably the
most renowned alternative is Westgard's Multirule Procedure
(4) . This approach reduces the high false rejection rate
that can accompany the strict use of ± 2 SD limits and provides
good sensitivity to clinically significant shifts, but it requires
multiple quality-control testing points (i.e., quality-control batches)
to achieve this sensitivity.
Recent developments in clinical quality control have focused attention
on using quality-control systems designed to detect clinically
significant changes in the measurement system
(5)(6) . As the observed variability in assay
systems continues to decrease, understanding clinically significant
errors and designing quality-control systems to specifically detect
these errors will become more and more important. Therefore, it is
necessary to identify quality-control approaches that have
predetermined average run lengths (ARLs) for detecting specifically
defined changes in the quality-control system. A multivariate
quality-control chart is an approach with that capability.
 |
Correlation in Clinical Quality-Control Data
|
|---|
The most compelling reason to examine multivariate quality-control
charts for monitoring in the clinical setting is the potential
correlation structure in laboratory quality-control data. When testing
control materials, laboratories will typically test two or three
concentrations of control materials at the same time (i.e., low, mid,
and high). The result is that there is often correlation between
control concentrations within an analytical run. The low, medium, and
high controls are typically related because they are affected by
essentially the same sources of variability at the same time. Examples
of sources of variation leading to correlation within an analytical run
include laboratory temperature and humidity, calibration curve shape,
electrical fluctuations in analyzers, and technician or pipetting
errors (particularly in pretreatment assays) to name a few.
It is not our purpose here to perform a definitive study with respect
to correlation in the clinical laboratory. Our experience, however,
provides undeniable evidence of correlation in the clinical setting.
Specifically, we have collected data from clinical chemistry and drug
monitoring assays and evaluated correlation in these data sets. The
data used in the evaluation included actual laboratory monitoring data
along with some data generated according to NCCLS EP-5 protocol for
evaluating precision. Estimated correlation coefficients between
concentrations across four assays were 0.05, 0.40, 0.67, and 0.70,
respectively. From this small number of assays, it is clear that the
presence of correlation in testing data is real; however, the degree of
correlation may vary from situation to situation.
In addition to using the aforementioned data for evaluating correlation
within an analytical run, the data were evaluated for the presence of
correlation from run to run within a control concentration.
Autocorrelation plots were analyzed for these data, and these plots
indicate no significant correlation from run to run within one
concentration of control material.
Because there are many possible sources of variability leading to
correlation, it is incumbent on laboratories to check quality-control
data for the presence of correlation. By calculating the sample
correlation coefficient r, a laboratory can readily quantify
the degree of correlation between concentrations. If substantial
correlation is evident, then the laboratory should consider the use of
quality-control approaches that will properly account for correlation
in the data.
Given the possibility of correlation in control data, the next step is
to determine its impact on the performance of quality-control methods.
If one considers the instance of two control materials, then the
difference between assuming independent and correlated control
concentrations can be demonstrated graphically, as shown in Fig. 1
. In the case where independence applies, the in-control region
determined by UCL = (mean + k SD) and LCL = (mean - k
SD) presents itself as a square. In the case where correlation is
present, the true in-control region presents itself as an ellipse
(7) . This means that assuming independence of control
concentrations and applying ± k SD limits as such when the
control concentrations are truly correlated will produce areas of
over-control and under-control, as shown in Fig. 1
. An over-control
area is the region in which independent control limits would indicate a
shift in the control materials when no true shift has occurred (i.e.,
commit an alpha error). The under-control area is the region in which
independent control limits would fail to indicate a true shift in the
control materials (i.e., commit a beta error).
As stated previously, the purpose here is not to definitively
characterize the nature of correlation between control concentrations
within a given analytical run. The purpose is to acknowledge the real
existence of this correlation in many instances and to provide a means
for dealing with this correlation. A method for dealing with
correlation, the
2 chart, is detailed in the next
section.
 |
The 2 Chart
|
|---|
Although a number of multivariate approaches are available for
quality-control application, the
2 chart
(8)(9) was used for this research. This
selection was based on the fact that the
2 chart assumes
a known covariance matrix, the same assumption made for evaluating
traditional quality-control approaches. In practice, a T2
chart (10)(11) can be used in the case of an
unknown covariance structure with results similar to the
2 chart.
Assuming a known covariance matrix for the control concentrations being
monitored, application of the
2 chart is very
straightforward. There is only an upper limit for the
2
chart, and it is based on the
2 distribution. The upper
limit for the chart is

,p2, where
p is the number of parameters being monitored and
is the
probability of a type I error (i.e., the probability of the
2 chart signaling a shift in the control materials when
none has occurred). In the clinical application, p would be
the number of control concentrations being monitored (typically two or
three). Given the upper limit for the chart, the statistic plotted is
simply n(x -
µ0)/
-1(x -
µ0), where n is the sample size (typically n = 1 for
the clinical application), x is the observed sample mean
vector (observed values from the quality-control batch),
µ0 is the original mean vector (control concentration
targets or historical means), and
-1 is the inverse of
the covariance matrix of the control concentrations. A simple algebraic
formula for calculating this statistic for two concentrations of
control material is as follows:
 |
 |
where x1 and x2 are
observations for control concentrations I and II, µ1 and
µ2 represent historical means for control concentrations
I and II, and
1 and
2 represent the SDs
of control concentrations I and II, respectively, with
12 the covariance between concentrations I and II. By
plotting this calculated statistic against the upper limit of the
2 chart, one can determine if the measurement system has
shifted.
 |
Method Comparison
|
|---|
To understand the performance of a multivariate approach, it is
necessary to compare such an approach to traditional methods of
clinical quality control. For the purposes of this research,
traditional methods of quality control include the strict application
of a single limit set at ± 2.75 SD limits
(12.75s), the use of an immediate retest for limits
set at ± 1.91 SD (21.91s), and Westgard's Multirule
Procedure as originally published. Although typical approaches would be
the strict application of ± 2 SD limits (12s)
and ± 2 SD limits with an immediate retest (22s),
these rules have been replaced with 12.75s and
21.91s to match the in-control ARL of the Westgard
Multirule Procedure. The multivariate approach chosen for application
is the
2 chart with alpha = 0.0119 to also match
the in-control ARL of the Westgard Multirule Procedure.
The comparison of the three traditional methods and the
2 chart is made for the typical clinical
application of N = 2 (i.e., n = 1, p = 2 from
the multivariate perspective). However, different shifts in the control
materials are considered in this research from those that generally
appear in the literature. The comparisons are made for the usual
instance when both concentrations of control material shift together
(i.e., concentration I and concentration II both shift the same
direction at the same time), denoted as N = 2/2. Additionally,
comparisons are made for the case when a single concentration of
control shifts while the other concentration of control remains on
target (i.e., concentration I experiences a shift while concentration
II stays centered or vice versa), denoted as N = 2/1, and the case
where the two control concentrations diverge (i.e., concentration I
shifts upward while concentration II shifts downward), denoted as
N = 2/-2. Fig. 2
graphically portrays the shifts investigated in this research.
Although all these shifts are not typically evaluated in the
literature, they may easily occur in practice and should be
considered.
In addition, the comparisons of the quality-control methods are made
based on ARLs, where the ARL is simply the number of quality-control
batches tested (on average) before the quality-control method will
signal. In the case where no change has occurred in the measurement
process, the ARL corresponds to the average number of quality-control
batches until a false rejection. In the case where a true shift has
occurred in the measurement process, the ARL corresponds to the average
number of quality-control batches until a true detection. The reason
for applying ARLs for comparison purposes as opposed to probabilities
is that the Westgard Multirule Procedure involves information over
multiple quality-control batches. Adding a run rule that requires four
quality-control batches (i.e., 41s) for evaluation in
no way improves the probability of detecting a true shift with the
first sample evaluated after the shift occurs. It does, however, reduce
the average number of quality-control batches required to detect the
shift in the long run. Therefore, the ARL is used for comparison
purposes in this research.
To generate the ARL information, the calculations for
12.75s and 21.91s are straightforward
probability calculations. The ARLs for the Westgard Multirule Procedure
are determined using computer simulation. For the
2
chart, ARLs are generated by integrating the non-central
2 distribution (12) . Additionally, a
correlation structure must be specified to evaluate the
2 chart, and the correlation between the concentrations
is assumed to be
= 0.60 and held constant for all analysis in
this research. Here, we consider only two concentrations of control
material; however, the modeling can be extended to the instance of
three concentrations of control with similar results.
The results from the statistical modeling are shown in Figs. 3
5. Fig. 3
shows the ARLs for the case when two concentrations
of control are both shifted together (N = 2/2). Note that both
means are shifted in terms of SDs (i.e., both control materials
experience a 1.0 SD shift in their means). Figs. 4
and
5 show the ARLs when the control concentrations are shifted
independently (i.e., N = 2/1 and N = 2/-2).
Examination of the graphs yields some valuable insights. The first is
that the Westgard Multirule Procedure is the most sensitive to changes
in both concentrations of control materials in the same direction
(N = 2/2). The other traditional approaches show performance close
to the Westgard Multirule Procedure, with the
2
chart being the least sensitive to this type of shift. However, the
2 is more sensitive to the other two types of shifts
considered in the research, N = 2/1 and N = 2/-2. In fact,
the
2 chart substantially outperforms the other
approaches for the case of diverging control concentrations.
Although the traditional methods have reduced sensitivity to a shift in
a single concentration of control or two diverging control
concentrations, the performance of the multivariate approach actually
improves. This is because of the nature of the in-control ellipse for
the
2 chart. For a shift common to two
concentrations of control, the direction of the shift moves along the
axis of the in-control ellipse. However, for a shift in a single
concentration of control, the direction of the shift moves away from
the axis of the ellipse, making the approach more sensitive to these
kinds of shifts. For two diverging concentrations of control material,
the
2 chart is even more sensitive to the change because
of the direction of the mean vector shift vs the in-control ellipse.
These results are also important from the standpoint of the perceived
error protection for the traditional quality-control approaches. When
the ARLs are examined for the different types of mean shifts, it is
clear that the error protection capabilities for the traditional
methods decrease. For the multivariate case, the error protection for a
shift in a single concentration or diverging concentrations actually
improves. This means that a
2 chart developed for the
case of a shift in both materials will provide that minimum degree of
error protection regardless of the type of shift encountered. This is
not the case for traditional quality-control approaches.
To further evaluate the performance of the
2 chart,
the sensitivities to changes in precision for all the quality-control
approaches are evaluated in Fig. 6
. The chart shows the ARLs for detection of changes in the
control concentration SDs by multiples of the SD. Only the case where
both SDs are inflated is considered. From an examination of the chart,
it is clear that all the methods considered have essentially the same
sensitivities to changes in random error or precision. Therefore, none
of the approaches has a particular advantage over the other methods in
terms of detecting changes in random error.
 |
Conclusion
|
|---|
The existence of correlation between control concentrations in
quality-control data is very real. Although this correlation may not
always be large and certainly varies from assay to assay, its presence
can play a large role in the performance of a selected quality-control
approach. To deal with correlated control concentrations in clinical
quality control, a multivariate approach has been proposed. This
approach utilizes the correlation structure of multiple concentrations
of control materials to determine the appropriate in-control region for
the monitoring application.
There are a number of advantages to the use of the multivariate
approach for monitoring clinical quality control. The first is that the
approach provides the appropriate control region for the application.
If the assumption of independence does not hold, then the assumed
performance of traditional approaches can be misleading. The
multivariate approach, however, can guarantee error protection for a
variety of different types of shifts in the control materials.
Another advantage of the
2 chart is that it moves
away from the application of runs rules. There is an inherent
uneasiness with detecting shifts that occurred four or five
quality-control batches ago and shutting down the system. What about
all the patient samples in between that were subjected to the same
shift in the measurement process? With the
2 chart, each
quality-control batch would be either acceptable or unacceptable on its
own merit. There is still a risk that a shift could go undetected;
however, that risk does not change from quality-control batch to
quality-control batch, and that risk is considered explicitly in the
design of the
2 chart.
Given the potential for correlation in clinical quality-control data,
it would be advisable for laboratories to check for correlation in
their data. If the assumption of independence holds, then traditional
methods should be continued. If substantial correlation is evident,
then a multivariate approach, such as the
2
chart, should be applied to the quality-control application.
 |
Footnotes
|
|---|
School of Industrial Engineering, University of Oklahoma, 202 West Boyd, Suite 124, Norman, OK 73019.
 |
References
|
|---|
-
Levey S, Jennings ER. The use of control charts in the clinical laboratory. Am J Clin Pathol 1950;20:1059-1066.[Web of Science][Medline]
[Order article via Infotrieve]
-
Westgard JO, Groth T, Aronsson T, de Verdier C. -H. Combined Shewhart-Cusum control chart for improved quality control in clinical chemistry. Clin Chem 1977;23:1881-1887.[Abstract/Free Full Text]
-
Parvin CA, Gronowski AM. Effect of analytical run length on quality-control (QC) performance and the QC planning process. Clin Chem 1997;43:2149-2154.[Abstract/Free Full Text]
-
Westgard JO, Barry PL, Hunt MR. A multi-rule Shewhart chart for quality control in clinical chemistry. Clin Chem 1981;27:493-501.[Free Full Text]
-
Linnet K. Choosing quality-control systems to detect maximum clinically allowable analytical errors. Clin Chem 1989;35:284-288.[Abstract/Free Full Text]
-
Petersen PH, Fraser CG. Setting quality standards in clinical chemistry: can competing models based on analytical, biological, and clinical outcomes be harmonized?. Clin Chem 1994;40:1865-1868.[Free Full Text]
-
Jackson JE. Quality control methods for two related variables. Ind Qual Control 1956;1:1-4.
-
Alt FB. Encyclopedia of statistical sciences 1985:110-122 John Wiley & Sons New York. .
-
Tracy ND, Young JC, Mason RL. Multivariate control charts for individual observations. J Qual Technol 1992;24:88-95.
-
Alt FB. Multivariate quality control: state of the art. Am Soc Qual Control Qual Congr Trans 1982;:886-893.
-
Jackson JE. Multivariate quality control. Commun Statistics 1985;14:2657-2688.
-
Anderson TW. An introduction to multivariate statistical analysis 1958:72-77 John Wiley & Sons New York. .