Vitreous humor is widely used for postmortem biochemical analysis because it is easier to collect than cerebrospinal fluid and its composition changes more slowly after death than the composition of blood [reviewed by Coe (1)(2)]. The concentrations of sodium, urea, and creatinine in the vitreous humor change little over time, which means that it is possible to use these specimens to make postmortem diagnoses of renal failure, severe dehydration, salt intoxication, or excessive water intake (2)(3)(4)(5). Many drugs, including ethanol, can also be measured conveniently after death in vitreous humor (2).
Unfortunately, the biochemical analysis of vitreous humor is not always straightforward, and several authors have reported problems with the reproducibility and accuracy of these measurements. These problems include unexplained differences in results obtained from left and right eyes (6)(7) and different results when specimens have been analyzed using different instruments (8). We encountered a problem with the reproducibility of vitreous humor electrolyte measurements made with two Hitachi random access analyzers and found that this could be avoided by heating the specimens at 100 °C for 5 min before they were analyzed. Specifically, vitreous humor specimens were heated in capped glass tubes in a heating block, cooled at room temperature for 30 min, and then centrifuged at 1800g for 5 min. Most of the specimens were obviously less viscous on pipetting after the treatment, and some contained a visible precipitate.
The effect of heat treatment was assessed using specimens collected
from either the left or right eyes of 42 different patients during
routine postmortem examinations. Each specimen was divided into two
aliquots, one of which was treated, and the other acted as a control.
Duplicate measurements of each aliquot were performed on Hitachi 737
and Hitachi 911 random access analyzers with precision assessed by
calculating the SD of duplicates (9) (Table 1
). The reproducibility of sodium and potassium measurements on
untreated vitreous humor specimens was significantly worse on the
Hitachi 911 (SD of duplicates, 20.8 and 1.97 mmol/L, respectively)
compared with the Hitachi 737 (SD of duplicates, 2.3 and 0.23 mmol/L,
respectively). Heat treatment improved the precision of sodium and
potassium measurements on the Hitachi 911 and of sodium and chloride
measurements on the Hitachi 737 (Table 1
). The heat treatment increased
the mean results for sodium and potassium measured on the Hitachi 911
but produced no change in the precision or mean values of urea or
creatinine measurements on either analyzer (Table 1
).
|
To determine whether the heat treatment degraded glucose in vitreous humor, glucose was added to four different specimens that had been heated once to reduce their viscosity and allow reproducible pipetting. Glucose was added because it is usually detectable only in vitreous humor from patients with marked hyperglycemia. The glucose concentrations of these specimens before and after repeating the heat treatment and centrifugation were 12.5 and 13.1 mmol/L, 11.8 and 12.0 mmol/L, 11.7 and 11.6 mmol/L, and 11.9 and 12.1 mmol/L, respectively. These results indicated that glucose, like urea and creatinine, was not degraded in vitreous humor subjected to these conditions. It is even possible that the heat treatment could slow the rate of change of vitreous humor glucose concentrations by denaturing key glycolytic enzymes.
Heating vitreous humor to 100 °C for 5 min was a simple and safe method for improving the precision of electrolyte measurements on the Hitachi 911 and 737 analyzers without any adverse effects on the measurement of urea, glucose, or creatinine. We found that centrifuging vitreous humor specimens alone, as suggested by others (2), did not have the same effect. We postulate that the beneficial effects of heating derived from the reduced viscosity of the treated specimens. Like previous workers (8), we found large differences in the reliability of vitreous humor measurements on different analyzers, presumably caused by differences in the mechanics of pipetting, pumping, and dilution. Heat treatment obviously would not be suitable before the measurement of volatile or labile substances. We are examining whether treatment with enzymes that break down mucopolysaccharides might be a suitable alternative in these situations (10).
It is interesting to speculate that some of the unusual results of biochemical testing of vitreous humor reported in the past may be explained by variable specimen viscosity and poor analytical precision. Coe and Apple (8) showed significant variations in sodium, urea, and glucose results for some specimens between different analyzers and commented on the diagnostic confusion that this could produce. They reported good precision (CV <4.3%) for all of their tests on a single pooled vitreous control. Their instruments may not have been as susceptible to the precision problems that we encountered, although it is also possible that their single pooled control may not have reflected the wide variation in viscosity that can occur between individual specimens. Balasooriya et al. (6) and Madea et al. (7) showed large differences in potassium concentrations between the left and right eyes of some patients and used this to argue that potassium measurements are not useful for calculating the postmortem interval. Neither of these groups performed repeat measurements on individual specimens; therefore, it is possible that some of the apparent eye-to-eye variation was the result of analytical variation.