|
|
||||||||
Articles |
2nd Department of Medicine, St. Johanns Spital, Salzburg General Hospital, Muellner Hauptstrasse 48, A-5020 Salzburg, Austria.
a Author for correspondence: Fax 43-662-651674; e-mail R.Weitgasser{at}lkasbg.gv.at
| Abstract |
|---|
|
|
|---|
Methods: On average, 248 glucose measurements were performed with two of each brand of meter on capillary blood samples from diabetic patients attending our outpatient clinic. Two to three different lots of strips were used. All measurements were performed by one experienced technician, using blood from the same sample for the meters and the comparison method (Beckman Analyzer 2). Results were evaluated by analysis of clinical relevance using the percentage of values within a maximum deviation of 5% from the reference value, by the method of residuals, by error grid analysis, and by the CVs for measurements in series.
Results: Altogether, 1987 blood glucose values were obtained with meters compared with the reference values. By error grid analysis, the newer devices gave more accurate results without significant differences within the group (zone A, 9898.5%). Except for the One Touch II (zone A, 98.5%), the other older devices were less exact (zone A, 8792.5%), which was also true for all other evaluation procedures.
Conclusions: New generation blood glucose meters are not only smaller and more aesthetically appealing but are more accurate compared with previous generation devices except the One Touch II. The performance of the newer meters improved but did not meet the goals of the latest American Diabetes Association recommendations in the hands of an experienced operator.
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
statistics
Values were analyzed for clinical relevance by determination of
the percentage of values within a maximum deviation of 5% from the
reference value, according to recommendations by the American Diabetes
Association (ADA) (1)(2). In addition, the error
grid analysis method of Clarke and co-workers
(7)(8) was used to assess accuracy. The error
grid defines the x-axis as the reference blood glucose and
the y-axis as the value generated by the glucose meter. The
graphic model describes clinically relevant deviations using
asymmetrically arranged areas for glucose ranges between 3.9 and 22.2
mmol/L. The agreement between glucose meter values and reference
glucose values is expressed by different zones and thus gives the
accuracy of the meters: zone A, clinically accurate; zone B, clinically
irrelevant deviation by >20% from the reference; zone C, unnecessary
overcorrection possible; zone D, "dangerous failure to detect and
treat" errors; and zone E, "erroneous treatment" danger. To
assess the overall deviation of the devices, we also calculated the
mean (SD) difference from the Beckman glucose oxidase results
(13). To determine within-run precision, the CVs for 10
measurements in series were calculated for three different clinically
relevant blood glucose ranges: 2.93.9 mmol/L, 9.110 mmol/L, and
1515.7 mmol/L.
| Results |
|---|
|
|
|---|
|
|
The error grid analysis is shown in Fig. 1
. For all newer meters, 98% of values were within zone A and
100% were within zones A + B compared with the Accutrend, Companion 2,
and Glucometer 3, which gave a few values in "risk zones" C, D, and
E. Glucometer 3 had estimations in a low glucose range (<4 mmol/L) in
zone E and in a high glucose range (>13 mmol/L) in zone C, Accutrend
and Companion 2 gave values only in a high glucose range (>13 mmol/L)
in zone D. Again, One Touch II was an exception among the older
devices, performing similar to the new generation meters.
|
The CVs for measurement in series, which were used to define precision
of the devices, are shown in Table 3
. The Glucometer Esprit performed the worst, especially with
respect to low glucose ranges, where only the old meter Companion 2
showed even more dispersion.
|
| Discussion |
|---|
|
|
|---|
25% of
measurements. The reason for this improved performance by the
newer glucose meters is probably attributable to both technical
improvements in the devices and the reduced blood volumes necessary for
measurement. The newer meters need only 35 µL compared with 1050
µL for previous systems, which makes mistakes in application of blood
drops to test strips unlikely. Strips such as those used for the
Glucocard take up only a limited amount of blood for measurement. These
advantages are combined with fast measurement within 2060 s; memory
function for up to 300 measurements; and smaller, more aesthetically
appealing devices. However, only the Glucotouch is still equipped with
test strips for visible control, which may help detect meter
dysfunction. We found no substantial difference with respect to the technical equipment of the tested meters using either reflectance or electronic sensor technique. Whether additional new techniques will help meet the goals of the latest ADA criteria is thus questionable. To justify these stringent criteria, one must also be aware of a broad variability in the skill of users. SMBG, meanwhile, is performed by so many patients and healthcare personnel in various settings that user errors may impair results in daily practice despite improvement in analytical performance. As implemented by the Diabetes Control and Complications Trial (3) and the United Kingdom Prospective Diabetes Study (15), current standards for diabetes care (4) include increasing the frequency of SMBG by an increasing number of intensively treated type 1 and type 2 diabetic patients. Therefore, in addition to improvements in technical accuracy, appropriate training of the patients and healthcare personnel using glucose meters is the a mainstay of well-established SMBG. In addition to regular calibration and maintenance of meters, frequent comparison of function with values obtained by a reference laboratory method seems advisable.
In summary, the performance of the newer portable glucose meters when clinically assessed under laboratory conditions was substantially improved compared with all previous generation devices except for the One Touch II. These results may be extrapolated but have yet to be demonstrated in daily use by patients and healthcare personnel. Until reliable noninvasive blood glucose measurement methods are available for everyday clinical use, further improvement of currently available devices to meet the ADA standards is necessary.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
M.-H. Wu, M.-Y. Fang, L.-N. Jen, H.-C. Hsiao, A. Muller, and C.-T. Hsu Clinical Evaluation of Bionime Rightest GM310 Biosensors with a Simplified Electrode Fabrication for Alternative-Site Blood Glucose Tests Clin. Chem., October 1, 2008; 54(10): 1689 - 1695. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. G. Niehoff, T. W. van Haeften, N. C. Onland-Moret, C. C. Elbers, C. Wijmenga, and Y. T. van der Schouw C-Reactive Protein Is Independently Associated With Glucose but Not With Insulin Resistance in Healthy Men Diabetes Care, June 1, 2007; 30(6): 1627 - 1629. [Full Text] [PDF] |
||||
![]() |
K. Dungan, J. Chapman, S. S. Braithwaite, and J. Buse Glucose Measurement: Confounding Issues in Setting Targets for Inpatient Management Diabetes Care, February 1, 2007; 30(2): 403 - 409. [Full Text] [PDF] |
||||
![]() |
W. L. Clarke, S. Anderson, L. Farhy, M. Breton, L. Gonder-Frederick, D. Cox, and B. Kovatchev Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose-Error Grid Analysis Diabetes Care, October 1, 2005; 28(10): 2412 - 2417. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D M Stork, H. Kemperman, D W. Erkelens, and T. F Veneman Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia Eur. J. Endocrinol., August 1, 2005; 153(2): 275 - 281. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Klonoff The Need for Separate Performance Goals for Glucose Sensors in the Hypoglycemic, Normoglycemic, and Hyperglycemic Ranges Diabetes Care, March 1, 2004; 27(3): 834 - 836. [Full Text] [PDF] |
||||
![]() |
P. Bohme, M. Floriot, M.-A. Sirveaux, D. Durain, O. Ziegler, P. Drouin, and B. Guerci Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade Diabetes Care, April 1, 2003; 26(4): 1170 - 1175. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Amin, K. Ross, C. L. Acerini, J. A. Edge, J. Warner, and D. B. Dunger Hypoglycemia Prevalence in Prepubertal Children With Type 1 Diabetes on Standard Insulin Regimen: Use of Continuous Glucose Monitoring System Diabetes Care, March 1, 2003; 26(3): 662 - 667. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Skeie, G. Thue, K. Nerhus, and S. Sandberg Instruments for Self-Monitoring of Blood Glucose: Comparisons of Testing Quality Achieved by Patients and a Technician Clin. Chem., July 1, 2002; 48(7): 994 - 1003. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Sacks, D. E. Bruns, D. E. Goldstein, N. K. Maclaren, J. M. McDonald, and M. Parrott Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus Clin. Chem., March 1, 2002; 48(3): 436 - 472. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Boyd and D. E. Bruns Quality Specifications for Glucose Meters: Assessment by Simulation Modeling of Errors in Insulin Dose Clin. Chem., February 1, 2001; 47(2): 209 - 214. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |