| Abstract |
|---|
|
|
|---|
Methods: The Tosoh HLC-723 GHb V, A1c2.2 glycohemoglobin analyzer is an HPLC instrument with primary blood tube sampling, bar-code reading, cap piercing, and the ability to chromatographically separate labile hemoglobin A1c (HbA1c). We evaluated two analytical protocols, 2.2 and 3.0 min, and compared results for blood samples collected from diabetic and nondiabetic subjects with those obtained with Bio-Rad Diamat and Variant analyzers.
Results: Within- and between batch-imprecision (CVs) was <2% with linearity to at least 15.9% HbA1c. Although some hemoglobinopathies were detected in the 2.2-min chromatography, clearer evidence of abnormality was visible in the 3.0-min version. Comparison with established methods showed good correlation (r = 0.993; n = 316 with Diamat; and r = 0.995; n = 133 with Variant) but highlighted calibration differences.
Conclusions: The problems of manual blood sample preparation, labile HbA1c, and carbamylated hemoglobin interference associated with the older instruments have been eliminated in the new Tosoh analyzer. The 3.0-min protocol is preferred for routine use.
| Introduction |
|---|
|
|
|---|
Various methods have been used for the measurement of glycohemoglobins (4). Analytical goals have included greater practical convenience with automated systems and greater specificity of measurement of the N-terminal valine glycation product [hemoglobin A1c (HbA1c)].1 The precision of HPLC methods has favored their use, and the column technology has been refined to achieve shorter analysis times. Along with this have come compromises in the chromatographic quality and resolution. For many HPLC systems using the ion-exchange principle, it has been necessary to remove the labile HbA1c Schiff base intermediate before chromatography, thereby adding to the labor involved, particularly if this constitutes an off-system analytical step.
We have evaluated the newest of the HPLC ion-exchange systems, the Tosoh HLC-723 GHb V, A1c2.2, in an Anglo-French collaboration. The system is offered with two different analysis protocols thought to be suitable for general diabetic monitoring. The principal novel feature of the analyzer is its quantitative separation of the labile glycohemoglobin on column. Our evaluations were performed according to the Valtec protocol (5) in France and local evaluation procedures in the United Kingdom.
| Materials and Methods |
|---|
|
|
|---|
Samples are placed in Sysmex® racks, which can be continuously fed onto the sample loader as processing takes place. Blood sampling is possible from either primary tubes with optional cap piercing or from small dilution vials, with combinations acceptable in a rack. A bar-code reader handles common code configurations for positive sample identification.
column and reagents
Separation of hemoglobin fractions is achieved by injection of 3
µL of automatically diluted hemolyzed sample through a prefilter onto
a nonporous glycidyl methacrylate-ethylene glycol methacrylate
copolymer cation-exchange column maintained at 25 °C in an aluminum
block oven. The minimum expected lives of the prefilter and column are
500 and 2500 injections, respectively. The hemolysis/wash reagent is an
aqueous EDTA solution containing Triton X-100. Fractions are eluted
from the TSKgel® Glyco HSi column by a
three-step increasing ionic strength buffered succinic acid gradient.
The three component buffers, which undergo continuous vacuum degassing,
are contained in foil pouches located on top of the instrument.
A novel feature of the Tosoh HPLC system is that the labile HbA1c
(L-A1c) fraction does not have to be removed by preanalysis degradation
or on-column chemical or other means because it is well resolved from
the stable HbA1c (S-A1c) and other Hb fractions of interest (Fig. 1
) as a result of the interaction of properties of the resin,
buffers, and gradient used. The separated fractions are monitored with
a diode detector system at 415 and 510 nm.
|
operation
A two-point calibration may be performed after every power-up or
only when specifically requested. Alternatively, the instrument may be
externally calibrated by input of factors to allow alignment with
Diabetes Control and Complications Trial (DCCT) calibrators for
example. Recalibration is not required from the standby state. The
calibration parameters are printed after the initial calibration and on
all subsequent sample chromatograms.
Calculation of results is based on the ratio of the S-A1c fraction chromatographic area to that of the total HbA, which is automatically adjusted using the calibration equation. The percentages of total HbA1 and S-A1c are printed on the report, which may include other information if desired, e.g., full sample identification and calibration and peak information along with the chromatogram.
A 2.2-min chromatographic run time provides basic rapid separation, whereas a 3.0-min version allows better detection of hemoglobin variants. Both protocols have a lag time of 4.4 min after the initiation of processing, with the first result available in 6.6 min for the 2.2-min protocol and 7.4 min for the 3.0-min protocol. After a batch of analyses, the instrument enters a 1015 min wash phase, which can be interrupted. The instrument then returns to stand-by for up to 3 h before automatic power-off. Additional samples may be analyzed at any time in the stand-by period without a requirement for further calibration.
comparison instruments
The Tosoh analyzer was compared with two others used routinely in
the authors' laboratories. In France, a Bio-Rad Diamat HbA1c analyzer
(Bio-Rad Laboratories) was operated according to the manufacturer's
instructions. Manually diluted samples were preincubated for 30 min at
37 °C (off-system) to remove the labile fraction. Calibration was by
means of Bio-Rad materials with locally accepted values that were not
confirmed as DCCT aligned. In the UK laboratory, a Bio-Rad Variant
HbA1c analyzer (Bio-Rad Laboratories) was operated according to Bio-Rad
protocols but with manually diluted samples preincubated off-system for
30 min at room temperature to remove the L-A1c fraction. Calibrators
supplied by Bio-Rad were used, and this analyzer had periodic
verification of calibration against a DCCT Secondary Reference
Laboratory in the University of Minnesota, Minneapolis, MN.
evaluation samples
Venous blood samples were collected into EDTA Vacutainer Tubes
(Becton Dickinson) in both the United Kingdom and France. In addition,
outpatient samples in France were collected into Bio-Rad sample
preparation vials (cat. no. 196-1026) containing aqueous EDTA and
potassium cyanide. Blood specimens were obtained from patients whose
diabetic control was being assessed routinely and were used in this
study according to local ethics guidelines. These were used for
comparison with established local methods.
Samples for assessment of the reference interval were collected from 40 nondiabetic, nonuremic, nonobese subjects (15 men and 25 women, ages 2255 years) with no known diseases.
Automatic 1:250 dilutions of all blood samples were prepared by the Tosoh analyzer and all manual dilutions were made in hemolysis/wash reagent according to the manufacturer's instructions.
Quality-control materials used throughout the evaluations were Lyphocheck Diabetes Control Levels 1 and 2 (Bio-Rad Laboratories).
labile fraction
Because the labile fraction is not removed but is separated as a
distinct peak on the column, the following steps were taken to confirm
the L-A1c peak characteristics.
Glucose incubation.
To confirm the position and
effective separation of the labile fraction peak, buffered
saline-washed red blood cells were incubated with glucose solutions to
increase this fraction and were analyzed at intermediate time points
for 6 h. Two samples of washed red blood cells with HbA1c values
of 5.1% and 10.1%, respectively, were incubated with glucose
solutions to achieve final glucose concentrations of 10, 20, and 50
mmol/L. These suspensions were incubated at 37 °C for 6 h.
Aliquots were removed after 15, 30, 60, 120, 240, and 360 min and
immediately diluted in hemolysis/wash reagent to prevent further
glycation.
Chemical removal.
To investigate the possible multicomponent
nature of the L-A1c peak, chemical elimination of the labile fraction
with a Bio-Rad EDTA-cyanide preparation was performed. Twelve samples
collected in Bio-Rad sample preparation vials were incubated with
Bio-Rad borate buffer (cat. no. 196-1052) for 30 min at 37 °C to
destroy L-A1c. The samples were analyzed from the primary blood tube
and also from the prepared sample before and after incubation.
Storage effect.
Consistency of the S-A1c and L-A1c fractions
over time was investigated by intermittent analysis of blood samples
over 21 days. Samples were stored in primary tubes at 4 °C
throughout.
sample carryover
Carryover between blood samples with high and low HbA1c
concentrations was checked using both the method of Broughton et al.
(6) and as recommended in the Valtec protocol
(5). The Valtec procedure involved analysis of 10
low-concentration samples, then 10 high-concentration samples, and then
these 20 samples intermixed. The results of both analyses may then
examined for interactions. The method of Broughton et al. involved
processing triplets of low- and high-concentration samples sequentially
and examining the first and third results of each triplet for
systematic deviation. The method of Broughton et al. was used for the
3.0-min protocol, and the Valtec procedure for the 2.2-min protocol,
both with cap piercing.
sample sedimentation
When a cold start and a full 90-place sample loader are used, the
maximum time to final sampling is 272.4 min for the 3.0-min protocol.
We therefore examined the stability of results obtained on primary
blood samples with HbA1c concentrations of 5.5%, 7.2%, and 10.3%,
which were allowed to stand unmixed for up to 5 h. They were
presented to the analyzer hourly along with a second aliquot of each
sample in which the cells were resuspended immediately before analysis.
hemoglobin load
The manufacturer's claim is that the optimized total
chromatographic area should be between 1000 and 4000 units. We
therefore investigated the effect of varying hemoglobin load on
chromatographic quantification. For the 2.2-min protocol, increasing
volumes of blood were diluted in a fixed volume of diluent to give a
range of final sample dilution ratios from 1:1000 to 1:50. For the
3.0-min protocol, a similar range of dilution ratios was prepared by
taking a blood sample and manually diluting decreasing volumes of the
freshly mixed sample in hemolysis/wash reagent.
hemolysate stability
Blood samples representing various levels of diabetic control were
manually diluted and then analyzed repetitively over a period of 7
h, standing at room temperature in the intervening periods. Because it
is sometimes advantageous to collect outpatient samples directly from a
finger puncture into hemolysis/wash reagent, the effect of delayed
analysis was also investigated. Samples collected in this manner were
stored at room temperature and at 4 °C for 7 days and analyzed
daily.
assessment of interferences
Lipemia.
The potential interference of lipemia was
investigated by overloading samples of a pooled blood having an HbA1c
concentration of 5% with saline-diluted Intralipid solution
(Kabivitrum SA). The volumes of Intralipid added augmented the
basal triglyceride concentration by 2.0, 3.0, 4.6, 5.0, and 7.0 mmol/L.
Analysis of the samples was carried out immediately after
preparation.
Bilirubin.
Dilutions of a bilirubin solution (cat. no. B4126;
Sigma) were mixed with a pooled blood that had an HbA1c concentration
of 10% to obtain six final concentrations of bilirubin ranging from 25
to 500 µmol/L. Measurement of HbA1c was performed immediately after
preparation.
Carbamylation.
A range of sodium cyanate solutions up to 10
mmol/L were mixed with a pooled blood that had an HbA1c concentration
of 5%. The samples were incubated for 1 h at 37 °C to achieve
in vitro carbamylation of the hemoglobin before analysis on the Tosoh
2.2-min protocol and on the Diamat.
Identification of hemoglobin variants.
Variant hemoglobins
were identified by means of electrophoresis at acid and alkaline pH
with further investigation using either standard hematological or
molecular biology techniques as appropriate.
| Results |
|---|
|
|
|---|
|
linearity
This was assessed by comparing expected and observed analytical
values using results from proportional mixtures of high and low HbA1c
samples. For the 2.2-min protocol, samples were prepared by combining
proportions of undiluted blood containing 4.7% and 15.9% HbA1c before
preparation of the hemolysates. For the 3.0-min protocol, samples
containing 5.1% and 14.0% HbA1c were hemolyzed and then
proportionally mixed. Linear regression analysis of the 2.2-min
protocol data gave a relationship of y =
1.0x - 0.62; r = 0.995; analysis of
the 3.0-min protocol gave a relationship of y =
0.986x + 0.15; r = 0.999.
method comparisons
Comparisons for the 2.2-min protocol were made against the local
established method, a Bio-Rad Diamat (Paris laboratory), using bloods
from 319 diabetic and nondiabetic subjects being routinely monitored.
Initial evaluation of the intermethod bias revealed three extreme
outliers (Fig. 2
). Further examination of the data showed that two of these
samples had the highest L-A1c fractions among these subjects, at 5.7%
and 6.1%. These outliers were omitted from Passing and Bablok
regression analysis, which gave a slope (95% confidence interval) of
0.97 (0.950.98) and an intercept of 0.16 (0.040.25).
|
Samples from 137 diabetic and nondiabetic subjects were analyzed using
the 3.0-min protocol, and the results were compared with the local
established method, a Bio-Rad Variant (Newcastle laboratory). Initial
evaluation of intermethod bias revealed four extreme outliers in the
group (Fig. 3
). Further examination of the data showed that all of these
samples had urea concentrations above the age-related reference
interval. These outliers were omitted from Passing and Bablok
regression analysis which gave a slope (95% confidence interval) of
1.00 (0.981.00) and intercept of -0.80 (-0.80 to -0.63).
|
reference interval
Reference values for healthy nondiabetic volunteers were
4.35.4%, with a mean of 4.9%. These were obtained by use of Tosoh
calibrators with values assigned by the Japanese Diabetic Society.
DCCT-aligned values for the calibrators are now also available, and
these are ~0.5% higher.
labile fraction
For bloods incubated with glucose concentrations up to 50 mmol/L
and for up to 6 h, the chromatographically identified S-A1c peak
did not show a change beyond method reproducibility. Peaks identified
as L-A1c showed increases as a function of both time and glucose
concentration, which after a 6-h incubation were 39233% higher than
the values of nonincubated samples.
When subjected to chemical modification by progressively longer incubation in borate-buffered solution, the L-A1c peak values were similar between primary and preincubation samples but diminished post incubation. However, in all cases a substantial remnant varying from 58% to 85% of the original peak value persisted, suggesting that the fraction is of a heterogeneous nature.
During storage over a 21-day period of undiluted blood samples with initial S-A1c percentages of 6.910.7%, the labile fraction diminished by 3263% of the original value. In the same samples, the absolute change in the stable fraction was no greater than 0.2%, which is within 2.8 SD of the initial values and, therefore, not significant.
carryover
Carryover was 1.2% in the 2.2-min protocol and 0.3% in the
3.0-min protocol. Both are within the achievable reproducibility of the
system under these conditions.
sample sedimentation
The largest absolute changes in measured HbA1c values for samples
with concentrations up to 10.2% and that were subjected to a maximum
5-h sampling delay period were 0.1% HbA1c (relative percentage changes
of 1.9% at 5.4% HbA1c, 1.0% at 10.2% HbA1c, and 0% at 7.1%
HbA1c). These differences fall within method reproducibility for the
3.0-min protocol.
hemoglobin load
Sample dilution ratios of 1:1000 to 1:50 gave total
chromatographic areas of 363-5284 units in the 2.2-min protocol. Areas
between 700 and 4000 provided good resolution and satisfactory
integration as judged subjectively. The same range of dilutions with
the 3.0-min protocol gave total chromatographic areas ranging from 338
to 8080 units. Qualitative assessment of these chromatograms indicated
that the acceptable area range was 500-6000 units.
hemolysate stability
All results obtained on hemolysates analyzed repeatedly over a 7-h
period agreed to within 0.1% HbA1c compared with the parent blood
samples analyzed at the same time. Longer storage at room temperature
was unsatisfactory. Stability of hemolysates was acceptable for up to
24 h if stored at 4 °C.
interferences
No interference was observed with up to 10 mmol/L triglyceride and
400 µmol/L bilirubin on a sample with 5% HbA1c.
carbamylation
On the Tosoh the S-A1c peak was unaffected; only the labile
fraction increased and this was by 20% L-A1c at the maximum cyanate
concentration. However, the peak measured as HbA1c on the Diamat
increased artifactually to an apparent 28.7% HbA1c at the same cyanate
concentration. In these prolonged incubation experiments, hemoglobin
degradation was indicated in the Diamat chromatograms by a marked
baseline rise, which may be the cause of some of the increases seen in
Diamat results.
hemoglobinopathies
The relative utility of the two protocols in demonstrating the
presence of hemoglobin variants was investigated using samples
containing HbAC, HbAE, HbAD, HbF, HbAS, and HbTyne (7) and
showed superior identification for the 3.0-min version (Fig. 4
).
|
| Discussion |
|---|
|
|
|---|
Although reproducibly measured glycohemoglobin is an essential prerequisite for diabetes monitoring, its clinical value is greatly enhanced by relating these measures to likely clinical outcome. This is best achieved by reference to the now well-established evidence base found in DCCT for type I diabetes (1) and to the more recent corroborative data from the UK Prospective Diabetes Study for type II (2). The American Diabetes Association recommends that diabetes monitoring using glycohemoglobin should be performed only with DCCT-traceable methods (11). Differences in bias observed in the relationship of the Tosoh instruments to the two different Bio-Rad systems were calibration dependent and serve to emphasize the need to use fully traceable methods with DCCT-aligned calibrator values. Achieving this requires easy calibration adjustments and flexible software control, which along with the wide linear analytical range achievable with the Tosoh facilitates this requirement. Although the reference interval derived in this study was based on Japanese Diabetes Society-assigned values, the calibrator materials supplied by the manufacturer now also have the more useful DCCT values and the analyzer has been certified as DCCT traceable by the National Glycohemoglobin Standardisation Program for both of the protocols we evaluated.
Although performance features such as we have described above contribute to "safe use" of the analyzer for patient investigation, design aspects such as cap piercing provide safety in use for the analyst. Primary samples are processed in a safe and convenient manner without precluding the use of secondary samples, separately or intermixed, when only finger pricks or other small volumes of blood are available. These characteristics combined with positive sample identification and rapid analysis capability, even from standby, give the Tosoh analyzer considerable potential for clinical, near-patient use. This use is greatly assisted by the small analyzer footprint. Reagents are held in foil pouches, which are less bulky than rigid containers. Although the liquid levels cannot be visually inspected in the bags, they are easily checked by palpation, and all reagents except one may be pooled from bag to corresponding bag. However, analyzing larger batches of samples in the laboratory setting may be the more usual role for this instrument. In this situation, red cells are afforded the opportunity to settle as tubes stand on the sample loader. The Tosoh sampling system aspirates from the bottom of the tube, thereby raising the question of whether satisfactory results may still be obtained from samples that have been on the sample loader for several hours. Sedimentation of red blood cells causes a relative increase in the total hemoglobin load presented to the column, especially for later samples in a batch. In practice, the system was tolerant to wide hemoglobin loads and was unaffected by the cell sedimentation process in routine use. Combining this with on-board stability of small sample hemolysates gives walk-away and overnight utility to the analyzer.
L-A1c, the concentration of which varies with blood glucose concentrations at or near the time of blood collection, is an intermediate in the production of the corresponding stable moiety. The isoelectric point of the labile fraction closely approximates that of its stable counterpart, which leads to little or no separation of the two, giving falsely increased results in many methods that rely on charge separation unless steps are taken to remove the more ephemeral indicator of glycemia (12). Although these removal steps have generally required sample pretreatment (13), more modern systems have utilized on-column degradation as part of the automatic analytical process (14). An unique feature of the Tosoh analyzer, using either protocol, is its ability to chromatographically resolve the labile fraction, thus obviating the need for pre- or on-column treatment to destroy this fraction before analysis. We have confirmed that the labile fraction is separated on the Tosoh system, irrespective of its concentration. The benefit of this is demonstrated in the comparison with the locally used routine methods. For example, with the Bio-Rad Variant, we had to use an extended preanalysis incubation to minimize this confounding factor. Others have also reported that standard Bio-Rad Variant procedures may be inadequate for some samples encountered in routine laboratory practice (15). Similar limitations may affect analyzers utilizing on-column labile fraction degradation but which lack labile/stable fraction chromatographic resolution.
The presence of uremia in patients whose long-term glycemic status is being assessed is not an uncommon finding in those whose renal function has deteriorated, and it can provide yet another potential interference with some, although not all, glycohemoglobin methods (16). Although the evidence from our labile fraction experiments suggests that the L-A1c peak does contain all of the labile entity, it also shows that it is not the sole component. We demonstrated that this peak includes, among other things, carbamylated hemoglobin. This raises the question of how other chromatographic systems with lower resolution capabilities handle these unidentified and potentially interfering components or whether they remain as occult effects (17). In our Bio-Rad Variant/Tosoh comparison, the extreme outliers with increased urea concentrations demonstrated the problem of coelution of carbamylated hemoglobin with HbA1c on the Variant. In the Bio-Rad Diamat/Tosoh study, we also demonstrated an artifactual effect of adduct formation on measured HbA1c, which was eliminated on the Tosoh analyzer by virtue of its superior chromatographic resolution.
An additional benefit of enhanced resolution is the ability to detect the presence of hemoglobinopathies, and in this respect the 3.0-min protocol was superior to the 2.2-min protocol for the hemoglobin variants we investigated. Abnormal concentrations of normal hemoglobins such as HbF are also clearly recognized on the Tosoh, which is important in persons with diabetes in whom there is an increased prevalence (18). In the 3.0-min mode, it is less likely that any hemoglobin structural abnormality will go unrecognized and that the apparent HbA1c will be inappropriately reported. The similarity between the chromatograms for the relatively common HbE and the very rare HbTyne show that the analyzer cannot be used with either of the protocols we evaluated to classify hemoglobinopathies but only to indicate their presence. Mathematical elimination of unusual peaks in the calculation of results does not guarantee that the calculated result is a valid or appropriate measure of HbA1c. What is important is that the presence of any abnormality is readily detected by the busy analyst for further consideration of its implications by the physician.
This joint evaluation has shown that the Tosoh A1c 2.2 performs to high analytical standards and has several advantages over other HbA1c HPLC systems available at this time, not the least of which is its high chromatographic resolution. Certain of its features, particularly its speed, lack of requirement for frequent calibration, low maintenance, and safe primary sampling capability, make it equally as well suited for diabetes clinic/near-patient testing applications as for intensive central laboratory use.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
Ser]. Hemoglobin 1994;18:333-336.
[ISI][Medline]
[Order article via Infotrieve]
The following articles in journals at HighWire Press have cited this article:
![]() |
The Juvenile Diabetes Research Foundation Continuo Continuous Glucose Monitoring and Intensive Treatment of Type 1 Diabetes N. Engl. J. Med., October 2, 2008; 359(14): 1464 - 1476. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Thomas, E. Hypponen, and C. Power Obesity and Type 2 Diabetes Risk in Midadult Life: The Role of Childhood Adversity Pediatrics, May 1, 2008; 121(5): e1240 - e1249. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Thomas, E. Hypponen, and C. Power Prenatal Exposures and Glucose Metabolism in Adulthood: Are effects mediated through birth weight and adiposity? Diabetes Care, April 1, 2007; 30(4): 918 - 924. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Thomas, E. Hypponen, and C. Power Type 2 diabetes mellitus in midlife estimated from the cambridge risk score and body mass index. Arch Intern Med, March 27, 2006; 166(6): 682 - 688. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. Beisswenger, K. S. Drummond, R. G. Nelson, S. K. Howell, B. S. Szwergold, and M. Mauer Susceptibility to Diabetic Nephropathy Is Related to Dicarbonyl and Oxidative Stress Diabetes, November 1, 2005; 54(11): 3274 - 3281. [Abstract] [Full Text] [PDF] |
||||
![]() |
The Diabetes Research in Children Network (DirecNe Eight-Point Glucose Testing Versus the Continuous Glucose Monitoring System in Evaluation of Glycemic Control in Type 1 Diabetes J. Clin. Endocrinol. Metab., June 1, 2005; 90(6): 3387 - 3391. [Abstract] [Full Text] [PDF] |
||||
![]() |
The Diabetes Research in Children Network (DirecNe A Randomized Multicenter Trial Comparing the GlucoWatch Biographer With Standard Glucose Monitoring in Children With Type 1 Diabetes Diabetes Care, May 1, 2005; 28(5): 1101 - 1106. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M.W. Spijkerman, M. F. Yuyun, S. J. Griffin, J. M. Dekker, G. Nijpels, and N. J. Wareham The Performance of a Risk Score as a Screening Test for Undiagnosed Hyperglycemia in Ethnic Minority Groups: Data from the 1999 Health Survey for England Diabetes Care, January 1, 2004; 27(1): 116 - 122. [Abstract] [Full Text] [PDF] |
||||
![]() |
T Lahousen, R E Roller, R W Lipp, and W J Schnedl Silent haemoglobin variants and determination of HbA1c with the HPLC Bio-Rad Variant II J. Clin. Pathol., September 1, 2002; 55(9): 699 - 703. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Bry, P. C. Chen, and D. B. Sacks Effects of Hemoglobin Variants and Chemically Modified Derivatives on Assays for Glycohemoglobin Clin. Chem., February 1, 2001; 47(2): 153 - 163. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||