The discipline of clinical chemistry is dynamic. Even a superficial glance at contents pages of the Journal over recent years demonstrates a marked evolution of the field. Appropriately, the Information for authors has also changed over the years to reflect new concepts of what constitutes an acceptable publication.
Readers of the Journal may have noticed an addition to the 1999 Information for authors (1), under the heading Description of Analytical Methods and Results, namely:
"Analytical quality. Results obtained for the performance characteristics should be compared objectively to well-documented quality specifications, e.g., published data on the state of the art, performance required by regulatory bodies such as CLIA `88, or recommendations documented by expert professional groups". In addition, quality specifications can be derived from analysis of performance on clinical decision-making.
Many manuscripts deal with the development of new analytical methods or evaluation of commercially available analytical systems. Experimental designs and statistical techniques used to derive data on imprecision and bias are generally more than satisfactory. This is hardly surprising in view of the very many published protocols for the evaluation of methods (2). In contrast, objective analysis of whether the imprecision and bias are satisfactory is often less well done.
This is not a new phenomenon. Although the idea of utilizing quality specifications in assessing the acceptability of method performance was firmly stated in 1974 (3), it was pointed out more than a decade ago that few evaluators actually did compare the performance achieved with preset quality specifications (4)(5). This seems rather difficult to understand because quality specifications based on the state of the art (6), the views of an expert individual (7), and biological variation (8) had been available for many years. Indeed, even a superficial reading of the more recent literature would demonstrate that many papers, reviews, conference proceedings, and book chapters deal with the generation and application of quality specifications (9).
It may be that there are too many contradictory published recommendations, and it might not be easy for authors to select the most appropriate. In addition, new strategies to set quality specifications continue to appear, which might suggest that there is no ubiquitous professional consensus. Moreover, industry does not appear to use professionally set quality specifications as major considerations in either development or marketing.
In spite of these difficulties, there are many quality specifications against which experimental data can be compared, at least for tests reported on ratio (and difference) scales, and we suggest the following approach, which like the types of evidence and grading of recommendations used in clinical practice guidelines (10) can be placed in a hierarchy of objectivity, the best being first and the worst being last.
assessment of the effect of analytical performance on clinical
decision-making
Quality specifications in specific clinical situations.
Ideally, quality specifications should be derived objectively from an
analysis of medical needs. Thus, the effect on clinical decision-making
of the analytical quality found in any evaluation should be the subject
of objective assessment. Unfortunately, this is very difficult.
Scandinavian approaches (11) have shown how the calculations
can be done for a variety of analytes in a number of different clinical
settings. Others, notably Klee (12), have performed very
detailed and useful studies using similar approaches. When such clear
strategies can be identified, this is probably the best possible
approach; however, a major disadvantage is that only a few tests are
used in single well-defined clinical situations with standard
well-accepted medical strategies directly related to the test result.
Another significant drawback is that the quality specifications
calculated depend very much on the assumptions made regarding how test
results are used by clinicians (13); therefore, they must be
related to well-characterized strategies. Quality specification have
been derived from questionnaires using vignettes submitted to
clinicians on the use of individual tests in specific clinical
situations: these studies have very serious flaws, which have been
discussed previously by us (9), and we do not recommend
their use.
General quality specifications based on medical needs.
Consideration of the two major clinical settings in which test results
are used, namely monitoring individual patients and diagnosis using
reference intervals, shows that generally applicable quality
specifications might best be based on the components of biological
variation, namely, within-subject (CVI) and
between-subject (CVG) variation (14).
A very widely held view is that imprecision should be <0.50 CVI and bias should be <0.25(CVI2 + CVG2)1/2. This strategy has advantages in that data on components of biological variation are easily available on >180 quantities (15). This concept has been expanded recently (16).
For imprecision, although desirable performance is defined as CVA < 0.50CVI, users of quality specifications based on biology might consider that optimum performance could be defined by CVA < 0.25CVI, and that the more stringent quality specifications generated using this formula should be used for those quantities for which the desirable performance standards were easily achieved with current technology and methodology, and that minimum performance could be defined by CVA < 0.75CVI, and that the less stringent quality specifications generated using this formula should be used for those quantities for which the desirable performance standards were not attainable with current technology and methodology.
An additional basic concept is that, ideally, laboratories throughout a homogeneous population area should use exactly the same reference intervals. This was first proposed by Gowans et al. (17), who showed that for this to be achieved, the bias (BA) should be <0.250(CVI2 + CVG2)1/2. Analogously to imprecision, this quality specification for bias can be termed desirable performance. However, users of quality specifications based on biology might consider that optimum performance could be defined by BA < 0.125 (CVI2 +CVG2)1/2, and that the more stringent quality specifications generated using this formula should be used for those quantities for which the desirable performance standards are easily achieved with current technology and methodology, and that minimum performance could be defined by BA < 0.375(CVI2 + CVG2)1/2, and that the less stringent quality specifications generated using this formula should be used for those quantities for which the desirable performance standards are not attainable with current technology and methodology.
professional recommendations
Guidelines from national or international expert groups.
A
small number of national or international professional groups have
proposed detailed quality specifications for imprecision and bias. The
recommendations made by the National Cholesterol Education Panel have
been used extensively as criteria of acceptability of new methods
(18). One European Working Group has proposed quality
specifications for use in the evaluation of analytical systems based
upon biological variation data as detailed above and the state of the
art attained by the best 20% of laboratories (19); these
have been widely used in Europe for judging the acceptability of
analytical systems for the past few years. Another European Working
Group has suggested quality specifications for reference methods when
used for validation of routine methods and for assigning values to
materials used in external quality assessment schemes (EQAS) or
proficiency testing (PT) programs (20). Such guidelines have
the major advantage of being based on very extensive laboratory and
clinical experience of a number of experts from different backgrounds
and with a variety of professional experiences. Moreover, they are
usually based on extensive discussion of existing or new scientific
theories or experimental data before publication. In addition, the
method by which the recommendations were generated is published in
the peer-reviewed literature, which allows users to evaluate the
objectivity of the process used to reach the conclusions.
Guidelines from expert individuals or institutional groups.
Quality specifications have been proposed in a number of sets of
published guidelines on what has become termed best practice or good
laboratory practice. These are often developed or presented at a single
consensus conference without significant discussion. Examples include
the recommendations made on analytes used in the assessment of
thyrometabolic status (21) and on therapeutic drug
monitoring (22). These guidelines have an advantage in that
they usually are based on the very extensive laboratory and clinical
experience of an expert or an expert group from a single institution.
Although a disadvantage is that these guidelines may be somewhat
subjective and not based on scientific theory or experimental data, at
least the procedure used to generate the recommendations is published,
which again allows users to evaluate the objectivity the conclusions.
quality specifications laid down by regulation or by eqas
organizers
Quality specifications laid down by regulation.
A number of
countries have detailed the exact levels of performance required for an
analytical technique to be judged as acceptable. The US CLIA `88
legislation (23) documents acceptable total error for a
number of commonly assayed analytes, and it is easy to calculate [for
example, by using the approaches of Westgard et al. (24)]
whether the performance characteristics achieved for a new method meet
the criteria to pass. Similar legislation exists in Germany
(25), but the quality specifications are very different. The
advantage of this approach is that these standards, particularly the
CLIA `88 approaches, are well known and understood. However, in
contrast to the specifications used in Germany, which are clearly based
on the theoretical grounds developed by Stamm (26), a
disadvantage of the CLIA `88 quality requirements is that they appear
to be based upon the state of the art and therefore reflect what is
achievable rather than what is desirable.
Quality specifications laid down by EQAS organizers.
EQAS
organizers use a variety of measures of location and allowable
dispersion. In Europe (27), some countries use statistical
analysis of the data returned from the participant laboratories;
however, fixed limits are increasingly used. Fixed limits of
acceptability are also widely used in other parts of the world. Again,
it would be easy to calculate, using the approaches of Westgard et al.
(24), whether the analytical performance achieved in an
evaluation of a method would allow the laboratory to perform inside or
outside the fixed limits of acceptability. The major disadvantage of
these quality specifications is that, although often based on expert
opinion, they tend to be empirical and are clearly influenced by what
is actually achievable at the time.
published data on the state of the art
Published data from external quality assessment and proficiency
testing schemes.
Comparison of analytical quality could be
accomplished through reference to the performance achieved by groups of
laboratories participating in EQAS and PT. This has an advantage in
that many data are often available. However, the documented analytical
performance may not truly reflect the state of the art because the
materials used in the challenges may not behave exactly as specimens
from patients because of matrix effects, and participants may adopt
special analytical techniques to ensure good performance. In addition,
a prerequisite for evaluation of bias is that the materials are as
genuine as possible (for example, fresh-frozen human serum) with
traceable concentration (or other) values assigned by a reference
method or transferred by a reliable method from a reference material
(certified when this is available). Moreover, the state of the art
inferred from such schemes will change with time (and not always for
the better). Furthermore, performance achieved analytically may bear no
relationship to actual medical needs.
Published methodology.
Comparison may be done by reference to
performance documented in original works on similar or other methods
for measurement of the analyte(s) under investigation. This also has an
advantage in that many data are often available, but has a disadvantage
in that the method evaluations are carried out under optimal conditions
and the performance documented in the laboratory of the originator or
the original evaluator may be the best possible rather than that
achieved in practice. Again, performance achieved analytically may bear
no relationship to actual medical needs.
Now that the requirement to compare analytical quality to some type of quality specifications has been published in the Information for authors, those preparing their work for submission for publication should follow it. Authors should recognize that it might be difficult to assess the scientific or clinical validity of some of the published quality specifications: it is hoped that in the future those who prepare recommendations, particularly regulators, make their methodology completely transparent. For now, authors should compare their results to quality specifications based on explicitly stated criteria.
References
The following articles in journals at HighWire Press have cited this article:
![]() |
T. S.M. Reijn, M. O. Rikkert, W. J.A. van Geel, D. de Jong, and M. M. Verbeek Diagnostic Accuracy of ELISA and xMAP Technology for Analysis of Amyloid {beta}42 and Tau Proteins Clin. Chem., May 1, 2007; 53(5): 859 - 865. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. S. Vasan Biomarkers of Cardiovascular Disease: Molecular Basis and Practical Considerations Circulation, May 16, 2006; 113(19): 2335 - 2362. [Full Text] [PDF] |
||||
![]() |
D. K. Talwar, M. K. Azharuddin, C. Williamson, Y. P. Teoh, D. C. McMillan, and D. St. J. O'Reilly Biological Variation of Vitamins in Blood of Healthy Individuals Clin. Chem., November 1, 2005; 51(11): 2145 - 2150. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Q. Hanson, J. H. Eckfeldt, K. Schwichtenberg, O. Aras, and M. Y. Tsai Interlaboratory Variation of Plasma Total Homocysteine Measurements: Results of Three Successive Homocysteine Proficiency Testing Surveys Clin. Chem., September 1, 2002; 48(9): 1539 - 1545. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. B. Sacks, D. E. Bruns, D. E. Goldstein, N. K. Maclaren, J. M. McDonald, and M. Parrott Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes Mellitus Clin. Chem., March 1, 2002; 48(3): 436 - 472. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Phillipov and P. J. Phillips Components of Total Measurement Error for Hemoglobin A1c Determination Clin. Chem., October 1, 2001; 47(10): 1851 - 1853. [Full Text] [PDF] |
||||
![]() |
J. C. Boyd and D. E. Bruns Quality Specifications for Glucose Meters: Assessment by Simulation Modeling of Errors in Insulin Dose Clin. Chem., February 1, 2001; 47(2): 209 - 214. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. Fuentes-Arderiu, J. Miro-Balague, P. Hyltoft Petersen, and C. G. Fraser State of the Art Instead of Biological Variation to Set Requirements for Imprecision Following is a reply to the above letter by X. Fuentes-Arderiu and J. Miro-Balague Clin. Chem., October 1, 2000; 46(10): 1715 - 1717. [Full Text] [PDF] |
||||
![]() |
E. Nexo, F. Engbaek, P. M. Ueland, C. Westby, P. O'Gorman, C. Johnston, B. F. Kase, A. B. Guttormsen, I. Alfheim, J. McPartlin, et al. Evaluation of Novel Assays in Clinical Chemistry: Quantification of Plasma Total Homocysteine Clin. Chem., August 1, 2000; 46(8): 1150 - 1156. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Talwar, H. Davidson, J. Cooney, and D. St. JO'Reilly Vitamin B1 Status Assessed by Direct Measurement of Thiamin Pyrophosphate in Erythrocytes or Whole Blood by HPLC: Comparison with Erythrocyte Transketolase Activation Assay Clin. Chem., May 1, 2000; 46(5): 704 - 710. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. H. Garde, A. M. Hansen, L. T. Skovgaard, and J. M. Christensen Seasonal and Biological Variation of Blood Concentrations of Total Cholesterol, Dehydroepiandrosterone Sulfate, Hemoglobin A1c, IgA, Prolactin, and Free Testosterone in Healthy Women Clin. Chem., April 1, 2000; 46(4): 551 - 559. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Cembrowski and R. N. Carey Adding Value to Proficiency Testing Programs Clin. Chem., January 1, 2000; 46(1): 7 - 8. [Full Text] [PDF] |
||||
![]() |
C. M. Cobbaert and T. K. J. Luderer Total Error Evaluation of Roche Direct HDL-Cholesterol Reagent and Calibrator across 31 Lot Combinations: A 2-Year Experience Clin. Chem., January 1, 2000; 46(1): 133 - 134. [Full Text] [PDF] |
||||
![]() |