Department of Laboratory Medicine and Pathology, Hennepin County Medical Center, Minneapolis, MN 55415.
2
Departments of Pathology and Medical Research
Technology, University of Maryland School of Medicine, Baltimore, MD
21201.
3
Department of Pathology and Laboratory Medicine,
University of Louisville, School of Medicine, Louisville, KY 40292.
4
Department of Pathology, Hartford Hospital, Hartford, CT
06102.
a Address correspondence to this author at: Hennepin County Medical Center, Clinical Laboratories No. 812, 701 Park Ave., Minneapolis, MN 55415. Fax 612-904-4229; e-mail fred.apple{at}co.hennepin.mn.us.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
To date, all quantitative laboratory-based instrumentation provides measurement of biochemical markers of myocardial injury as discrete single tests after separation of serum or plasma from whole blood. The use of a panel of tests that includes an early marker such as myoglobin or CK-MB mass, and a cardiac-specific marker such as cTnI, may offer clinicians more information for ruling in or ruling out acute MI. The purpose of this study was to clinically evaluate the Triage® Cardiac Panel (Biosite Diagnostics) for the quantitative simultaneous measurement of myoglobin, CK-MB mass, and cTnI in heparin-treated whole blood and plasma in patients with chest pain to rule in or rule out acute MI.
| Materials and Methods |
|---|
|
|
|---|
0.1 mV on at
least two contiguous leads) or new symmetric T wave inversions
0.1 mV
or both; and changes in serial plasma or serum protein markers (CK-MB
or cTnI) associated with MI (2)(7). Temporal changes in
protein markers were defined as follows: increased concentrations above
the decision thresholds limit or an increase of at least 50% over time
if within the reference limits. Each of the four participating sites
individualized their marker criteria, using the following
pre-established decision thresholds: cTnI (Dade Stratus), 1.5 µg/L
(4)(15); cTnI (Behring Opus), 2.5 µg/L (16);
cTnI (Beckman Access), 0.15 µg/L (15); CK-MB mass (Dade
Stratus), 5.0 µg/L (7)(15) or 4.0 µg/L (8);
CK-MB mass (Beckman Access), 7.5 µg/L (15); and myoglobin
(Dade Stratus), 110 µg/L (recommended in manufacturer's package
insert). Clinicians were blinded to the results of the Triage Panel. Patients (n = 192) experiencing symptoms of acute MI admitted through the ER were entered into the study. Specimens (n = 170) were obtained from 59 (31%) patients diagnosed with acute MI, of which 37 were Q-wave MIs. The median (25th75th percentile) time for admission following onset of chest pain was 7.0 h (3.810.8 h). An additional 262 samples from the 133 patients excluded from MI were also obtained. Adequate specimens for the Triage analysis were not available in 50 patients for more than one sampling in the non-AMI patients, a limitation of the study. This group included patients with unstable angina, coronary artery disease, and other causes for chest pain, but with MI ruled out. Serial whole blood or plasma specimens were collected at admission and at 6- to 8-h intervals for the first 12 to 24 h. In addition, collections were obtained randomly up to 72 h. Comparison diagnostic efficiency and correlation studies of the Triage device with the three other systems used the following number of patient samples: Dade Stratus, n = 366; Beckman Access, n = 252; and Behring OPUS, n = 332. To investigate possible nonspecificity in patients with skeletal muscle disease or renal disease, we examined an additional 189 specimens from 117 patients with acute or chronic skeletal muscle injury or disease and 80 specimens from 80 patients with end-stage renal failure undergoing chronic hemodialysis.
The device evaluated, the Triage Cardiac Panel, is a self calibrating fluorescence immunoassay system for the quantitative determination of myoglobin, CK-MB mass, and cTnI and is optimized for heparin-treated whole blood and plasma specimens (17)(18). Serum is not an acceptable specimen. Briefly as taken from the manufacturer's package insert, after addition of the sample (six drops) to the sample port, the cells are separated from the plasma via a filter contained in the device. A predetermined quantity of plasma is allowed to react with fluorescent antibody conjugates within the reaction chamber. After incubation, the reaction mixture flows down the device detection lane. Complexes of the analytes and fluorescent antibody conjugates are captured on discrete zones, producing binding assays that are specific for each analyte. The concentration of each analyte, directly proportional to the fluorescence detected, is measured by the Triage meter. All results are available in 15 min. If whole blood or plasma testing was not performed within 24 h, plasma was stored frozen at -20 °C until tested. No studies were performed to test for marker stability beyond 24 h in the present study. The external quality-control solutions provided consist of a Cardiac Calibration Verification Control Set at three concentrations in a plasma pool and a Cardiac Control Set at two concentrations in an artificial matrix solution. In addition to the external quality-control sets, each Triage device contains two positive device controls, which are immunoassays and perform in the same manner as the specific assays being measured. The positive device control immunoassays comprise an antibody fluorescent conjugate and a ligand fluorescent conjugate. These conjugates bind to complementary ligand and antibody solid-phase zones, respectively, to provide specific fluorescence values at each zone. The device low-control assay is set to correspond approximately to the fluorescence signal generated at the analyte decision cutoff. The device high-control assay corresponds to a fluorescence signal at ~75% of the maximum signal of the doseresponse curve. If either of the two internal controls fail, the test must be repeated. Furthermore, the meter has the following capabilities: quality-control requirements programmed into the instrument, lock-outs for unauthorized personnel, and a laboratory information system interface (as stated by the manufacturer). The lowest detectable limits for the Triage were as follows: myoglobin, 2.7 µg/L; CK-MB, 0.75 µg/L; and cTnI, 0.19 µg/L. Imprecision as determined in the present study of all analytes at the upper reference limit/decision cutoff concentration, was as follows: 9.9% for myoglobin; 12% for CK-MB; and 12% for cTnI. Although not tested in this study, the manufacturer claims that all types of cTnI (free and complexed) are measured by the cTnI assay. The Triage decision cutoff concentrations were 107 µg/L for myoglobin, 4.3 µg/L for CK-MB, and 0.4 µg/L for cTnI, as determined by ROC curve analysis (see Results).
Statistical comparisons of myoglobin, CK-MB, and cTnI between times and between analyzers were performed by one-way and two-way ANOVA. Statistical significance was set at P <0.05, using nonparametric MannWhitney statistics. ROC curve analysis was performed as described previously (19) using True Epistat, Ver. 5. The McNemar test was also used for comparison of diagnostic sensitivity and specificity data between markers or assays; P <0.05 was considered statistically significant. Diagnostic sensitivity and specificity data are presented with 95% confidence intervals (CIs), determined with the Epistat, Ver. 5. Spearman correlation coefficients were calculated between the Triage and comparison methods.
| Results |
|---|
|
|
|---|
|
|
|
Because results for all three markers become available simultaneously
on the Triage device, we examined whether the panel could provide
improved early (<4 h) diagnostic information when compared with
individual test results. When clinical sensitivities were determined on
peak concentrations or over the entire sampling period in serial (cTnI
plus CK-MB plus myoglobin) or parallel (cTnI, CK-MB, or myoglobin),
neither approach showed significant improvement compared with
individual test results for either cTnI or CK-MB (Table 1
).
Table 1
also shows that the Triage cTnI demonstrated 95100%
specificity over the entire 24-h period after admission. In comparison,
CK-MB demonstrated 8291% specificity, and myoglobin demonstrated
6781% specificity.
Fig. 3
shows time vs concentration profiles for acute MI patients,
comparing the Triage cTnI assay with the Dade and Behring cTnI assays.
Although serial specimen patterns demonstrated the parallel rise after
acute MI, the absolute concentrations showed differences between
assays, underlying the need to validate each system independently.
|
Fourteen patients (10.5% of all non-MI patients) with a non-MI final discharge diagnosis had at least one increased Triage cTnI concentration after admission, with cTnI ranging from 0.42 to 5.40 µg/L. These included patients with the following diagnoses: unstable angina (n = 4), coronary artery disease (n = 3), congestive heart failure (n = 3), and noncardiac chest pain (n = 4). In patients with skeletal muscle injury, cTnI showed a clinical specificity of 94% compared with 44% for CK-MB and 21% for myoglobin. The nine samples that demonstrated an increased Triage cTnI without documented cardiac injury were from patients admitted after skeletal muscle trauma caused by automobile accidents and noncardiac surgery. In patients with renal disease, cTnI showed a clinical specificity of 100%, compared with 79% for CK-MB and 16% for myoglobin.
| Discussion |
|---|
|
|
|---|
The diagnostic specificity of the Triage Cardiac Panel was also highly concordant with the other systems in this comparison study, demonstrating maximum clinical specificities for CK-MB and cTnI of 91% and 100%, respectively. When non-MI populations, including patients with skeletal muscle injury and patients with chronic renal disease, were evaluated, cTnI specificity was substantially improved compared with CK-MB. Of the non-MI patients with underlying cardiac pathologies, 14 patients showed increased cTnI concentrations. In this population, which included patients with unstable angina and congestive heart failure, increased concentrations of cTnI have become a risk stratification tool, prompting clinicians to consider a more aggressive management regimen because of poor prognostic outcomes at 3042 days after presentation (11)(12)(13)(14). It should be noted that some of these studies use admission serum values when assessing outcomes, whereas others use peak serum values after serial sampling of cardiac markers. The present study was not designed to assess outcomes, and no patient follow up was obtained.
In addition to CK-MB mass and cTnI, myoglobin concentrations are also obtained on the Triage Panel. Although not cardiac specific, myoglobin is one of the very early markers to increase after acute MI (3)(4)(20). Although studies have demonstrated >95% sensitivity within 4 h after admission for ruling in acute MI (13)(17), the sensitivity of the Triage Panel for myoglobin only approached 7580%. The most likely explanation for the lower sensitivity of the Triage myoglobin assay (as well as for the Dade Stratus myoglobin assay, data not shown) in this study might be related to the patient population studied and the accuracy of the length of time from onset of chest pain to admission, which was estimated on the basis of histories taken at admission, always a confounding issue. However, we have no definite explanation for the lower sensitivities observed. The poor specificity for myoglobin in large populations of chest pain patients presenting with noncardiac chest pain (3)(20) was substantiated in the present study with a specificity of <80%. Several patients in our study experienced skeletal muscle trauma and injury concomitant with chest pain. Furthermore, we are unable to explain the concordance of only >70% for the Triage myoglobin assay.
This study does not address either the role of cardiac markers in reinfarction patients or in patients with acute MI presenting more than 72 h after the onset of chest pain. Clinical trials need to be designed to address both of these issues to determine whether a multiple-marker approach would offer cost benefits for improved patient diagnosis and treatment.
The Triage Cardiac Panel and device (both Food and Drug Administration
approved) offer clinicians quantitative whole blood analysis of
multiple cardiac markers and can serve as a point-of-care testing
device in the ER, the coronary care unit, or at the bedside. The Triage
Panel will compete directly with qualitative positive/negative bedside
tests for cTnI and cTnT that have demonstrated high sensitivity for the
detection of acute MI after 6 h, with negative results associated
with low risk and safe discharge of patients with an episode of acute
chest pain (21). Neither serial nor parallel analysis of the
multiple markers provided increased sensitivity or specificity for
detection or ruling out of MI (Table 1
). However, this study was not
designed to address this issue. Outcome studies should be conducted to
confirm whether the multiple marker approach or noncentralized
laboratory measurements of these markers will be cost-effective or
time-saving devices that will assist clinicians in more rapid or
appropriate management and patient therapy.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
H. C. Chew Cardiac Troponin T in Acute Coronary Syndrome with Renal Insufficiency Asian Cardiovasc Thorac Ann, August 1, 2008; 16(4): 284 - 287. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Anderson, C. D. Adams, E. M. Antman, C. R. Bridges, R. M. Califf, D. E. Casey Jr, W. E. Chavey II, F. M. Fesmire, J. S. Hochman, T. N. Levin, et al. ACC/AHA 2007 Guidelines for the Management of Patients With Unstable Angina/Non-ST-Elevation Myocardial Infarction: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 2002 Guidelines for the Management of Patients With Unstable Angina/Non-ST-Elevation Myocardial Infarction) Developed in Collaboration with the American College of Emergency Physicians, the Society for Cardiovascular Angiography and Interventions, and the Society of Thoracic Surgeons Endorsed by the American Association of Cardiovascular and Pulmonary Rehabilitation and the Society for Academic Emergency Medicine J. Am. Coll. Cardiol., August 14, 2007; 50(7): e1 - e157. [Full Text] [PDF] |
||||
![]() |
D. Abramov, M. Abu-Tailakh, M. Frieger, A. Ganiel, D. Tuvbin, and A. Wolak Plasma Troponin Levels After Cardiac Surgery vs After Myocardial Infarction Asian Cardiovasc Thorac Ann, December 1, 2006; 14(6): 530 - 535. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Bassan, A. Potsch, A. Maisel, B. Tura, H. Villacorta, M. V. Nogueira, A. Campos, R. Gamarski, A. C. Masetto, and M. A. Moutinho B-type natriuretic peptide: a novel early blood marker of acute myocardial infarction in patients with chest pain and no ST-segment elevation Eur. Heart J., February 1, 2005; 26(3): 234 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. von Lode, J. Rainaho, and K. Pettersson Quantitative, Wide-Range, 5-Minute Point-of-Care Immunoassay for Total Human Chorionic Gonadotropin in Whole Blood Clin. Chem., June 1, 2004; 50(6): 1026 - 1035. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. McCord, R. M. Nowak, P. A. McCullough, C. Foreback, S. Borzak, G. Tokarski, M. C. Tomlanovich, G. Jacobsen, and W. D. Weaver Ninety-Minute Exclusion of Acute Myocardial Infarction By Use of Quantitative Point-of-Care Testing of Myoglobin and Troponin I Circulation, September 25, 2001; 104(13): 1483 - 1488. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Fischer, K. Filzmaier, H. Stiegler, J. Graf, S. Fuhs, A. Franke, U. Janssens, and A. M. Gressner Evaluation of a New, Rapid Bedside Test for Quantitative Determination of B-Type Natriuretic Peptide Clin. Chem., March 1, 2001; 47(3): 591 - 594. [Full Text] [PDF] |
||||
![]() |
F. S. Apple, F. P. Anderson, P. Collinson, R. L. Jesse, M. C. Kontos, M. A. Levitt, E. A. Miller, and M. M. Murakami Clinical Evaluation of the First Medical Whole Blood, Point-of-Care Testing Device for Detection of Myocardial Infarction Clin. Chem., October 1, 2000; 46(10): 1604 - 1609. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Braunwald, E. M. Antman, J. W. Beasley, R. M. Califf, M. D. Cheitlin, J. S. Hochman, R. H. Jones, D. Kereiakes, J. Kupersmith, T. N. Levin, et al. ACC/AHA guidelines for the management of patients with unstable angina and non-st-segment elevation myocardial infarction: A report of the american college of cardiology/ american heart association task force on practice guidelines (committee on the management of patients with unstable angina) J. Am. Coll. Cardiol., September 1, 2000; 36(3): 970 - 1062. [Full Text] [PDF] |
||||
![]() |
C. Heeschen, B. U. Goldmann, L. Langenbrink, G. Matschuck, and C. W. Hamm Evaluation of a Rapid Whole Blood ELISA for Quantification of Troponin I in Patients with Acute Chest Pain Clin. Chem., October 1, 1999; 45(10): 1789 - 1796. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H.B. Wu, F. S. Apple, W. B. Gibler, R. L. Jesse, M. M. Warshaw, and R. Valdes Jr. National Academy of Clinical Biochemistry Standards of Laboratory Practice: Recommendations for the Use of Cardiac Markers in Coronary Artery Diseases Clin. Chem., July 1, 1999; 45(7): 1104 - 1121. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |