The main developments in the "ligand assay" field in which I have been involved are traced. These include the original development of "first generation" competitive assays relying on radiolabeled analyte markers; the development of the first "second generation", noncompetitive (ultrasensitive) methods, which rely on the use of labeled (monoclonal) antibodies and high specific activity nonisotopic labels (leading to the transformation of the immunodiagnostic field in the 1980s); and the development of the first "third generation" miniaturized, chip-based, microarray methods, which permit the simultaneous ultrasensitive measurement of many analytes in the same small sample. The latterapplicable both to immunoassay and to DNA/RNA analysisare likely to revolutionize the diagnostic and pharmaceutical fields in the next decade.
Ligand Assays: General Concepts
It seemed to me to be appropriate, in the first of the Edwin F. Ullman Award presentations, to overview some of the key developments in the "ligand assay" field since its origins in the mid-1950s, particularly those in which I myself have been considerably involved. However, in a historical account of restricted length, it is clearly impossible to do justice to the work of the many scientists who have contributed to the field; this presentation's primary intention is therefore to identify and clarify only the main concepts that have driven methodological advances in this area since its origins and that underlie those occurring at the present time.
The term ligand assay is generally used to describe the class of assays that rely on observation of the product(s) of the binding reaction between the analyte and a specific binding reagent, the latter comprisingin the case of immunoassaysan antibody directed against antigenic sites on the analyte molecule. Early immunoassays depended on direct observation of the binding reaction product, this typically taking the form of a precipitate of the antibody-antigen complex. Likewise, many so-called "bioassays" have relied on the binding reaction between the analyte and specific receptors present in cells and tissues. However, the term ligand assay as generally understood, and as used in this article, is restricted to assays relying on high specific activity labels (e.g., radioisotopic labels) to directly reveal the products of the binding reaction, thereby (amongst other benefits) increasing assay sensitivity.
Antibodies are readily prepared against many substances of biological interest, and label-based immunoassays (such as radioimmunoassays) have achieved widespread popularity in the past 30 years. Nevertheless, they conform to analytical principles that govern a number of analogous ligand assays that differ only in the nature of the binding reagent on which they rely. These methods [which exploit the specific binding characteristics of serum and receptor binding proteins, enzymes, single-chain DNA fragments (i.e., oligonucleotide probes), and others] have not hitherto achieved as great prominence or industrial importance as immunoassays; however, burgeoning interest in DNA analysis suggests that oligonucleotide-based ligand assays will prove at least equally important in the future.
The three principal phases in the development of this field, and their
origins, are identified and summarized in Fig. 1
:
|
Phase 1, the independent development in the 1950s (by Yalow and Berson in New York and by me in London) of the first generation "competitive" radiolabeled analyte methods culminating, at the end of the decade, in the publication of the "radioimmunoassay" (RIA) for insulin (1) and of the "saturation assay" for thyroxine (2);
Phase 2a, the development [by Wide et al. (3) in Uppsala and Miles and Hales (4) in Cambridge] of radiolabeled antibody ("immunoradiometric") methods in the late 1960s;
Phase 2b, the collaborative development (by myself and Wallac Oy, commencing in 1976) of the first of a second generation of "ultrasensitive", nonisotopic, noncompetitive labeled monoclonal antibody methods, this being ultimately marketed as "DELFIA" (5); and
Phase 3, the development (by myself and my co-workers, commencing in 1986) of miniaturized, multianalyte "microspot" assays (6)(7), leading, in 19911992, to a collaboration with Boehringer Mannheim GmbH on the development of a third generation of ultrasensitive antibody and oligonucleotide microarray technologies. (Note: a similar long-term projectthe Genosensor Projectwas subsequently launched in the United States in 1992, albeit this is directed solely to DNA/RNA analysis.)
Since its origins, a principal factor driving methodological developments in this area has been the need for greater assay sensitivity and, concomitantly, a reduction in incubation and processing times (implying higher sample throughputs). In these respects, the earliest competitive radioligand assays clearly represented a major advance on methods for the assay of substances of biological importance available in the 1950s, permitting, for example, the determination of many hormones whose concentrations in body fluids (~1010 molecules/L, or 107 10 molecules/mL) lay below the sensitivity limits of contemporary analytical techniques. Because of their specificity, sensitivity, and ease of use, these methods rapidly penetrated into many other areas, e.g., hematology and pharmacology, emerging as the dominant microanalytical methodology used in medicine from the 1960s to the mid-1980s.
Although (radio)labeled antibody (immunoradiometric) assays were first described in the late 1960s, doubts regarding their sensitivities relative to those of "radiolabeled analyte" methods, coupled with the difficulties in, and cost of, producing relatively pure labeled antibodies, led to the virtually universal neglect of this approach by researchers and manufacturers until the mid-1980s. Thus, even the description by Köhler and Milstein (8) in 1975 of the in vitro synthesis of monoclonal antibodiesa technique that removed many of the logistic problems associated with the large-scale production of pure, labeled antibodiesgenerated little immediate interest among researchers and kit manufacturers. Indeed, as late as 1982 and even later, a prominent scientist in the field publicly asserted that monoclonal antibodies offered no substantial benefits, implicitly failing to recognize the potential advantages (in terms of improved sensitivity and reduced incubation times) of the labeled antibody approach. These failures permitted, inter alia, the use of monoclonal antibodies for immunoassay purposes to be successfully patented (in the United States) as late as 1980 [see (9)].
However, after the demonstration (theoretical and experimental) of the improved performance achievable using noncompetitive, labeled (monoclonal) antibody methods (particularly when accompanied by the use of high specific activity nonisotopic labels), the immunodiagnostic field was transformed in the mid-1980s, and virtually all kit manufacturers subsequently adopted this approach. The resulting improvement in assay performance provided the basis for the explosive development of the fast, high-throughput immunoanalyzers that now dominate the field.
The third phase identified in Fig. 1
promises as great an ultimate
transformation of the ligand assay field as followed the emergence of
the ultrasensitive methods in the early 1980s. Although the earliest
studies on the feasibility of an ultrasensitive microarray technology
commenced in my laboratory as long ago as 1986, it represents a far
greater challenge to manufacturers than posed by the earlier methods,
and the development time and costs involved are inevitably greater.
Miniaturized microarray technologies permit, in principle, the
simultaneous, ultrasensitive measurement of tens, hundreds, or even
thousands of substances in a small sample (e.g., a drop or two of
blood). Arrays of high capacity are clearly likely to prove especially
valuable in the context of DNA analysis; however, very large analyte
"menus" are unlikely to be required in the immunodiagnostic field
in the immediate future (although if methods are developed that permit
the isoforms of analytes of heterogeneous molecular composition to be
individually distinguished, a comparable need in this area may
ultimately arise). Nevertheless, even at the present time, miniaturized
microarray technology offers substantial advantages in certain obvious
immunodiagnostic applications (such as in transfusion blood screening
and allergy testing); moreover, the technology, once fully developed,
is likely to offer major cost advantages when the determination of more
than one analyte is required and could, therefore, replace most current
methods even in circumstances where analyte panels of relatively small
size are of diagnostic value.
On the Meaning of Assay "Sensitivity" and "Precision"
As indicated above, the need for high sensitivity underlay the original inception of ligand assay methods and has constituted a major factor driving their ongoing development. However, notwithstanding this word's long-established meaning as indicative of the ability of an assay system to determine (or "sense") low analyte concentrations, differences regarding the concept it represents, and the way in which the sensitivity of an assay system should therefore be correctly assessed, led to early and prolonged dispute regarding RIA design. This topic has been discussed at length in recent publications (10)(11)(12)(13) and hence need not be dwelt on in detail here. Nevertheless, it is appropriate to summarize the issues involved because misunderstandings regarding the assessment of assay performance have played a major part in the history of the ligand assay field.
Certain (although not all) relevant national and international
organizationsincluding, for example, IUPAChave, in recent years,
formally defined the sensitivity of a measuring system in terms of the
response/stimulus ratio or (equivalently) the slope of the
doseresponse curve (14). Of greater importance in the
present context, Berson and Yalow [in their many publications relating
to immunoassay theory, e.g., (15)(16)] likewise
defined the sensitivity of an assay in terms of the slope of the curve
relating the response variable [the bound-to-free ratio (B/F) in their
earlier papers, but later the fraction of labeled analyte bound (b)]
to analyte concentration ([H]), subsequently arguing that the
response curve slope (db/d[H]) is implicitly indicative of the
detection limit of an assay (16). In contrast, Ekins et al.
[e.g., (17)] explicitly defined the
sensitivity of an assay as represented by the (im)precision of
measurement of an analyte concentration of zero (i.e.,
Ho =
Ro/(dR/d[H])o, where R
is the response variable (however expressed) and
Ro is
the standard deviation of R and (dR/d[H])o is the
response curve slope, both at zero dose), this quantity being the key
determinant of the lower detection limit of any measuring system and
thus of the actual ability of the system to determine,
detect, or sense small amounts "of that which it is intended to
measure" (18). The crucial difference between these two
definitions is that the latter specifically requires detailed
statistical analysis of assay data (to determine, either directly or
indirectly, the magnitude of random errors in the measurement of the
response variable, represented by
R). In contrast, the
slope definition implicitly assumes constancy of such errors
irrespective of any alteration in assay design (such as a change in the
antibody concentration used in the system) or the manner in which the
response curve is plotted.
The invalidity of this assumption, and thus the inadequate and
misleading natureindeed untenabilityof the slope definition may be
illustrated in many ways. For example, it may readily be shown (using
either experimental data or the equations shown in Fig. 3
) that
plotting RIA data in terms of B/F implies that assay sensitivity (as
thus defined) is increased by increasing the antibody concentration;
however, the opposite conclusion is reached if the same data are
plotted in terms of F/B. Plotting the curve in terms of b leads to a
third conclusion, i.e., that maximal slope (sensitivity) is achieved
using an antibody concentration binding 33% of a trace amount of
labeled antigen (16) (this concentration being
0.5/K, where K is the affinity constant governing
the antibody/antigen reaction). In short, contradictory conclusions
will be drawn regarding the conditions yielding maximal assay
sensitivity in consequence of plotting identical experimental data in
different coordinate frames (i.e., in terms of different dose or
response variables) while disregarding random errors in the measurement
of the chosen response variable. [It should perhaps be emphasized that
this phenomenon is not a unique feature of RIA but is a characteristic
of all measuring systems. For example, which of two
balances is deemed the more sensitive (slope definition) is likely to
depend on whether the response of each balance to a weight placed on
the pan is expressed in terms of the angular rotation of the beam or
the movement of the pointer tip across the scale.]
|
Misunderstanding of this issue contributed, for example, to initial uncertainties regarding the relative sensitivities of labeled antigen and the (radio)labeled antibody [immunoradiometric (IRMA)] assay methods discussed below. Although the latter were originally claimed (19) to be inherently more sensitive than RIA, the claim was based on a consideration of response curves per se and was not substantiated either theoretically or experimentally. Moreover, its validity was challenged by Rodbard and Weiss who, in 1973, published detailed theoretical studies (20) purportedly demonstrating that both labeled analyte and labeled antibody methods possess essentially equal sensitivities (albeit these authors failed to perceive the crucial distinction between competitive and noncompetitive assay designssee below). Likewise, many observers were initially skeptical of the possibility of developing ultrasensitive microspot ligand assays (as described below in Principle Developments in the Ligand Assay Field) because the use of "vanishingly small" amounts of capture binding agent located on a small area contravenes widely accepted assay design rules stemming from the slope definition of sensitivity.
Although attracting less attention, disagreement also centered on the concept of precision adopted by Berson and Yalow (15)(16). This was similarly defined as the slope of the (b vs log[H]) response curve,1 without reference to the random errors incurred in the response measurement. Assays designed in conformity with this definition do not yield maximal precision according to the widely accepted scientific meaning of this term. In short, theoretical prediction or experimental observation of the effects of a change in assay components on the response curve slope in isolation (i.e., without consideration of concomitant effects on the errors incurred in the response measurement) provides a misleading indication of the effects of the change on assay sensitivity2 and precision and a fallacious basis for assay optimization and design. Nevertheless, an absence of appropriate statistical analysis of assay data has frequently characterized experimental studies in the ligand assay field, initially obscuringamong other consequencesthe potential value of technical developments (such as the in vitro synthesis of monoclonal antibodies) or of new analytical concepts that have ultimately led to revolutionary improvements in assay design and performance.
Principal Developments in the Ligand Assay Field
phase 1: radioimmunoassay and saturation analysis
As indicated in Fig. 1
, the publication of the insulin assay
method of Yalow and Berson (1) essentially coincided with
that of my method for serum thyroxine (T4) (2).
Dr. Yalow has revealed that the insulin assay emerged as a fortuitous
by-product of studies on the immunogenicity of insulin. In contrast,
the T4 assay that I developed represented the outcome of an
explicit quest for a general microanalytical method of sufficient
sensitivity to assay substances (such as hormones) present at very low
concentrations in body fluids. The reasons for, and a brief account of,
this search may be of historical interest.
As a junior member of the Department of Physics Applied to Medicine at the Middlesex Hospital Medical School in London in the 1950s, I had the opportunity to carry out spare-time research relating to thyroid physiology. I also had the good fortune to have as a colleague and friend J.F. Tait, then collaborating with S.A. Simpson (later S.A. Tait) of the Courtauld Institute of Biochemistry on the studies that led to the discovery of aldosterone (21). Both were hampered by the lack of analytical methods of sufficient sensitivity to measure hormone concentrations in blood; this issue therefore constituted an occasional topic of discussion, although Tait and his co-workers succeeded in the early 1950s in applying the "double-isotope derivative analysis" technique to the assay of certain steroids (22), a method thatalthough cumbersome and time-consuming to performultimately permitted many pioneering studies in the steroid field to be carried out.
Reflecting on this problem when walking home one evening in the spring
of 1954, I perceived a possible general solution to it (Fig. 2
3
) and a particular solution applicable to the assay of
thyroxine based on the use (as a "saturable" binding agent) of the
recently described specific thyroxine-binding globulin. Reaching home,
I carried out a simple analysis of the implications of the mass action
laws, deriving equations enabling the form of the doseresponse curves
in a system of this kind to be predicted (Fig. 3
). Regrettably, these concepts were viewed with skepticism by my
peers, and funding for their experimental implementation was denied.
The consequent lack of suitable equipment or means to purchase and
radiolabel the key reagents required led to the events that followed.
|
Although work on the proposed methodology was, for the above reasons,
precluded for the next 3 years, two fortuitous occurrences in 1957
enabled small-scale studies to commence without financial assistance.
The first was the admittance to the Middlesex Hospital of a patient
with thyroid carcinoma metastases who was treated with massive doses of
I throughout the year (i.e., 200 mC at 2-month
intervals), and whose blood radioactivity was being routinely
monitored. In the course of these measurements, I unexpectedly observed
that the radioactive material was largely protein bound, apparently
comprising I-T4. Having confirmed the latter
chromatographically, I constructed simple electrophoretic equipment
with which I was able to confirm the changes in distribution of labeled
T4 between specifically (i.e., thyroxine-binding globulin)
bound and "free" (albumin-bound) moieties following the addition of
varying amounts of exogenous, unlabeled T4, thus enabling
response curves relating the observed distribution to unlabeled
T4 concentration to be derived (Fig. 4
).
|
The second fortuitous event of 1957 was a 6-week visit to my department by A. Harjanne, a clinical chemist from Helsinki, with whom I discussed the principles of the proposed analytical methodology and its requirement for high specific activity labeled hormone preparations. Dr. Harjanne suggested the use of the Conway microdiffusion flask (a device used by clinical chemists to determine iodine in blood) to radioiodinate substances such as thyroxine and proteins to high specific activities, notwithstanding the restricted facilities available. During the next 2 years, using either residual blood from samples occasionally derived from the patient referred to above or T4 labeled by the Conway diffusion method, I slowly obtained sufficient data to merit submission of a paper in 1959 describing the T4 assay. However, it was recommended by my departmental director that the method should not be presented as exemplifying a general analytical principle until it had been applied to another analyte; it was therefore not until 1962having used the same approach to assay serum vitamin B12 (23)that I published a description of "saturation analysis" as a general microanalytical technique (24).
The widespread adoption of this approach (sometimes termed "competitive protein-binding assay", a term that clearly embraces RIA4 ) rapidly followed the publication of the studies by Yalow and Berson (1) and myself (2); however, it is unnecessary to elaborate further on these events, which are well known. The advantages and consequent popularity of the approach owed primarily to the specificity deriving from the "molecular recognition" features characterizing many binding reactions on which living systems rely, coupled with a high sensitivity stemming (in part) from the use of a high specific activity label (such as a radioisotope) to observe the reactions between small numbers of reactant molecules. However, the latter is not the sole factor contributing to the sensitivity of these methods; of equal importance is the choice of binding reagents with appropriate physicochemical properties and of optimal reagent concentrations. This choice is a key feature of ligand assay design, althoughas indicated earlierit emerged as a major topic of controversy arising from differing perceptions of the meanings of assay sensitivity and precision. Disagreement and debate on the meanings of these terms continues (10)(11)(12)(13). Unfortunately, this cannot be brushed aside as a trivial semantic issue of no practical importance; many workers in the field continue to rely on the response curve slope alone as an indicator of the ability of an assay system to determine low analyte concentrations, and select assay reagents and protocols in accordance with this specious concept.
phase 2a: immunoradiometric and other labeled antibody assays
Subsequent to the emergence of the first generation labeled
analyte techniques described above, Wide et al. (3) in
Sweden, followed shortly by Miles and Hales in the United Kingdom
(4), developed in the late 1960s radiolabeled antibody
methods (commonly referred to as immunoradiometric assays, or IRMAs),
the studies by Miles and Hales being distinguished by their development
of immunosorbent techniques for the preparation of relatively pure
labeled antibodies. These methods constituted a reversion to the
earlier "labeled reagent" techniques exploited by Avivi et al.
(22), although they were distinguished therefrom by the use
of reagents (i.e., antibodies) of much higher specificity than the
labeled organic substances (e.g., acetic anhydride and, later, pipsyl
chloride) used by Tait and his co-workers. The apparent motivation of
Miles and Hales for adopting this approach was a belief that the
labeled antibody approach would yield higher sensitivity
(19), although they provided neither experimental nor
rigorous theoretical evidence supportive of the claim, and as indicated
previously, it was later disputed by Rodbard and Weiss
(20).5
One disadvantage of the technique developed by Miles and Hales is a concomitant loss in specificity because of the increase in relative potency of cross-reacting antigens as the antibody concentration used in the system is increased (25); however, this effect is largely annulled in assays of "sandwich", or "two site", design as devised by Wide (26), in which the analyte must be simultaneously recognized by two antibodies (directed against different epitopes on the analyte molecule) for an assay response to be generated.
Largely because of the absence of evidence demonstrating any improvement in performance of labeled antibody assays, coupled with their heavy consumption of antibody, assays of this design were almost totally disregarded by researchers and kit manufacturers for the next 1015 years, albeit a commercial kit for the assay of hepatitis B antigen constituted a notable exception. However, the manufacturer's adoption of the labeled antibody (sandwich) approach in this case appeared to be unrelated to sensitivity considerations but stemmed from the unusually large size of the antigen molecule, which caused difficulties in distinguishing between free and antibody-bound antigen moieties in a conventional RIA. Other exceptions included certain systems based on the use of enzyme labels (27), commonly referred to as enzyme-linked immunosorbent assay (ELISA); however, the primary motivation for their development likewise appears to have been unrelated to a desire to increase sensitivity (such methods commonly are relatively insensitive) but was to avoid the legal and practical impediments to the use of radioisotopically based methods in many laboratories. In short, although labeled antibody methods were first developed in 1967, the improved assay performance that, in certain circumstances, they may yield was largely unrecognized for many years, notwithstanding their occasional adoption for practical and logistic reasons.
phase 2b: ultrasensitive assays
The uncertainty regarding the relative sensitivities of
conventional RIAs and IRMAs stemmed partly from the widespread
misunderstanding of the meaning of sensitivity as indicated above (and
hence of the experimental and statistical procedures required to
determine the ability of a method to measure low analyte
concentrations) and partly from a failure to recognize that the key
distinction in this context is not between labeled antibody and labeled
analyte methods per se but between competitive and noncompetitive assay
designs.6
For example, certain labeled antibody methods can be portrayed
as relying on the competition for a limited amount of labeled antibody
between the analyte and exogenous unlabeled analyte coupled to a solid
support, and can therefore also be described as competitive
(28).
Much of the confusion relating to this issue can be dispelled by viewing the basis of ligand assays in terms of the "binding site occupancy" principle. When a "sensor" or "capture" antibody (or other binding agent) is introduced into an analyte-containing medium, sensor antibody binding sites are occupied by analyte molecules to a fractional extent that reflects both the equilibrium constant governing the binding reaction and the final unbound analyte concentration in the mixture. All immunoassays depend on measurement of the "fractional occupancy" of sensor binding sites by analyte, from which the analyte concentration in the system is deduced. However, occupied binding sites may be measured either "directly" or "indirectly" (i.e., by determination, in the latter case, of unoccupied sites). The statistical error incurred in the measurement of binding site occupancy (a key determinant of the sensitivity of an assay) depends on which of these alternative approaches is adopted, as does the optimal concentration of antibody used in the system.
In short, the crucial distinction between so-called competitive and
noncompetitive methods does not center on which component of the system
(if any) is labeled but stems from the particular approach adopted to
determine binding site occupancy. For example, as shown in Fig. 5
, in single-site labeled antibody assays of the type developed
by Miles and Hales (4), the labeled antibody itself
constitutes the sensor antibody, which after reaction with analyte may
be separated into occupied and unoccupied fractions (using, for
example, an immunosorbent comprising antigen or anti-idiotypic antibody
linked to a solid support). If, after the separation of bound and free
labeled antibody fractions, the signal emitted by labeled antibody
bound to analyte (i.e., the "occupied" antibody fraction remaining
in solution) is measured directly, the assay can be classed as
noncompetitive. Conversely, if labeled antibody unbound to analyte
(i.e., attached to the immunosorbent) is measured, then the assay can
[as exemplified by Yorde et al. (28)] be described as
competitive.
|
Competitive and noncompetitive assay methods differ in many of their performance characteristics (including their potential sensitivities) in consequence of the differing statistical errors incurred in the measurement of binding site occupancy. In general, direct measurement of occupied sites represents the better strategy, generally yielding higher sensitivity. (Analogously, it is preferable to determine a 1-cm length by measuring it directly rather than subtracting measurements of two greater lengths, e.g., 1 m and 99 cm, each of which is subject to error).
These simple concepts are supported by detailed theoretical analysis.
Thus it can be shown (29) that, in a competitive assay
design, the maximum sensitivity (i.e., the lowest attainable detection
limit) that can achieved in a correctly optimized system is given by
/K (where
is the relative "manipulation"
error7
in the measurement of the assay response, and K is
the affinity constant governing the antibody-antigen reaction),
assuming that the label used is of infinite specific activity (i.e.,
that the statistical errors of the kind incurred in counting
radioisotopic disintegrations are zero). In contrast, the corresponding
expression relating to a noncompetitive assay is
/K
x bn (where bn is the fraction of labeled
antibody that is nonspecifically bound) (7). Thus if
bn = 0.01, a noncompetitive assay is potentially 100-fold
more sensitive than a competitive assay, other factors being equal.
However, a different picture emerges when statistical counting errors
are taken into account, as shown in Fig. 6
(7). For example, when a radioisotope such as
I is used to label the antibody, sensitivity is
increasingly constrained (with an increase in the affinity constant of
the labeled antibody) by counting errors consequent on the limited
specific activity of the label. In other words, assuming that
nonspecific binding of labeled antibody is kept to a very low level,
counting errors are likely to replace manipulation errors as the
principal limitation on assay sensitivity in a noncompetitive system.
|
These theoretical considerations indicate that the achievement of substantially higher sensitivities than are attainable by the "first generation" RIA techniques requires (a) use of a noncompetitive labeled antibody assay design, (b) use of a nonisotopic label of much higher specific activity than conventional radioisotopes, and (c) reduction of nonspecific binding of labeled antibody to the minimum practicable level (i.e., at least <1% and ideally <0.01%).
These conclusions led to my immediate recognition (9) of the potential importance (in an immunoassay context) of the Nobel Prize-winning development of in vitro techniques of monoclonal antibody production by Köhler and Milstein (8) and to the collaborative development by Ekins and co-workers (initiated ~1976) with Wallac Oy (then an instrument manufacturer) of the time-resolved fluorometric immunoassay methodology now known as DELFIA (5)(30).8 Based on the use of lanthanide chelate fluorophors and labeled monoclonal antibodies, this was the first commercial ultrasensitive nonisotopic immunoassay methodology to be developed. The same basic approach has subsequently been adopted by many other manufacturers using other high specific activity nonisotopic labels, as recently reviewed by Kricka (36). The use of such labels in immunoassays of noncompetitive design has revolutionized the field in the past decade and underlies continuing attempts to further improve assay sensitivities.
The achievement of high sensitivity is clearly of great clinical importance in the case of certain analytes, for example, thyrotropin, whose serum concentrations in thyrotoxic individuals not only fell below the limit of detection of the original RIA methods in common use in the early 1980s but were essentially indistinguishable from reference values. However, an equally important consequence of the development of ultrasensitive methodologies of this type has been a dramatic reduction in assay performance times, leading to the emergence of the fully automated immunoanalyzers that now dominate the immunodiagnostic field. Total incubation times in the order of minutes are typical, replacing the hours or days characterizing first generation competitive RIA methodologies.
To summarize this section, the use of nonisotopic labels of much higher specific activity than radioisotopes, when combined with noncompetitive strategies of binding site occupancy measurement,9 yield ligand assays of greatly enhanced sensitivity and that require much shorter incubation times than competitive techniques relying on isotopic labels. The general recognition of the advantages of this approach (coupled with the increasing availability of monoclonal antibodies) has caused a major transformation of the immunodiagnostic field in the past decade and is the principal reason for a diminishing commercial interest in the original, isotope-based methods.
Against this historical and conceptual background, we may now turn to the development of ultrasensitive, miniaturized, multianalyte ligand assays based on microspot arrays.
phase 3: miniaturized, microarray-based, multianalyte
assays
The need to determine analyte panels in blood and other biological
fluids has become increasingly evident in fields such as endocrinology,
where resolution of certain diagnostic problems may require
determination of the concentrations of several different hormones.
However, this need is even more conspicuous in other areas, such as
allergy testing, forensic medicine, the screening of transfusion blood
for viral and other contaminants, and environmental monitoring. A
complementary objective currently attracting much interest is assay
miniaturization, not only because this permits measurement of many
analytes in the same small sample but because of the concomitant
reduction in assay performance times, the possibility it offers of
"point-of-care" testing, and many other practical benefits.
Traditional methods of combining immunoassays within the same system have relied on the use of different labels permitting a number of individual binding reactions to be distinguished. However, the scope for extending the number of labels used in this way is limited. In contrast, spatial isolation of many immunoassays in the form of an array on a solid "probe" offers much greater flexibility, andin macroscopic formhas been adopted as a format in certain commercial multianalyte assay kits. This approach also underlies miniaturized "microarray" technology.
The perception of the possibility of such technology followed from the theoretical prediction that highly sensitive immunoassays can be developed using far smaller amounts of antibody than have traditionally been perceived as necessary, implying that an individual assay can be performed using numbers of antibody molecules so low as to be confinable within a microspot a few square micrometers in area. This permits, in principle, the construction of exceedingly small arrays, each microspot in the microarray being directed against a different analyte.
As indicated earlier, all binding assays rely on the determination of
binding site occupancy. Moreover, it may readily be shown (from the
mass action laws) that the fractional occupancy (F) of a
sensor antibody (or other specific binder) is given by:
![]() |
![]() | (1) |
![]() | (2) |
A close similarity exists between these concepts and the operation of a simple thermometer. When placed in a liquid in a container, a small thermometer extracts heat from the surrounding liquid until it reaches thermal equilibrium, at which point the temperature of the liquid is indicated. Provided the thermal capacity of the thermometer is insignificant, the temperature it finally attains will be independent of its size. However, the introduction of a large thermometer (of high thermal capacity) into the liquid will cause a drop in the temperature of the liquid; thus, in these circumstances, the final temperature recorded will depend, inter alia, on the size of the thermometer and the volume of liquid to which it has been exposed.
Analogously, a small amount of antibody located on a microspot on a solid support [the total number of antibody sites being less than v/K x 10-5 x N, where v is the sample volume (in milliliters) and N is the Avogadro number (6.023 x 10)] provides the basis of an ambient analyte immunoassay system sensing the analyte concentration to which it is exposed. For example, if v = 1 and K = 10 L/mol, the maximum number of binding sites causing negligible disturbance (<1%) to the ambient analyte concentration is 6 x 107 . Assuming antibody surface densities (when closely packed as a monolayer) in the order of 104 105 molecules/µm2 , antibody microspots of areas in the order of 100-1000 µm2 are sufficient to accommodate numbers of molecules falling within ambient analyte immunoassay limits.
After the exposure of a microspot to an analyte-containing sample,
occupancy of antibody located within the spot may be determined by its
exposure to a second (labeled) antibody reactive either with occupied
or unoccupied sites, these representing noncompetitive and competitive
strategies of binding-site occupancy measurement, respectively (Fig. 7
).
|
Moreover, because the system depends on the determination of the
fractional occupancy of sensor antibody binding sites within the
microspot area, a useful stratagem is to label the sensor antibody
itself with a second label and to observe the ratio of signals emitted
by the two labels. The two antibodies may be labeled with, for example,
a pair of radioisotopic, enzyme, or chemiluminescent markers. However,
fluorescent labels are especially useful because they enable arrays of
microspots distributed on the surface of a "chip" or sample holder
base (each microspot directed against a different analyte) to be
optically scanned, thereby permitting multianalyte assays to be
performed on the same sample. Thus, by scanning the array (using, e.g.,
a confocal microscope (Fig. 8
) or charge coupled device camera) the fluorescent signals
emitted from each spot can be determined, revealing the ambient
concentration of analyte against which the antibody in the spot is
directed. These principles underlie the Microspot®
microarray technology currently under collaborative development by
myself together with Boehringer Mannheim GmbH.
|
The proposition that microspot assays conducted in this fashion not only permit many different analytes to be simultaneously determined but may be more sensitive and rapid than conventional systems challenges accepted ideas in the field and needs brief explanation. Although it is often suggested (and appears intuitively obvious) thatto achieve the highest sensitivitymost if not all of the analyte present in a test sample should be captured by the capture (i.e., sensor) antibody (thus maximizing the signal generated from occupied sites), the increase in the background signal consequent on the use of an increased amount of antibody must also be considered. In short, increasing the area of a microspot (while maintaining the antibody surface density constant) decreases its fractional occupancy by analyte (albeit total occupancy, is increased) and increases the background signal generated from the spot. Thus the signal-to-background ratio decreases with increase in spot diameter, this phenomenon being likely to reduce sensitivity.
Clearly, if the microspot area were reduced to zero, no signal (other than instrument "noise") would be observed, and the system would be totally insensitive. In practice, other statistical effects come into operation when the number of individual events (e.g., photons) observed by the detecting instrument is small, prohibiting the reduction of the sensor-antibody concentration below a certain lower limit. An additional constraint is the number of analyte molecules captured by the sensor antibody within the microspot area, whichif exceedingly smallwill also cause statistical variations in the emitted signal. Thus miniaturization of the system beyond a certain limit reduces sensitivity. Nevertheless, using high specific activity labels, the optimal sensor-antibody concentration used in a noncompetitive assay design may be extremely low, although a reduction to below 0.01/K is likely to offer little or no benefit, assuming thermodynamic equilibrium is reached in the system.
Noncompetitive sandwich assay designs have generally been regarded as inapplicable to analytes of small molecular size (albeit certain workers, e.g., (38), have claimed to have overcome the difficulties involved), and the competitive approach may therefore often be obligatory. Although the sensitivity attainable is unlikely to be as high as in noncompetitive assays, theoretical analysis reveals that highest sensitivity is likewise achieved when a very small amount of antibody located on a microspot is used (thus maximizing the fractional occupancy of antibody binding sites).
Detailed theoretical consideration of (noncompetitive) microspot
immunoassay sensitivity (39) permits prediction of the
sensitivities attainable when sensor antibodies of varying affinities
are used as a function of Smin (Fig. 9
). The analysis indicates that the attainment of high microspot
assay sensitivity requires a detector capable of determining very low
surface densities of labeled developing antibodies, the close packing
of sensor-antibody molecules within the microspotby minimizing
backgroundcontributing to the fulfillment of this requirement. It
also suggests that, using high specific activity labels, sensitivities
of microspot assays are unlikely to be inferior and (depending on the
measuring system) may be superior to the sensitivities achieved in
macroscopic assays of conventional design. Researchers at Boehringer
Mannheim have claimed that 0.01 labeled developing
molecules/µm2
have been detected in their experimental
studies, suggesting that assay detection limits approximating
10-17 mol/L (i.e., 103
104
molecules/mL) are attainable using the microspot approach.
|
These conclusions assume the establishment of thermodynamic equilibrium
in the system; however, the velocities of binding reactions fall with a
reduction of the concentrations of one or both of the reactants, and
the time to reach equilibrium increases. Moreover, it is well known
that diffusion constraints reduce reaction velocities when capture
molecules are linked to a solid support. However, it is intuitively
evident that the smaller the microspot, the lower these constraints.
Theoretical analysis of the rate at which analyte molecules migrate
toward, and bind to, an antibody microspot reveals that the (initial)
antibody occupancy rate [per unit area of microspot (OR)]
is given by:
![]() | (3) |
rm2
kadAb
becomes small compared to 4Drm, implying that OR
approximates ka[An]dAb. In other
words, the rate at which sensor antibody molecules located within the
microspot become occupied increases with reduction in
rm, ultimately approximating that seen in a
homogeneous solution as rm approaches zero.
Detailed computer models reveal the full sequence of events following
introduction of antibody microspots of different diameters into an
analyte-containing solution; these models embrace inter alia the
kinetics of the antibody-analyte reaction per se and the initial
establishment of analyte concentration gradients in the solution. Such
studies confirm that the smaller the microspot, the lower the diffusion
constraints on the rate of analyte binding to antibody and the more
closely the reaction kinetics approximate those characterizing a
homogeneous liquid-phase reaction system (Fig. 10
). These considerations reveal that higher signal-to-noise
ratios are attained in a shorter time using a microspot format,
implying that microspot assays are likely to prove at least as rapid as
those of conventional macroscopic design.
|
Notwithstanding the validity of the above concepts, their exploitation
on an industrial scale represents a formidable challenge. The
industrial method of constructing microarrays developed by Boehringer
Mannheim relies on small disposable polystyrene carriers onto which
microspots are deposited using "inkjet" technology. Arrays of as
many as 100200 spots may be deposited in this manner on the flat
bottom (~3-mm diameter) of the carrier wells, each droplet of reagent
(antibody or antigen) solution having a volume of <1 nL and yielding a
spot of diameter approximating 80 µm. Microspot array production by
these means is presently effected in a fully automated unit at a speed
of 10 000 carrier chips per hour. Spot size and array pattern and
shape are controlled automatically by a digital imaging feedback
device, allowing very precise microspot application (Fig. 11
).
|
An alternative approach applicable to oligonucleotide and polypeptide microarray construction has been developed (originally for other purposes) by Fodor et al. (41), using combinatorial synthetic techniques. In this elegant approach, selected areas on a solid substrate are sequentially unmasked (by a photolithographic process) and exposed to a particular nucleoside (or amino acid in the case of polypeptide arrays). By these means, the set of all oligonucleotides of length n (numbering 4n in total) are produced in 4n cycles. Among other advantages, this technique permits construction of large, high-density arrays, although the oligonucleotides present within individual microspots are likely to be of lower purity than when spots are formed by inkjet techniques using presynthesized material. Such in situ combinatorial techniques are, of course, not applicable to the construction of antibody arrays or to arrays of other molecular species not of simple single-stranded form.
Oligonucleotide arrays are clearly of potential importance for DNA
analysis (42); however, such arrays can also be used as
standard array templates, permitting individual researchers to
construct antibody arrays of their own design, using antibodies to
which complementary oligonucleotide sequences have been linked (Fig. 12
).
|
Microarray technologies are still undergoing intensive development, and
it would therefore be premature to present results that are likely to
be unrepresentative of their ultimate performance. Suffice to say that
a variety of noncompetitive and competitive "immunoarray" systems
have been developed, comprising multiple sandwich assays, labeled
analyte back-titration assays for low molecular weight analytes, and
capture-antigen assays for the determination of serum antibodies.
Assays have related to analytes within the fields of endocrinology,
allergy, and infectious disease; similar techniques have also been used
for the screening of therapeutic drugs. Thyrotropin assays have been of
high sensitivity (detection limits <0.01 mIU/L), and good correlation
with results yielded by other techniques has been obtained. Likewise,
close correlations with the results obtained with the latest
commercially available test kits has been demonstrated for a number of
allergens (e.g., birch, cat epithelium, house dust mite,
-amylase,
bee venom, and total IgE), with assay precision and sensitivity
superior to those of these kits. For total IgE a lower detection limit
of <10 IU/L (<0.01 IU/mL) has been achieved. Similarly, microspot
assays relating to a number of infectious diseases (e.g., HIV,
hepatitis B surface antigen, anti-hepatitis C, and rubella) have been
developed and have been shown to yield improved performance when
compared with the latest commercially available methods and
instruments.
Summary and Conclusions
Ligand assays have made a major impact on medicine and related fields in the past 40 years. Initially time-consuming and laborious, the technology was transformed in the 1980s by the emergence of noncompetitive labeled antibody methods (relying on high specific activity nonisotopic labels), yielding higher sensitivities in much shorter times than the original techniques and opening the way to the development of the many automatic instruments that dominate the field at the present time.
For a number of reasons, international attention is increasingly focused on assay miniaturization, and large resources are currently being devoted to this topic in many countries and by many companies. Miniaturized microarray technologies of the kind described in this article are particularly suited to diagnostic applications becausealthough only requiring small sample volumes they do not merely constitute miniaturized versions of conventional assay methods and thus do not themselves rely on the micropumps, microchannels, micrometer-sized reaction compartments, and other micromachined structures embodied in many of the microanalytical devices currently under development (although such devices may be necessary for sample preparation in certain analytical applications). Nor, in consequence, do they rely on the measurement of submicroliter (or nanoliter) sample volumes of such minuscule size that assay sensitivities are severely reduced and large statistical variations are likely to be encountered in the numbers of molecules (if any) that are present in the measured sample. There can, therefore, be little doubt that, if their commercial viability can be demonstrated, microarray technologies represent the next major advance in the diagnostic field. Clearly, however, this prediction will only be fulfilled if they match current methods in regard to sensitivity, precision, and speed of performance while yielding substantial improvements in convenience and cost.
Microarray-based DNA analysis technologies have attracted particular attention in the United States, their development being initiated by the Genosensor Consortium (established in 1992 as part of the US National Institute of Standards and Technology Advanced Technology Program). My own studies initially related principally to immunoarrays (in which area of application very high sensitivity is generally a prime requirement and whose achievement thus formed a key objective). Nevertheless, the future importance of "oligoarrays" for DNA analysis was evident from the outset, and the studies of myself and my collaborators have increasingly focused on this area of application.
Genetic analysis raises well-known ethical issues, which currently are the subject of much discussion and which are inappropriate to address here. It nevertheless appears likely that DNA analysis relying on oligonucleotide arrays to detect specific gene defects or genetically determined responses to particular drugs will feature prominently in the diagnostic armamentarium in the foreseeable future. Moreover, DNA analysis is of importance for a number of other reasons, for example, for tissue typing for transplantation surgery, "DNA fingerprinting" for forensic purposes, and the detection or identification of infectious agents (e.g., viruses) in human blood and tissues.
Meanwhile, antibody-based microarrays permitting the simultaneous determination of multiple analytes are likely to prove of particular value in certain obvious areas, e.g., in the identification of specific allergies and of viral antigens and antibodies in transfusion blood. Moreover, the requirement of the technology for small blood samples and the savings in time and cost of laboratory personnel consequent on the ability to simultaneously determine all clinically relevant biochemical indicators are likely to prove compellingly attractive to hospital laboratories throughout the world, currently under increasing budgetary pressures.
Footnotes
Division of Molecular Endocrinology, University College London Medical School, Mortimer Street, London W1N 8AA, United Kingdom. Fax 44-580-2737; e-mail r.ekins{at}ucl.ac.uk.
Based on the Edwin F. Ullman Award Address presented by Dr. Ekins at the 1998 Oak Ridge Conference.
1 1 Berson and Yalow defined precision as db/(d[H]/[H]); this is equivalent to db/(dlog/[H]), the slope of the b vs log[H] response curve. ![]()
2 2 Although in recent published papers relating to the meaning of sensitivity Dr. Pardue (11, 13) and I (10, 12) have disagreed on the definition and use of the word, we are essentially in agreement on the need for statistical analysis of assay data to assess the magnitude of statistical errors in the response variable, and hence the ability of an assay to determine a low analyte concentration. ![]()
3 3
Figs. 24
drawn in the period 19571960 and presented at a meeting of The Middlesex Hospital Medical School Research Society in London during this period. ![]()
4 4 As indicated in Fig. 2
, saturation analysis was originally conceived as a general analytical approach, relying on any kind of specific binding agent (including antibodies). The term saturation analysis, although not ideal, was coined in an attempt to represent the generality of the analytical concept involved, avoiding any implied restriction either to the use of a particular class of binding reagent or to a type of label (although radioisotopic labels were, for many reasons, a natural choice). Antibodies had long been known (in the 1950s) to specifically bind antigens, and I regarded the possibility of their use in this context (as indicated in Fig. 2
) as, therefore, entirely obvious. I was, of course, entirely unaware of the work of Yalow and Berson (1) in this area at the time Fig. 2
was first drawn and publicly presented. ![]()
5 5 IRMAs maydepending on their design and factors such as the nonspecific binding of labeled antibodyexhibit slightly greater sensitivity than correctly optimized RIAs; however, the improvement is marginal. On the other hand, the opposition by Rodbard et al. (20) to the claim by Miles and Hales (19) was incorrectly based on the supposition that the sensitivities of competitive and noncompetitive labeled antibody assay designs are identical. ![]()
6 6 I dislike the terms competitive and noncompetitive because, inter alia, they fail to represent the true (statistical) basis of the distinction between the different assay designs discussed in this article. Moreover, although widely used, the terms are ill-defined and are often used in different senses by different authors. ![]()
7 7 The term manipulation error is here used to signify the statistical sum of the random experimental errors arising from pipetting, separation of reaction products, and other processing steps involved in the assay; however, it excludes the statistical "counting" errors arising in the determination of the assay response (e.g., the counting errors incurred in the measurement of radioactive disintegrations or fluorescent photons). ![]()
8 8 The collaboration was initiated, and its objective was defined, by myself (based on my then-unpublished theoretical analyses relating to assay sensitivity) and was actively continued until ~19831984. Regrettably, no reference to myself or those of my colleagues participating in the collaboration was made in a number of papers published by members of the Wallac research team commencing in 1982 [e.g., (3134)], nor (until 1984) was the "dissociation enhancement" approach to fluorescence measurement (on which the assays described in these papers relied and that had been developed by one of my colleagues, Dr. Salifu Dakubu, [see (35)]) disclosed. ![]()
9 9 High specific activity nonisotopic labels yield little or no sensitivity improvement in competitive assays. ![]()
References
The following articles in journals at HighWire Press have cited this article:
![]() |
M. Hartmann, M. Schrenk, A. Dottinger, S. Nagel, J. Roeraade, T. O. Joos, and M. F. Templin Expanding Assay Dynamics: A Combined Competitive and Direct Assay System for the Quantification of Proteins in Multiplexed Immunoassays Clin. Chem., June 1, 2008; 54(6): 956 - 963. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Dandy, P. Wu, and D. W. Grainger Array feature size influences nucleic acid surface capture in DNA microarrays PNAS, May 15, 2007; 104(20): 8223 - 8228. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Kusnezow, Y. V. Syagailo, S. Ruffer, N. Baudenstiel, C. Gauer, J. D. Hoheisel, D. Wild, and I. Goychuk Optimal Design of Microarray Immunoassays to Compensate for Kinetic Limitations: Theory and Experiment Mol. Cell. Proteomics, September 1, 2006; 5(9): 1681 - 1696. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. FitzGerald, J. V. Lamont, R. I. McConnell, and E. O. Benchikh Development of a High-Throughput Automated Analyzer Using Biochip Array Technology Clin. Chem., July 1, 2005; 51(7): 1165 - 1176. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. G. He, G. Gerona-Navarro, and S. R. Jaffrey Ligand Discovery Using Small Molecule Microarrays J. Pharmacol. Exp. Ther., April 1, 2005; 313(1): 1 - 7. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Saviranta, R. Okon, A. Brinker, M. Warashina, J. Eppinger, and B. H. Geierstanger Evaluating Sandwich Immunoassays in Microarray Format in Terms of the Ambient Analyte Regime Clin. Chem., October 1, 2004; 50(10): 1907 - 1920. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Roda, M. Mirasoli, S. Venturoli, M. Cricca, F. Bonvicini, M. Baraldini, P. Pasini, M. Zerbini, and M. Musiani Microtiter Format for Simultaneous Multianalyte Detection and Development of a PCR-Chemiluminescent Enzyme Immunoassay for Typing Human Papillomavirus DNAs Clin. Chem., October 1, 2002; 48(10): 1654 - 1660. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Sloane, J. L. Duff, N. L. Wilson, P. S. Gandhi, C. J. Hill, F. G. Hopwood, P. E. Smith, M. L. Thomas, R. A. Cole, N. H. Packer, et al. High Throughput Peptide Mass Fingerprinting and Protein Macroarray Analysis Using Chemical Printing Strategies Mol. Cell. Proteomics, July 1, 2002; 1(7): 490 - 499. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. P. Price Microarrays: The Reincarnation of Multiplexing in Laboratory Medicine, But Now More Relevant? Clin. Chem., August 1, 2001; 47(8): 1345 - 1346. [Full Text] [PDF] |
||||
![]() |
R. Wiese, Y. Belosludtsev, T. Powdrill, P. Thompson, and M. Hogan Simultaneous Multianalyte ELISA Performed on a Microarray Platform Clin. Chem., August 1, 2001; 47(8): 1451 - 1457. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Wiltshire, S. O'Malley, J. Lambert, K. Kukanskis, D. Edgar, S. F. Kingsmore, and B. Schweitzer Detection of Multiple Allergen-specific IgEs on Microarrays by Immunoassay with Rolling Circle Amplification Clin. Chem., December 1, 2000; 46(12): 1990 - 1993. [Full Text] [PDF] |
||||
![]() |
R Swaminathan and M Wheeler Robotics into the millennium J. Clin. Pathol., January 1, 2000; 53(1): 22 - 26. [Full Text] [PDF] |
||||
![]() |
C. P. Price The Evolution of Immunoassay as Seen Through the Journal Clinical Chemistry Clin. Chem., October 1, 1998; 44(10): 2071 - 2074. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |