|
|
||||||||
Poster Sessions |
a author for correspondence: fax 408-942-5600 e-mail calberts{at}lfsus.jnj.com
The American Diabetes Association (ADA) has estimated that in 1993 7.8 million people in the United States had type 1 or type 2 diabetes (1) . About 70% of these individuals self-monitor blood glucose, and ~2 million glucose meters were purchased in 19951996 (2) . The ADA's recent reduction in the diagnostic threshold for diabetes is expected to increase the number of diagnosed patients (3) . Electrochemical meter systems offer some advantages for self-monitoring: compact size, rapid test time, and small sample volume. In this study, the analytical performance of a prototype of the FastTake(TM) system was evaluated for providing accurate and reproducible blood glucose results.
The FastTake system is based on the enzymatic oxidation of glucose
(glucose oxidase) using a mediator (ferricyanide) and electrochemical
detection. The system produces rapid (15 s) plasma equivalent results
from capillary whole blood samples over a wide range of glucose
concentrations (2006000 mg/L). The test strip contains a carbon-based
working electrode, silver/silver chloride reference electrode, and dry
reagents. This strip has been optimized to reduce temperature and
hematocrit sensitivities, using nonconductive, partially silanized
silica fillers, selected to balance the hydrophobicity/hydrophilicity
of the reaction environment (4) . The silica filler forms a
gelatinous layer, which restricts large sample components such as red
blood cells from impairing the reaction at the electrode. The
screen-printed reagents and the multilaminar construction of the test
strip reduce the blood sample requirement (
2.5 µL), compared with
most other currently available systems.
One hundred three adults with diabetes completed the Institutional Review Board-approved, 4-week study at either Good Samaritan Hospital, Los Gatos, CA (site 1, 53 participants, sea level) or the International Diabetes Center, Lutheran Medical Center, Wheat Ridge, CO (site 2, 50 participants, 5400 ft elevation). The age of the participants ranged from 18 to 74 years. Forty-four participants had previously been diagnosed with type 1 diabetes, and 59 had been diagnosed with type 2. All participants had prior experience with self-monitoring of blood glucose (average years of monitoring, 7.1).
During each of three site visits, participants tested capillary whole blood with the FastTake system. Small volumes of blood from the same finger puncture(s) were also used to measure hematocrit (STAT-CRIT®, Wampole Laboratories) and plasma glucose values (YSI 2700 Biochemistry Analyzer, Yellow Springs Instruments). During the study, each participant tested two FastTake test strip lots. The participants presented a wide range of blood glucose concentrations (3904120 mg/L) and hematocrit concentrations (3059%).
Over the range of glucose concentrations tested from both sites,
FastTake correlated well with plasma YSI reference values
(FastTake = 0.984YSI + 13.3 mg/L, n = 284,
Sy
x = 123 mg/L, r2 =
0.98). By linear regression analysis, results for site 1 compared well
to those for site 2. For glucose concentrations measured at site 1,
FastTake correlated well with plasma glucose values measured by the YSI
(FastTake = 0.987YSI - 8.4 mg/L, n = 143,
Sy
x = 122 mg/L, r2 =
0.963). Site 2 demonstrated similar linear regression statistics
(FastTake = 0.98YSI + 35.6 mg/L, n = 141,
Sy
x = 122 mg/L, r2 =
0.975). Error grid analysis (5) (Fig. 1
a) was used to show the clinical accuracy of FastTake for both
sites. Over the range of glucose tested, 97.9% of the results fell
within zone A (defined as "clinically accurate"), and 2.1% fell
within zone B (defined as "deviating from the reference method by
>20% but would lead to benign or no treatment error").
Individually, the error grid results were very similar for the two
sites, with site 1 and site 2 yielding 97.2% and 98.6%, respectively,
in Zone A. For both locations, the remaining data points fell in zone
B. During the studies, the clinical performance of the two test strip
lots was observed to be identical based on error grid analysis (Fig. 1a
). Across the range of observed patient hematocrits (3059%), the
FastTake meter bias relative to the YSI reference was observed to be
independent of hematocrit concentration (Fig. 1b
).
|
Precision and sample volume testing were conducted at LifeScan to further evaluate the FastTake prototype. Venous whole blood samples were collected from healthy donors in heparinized tubes. Because the product is not intended for use with venous blood samples, to simulate capillary blood, the PO2 was measured (Ciba Corning Model 278 Blood Gas System, Chiron Diagnostics) and maintained between 60 and 70 mmHg. Precision was evaluated on 2 days using 10 meters and whole blood from two donors. Over the glucose concentrations tested (~500, 800, 1500, 3000, and 5000 mg/L), within-run precision (CV) ranged from 2.7% to 5.3% (n = 20) for day 1 and 2.8% to 4.6% (n = 20) for day 2.
Sample volume testing was also conducted over 2 days by use of five meters and whole blood from two donors. Replicates (n = 3) were tested on each meter at three glucose concentrations (~700, 1800, and 3600 mg/L) with different sample volumes of 2.5, 3, 4, 5, 10, and 25 µL. Testing at 2.5 µL indicated that 95.6% of the data fell within ±20% of the reference method at the three glucose concentrations tested. For blood volumes of 3.0 µL and above, >97.8% of the laboratory data fell within ±20% of the reference.
These studies indicate that the FastTake system provides rapid, accurate, reproducible results in both the laboratory and clinical settings. Although the product is not intended for use with samples containing lower concentrations of oxygen, such as venous blood, the system performance was consistent at sea level and at the lower atmospheric oxygen testing environment encountered in the Denver clinic. System performance in a clinical setting was also found to be consistent throughout the glucose and hematocrit ranges observed.
Footnotes
LifeScan, Inc., Clinical Research M/S 3D, 1000 Gibraltar Dr., Milpitas, CA 95035
References
The following articles in journals at HighWire Press have cited this article:
![]() |
G. B.B. Kristensen, N. G. Christensen, G. Thue, and S. Sandberg Between-Lot Variation in External Quality Assessment of Glucose: Clinical Importance and Effect on Participant Performance Evaluation Clin. Chem., September 1, 2005; 51(9): 1632 - 1636. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |