|
|
||||||||
Articles |
Departments of
1
Pathology and
2
Medical and Research Technology, University of Maryland School of Medicine, Baltimore, MD 21201.
3
Dade Behring, West Sacramento, CA 95691.
4
Department of Pathology, Washington University School of
Medicine, St. Louis, MO 63110.
5
Dade Behring, Glasgow, DE 19714.
6
Becton Dickinson Vacutainer Systems, Franklin Lakes, NJ
07417.
7
Department of Laboratory Medicine and Pathology,
Hennepin County Medical Center, Minneapolis, MN 55415.
8
Chiron Diagnostics, Inc., East Walpole, MA 02032.
9
Genzyme Diagnostics, Cambridge, MA 02139.
10
Department of Pathology and Laboratory Medicine,
Hartford Hospital, Hartford, CT 06102.
a Address correspondence to this author at: Clinical Pathology, University of Maryland School of Medicine, 22 S. Greene St., Baltimore, MD 21201. Fax 410-328-5880; e-mail rchriste{at}umaryland.edu
| Abstract |
|---|
|
|
|---|
Methods: Three protocols were used. In protocol I, various CK-MB materials prepared in different matrices were screened as candidate standards. In protocol II, participating manufacturers calibrated their systems with concentrates of human heart CK-MB and then tested 20 patient samples to evaluate calibration bias. In protocol III, participating manufacturers calibrated their immunoassay systems using recombinant CK-MB2 (rCK-MB2) diluted into their respective sample diluents and measured 50 samples.
Results: Candidate materials showed high recovery in stripped human serum, but bias improved only from 59% to 38%. These data led to the use of human heart CK-MB diluted in each manufacturer's sample diluent. This strategy reduced bias from 31% to 15%. Because human heart CK-MB is difficult to provide, a lyophilized source of CK-MB2 was identified. rCK-MB2 was shown by sodium dodecyl sulfate-polyacrylamide gel electrophoresis, reversed-phase HPLC, intrinsic protein fluorescence, circular dichroism, agarose gel electrophoresis, immunoreactivity studies, high and low temperature stability, and reconstituted stability to be equivalent to human heart CK-MB. Calibration of immunoassay systems with rCK-MB2 added into each respective manufacturer's sample diluent showed a 13% between-manufacturer bias.
Conclusion: Lyophilized rCK-MB2 was determined suitable for use as a reference material for CK-MB mass assays.
| Introduction |
|---|
|
|
|---|
|
Split patient sample correlation studies have shown that commercially available CK-MB immunoassays correlate well (r >0.98) but are biased from one another, which is reflected by differences in the slopes of comparative analyses. When Henderson et al. (5) monitored the analytical performance of total CK (~230 participants) and CK-MB (~160 participants), they reported a wide dispersion of results between different analyzers as well as among identical analyzers. A major reason for such between-method bias is the lack of a CK-MB mass standard. To address this issue, the CK-MB Mass Assay Standardization Subcommittee was formed in January 1992 by the AACC to "identify or develop a standard material using human CK-MB that can be used to reduce laboratory-to-laboratory variation in the accuracy of the CK-MB mass assays".
The goals of this committee were to identify and validate a source of CK-MB standard, to establish purification and protein determination protocols, and to investigate the effect that various matrices and CK-MB materials (liquid or lyophilized) have on standardizing CK-MB mass assays. With cooperation among manufacturers, it is hoped that this reference material will be used to standardize assay methods such that between-assay bias will be reduced or eliminated.
| Materials and Methods |
|---|
|
|
|---|
sodium dodecyl sulfate-polyacrylamide gel electrophoresis analysis
Native heart CK-MB and rCK-MB2 were analyzed by sodium dodecyl
sulfate-polyacrylamide gel electrophoresis (SDS-PAGE) to ascertain
protein purity on 12% Tris-HCl ready gels (Bio-Rad). Gels were run on
MiniProtein II systems (Bio-Rad) and stained with Coomassie blue.
agarose gel electrophoresis
Agarose gel electrophoresis was used to verify CK isoenzyme purity
via separation by overall protein charge. Native heart CK-MB and
rCK-MB2 were applied to Universal Gel/8 agarose gels and
electrophoresed according to the manufacturer's protocol (Chiron
Diagnostics). Gels were stained for CK activity using the
Creatine Phosphokinase Isoenzymes kit (Sigma Diagnostics).
reversed-phase hplc
The purity and hydrophobic nature of native heart CK-MB and
rCK-MB2 were analyzed by reversed-phase HPLC using triplicate 10-µg
sample injections on a 2.1 x 30 mm Brownlee RP-300
(C5) column on a Hewlett Packard 1090 HPLC system
with photodiode array detection (Hewlett Packard). Proteins were eluted
at a flow rate of 0.25 mL/min, using a linear gradient of 0.8
g/L trifluoroacetic acid in acetonitrile and detected at
214 nm. Hewlett Packard Chemstation software was used to integrate the
data.
intrinsic protein fluorescence
The structural environments surrounding tryptophan residues in the
proteins were assessed by measuring the intrinsic protein fluorescence
of native heart CK-MB and rCK-MB2 at ambient temperature with a
McPherson FL-750 spectrofluorescence detector (McPherson). Fluorescence
excitation spectra were obtained by scanning the sample, 10 µmol/L
CK-MB, from 200 to 300 nm at an emission maxima of 320 nm. The
respective fluorescence emission spectra were measured by scanning from
300 to 400 nm at an excitation maxima of 290 nm. The relative
fluorescence intensity data for both native heart CK-MB and rCK-MB2
were imported into a graphics program and displayed in overlapping
plots.
circular dichroism
The protein-folding conformation of native heart CK-MB and rCK-MB2
was analyzed by measuring their circular dichroism at 25 °C on a
JASCO J720 spectropolarimeter (Jasco). A far-ultraviolet circular
dichroism spectrum was obtained by scanning 0.5 g/L samples from 178 to
260 nm with a 0.005-cm pathlength.
immunoreactivity
The doseresponse of native heart CK-MB and rCK-MB2 added
to serum was assessed using an IMx immunoanalyzer. A 5 g/L working
stock solution of CK-MB was diluted into heat-inactivated newborn calf
serum (Genzyme Diagnostics) to 0, 3, 10, 30, 100, and 300 µg/L CK-MB.
These samples were assayed in duplicate in separate IMx analytical runs
using the same stored calibration curve. Their fluorescence reaction
rates were plotted against standard dilution factors. The native heart
CK-MB and rCK-MB2 response curves in the immunoassay were overlayed for
comparison.
accelerated stability studies of rCK-MB2
To assess the storage shelf-life of lyophilized rCK-MB2 at
28 °C, accelerated stability studies were conducted by stressing
vials of rCK-MB2 in isothermal incubators set at 45, 37, and 30 °C
at staggered time intervals. For each temperature, stressed vials were
reconstituted at the same time with Genzyme Diluent (Genzyme
Diagnostics), diluted into assay range with heat-inactivated newborn
calf serum (Genzyme Diagnostics), and assayed in duplicate in the same
analytical run, using the IMx CK-MB mass assay. The percentages of
recovery were determined relative to that of the value for non-stressed
CK-MB stored at 28 °C.
The rate constant for the loss of CK-MB mass at each incubation temperature was determined by linear regression analysis of the time plot data using a graphical program. An Arrhenius plot was constructed using the logarithm of the rate constants against the inverse of incubation temperature in degrees Kelvin. From the Arrhenius plot, the rate constant for the loss of rCK-MB mass at 8 °C can be extrapolated by linear regression analysis and used to calculate the time for rCK-MB mass to decrease 10% at 8 °C (9).
reconstitution stability of rCK-MB2
The reconstituted or opened vial stability for rCK-MB2 was
performed by reconstituting two vials with Genzyme CK-MB Diluent
(Genzyme Diagnostics), storing them at 28 °C, and removing
aliquots from these vials at different time intervals for assay in
duplicate, using an IMx immunoanalyzer. The reconstituted product
stability of rCK-MB2 at 28 °C was measured by the decrease in
CK-MB mass to 90% of its day 0 value.
shelf-life study
Lyophilized rCK-MB2 was stored at 4 and -20 °C. At each
selected time point over a period of ~500 days, rCK-MB2 was
reconstituted in aca Plus sample diluent, and its mass was
measured using the aca Plus immunoanalyzer. A 10% loss of
CK-MB mass in vials stored at 4 °C relative to that stored at
-20 °C was used as the stability criterion.
| Results |
|---|
|
|
|---|
phase 1: screening of ck-mb candidates
CK-MB candidates.
Four different forms of CK-MB were
considered for the development of a CK-MB standard: human heart CK-MB,
hybrid CK-MB, rCK-MB2 (tissue isoform), and rCK-MB1 (serum
isoform, missing the C-terminal lysine on the M subunit). The hybrid
CK-MB, prepared by reassociation of the respective M and B subunits of
denatured CK-MM and CK-BB, was obtained from Aalto Scientific.
Different recombinant CK-MB forms were provided by Dr. M. Benjamin
Perryman (University of Colorado Health Science Center, Denver, CO)
(10). In addition, during the later stage of evaluation
(phase 3), lyophilized rCK-MB2 was provided by Genzyme Diagnostics.
Purification of CK-MB.
Human heart CK-MB, rCK-MB2, and rCK-MB1
were purified using anti-CK-MB monoclonal antibody immunoaffinity
chromatography according to established procedures (11). The
final preparations of CK-MB were stored at -20 °C in 20 mmol/L
Tris-HCl, pH 7.2 containing 500 mL/L glycerol. The purity of all
CK-MB preparations was assessed by SDS-PAGE and agarose gel
electrophoresis.
Determination of CK-MB concentration.
From the protein
sequence of CK-MB, one can determine that the B subunit of CK contains
4 tryptophan, 9 tyrosine, and 2 cystine amino acid residues, whereas
the M subunit of CK contains 4 tryptophan, 10 tyrosine, and 2 cystine
amino acid residues (6)(10). Based on this amino
acid content, the molar absorptivity of CK-MB was estimated as 0.82
L/cm-mol (absorbance of 1 g/L solution of CK-MB at 280 nm using
1-cm pathlength) (12).
The concentration of the purified CK-MB was determined spectrophotometrically on a Model 8450A Spectrophotometer (Hewlett Packard).
Lyophilized and liquid forms of human heart native CK-MB, rCK-MB1, rCK-MB2, and hybrid CK-MB were evaluated in different matrices, which yielded equivalent recovery results in immunoassays. Lyophilized rCK-MB2 was selected for use in all subsequent studies because of its long-term stability and ease of shipping and handling.
phase 2: determination of optimum calibrator matrix
Matrix selection.
Three different potential sample matrices
were evaluated: an unprocessed CK-MB-free human serum pool, stripped
serum, and a synthetic matrix (50 mmol/L phosphate buffer containing 50
g/L bovine serum albumin and 0.15 mol/L NaCl, pH 7.4). The
stripped human serum was prepared by mixing a human serum pool with a
blend of anion- plus cation-exchange resin to extract residual CK. Some
other serum components may have been removed unintentionally during
this process. The stripped human serum was donated by Dade Behring
(Glasgow, DE). Manufacturers' sample diluents were subsequently
evaluated as the study progressed.
Formulation.
To optimize the formulation of a standard for use
in all CK-MB mass assays, a screening experiment was designed in which
the four sources of CK-MB and the three matrices were evaluated. In
this study, formulations containing 0, 10, or 100 µg/L of human heart
CK-MB, hybrid CK-MB, rCK-MB1, or rCK-MB2 were prepared in three
matrices. These samples were shipped to nine participating
manufacturers of CK-MB mass assays. At each manufacturer's site, each
vial was hydrated with 2 mL of deionized water, and the CK-MB masses of
the reconstituted candidate standards were determined on their
respective immunoassay systems.
Manufacturers' correlation study: protocol I.
Purified,
lyophilized human heart CK-MB candidate standards (0, 10, and 100
µg/L) prepared in stripped human serum were assayed on each
immunoassay analyzer and later used to correct original calibration
curves via regression analysis. Additionally, 69 patient samples were
analyzed for CK-MB mass. For these studies, the following seven
immunoassay systems were used: ACS:180 (Chiron Diagnostics), Stratus II
(Dade-Behring Diagnostics; analysis performed at the Glasgow, DE
site), Tandem-E (Hybritech), OPUS Plus (Dade Behring; analysis
performed at the Glasgow, DE site), aca Plus (Dade Behring;
analysis performed at the Wilmington, DE site), AIA-600 (TOSOH), and
IMx (Abbott Diagnostics). Assays were conducted according to respective
manufacturers' specifications.
Manufacturers' correlation study: protocol II.
In an
alternative approach for the standardization of CK-MB mass assays, the
manufacturers of these assays were provided 20 patient sample pools
(not used in the study described above) and purified human heart CK-MB
from which each participating manufacturer prepared CK-MB candidate
standards at 0, 10, 50, 100, and 200 µg/L in their sample diluent.
The standards and the samples were analyzed in duplicate in their
respective immunoassay systems. The slope bias of these methods against
IMx was determined by linear regression analysis.
For the different calibrator matrices, CK-MB recovery (data not shown)
varied over a wide range as follows: stripped human serum (60100%),
human serum (2360%), and synthetic medium (2988%). Although
stripped human serum produced the highest CK-MB recovery, the
utilization of this matrix for reconstitution of CK-MB reference
material only slightly reduced the slope range from ± 59%
to ± 38% (Fig. 1
). The slight reduction in between-manufacturer bias achieved
with candidate standards made with stripped human serum is probably
attributable to the variable effect of this matrix in different
immunoassays. After dilution of human heart CK-MB reference material
using the manufacturers, sample diluents, the range of slopes for the
patient correlation plots was reduced from ± 31% to ± 15%
(Fig. 2
).
|
|
Because all CK-MB isoforms were recovered similarly in different matrices and native human CK-MB preparations are difficult to obtain and prone to batch-to-batch variability, rCK-MB2 is proposed as a reference material.
phase 3. preparation and properties of rCK-MB2
After phases 1 and 2 were completed, lyophilized rCK-MB2 was
evaluated as a CK-MB reference standard. To validate its similarity to
native human heart CK-MB, this material was analyzed in structural,
immunological, and stability studies and compared to native CK-MB
obtained from either BioProcessing, Inc., or provided by Dr. Jack
Ladenson (Washington University, St. Louis, MO) (11).
rCK-MB2 was expressed and purified from E. coli. The
purified rCK-MB2 has an apparent molecular mass similar to
native heart CK-MB, as shown by SDS-PAGE analysis (Fig. 3
). The purity and structural similarity of rCK-MB2 to native
heart CK-MB was further demonstrated by agarose gel electrophoresis
(Fig. 4
) and reversed-phase HPLC (Fig. 5
) analyses.
|
|
|
The intrinsic protein fluorescence spectra for native heart CK-MB and
rCK-MB2 also are similar (Fig. 6
). Circular dichroism studies of the native and recombinant
CK-MB proteins revealed no distinguishable secondary structure (data
not shown).
|
The immunoreactivity responses of native heart CK-MB and rCK-MB2 were
examined using the IMx CK-MB mass assay, which showed that both
proteins have similar immunoreactivity (Fig. 7
).
|
Different studies were performed to evaluate the stability of rCK-MB2.
In accelerated stability studies (Fig. 8
), the percentages of recovery were determined relative to the
value for CK-MB stored at 28 °C. The rate constant for the loss of
rCK-MB mass at 8 °C was extrapolated by linear regression analysis
from the Arrhenius plot and used to calculate the time for rCK-MB mass
to decrease 10% at 8 °C. Studies of lyophilized rCK-MB2 showed that
the refrigerated product shelf-life was 19 years (Fig. 9
). The reconstituted product stability of rCK-MB2 at 28 °C
was at least 21 days as measured by the decrease in CK-MB mass to 90%
of its day 0 value (Fig. 10
). The shelf-life of rCK-MB2 at 4 °C, as determined relative
to its mass at -20 °C, was 555 days (data not shown).
|
|
|
phase 4: final rCK-MB2 EVALUATION
Each of the participating manufacturers was provided lyophilized
rCK-MB2 and its diluent and 50 patient sample pools. Manufacturers
reconstituted the rCK-MB2, prepared rCK-MB2 candidate standards at 0,
10, 50, 100, and 200 µg/L in their respective sample diluents, and
determined the CK-MB mass values of the rCK-MB2 candidate standards and
50 samples. Data were analyzed for between-manufacturer bias by
assessing patient sample correlation plots.
A final correlation study was performed to evaluate the lyophilized
rCK-MB2 on different CK-MB mass assays. As shown in Fig. 11
, before using rCK-MB2 to establish the calibration curves for
different CK-MB immunoassays, the range of slopes was ± 40%.
This slope range was reduced to ± 13% after rCK-MB2 was used to
establish the calibration curves for the immunoassays.
|
| Discussion |
|---|
|
|
|---|
Initially, the catalytic activity of CK in serum or plasma was measured to evaluate the diagnosis of AMI and to rule out cardiac necrosis. Recently, Gella et al. (13) described the preparation of lyophilized CK-MB purified from human heart for use as a reference material for CK-MB catalytic concentration measurements. Upon reconstitution, the catalytic concentration of the reference CK-MB material was reported to be ~67 U/L when measured at 30 °C by the recommended method of the International Federation of Clinical Chemistry (13).
CK-MB mass assays were developed to improve both the analytical and
clinical sensitivity and specificity for measuring CK-MB compared with
the enzymatic assay. These mass assays utilize monoclonal anti-CK-MB
antibody in conjunction with anti-M or anti-B antibodies to accurately
measure small changes in CK-MB concentrations during the early hours
following AMI (Table 1
). Despite the fact that CK-MB expression is not
limited to cardiac tissue and that its diagnostic accuracy is low
(~50%) at 3 h after AMI, CK-MB is still regarded as the
benchmark for diagnosis of AMI. In the Chest Pain Evaluation Center
environment, Gibler et al. (14) examined the utility of
assaying CK-MB mass in a strategy that included sampling at
presentation and subsequently after 3, 6, and 9 h in >1000
low-risk nondiagnostic electrocardiogram patients. Although
these patients were at low myocardial infarction risk, this
study documented a sensitivity of 100% and specificity of 98.3% for
myocardial infarction diagnosis in this population of chest pain
patients (14).
The field of biochemical markers of cardiac injury is in a dynamic state, in which assays for novel markers and current markers such as cardiac troponin I and cardiac troponin T continue to be developed to enhance the diagnosis of AMI (15). However, progress in the development of several different biochemical assays for new AMI markers has led to several analytical and interpretative issues, particularly those pertaining to comparisons with different CK-MB immunoassays. Therefore, it was deemed essential by the AACC to establish rigorous criteria for standardizing CK-MB mass assays. To accomplish this task, the AACC established a subcommittee to identify and develop a standard CK-MB material to reduce method-related variations in the analytic accuracy of CK-MB mass assays.
To demonstrate the similarity in composition among CK-MB mass assays,
the characteristics of 12 commercially available CK-MB immunoassays are
displayed in Table 1
. All of these assays use an anti-CK-MB monoclonal
antibody as either the capture or detection antibody with the exception
of the Hybritech Tandem-E assay, which utilizes anti-CK-B (capture) and
anti-CK-M (detection) monoclonal antibodies. Ten of the remaining 11
assays use the "Conan" CK-MB antibody (11). The Chiron
ACS:180 CK-MB Mark I mass assay uses antibodies developed by the
manufacturer. In contrast, the methods listed in Table 1
use various
labels and detection schemes. A detailed description of these assays
has been reported previously (1)(4).
The CK-MB standardization process was conducted in four phases: (a) identification of a CK-MB candidate reference material, (b) determination of an optimum assay matrix, (c) validation of the biochemical characteristics of the identified reference material, and (d) testing the selected material on different CK-MB mass assays.
Although four different forms of CK-MB were examined, the results indicate that rCK-MB2 is the most suitable as a reference material because it can be easily prepared in large, homogeneous quantities compared with the other CK-MB forms. Lyophilized rCK-MB2 was favored over that in a liquid format because of its stability and ease of handling. Furthermore, validation experiments in this study have shown that rCK-MB2 has physicochemical characteristics similar to, if not equivalent to, CK-MB purified from human heart (11).
Examination of three different assay matrices revealed that stripped
human serum showed the best recovery. However, utilization of this
matrix to reconstitute CK-MB material did not produce a significant
reduction in method bias, which may be related to different
manufacturers' assay response to this matrix. Further experiments
demonstrated that the use of each respective manufacturer's sample
diluent to reconstitute a CK-MB reference material substantially
reduced method bias to ± 15% (Fig. 2
).
The findings of this study show that lyophilized rCK-MB2 can be used as a reference material for standardizing CK-MB mass assays. The systematic bias between the CK-MB methods was reduced from ± 40% to ± 13% with the use of rCK-MB2. Manufacturers should use this material to adjust their CK-MB mass assay calibration curves so that method-related bias is minimized or eliminated for this important AMI diagnostic tool.
The next steps of this process include (a) production and purification of large quantities of rCK-MB2, and (b) performance of studies necessary to gain acceptance of lyophilized rCK-MB2 as a worldwide CK-MB standard. Finally, this study may serve as a model for other future standardization activities of new cardiac markers, such as troponin I and myoglobin. In particular, standardization of biochemical markers of cardiac injury will be essential to achieve cost-effectiveness of diagnostic testing of AMI.
| Acknowledgments |
|---|
| Footnotes |
|---|
2 Chairperson of the AACC CK-MB Mass Assay Standardization
Subcommittee. ![]()
3 Nonstandard abbreviations: CK, creatine kinase; CK-MB, CK-MM, and CK-BB, isoenzymes MB, MM, and BB of CK, respectively; AMI, acute myocardial infarction; CK-M and CK-B, M and B subunits of CK, respectively; rCK-MB2, recombinant CK-MB2 (tissue isoform); SDS-PAGE, sodium dodecyl sulfate-polyacrylamide gel electrophoresis; and rCK-MB1, recombinant CK-MB1 (serum isoform, missing the C-terminal lysine on the M subunit). ![]()
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
J. L. Anderson, C. D. Adams, E. M. Antman, C. R. Bridges, R. M. Califf, D. E. Casey Jr, W. E. Chavey II, F. M. Fesmire, J. S. Hochman, T. N. Levin, et al. ACC/AHA 2007 Guidelines for the Management of Patients With Unstable Angina/Non-ST-Elevation Myocardial Infarction: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 2002 Guidelines for the Management of Patients With Unstable Angina/Non-ST-Elevation Myocardial Infarction) Developed in Collaboration with the American College of Emergency Physicians, the Society for Cardiovascular Angiography and Interventions, and the Society of Thoracic Surgeons Endorsed by the American Association of Cardiovascular and Pulmonary Rehabilitation and the Society for Academic Emergency Medicine J. Am. Coll. Cardiol., August 14, 2007; 50(7): e1 - e157. [Full Text] [PDF] |
||||
![]() |
J. L. Anderson, C. D. Adams, E. M. Antman, C. R. Bridges, R. M. Califf, D. E. Casey Jr, W. E. Chavey II, F. M. Fesmire, J. S. Hochman, T. N. Levin, et al. ACC/AHA 2007 Guidelines for the Management of Patients With Unstable Angina/Non ST-Elevation Myocardial Infarction Executive Summary: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 2002 Guidelines for the Management of Patients With Unstable Angina/Non ST-Elevation Myocardial Infarction) Developed in Collaboration with the American College of Emergency Physicians, the Society for Cardiovascular Angiography and Interventions, and the Society of Thoracic Surgeons Endorsed by the American Association of Cardiovascular and Pulmonary Rehabilitation and the Society for Academic Emergency Medicine J. Am. Coll. Cardiol., August 14, 2007; 50(7): 652 - 726. [Full Text] [PDF] |
||||
![]() |
R. H. Christenson, S. H. Duh, F. S. Apple, G. S. Bodor, D. M. Bunk, M. Panteghini, M. J. Welch, A. H.B. Wu, S. E. Kahn, and for the American Association for Clinical Chemistr Toward Standardization of Cardiac Troponin I Measurements Part II: Assessing Commutability of Candidate Reference Materials and Harmonization of Cardiac Troponin I Assays Clin. Chem., September 1, 2006; 52(9): 1685 - 1692. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Apple Clinical Biomarkers of Cardiac Injury: Cardiac Troponins and Natriuretic Peptides Toxicol Pathol, January 1, 2006; 34(1): 91 - 93. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Panteghini and F. Pagani AACC Creatine Kinase MB (CK-MB) Standardization Material Used as Manufacturer's Working Calibrator Is Unable to Harmonize CK-MB Results between Two Commercial Immunoassays Clin. Chem., September 1, 2004; 50(9): 1711 - 1712. [Full Text] [PDF] |
||||
![]() |
Y. Sunahara, K. Uchida, T. Tanaka, H. Matsukawa, M. Inagaki, and Y. Matuo Production of Recombinant Human Creatine Kinase (r-hCK) Isozymes by Tandem Repeat Expression of M and B Genes and Characterization of r-hCK-MB Clin. Chem., March 1, 2001; 47(3): 471 - 476. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. Jaffe, J. Ravkilde, R. Roberts, U. Naslund, F. S. Apple, M. Galvani, and H. Katus It's Time for a Change to a Troponin Standard Circulation, September 12, 2000; 102(11): 1216 - 1220. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |