|
|
||||||||
Articles |
1
Research and Development, Ortho-Clinical Diagnostics, Cardiff Laboratories, Whitchurch, Forest Farm Estate, Cardiff CF4 7YT, Wales, UK.
a Author for correspondence. Fax 44 (0) 1222 526635; e-mail NCHRISTOFIDES{at}compuserve.com
| Abstract |
|---|
|
|
|---|
Methods: We used a direct equilibrium dialysis FT4 assay as the reference method and compared the results obtained with those of the FT4 assays under investigation, in patient sera having a wide range of sBC. We then compared the expected and observed FT4 results for sera diluted with an inert buffer. Because serum dilution causes a predictable decrease in sBC, an increasingly negative bias on progressive dilution is indicative of a sBC-dependent bias.
Results: The automated FT4 assay investigated (Vitros FT4) showed no demonstrable sBC-dependent bias by either test.
Conclusion: These two tests can be used to screen for sBC-dependent biases in FT4 assays.© 1999 American Association for Clinical Chemistry
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
The sBC and PBT4 concentrations of three sera were reduced in vitro by serially diluting (2- to 64-fold dilutions) sera in 10 mmol/L HEPES (Sigma Chemical; cat. no. H3375) buffer solution, pH 7.4. The sera chosen were from third-trimester pregnancies because these represented sera with high sBC and PBT4 concentrations. The undiluted and diluted samples were assayed in the two FT4 methods using standard protocols.
The automated FT4 method used was the Vitros Immunodiagnostic Products (Ortho-Clinical Diagnostics, Amersham UK). The manual direct ED method was the Nichols FT4 method (Nichols Institute Diagnostics). The physicochemical principles of the assays used are fundamentally different. The Vitros FT4 assay is a labeled antibody method (2)(14), and the Nichols assay is a direct equilibrium method (15).
The ED method is carried out by dialyzing the FT4 from 200 µL of serum into 2.4 mL of dialysis buffer (at 37 °C, over a 16- to 18-h incubation period). The FT4 in the protein-free dialysate is then quantified by a sensitive solid-phase RIA for T4. To minimize variability (e.g., assay-to-assay variation), all samples studied were analyzed on one occasion, with the samples randomized. The package insert quotes an intraassay CV at doses falling within and above the euthyroid range as being <13%. The euthyroid FT4 range, which was the observed range with one outlier deleted from each end, quoted in the package insert is 10.334.7 pmol/L.
In the Vitros FT4 assay, 25 µL of sample is
pipetted into microwells coated with a triiodothyronine
(T3)-protein conjugate, followed by 100 µL of a
sheep anti-T4 antibody labeled with horse-radish
peroxidase in a 150 mmol/L phosphate buffer containing 1.0 g/L
bovine gelatin and 1.0 g/L bovine
-globulin. During the
16-min incubation at 37 °C, a proportion of the labeled antibody
binds to the serum FT4 and to the well surface,
with the amount binding to the well surface being inversely related to
the serum FT4 concentration. The well is then
washed, and a signal reagent that produces luminescence is then added;
the resulting light emitted is measured in a luminometer. All
procedures are carried out automatically by the Vitros ECi
Immunodiagnostic system. The samples were assayed in a batch mode, with
quality-control samples at three concentrations run at the
beginning and end of the assay. The within-run imprecision for
FT4 values in the euthyroid and hyperthyroid
range is quoted in the package insert as being <3%. The assay was
calibrated using an in-house ED assay. The euthyroid range (1 and 99
percentiles) quoted in the package insert is 1028.2 pmol/L.
In addition to the FT4 assays, the patient sera were also analyzed in the Vitros Immunodiagnostic Products Total T4 (TT4) and T3 uptake (T3U) assays. The euthyroid T4 range (2.5 and 97.5 percentiles) quoted in the package insert is 71.2141 nmol/L. The T3U assay is calibrated in %T3U units, which are inversely related to the serum binding capacity (i.e., a serum with a high %T3U has a low binding capacity). The euthyroid %T3U range (2.5 and 97.5 percentiles) quoted in the package insert is 23.540.6% uptake.
All assays were performed following the manufacturers' instructions.
statistics
Analysis of data was carried out by standard methods with
Microsoft Excel spreadsheets. The dependence of the
FT4 bias observed with any variable was
determined by correlating (linear regression analysis performed by the
least-squares method) the methodological difference (from ED
FT4) of each patient (y-axis) against
the variable studied (x-axis). The sBC was calculated by
dividing the Vitros TT4 concentration by the ED
FT4 concentration (9). The
concentration of PBT4 was estimated by
subtracting the concentration of FT4, as measured
by ED, from the concentration of TT4 (9). Because the
FT4 concentration relative to TT4 was small
(<0.1% of the TT4), the PBT4 concentration
showed little difference from the TT4 concentration. The agreement
between the Vitros FT4 and the manual ED
FT4 was tested by Deming regression
(16). A measure of the sBC was also obtained by the T3U,
where the %T3U is inversely related to the binding capacity. The
Student unpaired t-test was used to compare the biochemical
profiles between the patient groups (within each assay format), and the
paired t-test was used when the comparison was across
different assay methods.
| Results |
|---|
|
|
|---|
The relationship between the Vitros FT4 and ED
FT4 assays in all patients studied is shown in
Fig. 1
. A significant correlation (r = 0.96;
P <0.001) between the two assays was observed. The equation
describing the relationship was:
|
Vitros FT4 = 0.85 (± 0.03; P <0.001) ED FT4 + 3.67 (± 0.61; P <0.001).
Good agreement between the FT4 methods was achieved in all patient groups.
The relationship between the Vitros FT4 to ED FT4 in (a) the ambulatory and pregnant subjects (groups combined, n = 44), and (b) the hospitalized patients (n = 25) is described by the following equations:
(a) Vitros FT4 = 0.9 (± 0.09) ED FT4 + 3.09 (± 1.19); r = 0.83;P <0.001
(b) Vitros FT4 = 0.89 (± 0.07) ED FT4 + 2.33 (±1.9); r = 0.94; P <0.001
Table 1
shows the mean (± SD) and observed ranges for the two
FT4 methods under investigation
(FT4 in pmol/L), the TT4 (in nmol/L), %T3U, and
sBC (in nmol/pmol) in the ambulatory, pregnant, and hospitalized
patients. The FT4 concentration, as measured by
both FT4 methods, in the hospitalized group was
significantly (P <0.001) higher than the
FT4 concentration in the ambulatory group. None
of the patients in the hospitalized group had FT4
concentrations below the corresponding euthyroid range of the two
FT4 assays examined. The calculated sBC in the
hospitalized group was significantly lower (P <0.001) than
that found in the ambulatory group, whereas the corresponding %T3U
value was significantly higher (P <0.001). Thus, both
measures of serum T4 binding support the view
that hospitalized patients have decreased serum
T4 binding capacities. The TT4 concentration in
the hospitalized group was significantly lower (P <0.001)
than that obtained in the ambulatory group, and 12 of the 25
hospitalized patients had TT4 concentrations below the euthyroid range.
The FT4 concentration in the pregnancy group was
significantly lower (P <0.001) than in the ambulatory group
with both FT4 methods, whereas the sBC and TT4
concentrations were increased (P <0.001). As expected, the
%T3U was significantly (P <0.001) reduced.
|
The FT4 concentrations obtained in the pregnancy group by the Vitros method were higher than the corresponding concentration obtained with the ED method. These differences, although small (mean difference, 2.6 pmol/L; range of difference in 95% confidence intervals, 2.13.2 pmol/L), were statistically significant (P <0.001). Similar differences from ED FT4 were seen in the ambulatory group (mean difference, 1.6 pmol/L; range of differences in 95% confidence intervals, 0.7 to 2.6 pmol/L; both P <0.001). The Vitros FT4 concentration in the hospitalized group was not significantly different (P >0.05) from the mean concentration obtained in the ED method.
Fig. 2
depicts the relationship between the Vitros
FT4 bias and sBC in all the patient samples. The
regression equations derived in each individual category of subjects
are shown below:
|
Ambulatory group: Vitros FT4 bias = 0.77 (± 0.43) sBC - 3.01 (± 2.62); r = 0.34; P >0.05
Pregnant group: Vitros FT4 bias = 0.167 (± 0.077) sBC - 0.079 (± 0.167); r = 0.475; P <0.05
Hospitalized group: Vitros FT4bias = -0.31 (± 0.665) sBC + 0.29 (± 0.182); r = 0.098; P >0.05
The slopes and intercepts of all the relationships did not differ significantly from each other (P >0.05). The regression equation describing the relationship of the Vitros FT4 bias and sBC (the only individual patient category that yielded a significant slope and r value, both at P <0.05, was the pregnancy group) indicates that at the highest observed sBC (23.9 nmol/pmol), the Vitros FT4 will be positively biased by 4 pmol/L (95% confidence interval, 2.65.4 pmol/L).
There was no apparent relationship (P >0.05) between the PBT4 concentrations and the Vitros FT4 concentrations in any of the patient groups studies.
Fig. 3
depicts the relationship between the Vitros
FT4 bias and %T3U in all patients studied. The
equations derived in each of the individual patient categories are
summarized below:
|
Ambulatory group: Vitros FT4 bias = -0.04 (± 0.22) %T3U + 2.75 (± 6.46); r = 0.035; P >0.05
Pregnant group: Vitros FT4 bias = 0.003 (± 0.199) %T3U + 2.61 (± 3.78); r = 0.003; P >0.05
Hospitalized group: Vitros FT4 bias = 0.06 (± 0.066) %T3U - 3.37 (± 3.18); r = 0.19; P >0.05
None of these relationships reached statistical significance.
The results of the serum dilution experiment performed on the pregnancy
sera are shown in Fig. 4
. The theoretically derived FT4
concentrations suggest that the decrease after a 32-fold dilution will
be <1%. Similarly, the FT4 concentration as
measured by ED was only slightly affected by serum dilution. At
dilutions of 1:16 and 1:32 (i.e., dilutions that are expected to
decrease the sBC to levels usually seen in some hospitalized patients)
the ED FT4 was within 5% of the concentration
obtained in the undiluted sample. The Vitros FT4
concentrations at these dilutions did not vary significantly (<5%)
from the FT4 concentration obtained in the
undiluted sample.
|
| Discussion |
|---|
|
|
|---|
We have used two simple tests to examine the bias of an automated (Vitros) FT4 method. The first test involved the comparison of the FT4 results obtained in the Vitros method in patients having a wide range of binding capacities against those obtained in a commercial direct ED system. The main reason for choosing ED as the reference method is that it is one of the methods, in addition to ultrafiltration, generally considered as the "gold standard" method for free thyroid hormone measurement. However, even these gold standards have their own inconsistencies and technical weaknesses (15)(29)(30)(31)(32). Thus, the elevation of this particular ED method as the definitive gold standard method is arguable. Nonetheless, the validity of the ED method chosen has been well documented (15), although even this ED may be biased in some nonthyroidal illness sera because of dilution effects during the dialysis step (33). The ED method we used produces a 13-fold dilution of dialyzable substances in sera. The patient categories included in the study have been documented to have high (i.e., third- trimester pregnancy) and low (i.e., hospitalized patients) binding capacities (13). This binding capacity profile has been confirmed in the present study by two independent methods (the sBC, derived by dividing the TT4 result by the ED FT4 result, and the %T3U). The second test used examined the effect of serum dilution on the FT4 concentration obtained by the methods under investigation (Vitros and ED FT4). The sample dilution test was chosen for several reasons: (a) one can alter the sBC in a predictable fashion (e.g., a twofold dilution will reduce the sBC by ~50%); (b) one can readily predict the ideal performance of an assay and thus eliminate the need to compare results with those obtained by a reference method; (c) one can "manufacture" serum dilution pools whose sBCs mimic the range of sBCs found in patients [e.g., the sBC concentration range in patients undergoing thyroid testing has been shown to be ~30-fold (13); thus dilution of a pregnant serum by this factor will encompass the sBC likely to be experienced by a laboratory]; and (d) this test can easily be performed by any user.
The law of mass action (2)(3) dictates that the
concentration of FT4 in serum depends on the
equilibrium that exists between the PBT4 and the
concentration and affinity (i.e., the sBC) of the free binding sites:
![]() |
In estimating the sBC (i.e., PBT4/FT4) we made the same assumption as Nelson and co-workers (9)(13), which is that the FT4 as measured by ED gives an unbiased estimate of the true FT4. As discussed previously, this assumption is arguable. The highest and lowest individual sBC values in our study population were 23.9 and 1.2 nmol/pmol, which represent a range of 19.9-fold. Thus, it is clear from this and other studies (9)(13) that patients undergoing thyroid function tests possess a very wide range of serum T4 binding (and consequently wide ranges of PBT4). The results presented here show that the FT4 concentrations obtained by the Vitros assay (relative to ED), were not dependent on T4 binding (sBC or T3U) or PBT4. The small positive bias (from ED) seen in pregnancy is likely because of calibration differences. This is supported by the fact that the corresponding FT4 concentrations in the ambulatory group were also positively biased to a similar degree.
In the second test, we examined the FT4 biases of the two methods (Vitros and ED FT4) by analyzing sera whose protein concentrations were decreased in vitro. Lowering of the sBC by decreasing the concentration of the binding proteins was accomplished by diluting sera in an inert buffer (HEPES). The FT4 concentrations as measured by both the ED and Vitros methods were not reduced by the serum dilution.
The estimation of serum FT4 by immunoassay, irrespective of the methodology used will invariably disturb the normal equilibrium between PBT4 and sBC (2)(3). This will come about as a result of the addition of the antibody and the dilution of the sample in assay reagents that may also contain T4 binders such as albumin or animal sera, which are often added to protect assays from interferences from heterophilic antibodies. These additions will lead to the establishment of a new equilibrium and a new "in vitro" FT4 concentration. The concentration of this in vitro FT4 will be dictated by the in vitro bound T4 and unbound binding site concentrations. The in vitro bound T4 will be the net sum of PBT4 and T4 bound by the immunoassay reagents (iPBT4), and the in vitro unbound binding sites will include, in addition to sBC, the free antibody-binding sites and the free binding sites of other binders included in the reagents [i.e., immunoassay binding capacity (iBC), which will be equal to the affinity x concentration of the free immunoassay binding sites]. Thus, the in vitro FT4 will be equal to (PBT4 + iPBT4)/(sBC + iBC). From this formula, one can predict that a high iBC will cause biases that will be dependent on the magnitude of the endogenous sBC. Thus, as the sBC decreases (as seen in hospitalized patients and mimicked by serum dilution), the assay will yield increasingly negative results. The results presented show a lack of sBC-dependent biases in both the Vitros and the ED FT4 assays, suggesting that the in vitro disturbance of the T4/protein equilibrium induced by both assays is negligible.
The tests described in the present study can be used to assess the presence and magnitude of sBC-dependent biases of other commercial FT4 methods.
| Footnotes |
|---|
1 Nonstandard abbreviations: T4, thyroxine; FT4, free thyroxine; ED, equilibrium dialysis; PBT4, protein-bound thyroxine; sBC, serum binding capacity; T3, triiodothyronine; TT4, total thyroxine; T3U, triiodothyronine uptake; iPBT4, protein-bound thyroxine in the immunoassay; and iBC, immunoassay binding capacity. ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. Waise and H. C Price The upper limit of the reference range for thyroid-stimulating hormone should not be confused with a cut-off to define subclinical hypothyroidism Ann Clin Biochem, March 1, 2009; 46(2): 93 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Sapin and M. d'Herbomez Free Thyroxine Measured by Equilibrium Dialysis and Nine Immunoassays in Sera with Various Serum Thyroxine-binding Capacities Clin. Chem., September 1, 2003; 49(9): 1531 - 1535. [Full Text] [PDF] |
||||
![]() |
R. Sapin, J.-L. Schlienger, F. Gasser, E. Noel, B. Lioure, F. Grunenberger, B. Goichot, and D. Grucker Intermethod Discordant Free Thyroxine Measurements in Bone Marrow-transplanted Patients Clin. Chem., March 1, 2000; 46(3): 418 - 422. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |